Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K092649
    Date Cleared
    2010-02-04

    (160 days)

    Product Code
    Regulation Number
    862.1117
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Immunoassay for the in vitro quantitative determination of N-terminal pro-brain natriuretic peptide in human serum and plasma. The Elecsys proBNP II STAT assay is used as an aid in the diagnosis of individuals suspected of having congestive heart failure. The test is further indicated for the risk stratification of patients with acute coronary syndrome and congestive heart failure. The test may also serve as an aid in the assessment of increased risk of cardiovascular events and mortality in patients at risk for heart failure who have stable coronary artery disease.

    The electrochemiluminescence immunoassay "ECLIA" is intended for use on the cobas e 601 immunoassay analyzer.

    Device Description

    The proBNP II Assay is a sandwich immunoassay with two antibodies directed towards epitopes within the N-terminal portion of the proBNP molecule. The capture antibody is biotinylated to react with streptavidin-coated microparticles. The signal antibody is tagged with ruthenium. Both antibodies are monoclonal. The electrochemiluminescence immunoassay "ECLIA" is intended for use on the cobas e 601 immunoassay analyzer.

    AI/ML Overview

    Acceptance Criteria and Device Performance Study for Elecsys proBNP II STAT Immunoassay

    This document describes the acceptance criteria and the study that demonstrates the Elecsys proBNP II STAT Immunoassay meets these criteria, based on the provided 510(k) summary.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Elecsys proBNP II STAT Immunoassay are primarily demonstrated by showing substantial equivalence to the predicate device, the Elecsys proBNP II Assay (K072437), with key performance characteristics either matching or falling within acceptable ranges. The primary difference is the faster "STAT" application time and limited instrument platform.

    FeatureAcceptance Criteria (Predicate Device K072437)Reported Device Performance (Elecsys proBNP II STAT Assay)
    Intended Use/Indications for UseImmunoassay for the in vitro quantitative determination of N-terminal pro-brain natriuretic peptide in human serum and plasma. Aid in diagnosis of individuals suspected of congestive heart failure. Risk stratification of patients with acute coronary syndrome and congestive heart failure. Aid in assessment of increased risk of cardiovascular events and mortality in patients at risk for heart failure with stable coronary artery disease. Intended for use on Elecsys and cobas e immunoassay analyzers.Same, except: Assay name changed to "Elecsys proBNP II STAT assay". Intended for use only on the cobas e 601 immunoassay analyzer.
    Assay ProtocolSandwich assaySame
    Detection ProtocolElectrochemiluminescent ImmunoassaySame
    Application Time18 MinuteSTAT (9 Minute) - This is a key improvement/change.
    Instrument PlatformRoche Elecsys 2010/cobas e 411 and MODULAR ANALYTICS E170/cobas e 601cobas e 601 - Limited to this specific platform.
    Sample Volume15 µLSame
    Sample TypeHuman serum and plasma treated with K2-EDTA, K3-EDTA, lithium heparin and Na-heparin plasma.Same
    ReagentsSandwich immunoassay with two monoclonal antibodies (biotinylated capture (streptavidin-coated microparticles) and ruthenium-tagged signal) directed towards epitopes within the N-terminal portion of the proBNP molecule.Same antibodies and same epitopes.
    Traceability/StandardizationStandardized against the Elecsys proBNP assay (K022516), which was standardized against reference standards by weighing pure synthetic NT-proBNP (1-76 amino acids) into equine serum matrix.Same.
    CalibratorElecsys proBNP II CalSet (K072437)Elecsys proBNP II STAT CalSet (different material number, identical stability, value assignment, and matrix).
    Calibration IntervalOnce per reagent lot (fresh reagent, <24h since registration). Renewed calibration recommended: MODULAR ANALYTICS E170, Elecsys 2010 and cobas e analyzers: After 1 month (28 days) when using same reagent lot; After 7 days (when using same reagent kit). Elecsys 1010 analyzer: With every reagent kit; After 7 days (ambient 20-25°C); After 3 days (ambient 25-32°C).Once per reagent lot (fresh reagent, <24h since registration). Renewed calibration recommended: cobas e 601: After 1 month (28 days) when using same reagent lot; After 7 days (when using same reagent kit).
    ControlsElecsys PreciControl Cardiac II (K072437).Same.
    Reagent Stability/StorageUnopened: 2-8°C until expiration. After opening: 2-8°C—12 weeks. On MODULAR ANALYTICS E170 and cobas e 601—8 weeks. On Elecsys2010 and cobas e 411—8 weeks.Same, but only reported on the cobas e 601.
    Measuring Range5-35,000 pg/mLSame
    Precision (cobas e 601 - predicate)Within-run (Repeatability): 1.9% CV @ 64.0 pg/mL, 1.5% CV @ 124.0 pg/mL, 1.3% CV @ 14142.0 pg/mL, 1.8% CV @ 77.0 pg/mL, 1.2% CV @ 2105.0 pg/mL. Total (Intermediate): 3.1% CV @ 46.0 pg/mL, 2.7% CV @ 125.0 pg/mL, 2.7% CV @ 32805.0 pg/mL, 2.7% CV @ 77.0 pg/mL, 2.7% CV @ 2170.0 pg/mL.Precision (cobas e 601 - STAT assay)Within-run (Repeatability): 2.4% CV @ 130.0 pg/mL, 2.2% CV @ 4942.0 pg/mL, 3.5% CV @ 59.0 pg/mL, 2.0% CV @ 142.0 pg/mL, 1.8% CV @ 522.0 pg/mL, 1.9% CV @ 934.5 pg/mL, 2.0% CV @ 6552.0 pg/mL, 2.9% CV @ 30,870.0 pg/mL. Total (Intermediate): 2.5% CV @ 130.0 pg/mL, 2.6% CV @ 4942.0 pg/mL, 3.5% CV @ 59.0 pg/mL, 2.5% CV @ 142.0 pg/mL, 2.0% CV @ 522.0 pg/mL, 2.5% CV @ 934.5 pg/mL, 2.3% CV @ 6552.0 pg/mL, 5.4% CV @ 30,870.0 pg/mL. Performance is comparable or improved across the measuring range, especially considering the STAT nature.
    Analytical SensitivityLimit of Detection (LoD): 5.00 pg/mL. Functional Sensitivity (LoQ): 50.00 pg/mL. (LoB not reported)Limit of Blank (LoB): 5.00 pg/mL. Limit of Detection (LoD): 5.00 pg/mL. Functional Sensitivity (LoQ): 50.00 pg/mL. This demonstrates equivalent or improved analytical sensitivity.
    Hook EffectNo high-dose hook effect up to 300,000 pg/mL (300 ng/mL).Same.
    Heterophilic Antibody/InterferenceUnaffected by: Hemoglobin ≤ 0.1 g/dL, Bilirubin up ≤ 25 mg/dL, Triglycerides ≤ 1,500 mg/dL, Biotin ≤ 30 ng/mL, Rheumatoid factors ≤ 1,500 IU/mL. No interference with 51 commonly used pharmaceuticals. Minimized risk of interference from immunological interactions with rare sera and monoclonal mouse antibodies. Measures taken to minimize interference from high titers of antibodies to streptavidin and ruthenium.Same.
    Method Comparison (Correlation between STAT and Predicate)N/A (this is the study's objective)Passing/Bablok: Slope 0.992, Intercept -2.141, Tau/r/rho 0.987. Linear Regression: Slope 0.986, Intercept 13.964, r 0.998.Deming Regression: Slope 0.988, Intercept 8.83, rho 1.00.These results demonstrate strong correlation between the two assays, indicating substantial equivalence in measurement.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: n = 120 patient samples were used for the method comparison study.
    • Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. However, based on the context of a 510(k) submission seeking equivalence to a previously approved device, these data are typically generated in controlled laboratory settings (often prospective testing of patient samples) by the manufacturer or authorized agents.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Number of Experts: Not applicable. This device is an immunoassay measuring a biomarker (NT-proBNP) quantitatively. The "ground truth" for the test set values is considered the measurements obtained by the predicate device itself, not an expert consensus on a diagnosis.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. As this is a quantitative immunoassay comparison, there is no "adjudication" in the sense of independent expert review of cases. The comparison is statistical, comparing the results of the new device against the predicate device.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study: No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging or subjective interpretation tasks where human readers' performance is being evaluated, possibly with AI assistance. For a quantitative immunoassay, the performance is assessed by comparing its measurements to a reference method (the predicate device) and evaluating its analytical characteristics (precision, sensitivity, range).
    • Effect Size of Human Readers with/without AI assistance: Not applicable.

    6. Standalone Performance Study

    • Standalone Study: Yes, a standalone performance study was done for the Elecsys proBNP II STAT Assay. The "method comparison" section directly shows the performance of the new device alone against the predicate device on patient samples (n=120) and demonstrates its analytical performance characteristics (precision, analytical sensitivity, measuring range, hook effect, limitations) independently. The core of this 510(k) summary is to demonstrate that the device itself performs equivalently to the predicate to warrant a "substantially equivalent" classification.

    7. Type of Ground Truth Used

    • Type of Ground Truth: The "ground truth" for the method comparison study was the measurements obtained from the existing, legally marketed predicate device (Elecsys proBNP II Assay, K072437). This is a common approach in 510(k) submissions for in vitro diagnostic (IVD) devices seeking substantial equivalence where a well-established predicate device exists. The predicate device's measured values are treated as the reference standard for comparison.

    8. Sample Size for the Training Set

    • Training Set Sample Size: The document does not explicitly mention a separate "training set" or its size in the context of typical machine learning models. This is an immunoassay, not an AI/ML-based device in the common understanding. The development and optimization of such assays typically involve iterative laboratory testing and refinement of reagents, protocols, and instrument parameters, which could be considered an internal "training" process, but not in the sense of a distinct, labeled dataset for algorithm training.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable in the conventional sense of AI/ML training. The "ground truth" for developing and optimizing the assay itself would be based on established analytical chemistry principles, characterization of assay components (antibodies, calibrators), and performance against known reference materials or other validated laboratory methods. This iterative process leads to the final assay formulation and protocol, which is then formally validated using the test set described above against the predicate.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1