510(k) Data Aggregation

    K Number
    K971738
    Date Cleared
    1997-08-07

    (87 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    ENDOPATH RESPOSABLE TROCAR SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ENDOPATH® Resposable Trocar System has application in abdominal, thoracic, and gynecologic minimally invasive surgical procedures to establish a path of entry for endoscopic instruments and endoscopes.

    Device Description

    The ENDOPATH® Resposable Trocar System consists of three main components: a Single Use obturator, a Single Use trocar sleeve housing and a reusable trocar sleeve. The obturator consists of a sharp flat blade tip or pyramidal tip and a spring loaded shield. The shield is designed to cover the flat blade or pyramidal tip to reduce the likelihood of injury to internal structures from puncture or laceration once the abdominal or thoracic cavity has been entered. The trocar sleeve housing has an inner and outer seal to maintain pneumoperitoneum when instrumentation is inserted and withdrawn through the cannula during a surgical procedure. Some trocar sleeve housings are provided with a stopcock for insufflating the operative space. Integral threads along the outside diameter of the cannula portion of the trocar sleeve provide a retention mechanism to stabilize the trocar sleeve in tissue. The reusable trocar sleeve, supplied smooth or threaded, is manufactured so that it may be cleaned, sterilized, and reused. The ENDOPATH® Resposable Trocar System shall be provided in a variety of sizes from 5mm to 12mm in diameter and 50mm to 150mm in length. In addition to the size variety they are supplied with or without stopcock for insufflation.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the ENDOPATH® Resposable Trocar System, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The provided text describes pre-clinical laboratory evaluations to demonstrate acceptable performance. However, it does not explicitly state specific quantitative acceptance criteria or detailed performance metrics. Instead, it indicates the device "demonstrated acceptable performance" in comparison to predicate devices across several functional aspects.

    Acceptance Criteria (Implied)Reported Device Performance
    Mating Obturator with SleeveAcceptable performance demonstrated.
    Insertion into Operative CavityAcceptable performance demonstrated.
    Shield Action to Cover BladeAcceptable performance demonstrated.
    Removal of Obturator from SleeveAcceptable performance demonstrated.
    Security of Sleeve in TissueAcceptable performance demonstrated.
    Maintenance of Pneumoperitoneum of Operative SpaceAcceptable performance demonstrated.

    Study Information

    1. Sample Size Used for the Test Set and Data Provenance:

      • Sample Size: Not specified. The document only mentions "Pre-clinical laboratory evaluations."
      • Data Provenance: Not specified. It's likely these were in-house laboratory tests, not human or animal studies, given the nature of the device and the lack of ethical review board mentions. The country of origin for the data is implied to be the US since the submitting company (Ethicon Endo-Surgery, Inc.) is based in Cincinnati, Ohio, and the submission is to the FDA. The studies were retrospective in the sense that they were conducted for this 510(k) submission, but not "retrospective" in terms of analyzing pre-existing patient data.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

      • Not applicable. The pre-clinical laboratory evaluations described likely involved engineering and technical testing against established performance benchmarks or predicate devices, rather than expert interpretation of a 'ground truth' as would be seen in diagnostic imaging.
    3. Adjudication Method for the Test Set:

      • Not applicable. As described, this was a technical performance study, not one requiring expert adjudication of results.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

      • No, an MRMC comparative effectiveness study was not done. The document describes pre-clinical laboratory evaluations, not studies involving human readers or clinical cases.
    5. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Not applicable. This device is a surgical instrument, not an algorithm or AI system. Its performance evaluation is based on its physical and functional characteristics.
    6. The Type of Ground Truth Used:

      • The "ground truth" for these pre-clinical evaluations would have been based on engineering specifications, established performance standards for similar predicate devices, and internal quality control benchmarks for mechanical and functional attributes.
    7. The Sample Size for the Training Set:

      • Not applicable. This is not an AI/machine learning device, so there is no concept of a "training set."
    8. How the Ground Truth for the Training Set Was Established:

      • Not applicable, as there is no training set mentioned or implied.