Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K994032
    Date Cleared
    2000-07-07

    (224 days)

    Product Code
    Regulation Number
    870.3470
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fluoro-Tex™ Cardiovascular Patch is intended to be used for the repair of the cardiovascular system.

    Device Description

    Fluoro-Tex Cardiovascular Patch consists of a sheet of porous expanded polytetrafluoroethylene (ePTFE) internally reinforced with fluorinated ethylene propylene (FEP).

    AI/ML Overview

    Here's an analysis of the provided text regarding the Fluoro-Tex™ Cardiovascular Patch, structured according to your request:

    In this case, the device is a cardiovascular patch, and the testing focuses on mechanical and biocompatibility properties rather than algorithm performance. Therefore, questions related to AI/algorithm performance (such as sample size for test/training sets, experts for ground truth, adjudication methods, MRMC studies, standalone performance, and ground truth for training data) are not applicable.


    Acceptance Criteria and Device Performance for Fluoro-Tex™ Cardiovascular Patch

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" but rather presents a comparison of the Fluoro-Tex™ Cardiovascular Patch to predicate devices. The implicit acceptance criterion is that the Fluoro-Tex™ patch performs comparably or better than the predicate devices across the measured mechanical properties.

    Test / Performance MetricPredicate Device 1 (Gore-Tex® Cardiovascular Patch)Predicate Device 2 (IMPRA® Cardiovascular Patch)Fluoro-Tex™ Cardiovascular Patch (Reported Performance)Implicit Acceptance Criteria (Comparative)
    Tensile Strength (kg/cm²)26.7 (n=4)20.5 (n=4)20.7 (n=6)Comparable to or exceeding predicate devices
    Suture Strength (kg/pin)2.1 (n=4)1.8 (n=4)2.0 (n=6)Comparable to or exceeding predicate devices
    Burst Strength (psi)Na (Not applicable/not provided)294 (n=5)284 (n=5)Comparable to or exceeding the IMPRA® patch (as Gore-Tex® data is not available)

    2. Sample size used for the test set and the data provenance:

    • Tensile Strength:
      • Gore-Tex®: n=4
      • IMPRA®: n=4
      • Fluoro-Tex™: n=6
    • Suture Strength:
      • Gore-Tex®: n=4
      • IMPRA®: n=4
      • Fluoro-Tex™: n=6
    • Burst Strength:
      • IMPRA®: n=5
      • Fluoro-Tex™: n=5
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). This appears to be laboratory-based mechanical testing data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable, as this is laboratory mechanical testing data, not expert-derived ground truth.

    4. Adjudication method for the test set:

    Not applicable, as this is laboratory mechanical testing data.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable, as this device is a physical medical patch, not an AI software.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable, as this device is a physical medical patch, not an algorithm.

    7. The type of ground truth used:

    • Mechanical Tests: The "ground truth" implicitly refers to the measured physical properties of the materials and devices under controlled laboratory conditions, using established engineering test methods.
    • Biocompatibility: Demonstrated through cytotox, FTIR, and in-house cleanliness testing, relying on the known safety profile of PTFE, the material used in the device and its predicates.

    8. The sample size for the training set:

    Not applicable, as this is a physical medical device, not an AI/machine learning model.

    9. How the ground truth for the training set was established:

    Not applicable, as this is a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1