    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    HG PRO® Nitrile Powder Free Examination Gloves are substantially equivalent to the Class 1 patient examination gloves bearing the product code Nitrile - LZA (21CFR 880.6250). They meet all the current specifications listed under the ASTM Specification D 6319 -10, Standard Specification for Nitrile Examination Gloves for Medical Application. They are made from acrylonitrile-butadiene copolymer dispersion. These gloves are blue in color and are powder free.

    AI/ML Overview

    The provided document describes the acceptance criteria and performance of the HG PRO® Nitrile Powder Free Examination Gloves, primarily comparing them to a predicate device (K022548) and relevant ASTM standards.

    Here's a breakdown of the requested information:

    1. Table of acceptance criteria and the reported device performance

    CharacteristicAcceptance Criteria / StandardsReported Device Performance (Subject Device: HG PRO® Nitrile Powder Free Examination Gloves)
    MaterialASTM D6319-10Nitrile Synthetic Rubber
    Color-Blue
    Texture-Textured Fingers
    SizeMedical Glove Guidance Manual - Labeling - Issued on January 22, 2008Extra Small, Small, Medium, Large, Extra Large
    Single UseMedical Gloves Guidance Manual - Issued on January 22, 2008Single use
    DimensionASTM D6319-10Meets ASTM D6319-10
    Length:
    Extra Small = 220 mm minimum
    Small = 220 mm minimum
    Medium = 230 mm minimum
    Large = 230 mm minimum
    Extra Large = 230 mm minimum
    Palm Width:
    Extra Small = 60-80 mm
    Small = 70-90 mm
    Medium = 85-105 mm
    Large = 100-120 mm
    Extra Large = 110-130 mm
    ThicknessASTM D6319-10Meets ASTM D6319-10
    Finger: 0.05 mm min
    Palm: 0.05 mm min
    Physical PropertiesASTM D6319-10Meet ASTM D6319-10
    Tensile Strength:
    14 MPa min (before aging)
    14 MPa min (after aging)
    Ultimate Elongation:
    500% min (before aging)
    400% min (after aging)
    Watertight test (1000 ml)ASTM D5151-06Pass AQL 1.5
    Residual PowderASTM D6124-06Meet ≤ 2.0 mg/glove
    BiocompatibilityPrimary Skin Irritation - ISO 10993-10: 2010
    Dermal Sensitization - ISO 10993-10: 2010Pass
    Not a primary skin irritant under the conditions of the study
    Not a contact sensitizer under the conditions of the study

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes used for each specific test (e.g., watertight test, physical properties). It refers to compliance with ASTM standards, which would dictate the sample sizes for those tests. The data provenance is not specified, but the manufacturer is based in Thailand (Healthy Glove Co., Ltd.). The studies appear to be prospective, laboratory-based tests conducted to meet established standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable to the type of device and testing presented. The "ground truth" for examination gloves is established by meeting predefined physical, chemical, and biological standards (ASTM, ISO), not by expert consensus on clinical findings.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This device is an examination glove, and its performance is evaluated against objective, measurable standards, not through adjudication by experts on subjective clinical interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-powered diagnostic device, but a physical medical device (examination glove). Therefore, MRMC studies and AI assistance are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is an examination glove, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device is established by objective, measurable physical, chemical, and biological performance standards as defined by organizations like ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization). For example:

    • Dimensions: Measured against specified ranges (e.g., length, palm width).
    • Thickness: Measured against minimum requirements.
    • Physical Properties: Tensile strength and ultimate elongation measured against minimum requirements (before and after aging).
    • Watertightness: Tested against leakage rates (AQL 1.5).
    • Residual Powder: Measured against maximum limits.
    • Biocompatibility: Tested for primary skin irritation and dermal sensitization using established biological evaluation standards (ISO 10993-10).

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As above, this is not an AI/ML device.