Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K152526
    Manufacturer
    Date Cleared
    2015-10-02

    (29 days)

    Product Code
    Regulation Number
    862.1150
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Elecsys Progesterone III CalSet

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Progesterone III CalSet is used for calibrating the quantitative Elecsys Progesterone III assay on the Elecsys and cobas e immunoassay analyzers.

    Device Description

    The Elecsys Progesterone III CalSet is a lyophilized product consisting of plant-derived progesterone in a human serum matrix. It is traceable via ID-GC/MS (isotope dilution gas chromatography/mass spectrometry) to highly purified progesterone by weight analogous to BCR-348R and ERM-DA347.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Elecsys Progesterone III CalSet, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes three stability studies with a consistent acceptance criterion.

    Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
    CalSet Value AssignmentPassed pre-defined acceptance criteria (specific values not detailed but stated as met). PreciControl Universal (Level 1 & 2) acceptance criteria must be met.Passed. Assigned values are the median value over at least six runs on at least three analyzers. PreciControl Universal acceptance criteria met.
    Stability (Reconstituted)95 to 105% signal recovery of reference valueAll tested conditions (2-8°C for 25 hours, -20°C for 32 days, 20-25°C open vial for 3 hours, 2 freeze/thaw cycles) met the acceptance criterion.
    Accelerated Stability (Lyophilized)95 to 105% recovery of reference valueOne lot evaluated in duplicate met the acceptance criterion.
    Real-Time Stability (Lyophilized)95-105% recovery of the reference valueOngoing study, but the testing will continue until data supports an 18-month claim. Implied to be meeting acceptance criteria during the evaluation period to support the claim.

    2. Sample Size Used for the Test Set and Data Provenance

    • CalSet Value Assignment: Tested on "at least three (3) cobas e 411 analyzers and at least three (3) cobas e 601/MODULAR ANALYTICS E170 analyzers with all Progesterone III reagent lots available." The assigned value is the median of at least six runs of the respective calibrator. The testing also involved PreciControl Universal levels 1 & 2.
    • Stability Studies (Reconstitution, Accelerated, Real-Time): Specific sample sizes for each test condition (e.g., number of vials, number of runs per vial) are not explicitly stated as a single number. However, for recombinant stability and accelerated stability, it mentions "on-test materials" and "one Progesterone III CalSet lot was evaluated in duplicate" for accelerated stability. The real-time study is ongoing with "Progesterone III CalSet test material" tested in duplicate.
    • Data Provenance: The studies were conducted at Roche Diagnostics GmbH in Mannheim, Germany, and Penzberg, Germany, and Roche Diagnostics United States. The data appears to be prospective as it's part of the premarket notification for a new device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This device is a calibrator for an immunoassay. The concept of "experts establishing ground truth" in the diagnostic imaging/CAD sense does not directly apply here. Instead, the ground truth for the analyte concentration in the calibrator is established through traceability to highly purified progesterone by weight analogous to BCR-348R and ERM-DA347 using ID-GC/MS (isotope dilution gas chromatography/mass spectrometry). This is a metrological approach to establishing a reference value, rather than expert interpretation of a clinical finding.

    4. Adjudication Method for the Test Set

    Not applicable in the context of a calibrator's analytical performance.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not performed. This type of study is designed for diagnostic tools that assist human readers (e.g., CAD systems for mammography), whereas the Elecsys Progesterone III CalSet is a calibrator for an automated immunoassay system.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the performance studies described (value assignment and stability) are standalone evaluations of the calibrator's analytical characteristics. The calibrator itself is a component used by an automated immunoassay system (Elecsys and cobas e immunoassay analyzers) and does not involve human interpretation in its direct function.

    7. The Type of Ground Truth Used

    The ground truth for the Elecsys Progesterone III CalSet is established through traceability via ID-GC/MS (isotope dilution gas chromatography/mass spectrometry) to highly purified progesterone by weight analogous to certified reference materials (BCR-348R and ERM-DA347). This provides a highly accurate, metrologically sound reference for the concentration of progesterone in the calibrator.

    8. The Sample Size for the Training Set

    The concept of a "training set" is not directly applicable in the documentation for this calibrator. Calibrators are manufactured to specific specifications and then undergo verification and validation testing, not a machine learning training process. The value assignment process involves running calibrators on multiple instruments and reagent lots to determine assigned values, which is more akin to a robust validation process than machine learning training.

    9. How the Ground Truth for the Training Set was Established

    As noted above, a "training set" in the machine learning sense is not used. The reference values (ground truth) for the calibrator were established through the metrological traceability mentioned in point 7.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1