Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K123582
    Device Name
    ENDO A CLASS
    Date Cleared
    2013-08-16

    (269 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ENDO A CLASS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This application area extends to endodontic procedures using a root canal instrument which is intended by the manufacturer for use in the mechanical and rotary preparation of root canals.

    Device Description

    The motor turned by the power converted into DC2.4V by controller delivers its turning power to the file through spin to perform punching, cutting and removing functions. The hand-piece can be operated, stopped and set/adjusted on/in its speed, torque and turning direction by handling of the controller.

    AI/ML Overview

    The provided text describes Saeyang Microtech's ENDO a class device, a cordless endodontic handpiece. The 510(k) summary (K123582) details its intended use, substantial equivalence to predicate devices, and the non-clinical tests performed.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" in the way one might expect for a diagnostic or AI device (e.g., sensitivity, specificity thresholds). Instead, the performance is demonstrated through compliance with established international standards and a comparison of technical specifications with predicate devices to establish substantial equivalence.

    Based on the provided information, the acceptance criteria relate to compliance with regulatory standards and functional equivalence to predicate devices.

    Acceptance CriterionReported Device Performance
    Regulatory Compliance & Safety
    IEC 60601-1: 1988 +A1 1991,+A2 1995 (Medical electrical equipment)Complied
    ANSI/AAMI/IEC 60601-1-2: 2007 (Electromagnetic compatibility)Complied
    ISO3964:1982Complied
    ISO7494-1:2004Complied
    ISO7785-2:1995Complied
    ISO11498:1997Complied
    Sterilization ValidationComplied
    Software ValidationComplied
    Speed Accuracy TestingComplied
    Temperature Rise TestingComplied
    General Safety HazardsNone demonstrated that violated requirements or resulted in hazards.
    Functional Equivalence to Predicate Devices
    Intended UseSubstantially equivalent to predicate devices (Endodontic procedures using a root canal instrument for mechanical and rotary preparation).
    OperationSubstantially equivalent to predicate devices.
    Speed Adjustment Range140-500 rpm (Predicate devices: 125-625 rpm, 280 rpm, 300 rpm) - Comparable range
    Torque Adjustment Range0.1-4.0 Ncm (Predicate devices: 1.0-5.0 Ncm, Low/High mode, 0.5-4.0 Ncm) - Comparable range
    Gear Ratio Selection4:1, 10:1, 16:1, 20:1 (Predicate devices: 17:1, 32:1, 16:1) - Offers more options or overlaps with predicates
    Forward/Auto Reverse DriveYES (Matches predicate devices)
    Reciprocation DriveYES (Matches predicate devices)
    User-defined Presets (Speed/Torque)YES (Matches predicate devices)
    Auto Power OffYES (Matches predicate devices)
    Product MaterialComparable to predicate devices (ABS, silicon, SUS)
    Principle of OperationSubstantially equivalent to predicate devices (motorized handpiece for punching, cutting, removing functions in root canals).
    Rechargeable BatteryNi_Mh 1.2V2 (Comparable to predicate Li_ion 3.7V, Ni_Mh 1.2V3)
    Motor VoltageDC2.4V (Comparable to predicate DC3.6V, DC3.6V, DC2V)
    Physical Dimensions & WeightComparable to predicate devices.

    2. Sample size used for the test set and the data provenance

    The document describes non-clinical bench tests performed to ensure compliance with relevant ISO and IEC standards. These tests typically do not involve "test sets" in the sense of patient data. Instead, they involve testing of physical devices or components against specified technical requirements and protocols.

    • Sample Size: Not explicitly stated as a number of devices, but the testing would have been conducted on a representative sample of the ENDO a class device.
    • Data Provenance: The tests were conducted internally by Saeyang Microtech in Korea, with results being "Complied" for each standard. Given the nature of bench testing for regulatory submission, this would be considered prospective testing for the purpose of demonstrating compliance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable to this submission. The device is a direct drive, AC-powered handpiece for endodontic procedures, not an AI or diagnostic imaging device that requires expert-established ground truth from clinical data. The "ground truth" here is the adherence to engineering and safety standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies for diagnostic accuracy involving multiple human readers, which is not relevant for this device. The non-clinical tests involved verifying physical and functional specifications against established standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-assisted diagnostic tools, not for a dental handpiece. The device is a mechanical tool, and the submission explicitly states: "No clinical testing was conducted."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The ENDO a class is a physical device, a dental handpiece, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" in this context refers to compliance with established international standards (ISO, IEC) for medical devices, specifically dental handpieces, and the functional specifications of the device meeting pre-defined thresholds or being comparable to predicate devices. It is based on engineering and performance testing protocols.

    8. The sample size for the training set

    This question is not applicable. There is no "training set" as this is not an AI or machine learning device.

    9. How the ground truth for the training set was established

    This question is not applicable. There is no "training set."

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1