Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K201049
    Device Name
    FINDER G6PD
    Manufacturer
    Date Cleared
    2022-09-14

    (876 days)

    Product Code
    Regulation Number
    864.7360
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    FINDER G6PD

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FINDER G6PD test is intended for semi-quantitative measurement of glucose-6-phosphate dehydrogenase in venous whole blood specimens collected in lithium heparin tubes, for the identification of G6PD deficient samples. The FINDER G6PD test is intended to be used with the FINDER Instrument in point of care or clinical laboratory settings.

    Device Description

    The FINDER G6PD test system measures G6PD quantitatively from a 50uL venous whole blood specimen. The blood specimen should be collected in lithium heparin anticoagulant. The G6PD test system is suitable for use in both a point-of-care setting and a clinical laboratory. The time to result is around 16 minutes from sample introduction. The FINDER G6PD test system consists of the FINDER G6PD Test Cartridge and the FINDER Instrument. The cartridge uses electrowetting-based digital microfluidics to integrate and automate all the sample and reagent handling steps. The instrument contains all the hardware and software required to operate the cartridge, providing electrowetting control, thermal control and detection capability, a touch-screen user interface and software necessary to perform the test and report results.

    AI/ML Overview

    FINDER G6PD Test: Acceptance Criteria and Performance Study

    The FINDER G6PD Test is intended for semi-quantitative measurement of glucose-6-phosphate dehydrogenase in venous whole blood to identify G6PD deficient samples. The device is used with the FINDER Instrument in point-of-care or clinical laboratory settings.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" as a separate section with specific numerical targets. However, the performance data presented from precision, linearity, interference, sensitivity, and method comparison studies collectively demonstrate the device's acceptable performance in comparison to the predicate device and established guidelines. For the purpose of this analysis, key performance metrics from the non-clinical and clinical studies will be presented as the "reported device performance."

    CategorySpecific MetricReported Device Performance
    PrecisionSingle Site Precision (CV%)
    (Hemolysate Controls)Low (1.4 U/gHb)Repeatability: 4.1%, Reproducibility: 6.5%
    Intermediate (7.0 U/gHb)Repeatability: 3.6%, Reproducibility: 5.3%
    High (17.4 U/gHb)Repeatability: 2.4%, Reproducibility: 5.2%
    Reproducibility Precision (CV%)
    (Fresh Whole Blood)Low (1.1 U/gHb)Repeatability: 5.9%, Reproducibility: 6.0%
    Intermediate (3.5 U/gHb)Repeatability: 3.4%, Reproducibility: 4.4%
    High (11.2 U/gHb)Repeatability: 6.6%, Reproducibility: 7.0%
    LinearityLinear Range0.8 to 19.7 U/g Hb
    InterferenceEndogenous SubstancesNon-interfering at maximum tested concentrations (e.g., Bilirubin 50 mg/dL, Hemoglobin 5 g/L)
    Common DrugsNon-interfering at maximum tested concentrations (e.g., Ampicillin 0.16 mM, Ibuprofen 2.5 mM)
    HematocritBias of -15.3% at 29% Hct (normal sample), -11.1% at 30% Hct (MDL sample) compared to 50% Hct.
    SensitivityLimit of Blank (LoB)0.2 U/g Hb
    Limit of Detection (LoD)0.4 U/g Hb
    Limit of Quantitation (LoQ)1.1 U/g Hb
    Method ComparisonSlope (95% CI)0.92 (0.89, 0.95)
    Intercept (95% CI)0.28 (0.12, 0.44)
    Correlation Coefficient0.9
    Bias at 3.1 U/gHb (30% AMM)0.0 U/gHb (-0.1, 0.2) or 1.4% (-2.5%, 5.4%)
    Bias at 8.4 U/gHb (80% AMM)-0.4 U/gHb (-0.5, 0.2) or -4.3% (-6.5%, -2.0%)
    **Categorical Agreement (
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1