Search Results
Found 1 results
510(k) Data Aggregation
(197 days)
EVITA 4 (8411900)
Evita 4 is a time cycled, constant volume, long term, intensive care ventilator for adults and children with a body weight of at least 3 kg.
Intended Environment for the device:
- hospital use
- inner clinical transport
The Evita 4 with the Software version 3.0 is a time cycled, constant volume ventilator The Evita + with the Oct.ware is microprocessor controlled and provides the following ventilation modes:
- CMV/assist
- SIMV
- MMV
- SB
- CPAP
- ASB/PSV
- PCV+
- APRV
- Apnea Ventilation
- ILV
This is a regulatory submission for a medical device (ventilator) and does not contain the kind of detailed study information (acceptance criteria, sample sizes, ground truth establishment, etc.) that would typically be found in a clinical study report or a 510(k) submission for a diagnostic AI/ML device.
The provided document, a 510(k) Summary, primarily focuses on demonstrating substantial equivalence to predicate devices, not on proving device performance against specific clinical acceptance criteria through a dedicated study with experts establishing ground truth.
Therefore, many of the requested fields cannot be extracted directly from this document. I will fill in what can be inferred or stated as "Not applicable/Provided" based on the content.
Here's an analysis based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state acceptance criteria in a quantitative manner as one would expect for a diagnostic device. Instead, it relies on demonstrating "substantial equivalence" to predicate devices. The "performance" is implicitly deemed acceptable if it is equivalent to the legally marketed predicate devices.
Acceptance Criteria (Inferred from Substantial Equivalence Claim) | Reported Device Performance (Summary of Equivalence) |
---|---|
Classification: Intensive Care Ventilator | Yes |
Operating Principle: Time cycled, volume constant, pressure controlled | Yes |
Software controlled | Yes |
Gas delivery system: Fast "High Pressure Servo Valve", microprocessor controlled | Yes |
Intended Use: For adults and children with body weight >= 3 kg | Identical to predicate Evita 4 (K 961687) |
Modes of Ventilation | Identical to predicate Evita 4 (K 961687) |
Ventilator Settings | Identical to predicate Evita 4 (K 961687) |
Flow Trigger Sensitivity | Identical to predicate Evita 4 (K 961687) |
Parameter setting | Identical to predicate Evita 4 (K 961687) |
Monitors/Indicators | Identical to predicate Evita 4 (K 961687) |
Pressure Limitation | Identical to predicate Evita 4 (K 961687) |
Alarm Functions | Identical to predicate Evita 4 (K 961687) |
Displayed Values | Identical to predicate Evita 4 (K 961687) |
Displayed Curves | Identical to predicate Evita 4 (K 961687) |
Intermittent PEEP | Identical with Evita 2 dura (K 970165) |
Inspiration Hold | Identical with Evita 2 dura (K 970165) |
Apnea Ventilation | Identical with Evita 2 dura (K 970165) |
Nebulization | Identical with Evita 2 dura (K 970165) |
Nebulization in pediatric mode | Identical with Evita 2 dura (K 970165) |
Suction Help Mode | Identical with Evita 2 dura (K 970165) |
Switch over function | Identical with Evita 2 dura (K 970165) |
Leak-Compensation | Identical with Evita 2 dura (K 970165) |
Minute Volume-Leak | Identical with Evita 4 (K 961687) |
Overall Safety and Efficiency | Substantially equivalent to predicate devices |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: Not applicable/Provided. This document does not describe a clinical study with a "test set" in the context of diagnostic performance evaluation. The substantial equivalence claim is based on comparing technical specifications and intended use.
- Data Provenance: Not applicable/Provided. No specific "data" in terms of patient results is discussed. The device is manufactured by Dräger Medizintechnik GmbH, Lübeck, Germany, and sold in the United States.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Number of experts: Not applicable/Provided. This document does not pertain to an AI/ML diagnostic device requiring expert-established ground truth.
- Qualifications of experts: Not applicable/Provided.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Adjudication method: Not applicable/Provided.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC study: No. This is a ventilator, not an AI-assisted diagnostic device for human readers.
- Effect size: Not applicable.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone performance: Not applicable in the context of an AI/ML algorithm. This device is a ventilator, designed to be used by trained medical professionals. Its "performance" refers to its mechanical and software functions meeting specifications, not an algorithm's diagnostic output.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Type of ground truth: Not applicable in the traditional sense of diagnostic device evaluation. The "ground truth" for a ventilator's performance would be its adherence to established electrical, mechanical, and software specifications, and its ability to safely and effectively deliver ventilation as intended, as demonstrated through engineering tests and comparison to legally marketed predicate devices.
8. The sample size for the training set
- Sample size for training set: Not applicable/Provided. This document does not describe an AI/ML device that undergoes training on a dataset.
9. How the ground truth for the training set was established
- Ground truth for training set: Not applicable/Provided.
Ask a specific question about this device
Page 1 of 1