Search Results
Found 2 results
510(k) Data Aggregation
(189 days)
Gem FlowCoupler System
The VESSEL EVERTER System is indicated for use with the Microvascular Anastomotic COUPLER and FLOW COUPLER Device in the anastomosis of only arteries normally encountered in microsurgical procedures only in the peripheral vascular system. The VESSEL EVERTER System is indicated for use with COUPLER and FLOW COUPLER System sizes from 2.0 to 4.0 mm.
Two accessory devices comprise the Vessel Everter System: Vessel Everter and Sizing Guide. The Vessel Everter system is an accessory device to the GEM FLOWCOUPLER System, a pair of implantable rings that are used for end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures.
The Vessel Everter System are two non-implantable, sterile, single use hand-held devices that facilitate sizing and use of the GEM FLOWCOUPLER System. The Vessel Everter is comprised of a handle with two soft silicone end effectors (tips), one on each end. The instrument handle features two separate silicone end effectors, each of a different size, intended to be used at the surgeon's discretion. The Vessel Everter end effector is designed with a narrowing tip to allow for engagement with multiple sizes of vessels, corresponding with a coupler size of 2.0-4.0 mm rings. The Vessel Everter is used to press vessel tissue onto the GEM FLOWCOUPLER rings. To accomplish this, the surgeon presses the Vessel Everter into the tissue, flaring the tissue onto and over the Coupler locking pins.
The Sizing Guide is a hard, plastic handle featuring holes on either end through which the vessel's outer diameter may be measured. The holes' diameters are intended to match the inner diameters of the various GEM FLOWCOUPLER rings. The holes correlate to vessel diameters of 2.0 mm to 4.0 mm in 0.5 mm increments. The surgeon may use the sizing guide to guide the selection of which coupler they will use for each particular anastomosis. The Sizing Guide is used by placing the vessel on the sizing guide and comparing the vessel diameter to the holes in the Sizing Guide.
The Vessel Everter and Sizing Guide are sterilized by ethylene oxide and are provided sterile for-single-patient use. The Vessel Everter and Sizing Guide are not implantable and are disposed of after single use.
The parent device, GEM FLOWCOUPLER System, remains unchanged and is not addressed in this submission.
The provided text describes a 510(k) premarket notification for the GEM FlowCoupler System accessories (Vessel Everter and Sizing Guide). It does not contain information about a study proving the device meets acceptance criteria in the context of an AI/ML device. Instead, it details performance testing for a medical device's physical and biological properties.
Therefore, many of the requested fields cannot be filled as they pertain to AI/ML device validation studies. However, I can extract the available information regarding the device's performance testing for the physical accessories.
Here's a breakdown based on the provided text, acknowledging the limitations for AI/ML specific information:
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria (Test) | Reported Device Performance (Result) |
---|---|
Silicone tip retention - tensile force | Subject devices met acceptance criteria |
Retention Barb - Cantilever Loading | Subject devices met acceptance criteria |
Coupler Engagement and Disengagement Force | Subject devices met acceptance criteria |
Intimal Integrity (porcine vessels) | Subject devices met acceptance criteria |
Biocompatibility: Cytotoxicity (L929 MEM Elution) | Not cytotoxic per ISO10993-5 (PASS) |
Biocompatibility: Sensitization (Kligman Maximization test) | Non-sensitizer per ISO10993-10 (PASS) |
Biocompatibility: Irritation (intracutaneous injection in rabbit) | Negative per ISO10993-10 (PASS) |
Biocompatibility: Toxicity (systemic injection in mouse) | Negative for toxicity per ISO10993-11 (PASS) |
Biocompatibility: Rabbit blood hemolysis test | Non-hemolytic per ASTM F756 (PASS) |
2. Sample size used for the test set and the data provenance
The text mentions "porcine vessels" for the Intimal Integrity Verification Testing but does not specify the sample size. It does not mention any human data or specific provenance for these tests.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. The tests performed are physical and biological evaluations of the device's materials and mechanical properties, not an AI/ML diagnosis where expert ground truth is established.
4. Adjudication method for the test set
Not applicable. This concept is relevant for AI/ML performance evaluation against a human-adjudicated ground truth.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI/ML device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is not an AI/ML device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the performance tests outlined typically refers to established standards and measurement techniques for mechanical properties (e.g., force measurements), material characteristics (e.g., cytotoxicity assays), and biological responses (e.g., hemolysis tests). For the intimal integrity test, the "ground truth" would be the observed integrity of the porcine vessel intima after device use.
8. The sample size for the training set
Not applicable. This is not an AI/ML device.
9. How the ground truth for the training set was established
Not applicable. This is not an AI/ML device.
Ask a specific question about this device
(61 days)
Gem FlowCoupler System
The FLOW COUPLER Device is a single use, implantable device that is intended to be used in the end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures. The FLOW COUPLER Device includes a pair of permanently implanted rings which secure the anastomosis and a removable Doppler probe that is press-fit onto one of the rings. When the FLOW COUPLER Device is used in conjunction with the FLOW COUPLER Monitor, the FLOW COUPLER System is intended to detect blood flow and confirm vessel patency intra-operatively and post-operatively at the anastomotic site. Post-operatively, blood flow can be detected on an as needed basis for up to 7 days. The FLOW COUPLER Doppler probe is not intended to be a permanent implant and should be removed 3 to 14 days post- operatively.
The FLOW COUPLER Device and System consists of a FLOW COUPLER Device and a FLOW COUPLER Monitor. The FLOW COUPLER Monitor is a pulsed Doppler ultrasound system designed for the detection of blood flow in vessels. The FLOW COUPLER Device includes a 20MHz ultrasonic Doppler transducer (probe) attached to one of the FLOW COUPLER rings, and an external lead. The probe via the external lead connects to the monitor and emits a pulsed ultrasonic signal. A varying audible signal is produced when the probe detects flow.
The provided text describes the GEM™FLOW COUPLER™ Monitor (GEM2010M-2), a cardiovascular blood flowmeter, and its substantial equivalence to a predicate device. However, the document does not detail specific acceptance criteria, a study that quantifies device performance against those criteria, or the methodology for ground truth establishment, sample sizes, or multi-reader studies in the manner typically required for AI/algorithm-driven device evaluations. The information provided focuses on the regulatory submission process (510(k)) and comparisons to a predicate device, rather than a detailed performance study with quantifiable metrics.
Based on the provided text, here's what can be extracted and what information is not available:
1. A table of acceptance criteria and the reported device performance
The document states that "The testing was completed with passing results per the pass/fail criteria defined for each test case." However, the exact acceptance criteria (e.g., specific accuracy, sensitivity, specificity thresholds) and the reported device performance for each are not explicitly listed. Instead, it broadly states that the modified monitor "meets the same performance specifications as the predicate device" and that the "performance data described in this submission confirm that the modified monitor meets the same performance specifications." The performance data categories include:
Test Category | Reported Device Performance (as stated in the document) |
---|---|
EMC Testing | Passing results per the pass/fail criteria defined for each test case. |
Electrical Safety Testing | Passing results per the pass/fail criteria defined for each test case. |
Acoustic Output Testing | Passing results per the pass/fail criteria defined for each test case. |
AC Power Continuous Use | Passing results per the pass/fail criteria defined for each test case. |
Battery Life Testing | Passing results per the pass/fail criteria defined for each test case. |
User Interface Functional Testing | Passing results per the pass/fail criteria defined for each test case. |
Flow Velocity Testing | Passing results per the pass/fail criteria defined for each test case. |
Functional Performance Testing | Passing results per the pass/fail criteria defined for each test case. |
Packaging Validation Testing | Passing results per the pass/fail criteria defined for each test case. |
Software Verification | Passing results per the pass/fail criteria defined for each test case. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided in the document. The document refers to "bench testing" and "performance data described in this submission" but does not detail the size of any test sets or the origin/nature of the data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided in the document. The device is a "pulsed Doppler ultrasound system designed for the detection of blood flow in vessels" that produces an "audible signal" when flow is detected, and the modified monitor "adds a qualitative visual display as a secondary indicator of blood flow." Therefore, the ground truth would likely relate to the presence/absence or characteristics of blood flow, which would typically involve highly specialized experts or direct measurement. However, details of any human expert involvement are absent.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not provided in the document.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
A multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned in the document. The device is a monitor for detecting blood flow, not an AI-driven image interpretation system.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The document describes the "FLOW COUPLER Monitor" as a "pulsed Doppler ultrasound system designed for the detection of blood flow in vessels." While it includes "software for digital filtering of the returned audio signal for noise reduction" and a "qualitative visual display," it is presented as a medical device providing information to a human user (e.g., producing a "varying audible signal" and a "qualitative visual display"). The question of "standalone" algorithm performance in the context of typical AI algorithms is not directly applicable or addressed here, as the device is a measurement tool, not an autonomous diagnostic algorithm. Its function is to detect blood flow, which is then interpreted by a human user.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The type of ground truth used to validate the device's ability to "detect blood flow and confirm vessel patency" is not explicitly stated. Given the nature of a Doppler ultrasound, the ground truth would typically be established through direct physiological measurements or other highly reliable methods for confirming blood flow and vessel patency, rather than expert consensus on images. However, the document does not specify.
8. The sample size for the training set
The document does not describe any training sets for the device. The device is a monitor, not a machine learning model that undergoes training in the typical sense.
9. How the ground truth for the training set was established
Since there is no mention of a training set, this information is not provided.
Ask a specific question about this device
Page 1 of 1