Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K244036
    Date Cleared
    2025-02-26

    (58 days)

    Product Code
    Regulation Number
    878.4850
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Heel Incision Safety Lancet is intended for capillary blood sampling from the heel of newborns, preemies and toddlers.

    Device Description

    Heel Incision Safety Lancet is comprised of Outer cover, Press carrier and Blade holder (include the blade). The press carrier provides an elastic force to puncture and ensure the blade can shrink back to the covers. The blade can be fired after press the press carrier. The blade holder is to protect the blade from triggering before use and form a sterile barrier to maintain the blade sterile. The device was sterilized by Radiation.

    AI/ML Overview

    The SteriLance Medical (Suzhou) Inc. Heel Incision Safety Lancet (SteriHeel 2) underwent various performance tests to demonstrate substantial equivalence to its predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state acceptance criteria in quantitative terms for each test. Instead, it lists the types of tests performed to verify substantial equivalence. The reported device performance is that the device verifies or demonstrates performance characteristics and compliance.

    Acceptance Criteria CategoryReported Device Performance
    BiocompatibilityEvaluation conducted in accordance with ISO 10993-1.
    - Cytotoxicity test results
    - Skin Sensitization test results
    - Intracutaneous reactivity test results
    - Acute systemic toxicity test results
    Non-Clinical PerformanceBench testing performed to verify substantial equivalence in critical performance characteristics.
    - Appearance and type verified
    - Dimensions of product verified
    - Puncture function, Safety performance, and Single-use function verified
    - Cutting depth and cutting width verified
    - Hardness, Surface roughness, Corrosion resistance, and Sharpness verified
    - Assembly state, Cleanness, and Sterility verified
    Incision Length and DepthVerified in the performance test report.

    2. Sample Size and Data Provenance

    The document does not specify the sample sizes used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). It only mentions "bench testing performed" for non-clinical data and "biocompatibility evaluation" without detailing the number of units tested.

    3. Number of Experts and Qualifications for Ground Truth

    This information is not applicable. The device is a medical lancet, not an AI or diagnostic imaging device that requires expert review for establishing ground truth. The performance criteria are objective physical and biological properties.

    4. Adjudication Method

    This information is not applicable, as no expert adjudication was involved in establishing ground truth for the physical and biological performance tests of a medical lancet.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was mentioned or performed, as this device is a physical medical instrument, not a diagnostic AI system requiring human-in-the-loop performance evaluation.

    6. Standalone (Algorithm Only) Performance Study

    No standalone algorithm performance study was mentioned or performed, as the device is a physical medical instrument and does not incorporate an AI algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for the device's performance is established through objective physical measurements, chemical analyses, and biological assays based on recognized international standards and bench testing methods. These include:

    • Physical measurements (dimensions, hardness, surface roughness, sharpness)
    • Functional tests (puncture function, safety, single-use, cutting depth/width)
    • Biocompatibility tests (cytotoxicity, sensitization, systemic toxicity)
    • Sterility validation.

    8. Sample Size for the Training Set

    No training set is referenced or relevant, as this is a physical medical device and not an AI or machine learning model.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210745
    Date Cleared
    2021-08-26

    (167 days)

    Product Code
    Regulation Number
    878.4850
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Heel Incision Safety Lancet is intended for the collection of capillary blood from the heel of newborn, preemie, and toddler. The lancet has equipped with safety protection features.

    Device Description

    Heel Incision Safety Lancet is comprised of top upper cover, bottom cover, button, safety plug, slider, rod, holder, spring, blade. The spring provides an elastic force to puncture and ensure the blade can shrink back to the covers. The blade can be fired when the spring is under pressure. The safety plug is to protect the blade from triggering before use. Heel Incision Safety Lancet is single use, sterile medical devices designed to be used in collecting the blood sample. Heel Incision Safety Lancet is intended to be used by professionals. Heel Incision Safety Lancets are sterile and non-toxic. The product is intended for prescription (Rx) only.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for a medical device (Heel Incision Safety Lancet) and primarily focuses on demonstrating substantial equivalence to a predicate device. It is not a study proving a device meets acceptance criteria for an AI/ML-based medical device.

    Therefore, many of the requested details regarding acceptance criteria, AI/ML study design, sample sizes, expert involvement, and ground truth establishment (especially for AI/ML models) are not applicable and not present in this document. This document describes a traditional medical device (a lancet) and its non-clinical performance testing.

    However, I can extract the closest analogous information available from the document regarding acceptance criteria and performance for this non-AI/ML device:

    1. A table of acceptance criteria and the reported device performance

    The document provides a "Summary of non-clinical testing" with "Items" (which can be interpreted as the performance characteristics for which acceptance was sought) and "Results" (the performance shown). While explicit "acceptance criteria" values (e.g., "must be > X N Force") are not always stated, the "Results" implicitly demonstrate acceptance where they state "No failures observed", "Meet the requirement", or "Compared with equivalent devices, there was no statistical difference."

    Items (Performance Characteristic)Stated Results (Performance/Acceptance)Implied Acceptance Criteria (Based on Results)
    Material of bladeF type stainless steel materialsBlade material is F type stainless steel.
    AppearanceThe surface has no burr and scratchAbsence of burr and scratch on the surface.
    Dimensions of product (Height)The height of product(mm): 32±1.5Height within 32±1.5 mm.
    Dimensions of product (Width)The width of product(mm): 35±1.5Width within 35±1.5 mm.
    Blade corrosion resistanceNo blemishesNo blemishes due to corrosion.
    Force to activate Safety Screw buttonCompared with equivalent devices, there was no statistical differencePerformance of safety screw button activation force is statistically similar to equivalent devices.
    Trigger forceCompared with equivalent devices, there was no statistical differencePerformance of trigger force is statistically similar to equivalent devices.
    Safety overriding and unlocking force after activationCompared with equivalent devices, there was no statistical differencePerformance of safety overriding and unlocking force is statistically similar to equivalent devices.
    SterilitySterileDevice is sterile.
    Limits acidity and alkalinityMeet the requirementAcidity and alkalinity limits are met.
    Total heavy metalMeet the requirementTotal heavy metal requirements are met.
    Incision depth and lengthAll models have been tested within tolerancesIncision depth and length for all models are within specified tolerances.
    Safety self-lockingThere were no failures observed in a test run of 500 devicesNo failures in safety self-locking in 500 tests.
    Safety plug pulloutThere were no failures observed in a test run of 500 devicesNo failures in safety plug pullout in 500 tests.
    Shooting performanceThere were no failures observed in a test run of 500 devicesNo failures in shooting performance in 500 tests.
    Accidental access to sharp once in safe modeThere were no failures observed in a test run of 500 devicesNo accidental access to sharp in safe mode in 500 tests.

    Biocompatibility Testing:

    Item (Biocompatibility Test)Test Results (Performance/Acceptance)Implied Acceptance Criteria (Based on Results)
    In Vitro CytotoxicityNo CytotoxicityAbsence of cytotoxicity.
    Skin SensitizationNo Skin sensitizationAbsence of skin sensitization.
    Intracutaneous reactivityNo irritationAbsence of irritation.
    Acute Systemic Toxicity TestNo Acute Systemic ToxicityAbsence of acute systemic toxicity.
    Material mediated pyrogenicityAbsence of pyrogensAbsence of pyrogens.
    In vitro HemolyticNo HemolyticAbsence of hemolytic properties.

    2. Sample sizes used for the test set and the data provenance

    • Test Set Sample Sizes: For key functional tests like "Safety self-locking," "Safety plug pullout," "Shooting performance," and "Accidental access to sharp once in safe mode," a test run of 500 devices was used, with "no failures observed." For other physical property and material tests, specific sample sizes are not explicitly stated, but are implied to be sufficient for meeting the standard/requirement (e.g., "All models have been tested within tolerances").
    • Data Provenance: The tests are "bench testing" and "biocompatibility evaluations" conducted by the manufacturer, SteriLance Medical (Suzhou) Inc., based in Suzhou, Jiangsu, P.R.China. The data is prospective, generated specifically for this 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable as this is not an AI/ML device requiring expert-established ground truth from medical images or clinical data. The "ground truth" for this device's performance is established by direct measurement of physical properties (e.g., dimensions, forces) and standard laboratory tests (e.g., material analysis, sterility, biocompatibility).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies or AI/ML ground truth establishment where human readers might disagree. For bench testing of a physical device, results are typically objective measurements or pass/fail outcomes against defined specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable as this is a physical medical device, not an AI/ML-based diagnostic or assistive technology. No MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as this is a physical medical device, not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    As explained, for this physical device, the "ground truth" for performance is established by:

    • Direct physical measurements (e.g., dimensions, force values).
    • Standardized analytical tests (e.g., material composition, sterility).
    • Biocompatibility tests according to ISO standards.
    • Pass/fail criteria based on safety and functional requirements.

    8. The sample size for the training set

    This is not applicable as this is a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This is not applicable as this is a physical medical device, not an AI/ML algorithm that requires a training set and associated ground truth establishment.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1