Search Results
Found 10 results
510(k) Data Aggregation
(59 days)
Stryker Orthopaedics Hip Systems Labeling Update
DALL-MILES Cable System
The DALL-Miles System is indicated for reattachment of the trochanter in any hip procedure using the trochanteric osteotomy (total or partial) approach.
The DALL-MILES Mini Cleat is indicated for vertical reattachment or reinforcement of the trochanter in any situation where the surgeon feels that the trochanter is at risk for detachment. The Mini Cleat is intended for use with the DALL-MILES System for trochanteric reattachment only.
The DALL-MILES Cables and Cable Sleeves are indicated for trochanteric reattachment and trauma surgery of the hip; to stabilize bone graft material; and for supplementary cerclage fixation with plates and screws for fracture fixation.
The DALL-MILES Trochanteric Grips and Grip Plates are indicated for use in the fixation of the greater trochanter due to trochanteric fracture or osteotomy with intramedullary fixation as the primary device.
The DALL-MILES Trochanteric Grip Plate is additionally indicated for use in the fixation of the greater trochanter due to extended trochanteric osteotomies.
Femoral Heads
The indications for use for total hip and hemi hip arthroplasty include:
- Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- Rheumatoid arthritis;
- Correction of functional deformity;
- Revision procedures where other treatments or devices have failed; and,
- Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Femoral Mesh
The indications for use for total hip and hemi hip arthroplasty include:
- Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- Rheumatoid arthritis;
- Correction of functional deformity;
- Revision procedures where other treatments or devices have failed; and,
- Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Surgical Mesh. Surgical Mesh is intended to be used to reinforce bone or tissue in any situation where additional strengthening and support is required due to poor bone/tissue quality.
Intramedullary Plug, Centralizer
The indications for use of total hip replacement prostheses include:
· noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
· rheumatoid arthritis;
· correction of functional deformity;
· revision procedures where other treatments or devices have failed; and,
· treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
The EXETER Centralizer is intended to be used to centralize the femoral stem within the intramedullary canal and is intended to be used with bone cement.
The EXETER 2.5mm Intramedullary Bone Plug is intended to be used to restrict the migration of bone cement down the femoral canal and permit cement pressurization during total hip arthroplasty and is intended to be used with bone cement.
Exeter X3 RimFit Cups
The indications for use for total hip arthroplasty include:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or inadequate for other reconstructive techniques, such as cementless fixation, as indicated by deficiencies of the acetabulum.
The Exeter X3 RimFit Cup is intended for cemented use only.
Femoral Stems
The indications for use for total hip and hemi hip arthroplasty include:
- noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- rheumatoid arthritis;
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and,
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
The Exeter V40 Femoral Stem is intended for use in total or hemi hip replacement. It is intended for cemented use only.
The devices covered by this bundled submission are Stryker Total Hip Systems which include Dall-Miles cable system components, femoral heads, femoral mesh, Intramedullary Plug, Centralizer, acetabular cups, and femoral stems. All devices are commercially available and have been cleared in prior 510(k) submissions.
This FDA 510(k) letter and summary (K240418) is for a "Stryker Orthopaedics Hip Systems Labeling Update." This submission is not for a new medical device or an AI/software device that requires performance evaluation. Instead, it is a "Change Being Effected" bundled submission to add a contraindication to the labeling of existing, already cleared Stryker Orthopaedics Hip Systems.
Therefore, the provided document does not contain information about acceptance criteria or a study proving the device meets acceptance criteria. It explicitly states under "Non-Clinical Testing" and "Clinical Testing" that these were not required as a basis for substantial equivalence because the technological characteristics of the subject devices are identical to those of the predicate devices.
In summary, the request for information regarding acceptance criteria, device performance, sample sizes, ground truth, expert qualifications, adjudication methods, MRMC studies, or standalone performance is not applicable to this document, as it pertains to a labeling update for previously cleared devices, not a new device or a device requiring new performance data.
Ask a specific question about this device
(51 days)
Stryker Orthopaedics Hip Systems Labeling Update
ADM/MDM X3 Inserts, MDM Acetabular Inserts, MDM Acetabular Liners
The indications for use for total hip arthroplasty include:
-
- Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
-
- Rheumatoid arthritis:
-
- Correction of functional deformity;
- Revision procedures where other treatments or devices have failed; and.
- Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
- Dislocation risks
MDM Liners are intended for cementless use only.
Universal Cement Restrictor, OmniFit Distal Cement Spacer
For cement spacers, mid-shaft restrictors and Cement Plugs:
· In cemented hip arthroplasty, when the cement spacer, restrictor and/or plug is thought to be advantageous.
Artisan Bone Plug
These bone plugs are intended to be placed in the femoral canal prior to the introduction of bone cemented hip procedure. The plug is placed distally to the femoral stem to help allow cement pressurization and to help prevent cement migration further down the femoral canal.
Trident® II Acetabular System (Trident II Cups (Clusterhold HA, Tritanium Clusterhole, Tritanium Multihole, Tritanium Solidback), 6.5mm Low Profile Hex Screw, Hex Dome Hole Plug
· Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatic arthritis or late stage avascular necrosis.
· Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
· Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
When used with MDM Liners
· Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
• Dislocation risks
When used with Constrained Liner:
• The Trident Constrained Acetabular Insert is indicated for use in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability.
The Trident® II Acetabular Shells are indicated for cementless use only.
Restoration® Modular Hip System
The Restoration® Modular Hip System is indicated for use in:
· Noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- Rheumatoid arthritis
- · Correction of functional deformity;
- Revisions procedure where other treatments or devices have failed; and
• Nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques
Additional indications specific to the Restoration Modular Hip System
The Restoration® Modular Hip System is intended to be used for primary and revision total hip arthroplasty, as well as in the presence of severe proximal bone loss. These femoral stems are designed to be press fit into the proximal femur.
Accolade II Femoral Stems, Anato Femoral Stem, Secur-Fit Advanced
The indications for use for total hip arthroplasty with stems include:
- · noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- rheumatoid arthritis;
- · correction of functional deformity;
- · revision procedures where other treatments or devices have failed; and,
· nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Additional indication specific to use of the Femoral Stem with compatible Howmedica Osteonics Constrained Liners:
· When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
The Stems are intended for cementless use only and are intended for total and hemiarthroplasty.
TRIDENT AND TRITANIUM® ACETABULAR COMPONENTS
• Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatic arthritis or late stage avascular necrosis.
- · Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
· Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
The HOWMEDICA OSTEONICS TRIDENT and TRITANIUM Acetabular Shells are intended for cementless use only. Dome hole plug is indicated for cementless hip arthroplasty, when an acetabular shell plug is thought to be advantageous
Acetabular Dome Hole Plug
The Dome Hole Plug is an optional device which is available to seal the Howmedica Osteonics Acetabular Shell components during cemented or cementless applications of the Howmedica Osteonics Done Hole Plug is threaded into the dome hole of the shell.
Indications
• In cemented or cementless hip arthroplasty, when an acetabular shell plug is thought to be advantageous.
DALL-MILES® SYSTEMS
The DALL-MILES System is indicated for reattachment of the trochanter in any hip procedure using the trochanteric osteotomy (total or partial) approach.
The DALL-MILES Mini Cleat is indicated for vertical reattachment of the trochanter in any situation where the surgeon feels that the trochanter is at risk for detachment.
The Mini Cleat is intended for use with the DALL-MILES System for trochanteric reattachment only.
The DALL-MILES Cables and Cable Sleeves are indicated for trochanteric reattachment and trauma surgery of the hig; to stabilize bone graft material; and for supplementary cerclage fixation with plates and screws for fracture fixation.
The DALL-MILES Trochanteric Grips and Grip Plates are indicated for use in the greater trochanter due to trochanteric fracture or osteotomy with intramedullary fixation as the primary device.
The DALL-MILES Trochanteric Grip Plate is additionally indicated for use in the greater trochanter due to extended trochanteric osteotomies.
Exeter® V40™ Hip System (includes Orthinox V40 Femoral heads)
The Exeter® V40™ Femoral Hip System is indicated for:
· noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
• rheumatoid arthritis;
- correction of functional deformity;
- · revision procedures where other treatments or devices have failed; and,
· treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
The Exeter® V40™ Femoral Stem Hip System is intended for use in total or hemi hip replacement. It is intended for cemented use only.
Exeter Centralizer, EXETER Intramedullary Plug
- noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
-
- rheumatoid arthritis;
-
- correction of functional deformity;
-
- revision procedures where other treatments or devices have failed: and,
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
The Exeter Centralizer is intended to be used to centralize the femoral stem within the intramedullary canal. The Exeter Centralizer is intended to be used with bone cement.
The Exeter Intramedullary Bone Plug is intended to be used to restrict the migration of bone cement down the femoral canal and permit cement pressurization during total hip arthroplasty. The Exeter Intramedullary Bone Plug is intended to be used with bone cement.
Exeter X3 RimFit Cups
The indications for use for total hip arthroplasty include:
- Painful, disabling joint disease of the hip from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or inadequate for other reconstructive techniques, such as cementless fixation, as indicated by deficiencies of the acetabulum.
Restoration GAP II Acetabular Shell
Painful, disabling joint disease of the hip resultive arthritis, rheumatoid arthritis, post- traumatic arthritis or late stage avascular necrosis.
Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
Where bone stock is of poor quality or inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
Segmental and/or cavitary acetabular defects which make it difficult to restore normal hip biomechanics or to reconstitute the normal structural continuity and integrity of the acetabulum, using standard total hip replacement acetabular components and procedures.
Gap Screws, Torx Screws, and Osteolock Bone Screws
HOWMEDICA OSTEONICS Torx Cancellous Bone Screws are intended for supplemental fixation of associated HOWMEDICA OSTEONICS cementless Acetabular Shells.
· HOWMEDICA OSTEONICS RESTORATION GAP Plate Screws are intended for fixation of the dome and illac plates of the associated HOWMEDICA OSTEONICS RESTORATION GAP Acetabular Shell, TRIDENT TRITANIUM Hemispherical Multihole Acetabular Shells, restoration Acetabular Augments, and Restoration Anatomic shells.
Insignia Hip Stem
Hip Arthroplasty Indications:
• Painful, disabling joint disease of the hip resulting from: noninflammatory degenerative joint disease (including osteoarthritis or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
· Revision of previous unsuccessful femoral head replacement, hip arthroplasty or other procedure.
· Correction of functional deformity
· Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Additional indication specific to use of Insignia Hip Stems with compatible Howmedica Osteonics Constrained Liners: · When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
Additional indication specific to use of Insignia Hip Stems with compatible ADM and MDM Acetabular Components: • When the stem is to be used with compatible Howmedica Osteonics ADM and MDM Acetabular Components, the device is indicated for Dislocation risks
Insignia Femoral Stems are intended for cementless use only and are intended for total and hemiarthroplasty procedures.
Omnifit HFX Femoral Stems
For use as a Bipolar Hip Replacement:
- Femoral head/neck fractures or non-unions.
- · Aseptic necrosis of the femoral head.
· Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular involvement or distortion.
Other Considerations:
· Pathological conditions or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
• Salvage of failed total hip arthroplasty.
· Femoral neck fractures.
For use as a Total Hip Replacement:
• Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatic arthritis or late stage avascular necrosis.
· Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
· Clinical circumstances which require an altered femoral resection level due to a proximal fracture, bone loss or calcar lysis.
Omnifit EON Cemented Femoral Stems
For use as a Bipolar Hip Replacement:
· Femoral head/neck fractures or non-unions.
- · Aseptic necrosis of the femoral head.
· Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular involvement or distortion. Other Considerations:
· Pathological conditions or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
• Salvage of failed total hip arthroplasty. - For use as a Total Hip Replacement:
• Painful, disabling joint disease of the hip resulting from: degenerative arthritis, post-traumatic arthritis or late stage avascular necrosis. - · Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
Indications for use as a Total Hip Replacement include:
· noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; - rheumatoid arthritis;
- · correction of functional deformity;
- · revision procedures where other treatments or devices have failed; and,
· treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Restoration Anatomic Shell
• Painful, disabling joint disease of the hip from: degenerative arthritis, rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
· Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
· Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
When used with MDM Liners
· Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
• Dislocation risks
When used with Constrained Liner:
• The Trident Constrained Acetabular Insert is indicated for use in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability.
Trident Constrained Acetabular Insert/ Constrained Acetabular Insert
The Constrained Acetabular Insert is intended for use as a component of a total hip prosthesis in primary or revision patients at a high risk of hip dislocation due to a history of dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability.
Trident X3, Trident Crossfire, and Trident X3/Crossfire Elevated Rim Acetabular Liners
· Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatic arthritis or late stage avascular necrosis.
· Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
· Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
UHR Bipolar
- · Femoral head/neck fractures or non-unions.
- · Aseptic necrosis of the femoral head.
· Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular involvement or distortion.
Other Considerations:
· Pathological conditions or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum. - Salvage of failed total hip arthroplasty.
Unitrax V40 Head, V40 Adapter Sleeve and C-Taper Adapter Sleeve
The HOWMEDICA OSTEONICS Unitrax Endoprosthesis, and the V40 Modular Adaptor sleeves are used as a hemiarthroplasty device for the following indications: femoral neck fractures, idiopathic avascular necrosis, and nonunions. The C-taper sleeves are intended for use as a Hemi-Hip Replacement with the following indications: femoral head/neck fractures or nonunions, aseptic necrosis of the femoral head/neck and oster- and post traumatic arthritis. The patient's acetabular bone stock must be adequate to support articulation with the ad of the endoprosthesis.
BIOLOX Delta Ceramic Heads (V40, C-Taper and Universal and C-Taper to Universal Taper Adapter Sleeve)
The femoral heads are intended fixation to their mating hip stems, and can be used in cemented or cementless hip arthroplasty procedures.
For Use as a Total Hip Replacement:
· Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatic arthritis or late stage avascular necrosis.
- · Revision of previous cup arthroplasty or other procedures
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
· Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies in the acetabulum.
For Use as a Bipolar Hip Replacement - Femoral head/neck fractures or non-unions.
- Aseptic necrosis of the femoral head.
- · Osteo-, rheumatoid, and post traumatic arthritis of the hip with minimal acetabular involvement or distortion.
· Pathological considerations or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
The general intended use for these devices is "In general, these devices are intended for use in primary or revision hip arthroplasty."
Alumina C-Taper and V40 Ceramic Heads
For Use as a Total Hip Replacement:
· Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
· Revision of previous cup arthroplasty or other procedures
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
· Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies in the acetabulum.
Additional indications for the Alumina C-Taper Ceramic Heads
For Use as a Bipolar Hip Replacement
- Femoral head/neck fractures or non-unions.
- · Aseptic necrosis of the femoral head.
- · Osteo-, rheumatoid, and post traumatic arthritis of the hip with minimal acetabular involvement or distortion.
· Pathological considerations or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
· Salvage of failed total hip arthroplasty
Co Cr Femoral Heads (C-Taper and V40 Taper, LFIT and non-LFIT)
· Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatic arthritis or late stage avascular necrosis.
· Revision of previous cup arthroplasty or other procedures
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
· Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies in the acetabulum.
Accolade C Femoral Stems, Accolade HFx Femoral Stems
Indications for use as a Total Hip Replacement include:
· noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- rheumatoid arthritis;
- · correction of functional deformity;
- · revision procedures where other treatments or devices have failed; and,
· treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Accolade TMZF and Accolade TMZF Plus Femoral Stems
The subject hip stem is a single-use device intended for use in total hip replacement. It is intended for the reconstruction of the head and neck of the femoral joint. This hip stem is intended for primary reconstruction of the proximal femur or revision total hip arthroplasty. This device is intended for use with any currently available Howmedica Osteonics acetabular component and V40TM femoral heads that can be mated with a TMZF 5° 40'trunnion.
Indications:
· Cementless primary hip surgery in cases of non-inflammatory degenerative joint disease including osteoarthritis.
avascular necrosis, rheumatoid arthritis, and correction of functional deformity.
• Treatment of nonunion, and femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
· Revision procedures where other treatments or devices have failed.
The devices covered by this submission are Stryker Total Hip components which include femoral stems, acetabular shells, liners, femoral heads, acetabular augments, acetabular bone screws, acetabular plugs, cables, trochanteric grips, cement restrictors, and distal femoral spacers. All devices are commercially available and have been cleared in prior 510(k) submissions.
All the subject devices have been cleared for MR conditional in previous 510(k)s. The purpose of this submission is to modify the MR conditional information in the instructions for use to update the parameters in which a patient who has the device can be safely scanned, per testing conducted accordance to updated FDA guidance. There have been no changes made to the devices requiring 510(k) clearance - only the MR conditional information in the instruction for use is being modified.
This document is a 510(k) Premarket Notification from the FDA regarding a labeling update for Stryker Orthopaedics Hip Systems. It explicitly states that no changes have been made to the devices themselves, only to the MR conditional information in the instructions for use.
Therefore, the usual format for acceptance criteria and studies proving the device meets those criteria (which would typically involve performance metrics for an AI/software-based device) is not applicable here. This submission is about demonstrating that the updated MR Conditional labeling information is safe, not about a new or modified device's performance against clinical endpoints.
However, I can extract information related to the testing performed to support the updated labeling, which serves as the "study" in this context.
Here's a breakdown of the relevant information:
1. Table of Acceptance Criteria (for MR Conditional Labeling Safety) and Reported Device Performance:
Acceptance Criteria Category (for MR Conditional Labeling Safety) | Reported Device Performance (as described in the document) |
---|---|
Safety in Magnetic Resonance (MR) Environment | New testing performed to comprehensively assess the RF-related heating effects induced by the subject devices when implanted into bone. |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: Not specified in terms of number of devices or scenarios tested, but the document mentions that testing was "comprehensively" performed.
- Data Provenance: The testing was conducted by the manufacturer, Stryker Orthopaedics, to obtain data for updating the MR conditional information in the labeling. This is non-clinical testing.
- Retrospective or Prospective: This would be considered prospective testing, as new tests were conducted specifically to gather data for the labeling update.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- This concept is not applicable here. The "ground truth" for MR Conditional labeling is established through physical testing and engineering analysis according to regulatory standards and guidance documents, not expert consensus on medical images or clinical outcomes.
4. Adjudication Method for the Test Set:
- Not applicable. As the testing is non-clinical (device safety in an MR environment), there's no "adjudication" in the sense of clinician agreement on an outcome. The results are based on direct measurements from physical tests.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This type of study is for evaluating the clinical effectiveness of AI/software in assisting human readers (e.g., radiologists, pathologists). This submission is for a labeling update of a physical medical device (hip implants) regarding its MR compatibility and does not involve AI or human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- No. This is not an AI/algorithm-based device. It is a physical orthopedic implant.
7. The Type of Ground Truth Used:
- The "ground truth" for this submission is based on non-clinical testing results that demonstrate the safety of the device in an MR environment, specifically regarding RF-related heating effects. This is a combination of physical measurements and engineering analysis in accordance with FDA guidance documents. It is not expert consensus, pathology, or outcomes data.
8. The Sample Size for the Training Set:
- Not applicable. This product is a physical device, not an AI/machine learning algorithm requiring a training set.
9. How the Ground Truth for the Training Set Was Established:
- Not applicable. See point 8.
In Summary:
This FDA 510(k) notification is for a labeling change related to MR compatibility of a hip implant system, not for a new or modified device's clinical performance or an AI/software device. The "study" referenced is non-clinical testing to ensure the safety parameters provided in the updated labeling are accurate and supported by data, specifically related to RF-induced heating. The acceptance criteria and "performance" are therefore framed around meeting the requirements of the FDA guidance document for MR safety testing.
Ask a specific question about this device
(437 days)
Smith & Nephew Hip Systems
The components are indicated for the following:
- Non-inflammatory degenerative joint disease including osteonecrosis, avascular necrosis and post traumatic arthritis;
- Rheumatoid arthritis;
- Arthritis secondary to a variety of diseases and anomalies and correction of functional deformity such as congenital hip dysplasia or ankylosing spondylitis;
- Revision procedures where other treatment or devices have failed; and
- Treatment of proximal femoral non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement.
Hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma or noninflammatory degenerative joint disease (NID/D) or any of its composite diagnoses of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip. fracture of the pelvis, and diastrophic variant.
Hip components are also indicated for inflammatory degenerative joint disease including rheumatoid arthritis secondary to a variety of diseases and anomalies, and congenital dysplasia; treatments of nonunion, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques: endoprosthesis, femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity.
The indications for use in total hip arthoplasty include rheumatoid arthritis; post-traumatic arthritis; avascular necrosis and femoral neck fractures.
- Advanced degeneration of the hip joint as a result of degenerative, post-traumatic or rheumatoid arthritis.
- Fracture or avascular necrosis of the femoral head
- Failure of previous hip surgery: joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement
- Advanced degeneration of the hip joint as a result of degenerative, post-traumatic or rheumatoid arthritis.
- Fracture or avascular necrosis of the femoral head
- Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement
- All forms of osteoarthritis
- Patients with hips at risk of dislocation
- Femoral neck fracture or proximal fracture to hip joint
The titanium plasma and titanium/HA coated implants are intended to be implanted without bone cement. The uncoated implant is intended to be implanted with bone cement> The POLARCUP Dual Mobility System is intended for single use only.
The components are intended for primary or revision patients at hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, and intra-operative instability and for whom all other options to constrained acetabular components have been considered.
The BIRMINGHAM HIP Dual Mobility Insert is intended for use in BIRMINGHAM HIP Resurfacing (BHR) System revision surgeries in cases where an acetabular cup is retained and the femoral component revised.
The purpose of this Traditional 510(k) is to add "MR Conditional" labeling to the subject Smith & Nephew Hip Systems. The technological characteristics, function of the devices, packaging and sterilization remain unchanged. No modifications have been made to the devices' design, material, sterilization and the manufacturing processes of the previously cleared devices.
This document is a 510(k) Premarket Notification from Smith & Nephew, Inc. for various Hip Systems. The purpose of this submission is to add "MR Conditional" labeling to these hip systems.
Here's an analysis based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The document describes MR Conditional labeling, meaning the devices have been tested and found to be safe for use in specific MRI environments. The acceptance criteria are based on established standards for MRI compatibility.
Test Parameter / Acceptance Criteria (Standard) | Reported Device Performance (Summary from study) |
---|---|
Magnetically Induced Displacement Force (ASTM F2052-15) | Not explicitly detailed, but implied to be within safe limits for "MR Conditional" labeling. |
Magnetically Induced Torque (ASTM F2213-2017) | Not explicitly detailed, but implied to be within safe limits for "MR Conditional" labeling. |
Radiofrequency (RF) Induced Heating (ASTM F2182-19e2) | Not explicitly detailed, but implied to be within safe limits for "MR Conditional" labeling. |
MR Image Artifacts (ASTM F2119-07 (2013)) | Not explicitly detailed, but implied to be assessed and within acceptable limits for "MR Conditional" labeling. |
General MR Safety for Active Implantable Medical Devices (ISO/TS 10974:2018(E)) | Implied the devices meet relevant safety aspects, although these are passive implants. |
Basic Safety and Essential Performance of MRI Equipment (IEC 60601-2-33 (Ed 3.2)) | Implied the devices are compatible with equipment meeting this standard. |
Key takeaway: The document confirms that "Magnetic Resonance Imaging (MRI) compatibility testing was conducted as per the FDA's guidance and the Standards listed" to support the "MR Conditional" labeling. However, it does not provide specific data points (e.g., maximum temperature rise, angular deflection, or translational force metrics) for each test. Instead, it makes a general statement that the testing supports the MR Conditional claim.
2. Sample Size Used for the Test Set and Data Provenance:
The document does not specify a "test set" in terms of patient data or clinical images, as the study is focused on the physical devices' compatibility with MRI environments, not clinical performance or diagnostic accuracy.
The studies were conducted on the devices themselves (hip systems and their components). The provenance is implied to be laboratory testing based on the referenced ASTM and ISO standards for MRI compatibility. The data is prospective in the sense that the tests were performed to demonstrate compliance for new labeling.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
This information is not applicable to this type of submission. The "ground truth" here is established by the scientifically validated test methods and established standards (e.g., ASTM, ISO, IEC) for MRI compatibility, not by expert consensus on clinical cases. The tests themselves define the "ground truth" for the physical properties being measured (e.g., magnetic susceptibility, RF heating).
4. Adjudication Method for the Test Set:
This is not applicable as the studies are physical/engineering tests against objective standards, not interpretation of clinical data requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
No, an MRMC comparative effectiveness study was not done. This type of study is used to assess the diagnostic performance of an AI algorithm in comparison to or in conjunction with human readers. This submission is for the "MR Conditional" labeling of hip systems, which is a safety and functional characteristic of the physical implant, not a diagnostic AI device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
No, a standalone algorithm performance study was not done. This submission is not for an algorithm or AI device.
7. The Type of Ground Truth Used:
The ground truth used for these "MR Conditional" studies is objective physical measurements conducted according to recognized international consensus standards (ASTM, ISO, IEC) and FDA guidance documents. These standards define the acceptable limits for parameters like magnetically induced displacement, torque, and RF-induced heating.
8. The Sample Size for the Training Set:
This information is not applicable as this is not an AI/machine learning device requiring a training set. The "training" for the hip systems refers to their design and manufacturing to meet performance specifications.
9. How the Ground Truth for the Training Set was Established:
This information is not applicable as this is not an AI/machine learning device. The "ground truth" for the design and manufacturing of the hip systems is established through engineering principles, material science, biomechanical testing, and adherence to quality systems and regulatory requirements, which are inherent to their previous clearances. The ground truth for the MR Conditional labeling is established by the referenced MRI compatibility test standards.
Ask a specific question about this device
(405 days)
Trinity Acetabular System, Trinity PLUS Acetabular Shell, MetaFix Hip System, TriFit CF and TS Hip
Systems, TaperFit Hip System, Revival Modular Hip System, MiniHip, Trinity Dual Mobility, MobiliT
, BiPolar-i, OMNI MOD Hip System, OMNI K1 and K2 Hip Systems, OMNI Bipolar Heads, Corin Biolox Delta
The BiPolar-i is intended for use in the following indications: Non-inflammatory degencrative joint disease including osteoarthritis and avascular necrosis in which the acetabulum does not require replacement, Treatment of non-union, femoral neck and trochanteric fractures of the proximal femur, Revision of failed partial hip replacements in which the acetabulum does not require replacement. The BiPolar-i is indicated for cementless use only.
The Trinity Acetabular System is indicated for use in non-intlammatory degenerative joint disease including osteoarthritis and avascular necrosis, rheumatoid arthritis, correctional deformity, developmental dysplasia of the hip (DDH) or congenital dysplasia of the hip (CDH). The Trinity Acctabular System is intended for cementless, single use only.
The indications for the Corin MctaFixM Hip Stem as a total hip arthroplasty, and when used in combination with a Corin hemi arthroplasty head, as a hip hemi-arthroplasty, include: Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis, Rheumatoid arthritis, Correction of functional deformity, Treatment of non-union and femoral neck fractures, Developmental dysplasia of the hip (DDH) and congenital dysplasia of the hip (CDH). The Corin MetaFixTM Hip Stem is indicated for cementless use only.
The indications for the MiniHip Stem as a total hip arthroplasty include: Non-inflammatory degenerative joint disease including osteoarthriis and avascular necrosis Rheumatoid arthritis Correction of functional deformity. Developmental dysplasia of the hip (DDH) and congenital dysplasia of the hiniHip Stem is indicated for cementless use only.
The indications for the TrinityTM Accabular System as a total hip arthroplasty include: Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis, Rheunatoid arthritis, Correction of functional deformity, Developmental dysplasia of the hip (DDH), and congenital dysplasia of the hip (CDH). The Trinity Acctabular System is intended for cementless, single use only.
The Trinity Dual Mobility System is intended for use in the following indications: 1. Non-inflammatory degenerative joint disease, including osteoarthritis & avascular necrosis 2. Rheumatoid Arthritis 3. Correction of functional deformity 4. Revision of previously failed total hip arthroplasty 5. Patients at increased risk of dislocation 6. Developmental dysplasia of the hip (DDH). The Trinity TM Dual Mobility System is indicated for cementless use only.
The indications for the Corin Trinity™ PLUS Accabular Shell as a total hip arthroplasty include: Non-inflammalory degenerative joint disease including ostoarthritis and avascular necrosis. Rheumatoid arthritis. Correction of functional deformity, Revision of previously failed total hip arthroplasty, Developmental dysplasia of the hip (DDH). The Trinity TM PLUS Acctabular Shell is indicated for cement less use only.
The MobiliT Cup, for cemented and cementless use, are indicated for primary replacement of the hip joint: - In degenerative pathologies: primary, secondary or post-traumatic osteoarthritis, rheumatoid arthritis - For patients who have a high risk of dislocation - In cases of necrosis of the femoral head - In cases of fracture of the neck of the femur - In cases of congenital luxation. The MobiliT Cup, for cemented and cementless use, are indicated for revision when the bone tissue remains sufficient after the removal of the previous acetabular cup. The cementless MobiliT standard Cup, with flanges or with flanges and hook are indicated for cementless use only. The cemented MobiliT Cup is indicated for cemented use only.
The Apex Hip System Bipolar Head is intended for use in combination with an Apex Hip System for uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate: Femoral neck and trochanteric fractures of the proximal femur. Osteonecrosis of the femoral head, Revision procedures where other devices or treatments for these indications have failed.
The OMNI Hip system Ceramic Femoral Heads are intended for use in combination with the OMNI Hip System Stems as the femoral component in total hip replacement procedures. This ceramic head is intended to articulate with the OMNI Interface Acctabular System or bipolar component. This prosthesis is intended for single use may be used for the following conditions, as appropriate: Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; Rheumatoid arthritis; Correction of functional deformity; Congenital dislocation; Revision procedures where other treatments or devices have failed; Femoral neck and trochanteric fiactures of the proximal femur.
The indications for use of the OMNI Modular Hip Stems in hip arthroplasty include the following conditions, as appropriate: Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; Rheumatoid arthritis: Correction of functional deformity; Congenital dislocation: Revision procedures where other treatments or devices have failed; Femoral neck and trochanteric fractures of the proximal femur. The OMNI Modular Hip stems are indicated for cementless use only and single use implantation.
The indications for use of the OMNI Modular Hip Stems in hip arthroplasty include the following conditions, as appropriate: Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; Rheumatoid arthritis: Correction of functional deformity; Congenital dislocation: Revision procedures where other treatments or devices have failed; Femoral neck and trochanteric fractures of the proximal femur. The OMNI Modular Hip stems are indicated for cementless use only and single use implantation.
The Revival Modular Revision Hip Stem is indicated in revision surgery of femoral components, following failure of primary cemented or un-cemented prosthesis. The REVIVAL™ Hip Stem 100mm distal component is also indicated in primary total hip arthroplasty. The indications for the Revival TM Modular Revision Hip Stem include: Non-inflammatory degenerative joint disease including primary and secondary osteoarthritis. Rheumatoid arthritis, Correction of functional deformity, Treatment of non-union and femoral neck fractment of traumatic dislocations of the hip, Failures of osteotomy, Treatment of arthrodesis. The Revival ™ Revision Hip Stem is indicated for cementless, single use only.
TaperFirM Hip Stem is indicated for the relief of pain and restoration following the effects of femoral neck fracture, osteo, theumatory arthritis, post- traumatic disease effects, avascular necrosis and total hip revision. The Taper it Hip Stem is indicated for hemi-arthroplasty when used in combination with Corin hemiarthroplasty femoral heads. The TaperFitTM Hip Stem is indicated for cemented, single use only.
The indications for use of the K 1 Hip arthroplasty include the following conditions, as appropriate: Noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis: Rheumatoid arthritis: Correction of functional deformity: Congenital disfocation: Revision procedures where other treatments or devices have failed; Femoral neck and trochanteric fractures of the proximal femur. The K 1 Hip Stem is indicated for cementless use only and single use implantation.
The indications for the TriFit CF Hip Stem as a total hip arthroplasty and as a hip hemiarthroplasty include: Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis, Rheumatoid arthritis, Correction of functional deformity, Treatment of non-union and femoral neck fractures, Developmental Dysplasia of the Hip (DDH), Previously failed hip surgery. The Trifit CF Hip Stem is indicated for cementless use only.
The indications for the Corin TriFit TSTM Hip atthroplasty, and when used in combination with a Corin hemi arthroplasty head, as a hip hemi-arthroplasty, include: Non-inflammatory degenerative joint disease including osteoadhritis and avascular necrosis, Rheumatoid arthritis, Correction of functional deformity, Treatment of non-union, femoral neck and trochanteric fractures of the proximal femur, Developmental dysplasia of the hip (DDH) or congenital dysplasia of the hip (CDH). The Tril'it TS Hip is intended for cementless use only.
The subject and predicate devices are one in the same and are comprised of several legally marketed Corin Ltd. hip products, which include OMNIIfe Sciences and Apex Surgical hip products. The subject devices include acetabular cups and liners, bone fixation screws, screw hole occluders, cemented and cementless femoral hip stems for primary and revision hip arthroplasty, fixation screws, modular necks, CoCrMo alloy and ceramic femoral heads, dual mobility acetabular systems, and bipolar heads. The purpose of this 510(k) is to notify the FDA of Corin's engineering assessment of the cross-compatibility of the subject devices, identification of conflicts, and updates to the product labeling. The subject hip devices components are manufactured from a variety of materials which include cobalt-chromium-molybdenum alloy, stainless steel alloy, unalloyed titanium, calcium phosphate (Bonit™ coating) Alumina Matrix Composite ceramic (Biolox Delta), and ultrahigh molecular weight polyethylene (UHMWPE), all of which conform to ASTM or ISO standards, or internal standards. The subject femoral hip stems and heads possess the same 12/14 taper design and reference system for determining head and neck offsets.
I am sorry, but the provided text does not contain information about the acceptance criteria and study proving device performance as typically seen in a medical device submission beyond general statements of substantial equivalence. The document is a 510(k) summary for hip replacement components, primarily focusing on cross-compatibility of various existing devices.
The text outlines:
- Device identification and manufacturer information.
- List of numerous predicate devices.
- Detailed descriptions of the indications for use for many hip components (e.g., Corin BiPolar-i, Trinity™ Acetabular System, MetaFix™ Hip, MiniHip™, Trinity™ Dual Mobility, MobiliT™ Cup, OMNI Bipolar Head, OMNI Delta Ceramic Femoral Head, OMNI MOD Hip System, OMNI K1/K2 Hip Systems, Revival™ Modular Hip Stem, TaperFit™, TriFit™ CF/TS Hip). These indications primarily relate to non-inflammatory degenerative joint disease, rheumatoid arthritis, correction of functional deformity, avascular necrosis, fractures, and revision surgeries.
- A "Performance Data" section which describes the types of engineering analyses and bench testing performed to establish cross-compatibility (e.g., ceramic head burst testing, head pull-off, fretting-corrosion, impingement testing, range of motion assessment, comparison of taper geometries, fatigue strength assessment, and contact stress/wear potential).
However, it does not provide:
- A specific table of acceptance criteria and reported device performance for the types of tests mentioned (e.g., what burst pressure was required vs. achieved). It only lists the types of tests done.
- Sample sizes used for test sets or data provenance.
- Number of experts and their qualifications for establishing ground truth (as this pertains to clinical studies, which are not detailed here for performance).
- Adjudication method.
- Multi-reader multi-case (MRMC) comparative effectiveness study information.
- Standalone algorithm performance (as this is not an AI/algorithm-driven device).
- Type of ground truth used (again, this is not a diagnostic device with "ground truth" in the typical sense).
- Sample size for training set.
- How ground truth for the training set was established.
The document's purpose is to demonstrate substantial equivalence of a range of hip components, including their cross-compatibility when used together. The "performance data" refers to the engineering and bench testing conducted to ensure this compatibility rather than clinical performance against specific metrics as one would find for a diagnostic or AI-driven device.
Ask a specific question about this device
(272 days)
MicroPort Orthopedics Total Hip Systems MR Labeling
MicroPort total hip systems are intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.
- non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
- inflammatory degenerative joint disease such as rheumatoid arthritis;
- correction of functional deformity; and,
- revision procedures where other treatments or devices have failed
The MicroPort total hip systems devices are single use only devices.
The subject devices are a variety of hip joint replacement prostheses. The components for these systems include an acetabular shell, acetabular liner, fixation screws, femoral head, femoral stem, modular neck, proximal body, centralizers, bone plugs, and neck sleeves. These components can be utilized in a variety of configurations to assemble the final construct. The femoral and acetabular components are manufactured from a variety of materials which include cobalt-chromium-molybdenum alloy, titanium alloy, unalloyed titanium, alumina ceramic, Alumina Matrix Composite ceramic (Biolox Delta), polymethylmethacrylate (PMMA), and ultra high molecular weight polyethylene (UHMWPE), all of which conform to ASTM or ISO standards, or internal standards. The only changes to the subject hip systems are updates to the labeling of the devices. Specifically, the package inserts and package labels are being updated to include MR Conditional language and symbols. The subjects are identical to the predicates in all aspects except for the labeling updates.
The information provided is about the MicroPort Orthopedics Inc. Hip Systems MR Labeling, which involves updating the labeling of existing hip systems to include MR Conditional language and symbols. This submission is for a labeling change and not for a new medical device or a change in the device's fundamental design or intended use.
Here's an analysis of the acceptance criteria and study information based on the provided document:
1. A table of acceptance criteria and the reported device performance
Since this submission is a labeling change for existing predicate devices to include MR Conditional language and symbols, the "acceptance criteria" and "reported device performance" are related to satisfying the requirements for safe use in an MR environment. The performance is documented through non-clinical testing.
Acceptance Criteria (for MR Safety and Compatibility) | Reported Device Performance (Summary) |
---|---|
Magnetic Field Interactions (displacement force): Demonstrate no unsafe magnetically induced displacement force in a specified MR environment (per ASTM F2052-6). | The tests determined that there are no safety issues related to magnetic field interactions under specific conditions identified in the labeling. This implies that the devices met the criteria for displacement force. |
MR Image Artifacts: Evaluate the extent of MR image artifacts (per ASTM F2119-7). | The tests determined the effects of the implants on the image quality, implying image artifacts were evaluated and deemed acceptable within the context of labeling for safe use. |
Radio Frequency (RF) Induced Heating: Demonstrate no unsafe RF induced heating in a specified MR environment (per ASTM F2182-11a). | The tests determined the effects of the MRI on the implants (including RF Heating) by evaluating the worst-case components and constructs. The conclusion indicates no safety issues, suggesting RF heating was within acceptable limits for specific conditions. |
Marking and Labeling for MR Safety: Ensure medical devices are appropriately marked with MR safety information (per ASTM F2503-13) and include MR Conditional language. | The only changes to the subject hip systems are updates to the labeling of the devices, specifically "to include MR Conditional language and symbols." This directly addresses the criterion for proper marking and labeling. The overall conclusion states that "All the information provided in this submission adequately supports the substantial equivalence of the labeling change," implying this criterion was met. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document does not explicitly state a numerical "sample size" for the test set or the "country of origin of the data." It mentions that "worst case components and constructs" were evaluated for non-clinical testing.
The data provenance is non-clinical testing, performed according to recognized ASTM standards (F2052-6, F2119-7, F2503-13, F2182-11a) and FDA guidance documents. This type of testing is generally performed in a controlled laboratory environment. It is not "retrospective or prospective" in the sense of clinical patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This information is not provided in the document. For non-clinical testing of MR compatibility, "ground truth" is typically established by adhering to standardized testing protocols and measurement methodologies, rather than expert consensus on patient data. The experts involved would be engineers and scientists specialized in MR safety testing.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
This information is not applicable or not provided. Adjudication methods like "2+1" or "3+1" are relevant for clinical studies where multiple human readers interpret medical images and their interpretations need to be reconciled to establish a consensus ground truth. This submission is based on non-clinical, objective laboratory testing for MR safety and compatibility, not human interpretation of images.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This submission is not for an AI device but for a labeling change for hip implant systems to ensure their safety and compatibility in an MR environment. Therefore, the concept of human readers improving with or without AI assistance is not relevant to this submission.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
No, a standalone (algorithm only) performance study was not done. This submission is for mechanical implants and their MR compatibility, not for a software algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for this submission's non-clinical testing is based on objective measurements and adherence to recognized industry standards (ASTM standards) for MR safety and compatibility. This includes physical measurements of magnetic field interaction, temperature changes due to RF heating, and visual/quantitative assessment of image artifacts in controlled laboratory settings. It's not based on expert consensus of clinical data, pathology, or outcomes data.
8. The sample size for the training set
Not applicable. This submission is for a labeling change for existing mechanical hip implant systems, not for a machine learning or AI device that requires a training set.
9. How the ground truth for the training set was established
Not applicable. As a non-AI device submission for labeling changes, there is no training set or ground truth in that context.
Ask a specific question about this device
(188 days)
Stryker Orthopaedics Hip Systems Labeling Update
In general, these devices are intended for use in primary or revision hip arthroplasty. Specific indications appear below:
Trident PSL Shell, Trident Tritanium Shell, Trident Hemispherical Shell
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
The Trident shells are intended for cementless fixation within the prepared acetabulum.
Tritanium Acetabular Shell System
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
The Tritanium Acetabular Shell System is intended for cementless use only.
Accolade II Femoral Stem, Secur-Fit Advanced Femoral Stem, Anato Femoral Stem
The indications for use for total hip arthroplasty with stems include:
- noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- rheumatoid arthritis;
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and,
- nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Additional indication specific to use of the Femoral Stem with compatible Howmedica Osteonics Constrained Liners: - When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
Accolade II, Secur-Fit Advanced and Anato Femoral Stems are intended for cementless use only and are intended for total and hemiarthroplasty.
Exeter® V40TM Hip System (includes Orthinox V40 Femoral heads)
The Exeter® V40™ Hip System is intended for use in total hip replacement. It is intended for cemented use only.
The Exeter® Hip is indicated for:
- noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- rheumatoid arthritis:
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and,
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Restoration Anatomic Shell
Indications for Use
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
When used with MDM Liners
- Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
- Dislocation risks
When used with Constrained Liner:
- The Trident Constrained Acetabular Insert is indicated for use in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability.
The Restoration® Anatomic Shell is indicated for cementless use only.
Omnifit HFX Femoral Stems
Indications:
For use as a Bipolar Hip Replacement:
- Femoral head/neck fractures or non-unions.
- Aseptic necrosis of the femoral head.
- Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular involvement or distortion.
Other Considerations:
- Pathological conditions or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty.
- Femoral neck fractures.
For use as a Total Hip Replacement:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Clinical circumstances which require an altered femoral resection level due to a proximal fracture, bone loss or calcar lysis.
Omnifit EON Cemented Femoral Stems
Indications
For use as a Bipolar Hip Replacement:
- Femoral head/neck fractures or non-unions.
- Aseptic necrosis of the femoral head.
- Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular involvement or distortion.
Other Considerations:
- Pathological conditions or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty.
For use as a Total Hip Replacement:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
Secur-Fit Max and Secur-Fit Max Plus Hip Stems
Secur-Fit Max and Secur-Fit Max Plus Hip stems are single use devices and are intended for cementless fixation within the prepared femoral canals of patients requiring hip arthroplasty.
Indications for use as a Total Hip Replacement include:
- noninflammatory degenerative joint disease. including osteoarthritis and avascular necrosis:
- rheumatoid arthritis;
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and,
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Accolade C Femoral Stems
The Accolade C Femoral Stems are single-use devices intended for cemented fixation.
Indications for use as a Total Hip Replacement include:
- noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- rheumatoid arthritis;
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and,
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Accolade HFx Femoral Stems
- noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- rheumatoid arthritis:
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and,
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Accolade TMZF and Accolade TMZF Plus Femoral Stems
The subject hip stem is a single-use device intended for use in total hip replacement. It is intended for the reconstruction of the head and neck of the femoral joint. This hip stem is intended for primary reconstruction of the proximal femur or revision total hip arthroplasty. This device is intended for use with any currently available Howmedica Osteonics acetabular component and V40™ femoral heads that can be mated with a TMZF 5° 40' trunnion.
Indications:
- Cementless primary hip surgery in cases of non-inflammatory degenerative joint disease including osteoarthritis, avascular necrosis, rheumatoid arthritis, and correction of functional deformity.
- Treatment of nonunion, and femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
- Revision procedures where other treatments or devices have failed.
Restoration Acetabular Wedge Augments
The indications for use of the Restoration Acetabular Wedge Augments:
General Indications for Total Hip Replacement Components:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
Indications Specific to the Acetabular Wedges:
- As an alternative to structural allograft in cases of superior and superior/posterior segmental acetabular deficiencies.
Acetabular Augments are intended for cementless use only to the bone interface, and are affixed to the mating cup using bone cement
Trident AD Acetabular shells
The Trident AD Acetabular shells are single use devices. The shell is intended for cemented or cementless fixation within the prepared acetabulum. The Trident AD Acetabular Component System is compatible with any appropriately selected Howmedica Osteonics hip stem/femoral head combination.
Indications:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
Trident T Acetabular shells
The Trident Acetabular shells are single-use devices intended for cementless fixation within the prepared acetabulum. They are compatible with Trident polyethylene acetabular bearing insert. If additional fixation is desired, the dome screw holes, if present, have been designed to accept Stryker Orthopaedics 6.5mm or 5.5mm bone screws.
General Indications for Total Hip Replacement Components:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
Restoration ADM System, Modular Dual Mobility (MDM) Liner and X3 Acetabular Insert
The indications for use of the total hip arthroplasty prostheses include:
- Noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- Rheumatoid arthritis:
- Correction of functional deformity; .
- Revision procedures where other treatments or devices have failed;
- Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
- Dislocation risks.
These devices are intended for cementless use only
Acetabular Dome Hole Plug
The Dome Hole Plug is an optional device which is available to seal the Howmedica Osteonics Acetabular Shell components during cemented or cementless applications of the acetabular cup. The Howmedica Osteonics Done Hole Plug is threaded into the dome hole of the shell.
Indications
- In cemented or cementless hip arthroplasty, when an acetabular shell plug is thought to be advantageous.
Trident X3, Trident Crossfire, Trident X2Vac and Trident X3/Crossfire Elevated Rim Acetabular Liners - Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated bv deficiencies of the acetabulum.
Trident® Constrained Acetabular Insert
The Trident® Constrained Acetabular Insert is intended for use as a component of a total hip prosthesis in primary or revision patients at a high risk of hip dislocation due to a history of dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability.
BIOLOX Delta Ceramic Heads (C-Taper, Universal Taper, V40 to Universal Taper Adapter Sleeve, and C-Taper to Universal Taper Adapter Sleeve)
The femoral heads are intended for mechanical fixation to their mating hip stems, and can be used in cemented or cementless hip arthroplasty procedures.
For Use as a Total Hip Replacement:
- Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous cup arthroplasty or other procedures
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies in the acetabulum.
For Use as a Bipolar Hip Replacement
- Femoral head/neck fractures or non-unions.
- Aseptic necrosis of the femoral head.
- Osteo-, rheumatoid, and post traumatic arthritis of the hip with minimal acetabular involvement or distortion.
- Pathological considerations or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty
V40 BIOLOX Delta Ceramic Heads
The femoral heads are intended for mechanical fixation to their mating hip stems, and can be used in cemented or cementless hip arthroplasty procedures.
For Use as a Total Hip Replacement:
- Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous cup arthroplasty or other procedures
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
For Use as a Bipolar Hip Replacement
- Femoral head/neck fractures or non-unions. .
- Aseptic necrosis of the femoral head.
- Osteo-. rheumatoid, and post traumatic arthritis of the hip with minimal acetabular involvement or distortion.
Other considerations:
- Pathological considerations or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty
Alumina C-Taper Ceramic Heads
For Use as a Total Hip Replacement:
- Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage vascular necrosis.
- Revision of previous cup arthroplasty or other procedures
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
For Use as a Bipolar Hip Replacement
- Femoral head/neck fractures or non-unions.
- Aseptic necrosis of the femoral head.
- Osteo-, rheumatoid, and post traumatic arthritis of the hip with minimal acetabular involvement or distortion.
Other Considerations:
- Pathological considerations or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty .
Alumina V40 Ceramic Heads
- Painful, disabling joint disease of the hip resulting from: Non-inflammatory degenerative arthritis (osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, pelvic fracture, failed fracture fixation, or diastrophic variant);
- Rheumatoid arthritis
- Correction of functional deformity
- Where bone stock is of poor quality or inadequate for other reconstructive techniques as indicated by the deficiencies of the acetabulum;
- Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques;
- Revision procedures where other treatments or devices have failed.
C-Taper CoCr Femoral Heads (LFIT and non-LFIT)
For use as a Bipolar Hip Replacement:
- Femoral head/neck fractures or non-unions 1.
-
- Aseptic necrosis of the femoral head.
- Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular involvement 3. or distortion.
Other Considerations:
-
- Pathological conditions or age considerations which indicate a more conservative approach to the acetabulum and the avoidance of the use of bone cement in the acetabulum.
-
- Salvage of failed total hip arthroplasty.
Indications for use as part of a Total Hip Replacement include:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid 1. arthritis, post-traumatic arthritis or late stage avascular necrosis.
-
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty, or other procedure.
-
- Clinical management situations where arthrodesis or alternate reconstructive techniques are less likely to achieve satisfactory results.
V40 CoCr Femoral Heads (LFIT and non-LFIT)
- Clinical management situations where arthrodesis or alternate reconstructive techniques are less likely to achieve satisfactory results.
- Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- Rheumatoid arthritis
- Correction of functional deformity
- Revision procedures where other treatments or devices have failed
- Nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Unitrax Endoprosthesis, Unitrax V40 Adapter Sleeve, Unitrax C-Taper Adapter Sleeve
- The Howmedica Osteonics Unitrax Endoprosthesis, and the Unitrax V40 Modular Adaptor sleeves are used as a hemiarthroplasty device for the following indications: femoral neck fractures, idiopathic avascular necrosis, and non-unions. The Unitrax C-taper sleeves are intended for use as a Hemi-Hip Replacement with the following indications: femoral head/neck fractures or non-unions, aseptic necrosis of the femoral head/neck and osteo- and post traumatic arthritis. The patient's acetabular bone stock must be adequate to support articulation with the head of the endoprosthesis.
Universal Distal Spacer
Howmedica Osteonics Corp.'s accessory products for cemented arthroplasty are optional devices intended to assist in the preparation, implantation and/or positioning of a femoral implant intended for cemented application.
Indications
For cement spacers, mid-shaft restrictors and Cement Plugs:
- In cemented hip arthroplasty, when the cement spacer, restrictor and/or plug is thought to be advantageous.
Torx Screws, GAP Plate Screws and Osteolock Bone Screws - Howmedica Osteonics Torx Cancellous Bone Screws are intended for supplemental fixation of associated Howmedica Osteonics cementless Acetabular Shells or Howmedica Osteonics Tibial Tray components.
- Howmedica Osteonics Osteolock Bone Screws are intended for supplemental fixation of associated cementless Trident Tritanium Multihole Acetabular shells, Restoration Acetabular Augments, and Restoration Anatomic Shell.
- Howmedica Osteonics Restoration GAP Plate Screws are intended only for fixation of the dome and iliac plates of the associated Howmedica Osteonics Restoration GAP Acetabular Shells, Trident Tritanium Hemispherical Multihole Acetabular Shells, Restoration Acetabular Augments and Restoration Anatomic Shell.
V40 to C-Taper Adapter Sleeve For Use as a Total Hip Replacement
- Painful disabling joint disease of the hip resulting from: degenerative arthritis. rheumatoid arthritis, post-traumatic arthritis or late stage vascular necrosis.
- Revision of previous cup arthroplasty or other procedures.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
For Use as a Bipolar Hip Replacement
- Femoral head/neck fractures or non-unions.
- Aseptic necrosis of the femoral head.
- Osteo-, rheumatoid, and post traumatic arthritis of the hip with minimal acetabular involvement or distortion.
Other Considerations:
- Pathological considerations or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty.
All of the subject devices have been found substantially equivalent in previous premarket indications. The purpose of this submission is to modify the labeling to add MR compatibility to the labeling for the devices indicated above, and to remove a contraindication for selected devices. There have been no changes made to the devices requiring 510(k) clearance - only the labeling is being modified.
The provided text is a 510(k) Summary for a labeling update of Stryker Orthopaedics Total Hip Systems, not a study describing acceptance criteria and device performance. Therefore, most of the requested information regarding acceptance criteria, device performance, sample sizes, expert qualifications, adjudication, MRMC studies, standalone performance, and ground truth establishment is not available in this document.
The document primarily focuses on demonstrating substantial equivalence to predicate devices for a labeling update to include MR compatibility and remove a contraindication. It mentions non-clinical testing for MR compatibility but does not provide specific acceptance criteria or performance numbers in the requested format.
Here's a breakdown of what can be extracted and what is not available:
Acceptance Criteria and Study Supporting Device Performance
Not directly available. This document is a 510(k) summary for a labeling update, not an original submission demonstrating device performance against specific clinical or non-clinical acceptance criteria with quantitative results. It describes non-clinical testing performed to update MR compatibility labeling, but it does not present acceptance criteria in a table format with corresponding device performance metrics.
What is present regarding testing:
Non-clinical laboratory testing was performed to characterize the MR compatibility of the devices. The tests were conducted according to specific ASTM standards:
- ASTM F2052-06 and ASTM F2052-14 for Magnetically Induced Displacement Force.
- ASTM F2213-06 (Reapproved 2011) for Magnetically Induced Torque.
- ASTM F2119-07 (Reapproved 2013) for Image Artifact.
- ASTM F2182-11a for Heating by RF Fields.
The document states that the labeling was modified to include the MR conditional symbol and provide parameters for safe scanning, and for certain devices, a contraindication was removed. This implies that the devices met the requirements of these ASTM standards for MR compatibility, allowing for the labeling update. However, no specific numerical "acceptance criteria" or "reported device performance" are provided in a table.
Detailed Breakdown of Requested Information:
-
A table of acceptance criteria and the reported device performance
- Not available in this document. The document describes the type of non-clinical tests performed (e.g., Magnetically Induced Displacement Force per ASTM F2052-06) for MR compatibility, but it does not provide a table with specific numerical acceptance criteria (e.g., "Displacement less than X") and the corresponding measured performance (e.g., "Measured displacement Y"). It only states that the labeling was updated to reflect MR compatibility based on these tests.
-
Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not available. The document mentions non-clinical laboratory testing but does not specify sample sizes for these tests (e.g., how many hip components were tested for MR compatibility). It is a regulatory submission, not a research paper detailing experimental methodology. Data provenance for non-clinical lab tests is not typically documented in this way in a 510(k) summary.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable / Not available. The testing described is non-clinical "MR compatibility" testing performed in a laboratory setting, not clinical evaluation requiring expert interpretation or ground truth establishment in a diagnostic context.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable / Not available. As the testing is non-clinical (laboratory-based MR compatibility), there is no adjudication process involving human interpretation of results in the way it would be for a clinical diagnostic study.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This submission is for a labeling update for existing orthopedic implants and does not involve AI or any diagnostic imaging interpretation requiring an MRMC study.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This submission is for a labeling update for existing orthopedic implants and does not involve an algorithm or software device.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not applicable / Not available. For the non-clinical MR compatibility testing, the "ground truth" would be the physical properties and responses of the hip system components under MR conditions, as measured by the ASTM standards, not a clinical "ground truth" like pathology or expert consensus. These are engineering measurements.
-
The sample size for the training set
- Not applicable. This submission relates to physical orthopedic implants and MR compatibility testing, not a machine learning algorithm that requires a training set.
-
How the ground truth for the training set was established
- Not applicable. This submission relates to physical orthopedic implants and MR compatibility testing, not a machine learning algorithm that requires a training set.
In summary, the provided document is a regulatory submission focusing on demonstrating substantial equivalence for a labeling change for established orthopedic implants, particularly concerning MR compatibility. It outlines the types of non-clinical tests performed but does not present a detailed study report with specific acceptance criteria, performance metrics, or clinical study details as requested.
Ask a specific question about this device
(90 days)
HIP SYSTEMS
The hip system devices included in this submission are sterile, single-use devices intended for use in primary and revision total hip arthroplasty to alleviate pain and restore function.
The overall indications for use for the subject total and hemi hip replacement prostheses include:
- noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- rheumatoid arthritis;
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and,
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
The Rejuvenate Hip System is intended for cementless use only.
Additional indications specific to the Hipstar Stem:
The Hipstar Femoral Stem is intended for cementless use only.
When mated with a constrained acetabular liner the Hipstar Stem is indicated for use in primary and revision total hip arthroplasty for patients at high risk of hip dislocation due to a history of prior dislocations, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability.
Additional indications specific to the Restoration Modular Hip System:
The Restoration Modular Hip System is intended for primary and revision total hip arthroplasty as well as in the presence of severe proximal bone loss. These femoral stems are designed to be press fit into the proximal femur.
The devices included in this submission are femoral heads, acetabular shells, acetabular inserts, modular necks, modular stems and accessory components used in hip arthroplasty procedures. All devices have been previously deemed substantially equivalent in prior 510(k) submissions and are commercially available. The purpose of this submission is to modify the labeling of these devices to include safety information in the Instructions for Use (IFUs) regarding modular hip stem junctions. In addition to this specific update, general revisions have been made to the IFUs to harmonize the language between the Howmedica Osteonics' hip stem labeling
This K121308 submission is for a modification to the labeling of existing hip system devices, not for a new device requiring performance testing against specific acceptance criteria. The submission states: "The subject hip system devices have been deemed substantially equivalent to other commercially available hip arthroplasty systems based upon an evaluation of intended use, design, materials, and operational principles in prior 510(k) submissions. The purpose of this submission is solely to revise the labeling to add safety information regarding modular hip stem taper junctions."
Therefore, the requested information regarding acceptance criteria and a study proving the device meets them is not applicable to this specific 510(k) submission.
To address the prompt directly based on the provided text, the relevant section is "Substantial Equivalence," which implies that the performance of the device has already been established and accepted in prior submissions.
Here's a breakdown of why the requested information cannot be provided from the given document:
- 1. A table of acceptance criteria and the reported device performance: Not applicable. This submission is for a labeling change, not a new device requiring performance criteria to be met. The substantial equivalence is based on prior 510(k)s.
- 2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable. No performance study data is presented in this submission.
- 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. No performance study data where ground truth is established is presented.
- 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. No performance study requiring adjudication is presented.
- 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a hip implant, not an AI-assisted diagnostic tool.
- 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is a hip implant, not an algorithm.
- 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. No ground truth is established for this submission.
- 8. The sample size for the training set: Not applicable. No AI/machine learning training set is relevant to this device.
- 9. How the ground truth for the training set was established: Not applicable. No AI/machine learning training set is relevant to this device.
Conclusion based on the provided text:
The device's performance and acceptance criteria were established in previous 510(k) submissions, which deemed them "substantially equivalent to other commercially available hip arthroplasty systems based upon an evaluation of intended use, design, materials, and operational principles." This specific submission (K121308) does not provide details on the acceptance criteria or a study proving those criteria are met, as its sole purpose is to "revise the labeling to add safety information regarding modular hip stem taper junctions."
Ask a specific question about this device
(196 days)
METAL-ON-METAL HIP SYSTEMS-ADDITIONAL CONTRAINDICATIONS
Biomet Metal-On-Metal Total hip replacement components are intended for either cemented or uncemented use to replace the articulating portions of the hip during total hip arthroplasty.
Indications For Use:
- Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- Rheumatoid arthritis.
- Correction of functional deformity.
- (a) Revision procedures where other treatment or devices have failed (M2a-Taper™ and M2a-RinglocTM).
(b) Revision of previously failed total hip arthroplasty (M2a-Magnum™ and M2a-38™). - Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
The components of the M2a-Magnum™ system were also cleared for diastrophic variant, fracture of the pelvis, fused hip, Legg Perthes, slipped capital epiphysis, subcaptial fractures, and traumatic arthritis indications. The M2a-Taper™ Metal-on-Metal Hip Joint Replacement Prosthesis are intended for use in cemented and non-cemented primary and revision hip joint arthroplastic procedures.
Metal-On-Metal Hip prosthesis components provide for total hip replacement with a direct metal to metal articulation. They consist of a series of metal femoral head prostheses (either monolithic or modular) that articulate with highly congruent mating acetabular components without the need for a ceramic or polyethylene liner.
This submission is intended to notify FDA that Biomet has included additional Contraindications to the labeling for Metal-On-Metal total hip prosthesis components to assure the safe and effective use of Biomet Metal-On-Metal components for the appropriate patient populations. These changes are incorporated into one IFU that will be applicable for all Metal-On-Metal components currently cleared by FDA.
The provided text is a 510(k) summary for a medical device. It explicitly states that no non-clinical or clinical testing was performed or included to support substantial equivalence. The submission is solely for updating the labeling (specifically, adding contraindications) for previously cleared metal-on-metal hip prosthesis components.
Therefore, the requested information regarding acceptance criteria, study details, sample sizes, ground truth, and expert involvement cannot be extracted from this document, as no such studies were conducted for this particular 510(k) submission.
The document indicates:
- "Non-Clinical mechanical testing is not included to support a substantial equivalence determination since this submission is intended only to update labeling for previously cleared medical devices."
- "Clinical testing is not included to support a substantial equivalence determination since this submission is intended only to update labeling for previously cleared medical devices."
As a result, a table of acceptance criteria and reported device performance cannot be generated, nor can any details about studies, sample sizes, experts, or ground truth.
Ask a specific question about this device
(379 days)
SMITH & NEPHEW HIP SYSTEMS WITH HA
Total hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity.
Smith & Nephew Hip System components with HA are intended for single use only and are intended to be implanted without bone cement.
This 510/k) submission seeks FDA clearance to have the calcium phosphate coating (plasma sprayed from hydroxyapatite (HA) powder) on the subject hip components applied by an alternate coating supplier. The calcium phosphate coating is referred to as "HA coating." The hip system components included in this premarket notification are:
- . Porous Synergy Hip Stems
- . Non-Porous Synergy Stems
- Porous Anthology Hip Stems .
- . Porous Reflection Acetabular Shells
The Smith & Nephew Hip Stems with HA coating have been cleared for market by FDA via K052792, K002996 and K970337. The Reflection Acetabular cups with HA coating have been cleared for market by FDA via K990666 and K022556. The design, principle of operation, type of substrate materials, dimensional characteristics and indications for use will not change as a result of the implementation of an alternate HA coating supplier.
The provided document is a 510(k) summary for a medical device seeking FDA clearance. It focuses on demonstrating substantial equivalence to previously cleared devices, rather than establishing de novo safety and effectiveness through clinical trials with specific acceptance criteria and detailed performance metrics.
Therefore, many of the requested elements regarding acceptance criteria, study design, statistical details, and ground truth establishment are not applicable or available in this type of regulatory submission. The document primarily describes how the new HA coating supplier's product matches the characteristics and performance of the predicate device's coating.
Here's a breakdown based on the provided information:
Acceptance Criteria and Reported Device Performance
The concept of "acceptance criteria" in this context refers to the successful demonstration of substantial equivalence to predicate devices. This is achieved by showing that the new HA coating produced by the alternate supplier is chemically and physically comparable to the original, and that the HA-coated components maintain the equivalent performance characteristics.
Acceptance Criteria (Implied by Substantial Equivalence for Medical Devices) | Reported Device Performance |
---|---|
Material Characterization of HA Coating: | All characterization parameters (composition, crystallinity, etc.) on the starting HA powders and final HA coating were provided and reviewed. NIST SRM 2910 |
Porous Coating Characterization (Pre-HA): | The sintered, metallic beaded porous coatings (prior to HA application) were characterized per "Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement." The results were acceptable. |
Dual Coating Characterization (HA on Porous): | Additional characterization was performed to demonstrate that the definition of porosity (per 21 CFR 888.3358) was met for the dual coating of calcium phosphate on the metallic, sintered beaded coatings. The results were acceptable. |
Maintenance of Design, Material, and Indications: | The design, principle of operation, type of substrate materials, dimensional characteristics, and indications for use remain unchanged from the predicate devices. |
Overall Substantial Equivalence: | "All of the observed results indicate that the Smith & Nephew Hip Systems with HA are substantially equivalent to devices currently cleared for market." |
Study Details (Applicable to 510(k) Substantial Equivalence, Not Clinical Studies)
-
Sample size used for the test set and the data provenance:
- Test Set: Not a traditional "test set" in the sense of clinical samples. The "test set" here refers to the actual device components and HA coating samples produced by the alternate supplier, which were subjected to various material and coating characterization tests. The specific number of samples for each test (e.g., how many HA powder lots, how many coated implants) is not specified in this summary.
- Data Provenance: The tests were conducted on components manufactured by Smith & Nephew using the alternate HA coating supplier. The testing described appears to be laboratory-based (e.g., material characterization, physical property testing), not clinical data from patients. The document explicitly states: "Clinical data was not needed to support the safety and effectiveness of the subject HA coated devices."
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. This submission relies on objective material and physical property testing, not expert interpretation of results to establish "ground truth" in a diagnostic sense. The "ground truth" (or benchmark) for comparison was the characteristics of the predicate device's HA coating and accepted industry standards/guidance documents.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable. Adjudication methods are used for interpreting subjective data (e.g., image reads, clinical endpoints). The data here are quantitative material and coating characteristics, which are objectively measured and compared to predefined specifications or predicate device characteristics.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is a hardware/material modification submission, not an AI/software device. No human reader studies were conducted or are relevant to this type of device.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a hardware/material modification submission, not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The "ground truth" for demonstrating substantial equivalence was a combination of:
- Specifications and characteristics of the legally marketed predicate devices.
- Relevant guidance documents (e.g., "510(k) Information Needed for Hydroxyapatite Coated Orthopedic Implants," "Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement").
- Industry standards and established test methods (e.g., comparing to NIST SRM 2910|a) for certain parameters).
- Regulatory definitions (e.g., 21 CFR 888.3358 for porosity).
- The "ground truth" for demonstrating substantial equivalence was a combination of:
-
The sample size for the training set:
- Not applicable. This is not a machine learning or AI device that requires a training set. The "training" for the device would be the manufacturing process development and quality control of the alternate coating supplier.
-
How the ground truth for the training set was established:
- Not applicable. As above, no training set in the AI sense was used. The manufacturing process for the HA coating would have established its own internal quality control standards and specifications based on the desired material properties comparable to the predicate device.
Ask a specific question about this device
(51 days)
BIOMET METAL-ON-METAL HIP SYSTEMS- EXPANDED CONTRAINDICATIONS
K993438, K011110, K042037, K043537, K061423, K062995:
- Non-inflammatory degenerative joint disease including avascular necrosis, diastrophic variant, fracture of the pelvis, fused hip, Legg Perthes, osteoarthritis, slipped capital epiphysis, subcapital fractures, and traumatic arthritis.
- Rheumatoid arthritis.
- Correction of functional deformity.
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
- Revision of previously failed total hip arthroplasty.
- Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- Rheumatoid arthritis.
- Correction of functional deformity.
- Revision procedures where other treatment or devices have failed.
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
The systems consist of acetabular cups, femoral heads, and adapters and inserts. The Co-Cr-Mo head articulates with Co-Cr-Mo.
This 510(k) summary is for the Biomet® Metal-on-Metal Hip Systems-Expanded Contraindications. Based on the provided document, the device did not undergo clinical testing to prove its acceptance criteria because it was determined to be "substantially equivalent" to predicate devices already on the market.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Not applicable | Not applicable |
Explanation: The document explicitly states: "Non-Clinical Testing: Non-clinical laboratory testing is not provided as a basis for substantial equivalence. Clinical Testing: None provided as a basis for substantial equivalence." The basis for clearance was "substantial equivalence" to legally marketed predicate devices, meaning that the technological characteristics (materials, design, sizing, and indications) are considered identical. The only change highlighted is "Contraindications have been expanded to include patients with chronic renal failure and patients who are pregnant or may become pregnant." This indicates a change in labeling rather than a change requiring new performance data.
2. Sample size used for the test set and the data provenance
Not applicable, as no new clinical or non-clinical testing was performed as a basis for substantial equivalence. The clearance was based on demonstrating equivalence to existing predicate devices.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable, as no new clinical or non-clinical testing was performed.
4. Adjudication method for the test set
Not applicable, as no new clinical or non-clinical testing was performed.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a hip prosthesis, not an AI-assisted diagnostic tool or imaging system.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a hip prosthesis, not an AI algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Not applicable, as no new clinical or non-clinical testing was performed. The "ground truth" for the clearance was the existing performance and safety profile of the predicate devices.
8. The sample size for the training set
Not applicable, as no new development or testing using training sets (as in machine learning) was conducted. The device is a physical hip prosthesis.
9. How the ground truth for the training set was established
Not applicable, as no training set was used.
Ask a specific question about this device
Page 1 of 1