Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K964694
    Device Name
    ELECSYS LH ASSAY
    Date Cleared
    1997-01-24

    (63 days)

    Product Code
    Regulation Number
    862.1485
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ELECSYS LH ASSAY

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Immunoassay for the in vitro quantitative determination of human luteinizing hormone in human serum and plasma.

    Device Description

    Sandwich principle. Total duration of assay: 18 minutes.
    • 1st incubation (9min.): 20 µL of sample, a biotinylated monoclonal LH-specific antibody (75 µL) and a monoclonal LH-specific antibody labeled with a ruthenium complex (75 µL)** react to form a sandwich complex.
    • 2nd incubation (9min.): after addition of streptavidin-coated microparticles (30 µL), the complex becomes bound to the solid phase via interaction of biotin and streptavidin.
    **Tris(2,2'-bipyridyl)ruthenium(II) complex (Ru(bpy)3)2+)
    •The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier (0.4 second read frame).
    •Results are determined via a calibration curve which is instrument-specifically generated by 2-point calibration and a master curve provided via the reagent bar code.

    AI/ML Overview

    The provided text describes the performance characteristics of the Elecsys® LH Assay and compares it to a predicate device, the Enzymun-Test® LH. The document acts as a 510(k) Summary, indicating that the study is for regulatory submission to establish substantial equivalence.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state "acceptance criteria" for each performance metric but rather presents "Performance Characteristics" for both the new Elecsys® LH and the predicate Enzymun-Test® LH. The implication is that the Elecsys® LH's performance should be comparable to or better than the predicate device for substantial equivalence.

    FeatureAcceptance Criteria (Implied: Comparable to or better than Predicate)Elecsys® LH Reported Performance
    PrecisionSimilar %CV values to predicate across low, mid, and high levels.Low: Within-Run Mean 0.54 mIU/mL, %CV 1.82%; Total Mean 0.54 mIU/mL, %CV 5.17%
    Mid: Within-Run Mean 9.38 mIU/mL, %CV 1.13%; Total Mean 9.38 mIU/mL, %CV 1.97%
    High: Within-Run Mean 50.72 mIU/mL, %CV 0.81%; Total Mean 50.72 mIU/mL, %CV 1.99%
    Lower Detection LimitComparable to or lower than predicate (0.50 mIU/mL).0.10 mIU/mL
    LinearityComparable to predicate range (0.5 - 150 mIU/mL) with a deviation of ±10%.0.1 - 200 mIU/mL (with a deviation from a linear line of ±10%)
    Method ComparisonStrong correlation with the predicate device (r value close to 1).Vs Enzymun-Test® LH:
    Least Squares: y = 1.00x - 0.199, r = 0.993, SEE = 1.141
    Passing/Bablok: y = 0.964x + 0.040, r = 0.993, SEE = 0.456
    Interfering substancesNo interference at or above specific concentration limits for bilirubin, hemoglobin, lipemia, biotin, and rheumatoid factor.No interference at:
    Bilirubin: 25.0 mg/dL
    Hemoglobin: 1 g/dL
    Lipemia: 1500 mg/dL
    Biotin: 30 ng/mL
    Rheumatoid Factor: 1500 U/mL
    Specificity (Cross-reactivity)Low or no cross-reactivity with FSH, HCG, TSH, HGH, HPL (e.g.,
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1