Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K223040
    Device Name
    Endo SmartCap
    Manufacturer
    Date Cleared
    2023-03-01

    (153 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Endo SmartCap

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Endo SmartCap™ is intended to be used with an air or CO2 source and/or a pump along with a sterile water source to supply air or CO2 and sterile water to an endoscope during endoscopic procedures.

    Device Description

    The Endo SmartCap™ supplies sterile water and either air or CO2 to an endoscope during endoscopic procedures when connected to a sterile water source and an air or CO2 source. Two tubes are arranged coaxially; air or CO2 supplied through the outer tube pressurizes the sterile water container, forcing water up through the inner tube and to the endoscope. The proximal end of the tube set includes a bottle cap for attachment to a user-provided sterile water container; the Endo SmartCap™ is compatible with all major brands of sterile water containers. The proximal end of the tube set terminates in a tube weight which keeps the tip of the tubing submerged. A pinch clip prevents water from dripping from the tube's distal end when the device is detached from the endoscope. The distal end of the tube set terminates in a connector which mates with the corresponding brand of endoscope; Endo SmartCap™ models are available for all major brands of endoscopes. The device is provided sterile (EtO). The device is fabricated from metals, plastics, and elastomers. The Endo SmartCap™ is intended for 24 hour multi-patient use and should be discarded daily. No portion of the device comes in direct contact with patients.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for the "Endo SmartCap™" device. This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving that an AI/ML device meets specific performance criteria. The document describes a medical device accessory, not an AI/ML system.

    Therefore, many of the requested details about acceptance criteria for AI/ML performance, study design, sample sizes for test/training sets, expert involvement, and ground truth establishment cannot be found in this document because the device is a physical accessory (Endoscope and Accessories), not an AI/ML software.

    However, I can extract the information related to the device's functional performance testing.

    Here's what can be extracted from the document:

    1. Table of Acceptance Criteria and Reported Device Performance (for functional testing):

    TestAcceptance CriteriaReported Performance
    Confirmation of flow rateThe average mean flow of the 100551-production sample must be no lower than 10% of the flow of the same tubeset without a filterPASS
    Prevention of backflowThe 100551 shall not allow water to backflow into the CO2EFFICIENT insufflatorPASS

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated as a number of devices/tests, but implies multiple production samples for the flow rate test ("average mean flow of the 100551-production sample").
    • Data Provenance: The document does not specify the country of origin of the data. The tests conducted are functional performance tests of the device itself, not data-driven performance. The tests are retrospective in the sense that they were conducted for the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable as the device is a physical accessory undergoing functional engineering tests (flow rate, backflow prevention), not an AI/ML algorithm requiring expert interpretation of medical images or data for ground truth. The "ground truth" here is the objective measurement of physical properties.

    4. Adjudication method for the test set:

    • Not applicable for functional engineering tests. The results are objective measurements against predefined engineering specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a physical medical accessory, not an AI/ML system for diagnostic assistance. No human reader studies with or without AI were conducted or are relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an algorithm. The reported tests are for the standalone physical device's functional performance.

    7. The type of ground truth used:

    • For flow rate and backflow prevention tests, the "ground truth" is established by engineering specifications and direct physical measurement. For example, for flow rate, it would involve direct measurement of water/gas volume over time. For backflow, it's the observation of whether water backflows into the insufflator.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device, so there is no training set.

    9. How the ground truth for the training set was established:

    • Not applicable. As there is no training set for an AI/ML algorithm, there is no ground truth for a training set.

    In summary, this document is for a traditional medical device accessory (Endo SmartCap™) seeking 510(k) clearance based on substantial equivalence. The "study that proves the device meets the acceptance criteria" in this context refers to engineering performance testing rather than the clinical validation or AI algorithm performance studies typically associated with the detailed questions provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133298
    Device Name
    ENDO SMART
    Manufacturer
    Date Cleared
    2014-09-19

    (329 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ENDO SMART

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Endo Smart ES-100 Endo Motor is indicated for use in standard endodontic procedures using rotary endodontic files and rotary endodontic drills

    Device Description

    Endo Smart, Model ES-100 is a controller for micro motor and contra angle headpieces which is designed to assist dentists and dental surgeons perform standard endodontic procedures. It provides power to the hand-held micromotor and contra angle which holds the drill bit and/or file used in endodontic procedures.

    Endosmart is only intended to be used with motor hand piece (manufactured by Saeshin Precision Company; part name CUBE EP) and Contra angle headpiece (Manufactured by Saeshin Precision Company; part name ACL (B) - 42P) at 16:1 gear ratio.

    AI/ML Overview

    The Metabiomed, Inc. Endo Smart, ES-100 is an endo motor designed to assist dentists and dental surgeons in standard endodontic procedures using rotary endodontic files and drills.

    Here's an analysis of the acceptance criteria and the study presented:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance for Endo Smart ES-100
    Intended Use- Indicated for use in standard endodontic procedures using rotary endodontic files and rotary endodontic drills.- Endo Smart ES-100 Endo Motor is indicated for use in standard endodontic procedures using rotary endodontic files and rotary endodontic drills.
    Patient Contacting Elements- Patient contacting elements should have prior market clearance.- Patient contacting elements supplied with Endo Smart ES-100 (motor hand piece, contra angle headpiece, and files) have been previously cleared for marketing (K111616).
    Biocompatibility- Biocompatibility testing should be performed for patient-contacting components.- The Endosmart ES-100 Endo Motor (controller unit) does not contact patients, therefore no biocompatibility testing was performed on the controller. (This implies that the contacting elements, already cleared, met biocompatibility criteria under their previous clearance).
    Electromagnetic Compatibility (EMC) and Electrical Safety- Conformance to relevant IEC standards for medical electrical equipment.- Conforms with IEC 60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance.
    - Conforms with IEC 60601-1-2 Medical electric equipment, General requirements for safety collateral standard electromagnetic compatibility.
    - Conforms with IEC 61000-3-2 Electromagnetic Compatibility (EMC) Part 3: Limits -Section 3: Limitations of Voltage Fluctuations.
    Performance (Functional Equivalence to Predicate devices)- Similar technological characteristics to predicate devices (e.g., torque range, mode of operation, accessories, display, protection type against electric shock).- Torque Range: 0.6~6.4Ncm (Gear 16:1 basis) (Predicate E-CUBE: 0.6 to 5.2 Ncm (Gear ratio: 16:1)) - Mode of Operation: Rotary, Auto Stop/Forward and Reverse (Predicate E-CUBE: Rotary, Auto Stop/Forward and Reverse) - Micro motor Manufacturer: SAESHIN PRECESION CO., Ltd. (Same as predicate) - Micro motor Model: CUBE EP (Same as predicate) - Contra Angle: ACL(B)-42EP, 16:1 (Same as predicate) - Display: LCD (Same as predicate) - Protection type and level against electric shock: Class IIa /Type B applied part, internal powered device (Predicate E-CUBE: Class I/ Type BF) - Physical Characteristics (Size, Weight): (Close but not identical. ES-100: 110 x 134 x 116.5 mm, 443g. E-CUBE: 110 x 196 x 139 mm, 582g) - Power: Lithium Ion, DC 7.2V Rechargeable, 2.2 Amp (Predicate E-CUBE: AC or DC powered)
    Software Validation- Adherence to FDA Software Validation Guidance.- Software Validation FDA Software Validation Guidance mentioned as performed.

    2. Sample size used for the test set and the data provenance

    The document indicates that conformity to standards (IEC 60601-1, IEC 60601-1-2, IEC 61000-3-2, and FDA Software Validation Guidance) was used to establish performance and safety. There is no mention of a traditional "test set" in the context of clinical or diagnostic performance with a specified sample size of patients/cases. The studies performed are engineering tests demonstrating adherence to safety and performance standards for an electro-mechanical device.

    Therefore, information on data provenance (country of origin, retrospective/prospective) is not applicable in the context of this 510(k) submission for this device type, as it relies on engineering testing rather than clinical study data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The performance testing described relates to adherence to technical standards and the functional characteristics of the device, not a diagnostic or clinical assessment where ground truth would be established by experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This type of adjudication pertains to clinical studies where expert consensus often establishes ground truth for diagnostic accuracy, which is not the nature of the testing described for this endo motor.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is not an AI-based diagnostic tool. It is an electro-mechanical device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is not an algorithm, and its function inherently involves a human operator (dentist/dental surgeon).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the device's acceptable performance is established through adherence to recognized international and electrical safety standards (IEC 60601-1, IEC 60601-1-2, IEC 61000-3-2) and FDA Software Validation Guidance. For the functional aspects, the ground truth is comparative equivalence to legally marketed predicate devices in terms of intended use and technological characteristics (e.g., torque range, mode of operation, use of specific micro motor/contra angle).

    8. The sample size for the training set

    Not applicable. The device is not a machine learning or AI model that requires a "training set."

    9. How the ground truth for the training set was established

    Not applicable. No training set for an AI/ML model is mentioned.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140753
    Manufacturer
    Date Cleared
    2014-07-15

    (111 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ENDO SMART BOTTLE, RINSE AND INSUFFLATION SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EndoSmart Bottle is intended to be used with an air or CO2 source and/or a pump along with a sterile water source to supply air or CO2 and sterile water to an endoscope during endoscopic procedures.

    Device Description

    Endo Smart Bottle is designed to be attached to an endoscope, to help supply water, air/CO2. The device consists of a connector cap, a water bottle, a threaded cap, a small tube, and a male luer. The main function of the subject device is -

    • . To provide water for rinsing the lens.
    • To provide air or CO2 to insufflate the anatomical lumen, to help the end user to see . the inner wall more clearly.

    Similar to its predicate device, the subject device is provided sterile to the end user, fabricated from plastics, and elastomers, intended for daily (24 hour) multi-patient use and must be discarded daily. Both the subject device and its predicate device do not come in direct contact with patients.

    AI/ML Overview

    This document is a 510(k) summary for the Medivators Endo Smart Bottle, an endoscopic irrigation/suction system. It primarily focuses on demonstrating substantial equivalence to a predicate device, the Endo Smart Cap. The document describes the device, its intended use, and non-clinical performance data to support its safety and effectiveness.

    Here's an analysis of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a quantitative manner (e.g., minimum accuracy, sensitivity, or specificity thresholds). Instead, it lists types of performance tests designed to demonstrate equivalence to the predicate device. The "reported device performance" is described qualitatively as having passed these tests, indicating equivalence.

    Acceptance Criteria (Inferred from testing)Reported Device Performance (Qualitative)
    Lens Rinsing FunctionPerformed equivalently to predicate
    Insufflation FunctionPerformed equivalently to predicate
    Safety (Design Verification)Performed equivalently to predicate
    Cytotoxicity EvaluationMet biocompatibility requirements
    Intracutaneous Irritation TestMet biocompatibility requirements
    SensitizationMet biocompatibility requirements
    Functional Testing (Shelf-life)Maintained function over shelf-life
    Dye Penetration Testing (Shelf-life)Maintained integrity over shelf-life
    Peel Strength Testing (Shelf-life)Maintained integrity over shelf-life
    Visual Seal Inspection (Shelf-life)Maintained integrity over shelf-life
    Sterilization ValidationMet sterility requirements

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not provide details on the specific sample sizes used for each non-clinical performance test (e.g., how many bottles were tested for lens rinsing function). It also does not mention data provenance (country of origin or retrospective/prospective nature) as these are non-clinical engineering and bench tests, not clinical studies involving patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable. The evaluations described are "Design Verification," "Biocompatibility," "Shelf-life validation," and "Sterilization validation," which are engineering and laboratory tests based on established standards and protocols, not expert consensus on medical interpretation.

    4. Adjudication Method for the Test Set

    This is not applicable as the tests are non-clinical engineering and laboratory tests, not clinical performance studies requiring adjudication of results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No such study was conducted or is mentioned. This product is a physical medical device (endoscopic irrigation/suction system) and does not involve AI or human "readers" in the context of diagnostic interpretation.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. This device is a manual accessory for an endoscope and does not involve algorithms or AI.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the non-clinical performance tests would be defined by established engineering and biological standards, specifications, and test methodologies relevant to medical device performance (e.g., maintaining specific flow rates, successfully clearing a lens, demonstrating no cytotoxic effects, maintaining sterile barrier integrity). It is not based on expert consensus, pathology, or outcomes data in a clinical sense.

    8. The Sample Size for the Training Set

    This is not applicable. This device does not use machine learning or AI, and therefore does not have a "training set."

    9. How the Ground Truth for the Training Set was Established

    This is not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K093665
    Device Name
    ENDO SMARTCAP
    Manufacturer
    Date Cleared
    2009-12-15

    (20 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ENDO SMARTCAP

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Endo SmartCap™ is intended to be used with an air or CO2 source and/or a pump along with a sterile water source to supply air or CO2 and sterile water to an endoscope during endoscopic procedures.

    Device Description

    Not Found

    AI/ML Overview

    This document does not contain the detailed information necessary to complete the request regarding acceptance criteria and a study proving a device meets them. The provided text is a 510(k) clearance letter and summary for the "Endo SmartCap™," which primarily focuses on regulatory approval and substantial equivalence to predicate devices, rather than detailed performance study results or acceptance criteria.

    Specifically, the document does not include:

    • A table of acceptance criteria and reported device performance.
    • Sample size used for a test set, data provenance, or details about the ground truth establishment for a test set.
    • Information about expert involvement for ground truth, adjudication methods, or MRMC studies.
    • Results from a standalone algorithm performance study.
    • Sample size for a training set or ground truth establishment for a training set.

    The document primarily states the intended use of the device and its regulatory classification. It lists predicate devices but does not delve into their performance or comparison in detail.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1