Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K142654
    Manufacturer
    Date Cleared
    2014-10-28

    (40 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Fortrex 0.035 OTW PTA Balloon Catheter

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fortrex 0.035" OTW PTA Balloon Catheter is intended to dilate stenoses in the iliac, femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.

    Device Description

    The Fortrex™ 0.035" PTA Balloon Catheter (Fortrex Catheter) is an over the wire (OTW) 0.035"dual lumen catheter with a distally mounted semi-compliant inflatable balloon and an atraumatic tapered tip. The manifold includes a lumen marked "THRU". This is the central lumen of the catheter, which terminates at the distal tip. This lumen is used to pass the catheter over a guidewire with a maximum diameter of 0.035". The lumen marked "BALLOON" is the balloon inflation lumen, which is used to inflate and deflate the balloon with a mixture of contrast medium and saline solution. The balloon has two radiopague markers for positioning the balloon relative to the stenosis. The radiopaque marker bands indicate the dilating or working section of the balloon.

    The Fortrex Catheter is available with balloon diameters 4.0 mm to 12.00 mm, balloon lengths 20 mm, 40 mm, 80 mm, and 100 mm and with shaft lengths of 40 cm, 80 cm, and 135 cm.

    AI/ML Overview

    This document is a 510(k) summary for the Fortrex™ 0.035" OTW PTA Balloon Catheter, which is a medical device intended to dilate stenoses in various arteries and for stent post-dilatation in the peripheral vasculature. The summary outlines the device's characteristics, its similarities to predicate devices, and the performance testing conducted to demonstrate substantial equivalence.

    Here's an analysis of the provided text in response to your questions:

    1. A table of acceptance criteria and the reported device performance

    The document states: "The Fortrex catheter met all acceptance criteria for the performance testing. The results from these tests demonstrate that the technological characteristics and performance criteria of the Fortrex Catheter are comparable to the predicate EverCross Catheter and that the Fortrex Catheter performs in a manner equivalent to the predicate EverCross device currently on the market."

    However, specific quantitative acceptance criteria values are not provided, nor are the numerical results for each test. The table below lists the performance tests conducted, but without the specific criteria or reported performance data.

    Performance TestAcceptance Criteria (Not specified)Reported Device Performance (Not specified quantitatively)
    Balloon Dimensional VerificationMet acceptance criteriaMet all acceptance criteria
    Balloon Rated Burst PressureMet acceptance criteriaMet all acceptance criteria
    Minimum Balloon Burst StrengthMet acceptance criteriaMet all acceptance criteria
    Balloon FatigueMet acceptance criteriaMet all acceptance criteria
    Balloon ComplianceMet acceptance criteriaMet all acceptance criteria
    Catheter Dimensional VerificationMet acceptance criteriaMet all acceptance criteria
    Inflation / Deflation TimeMet acceptance criteriaMet all acceptance criteria
    Wire MovementMet acceptance criteriaMet all acceptance criteria
    RadiopacityMet acceptance criteriaMet all acceptance criteria
    Device TrackingMet acceptance criteriaMet all acceptance criteria
    Insertion ForceMet acceptance criteriaMet all acceptance criteria
    Balloon Pull-back ForceMet acceptance criteriaMet all acceptance criteria
    Reinsertion ForceMet acceptance criteriaMet all acceptance criteria
    Torque StrengthMet acceptance criteriaMet all acceptance criteria
    Tensile StrengthMet acceptance criteriaMet all acceptance criteria
    KinkMet acceptance criteriaMet all acceptance criteria
    Marker Band to Balloon BodyMet acceptance criteriaMet all acceptance criteria
    Biocompatibility (ISO 10993-1)Met acceptance criteriaMet all acceptance criteria
    - CytotoxicityMet acceptance criteriaMet all acceptance criteria
    - SensitizationMet acceptance criteriaMet all acceptance criteria
    - Intracutaneous reactivityMet acceptance criteriaMet all acceptance criteria
    - Acute systemic toxicityMet acceptance criteriaMet all acceptance criteria
    - HemolysisMet acceptance criteriaMet all acceptance criteria
    - PyrogenMet acceptance criteriaMet all acceptance criteria
    - Complement activationMet acceptance criteriaMet all acceptance criteria
    - ThromboresistanceMet acceptance criteriaMet all acceptance criteria
    - Partial thromboplastin timeMet acceptance criteriaMet all acceptance criteria
    - Platelet/leukocyte countsMet acceptance criteriaMet all acceptance criteria
    - MutagenicityMet acceptance criteriaMet all acceptance criteria
    - LymphomaMet acceptance criteriaMet all acceptance criteria
    - Micronucleus assayMet acceptance criteriaMet all acceptance criteria

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document lists performance tests but does not specify the sample sizes used for each test. It also does not provide information on the data provenance in terms of country of origin or whether the study was retrospective or prospective. The study appears to be an internal bench testing and biocompatibility testing conducted by the manufacturer, Covidien llc.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This document describes performance testing of a medical device (PTA balloon catheter), not a diagnostic algorithm that requires expert-established ground truth from medical images or clinical data. Therefore, this question is not applicable to the type of study presented here. The "ground truth" for these tests would be established by validated test methods and reference standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as this is bench and biocompatibility testing of a physical device, not an assessment requiring human adjudication of results.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document pertains to the 510(k) clearance of a physical medical device (balloon catheter), not an AI/software device that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This document is about a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the performance tests:

    • Bench Testing: The "ground truth" is established by the specifications and validated test methods for a medical device (e.g., burst pressure measured by a pressure gauge, dimensions measured by calibrated instruments).
    • Biocompatibility Testing: The "ground truth" is established by the accepted standards and protocols defined in ISO 10993-1, using biological assays and analytical methods.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning study, so there is no training set.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this type of device study.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1