K Number
K964694
Device Name
ELECSYS LH ASSAY
Date Cleared
1997-01-24

(63 days)

Product Code
Regulation Number
862.1485
Panel
CH
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Immunoassay for the in vitro quantitative determination of human luteinizing hormone in human serum and plasma.

Device Description

Sandwich principle. Total duration of assay: 18 minutes.
• 1st incubation (9min.): 20 µL of sample, a biotinylated monoclonal LH-specific antibody (75 µL) and a monoclonal LH-specific antibody labeled with a ruthenium complex (75 µL)** react to form a sandwich complex.
• 2nd incubation (9min.): after addition of streptavidin-coated microparticles (30 µL), the complex becomes bound to the solid phase via interaction of biotin and streptavidin.
**Tris(2,2'-bipyridyl)ruthenium(II) complex (Ru(bpy)3)2+)
•The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier (0.4 second read frame).
•Results are determined via a calibration curve which is instrument-specifically generated by 2-point calibration and a master curve provided via the reagent bar code.

AI/ML Overview

The provided text describes the performance characteristics of the Elecsys® LH Assay and compares it to a predicate device, the Enzymun-Test® LH. The document acts as a 510(k) Summary, indicating that the study is for regulatory submission to establish substantial equivalence.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not explicitly state "acceptance criteria" for each performance metric but rather presents "Performance Characteristics" for both the new Elecsys® LH and the predicate Enzymun-Test® LH. The implication is that the Elecsys® LH's performance should be comparable to or better than the predicate device for substantial equivalence.

FeatureAcceptance Criteria (Implied: Comparable to or better than Predicate)Elecsys® LH Reported Performance
PrecisionSimilar %CV values to predicate across low, mid, and high levels.Low: Within-Run Mean 0.54 mIU/mL, %CV 1.82%; Total Mean 0.54 mIU/mL, %CV 5.17%
Mid: Within-Run Mean 9.38 mIU/mL, %CV 1.13%; Total Mean 9.38 mIU/mL, %CV 1.97%
High: Within-Run Mean 50.72 mIU/mL, %CV 0.81%; Total Mean 50.72 mIU/mL, %CV 1.99%
Lower Detection LimitComparable to or lower than predicate (0.50 mIU/mL).0.10 mIU/mL
LinearityComparable to predicate range (0.5 - 150 mIU/mL) with a deviation of ±10%.0.1 - 200 mIU/mL (with a deviation from a linear line of ±10%)
Method ComparisonStrong correlation with the predicate device (r value close to 1).Vs Enzymun-Test® LH:
Least Squares: y = 1.00x - 0.199, r = 0.993, SEE = 1.141
Passing/Bablok: y = 0.964x + 0.040, r = 0.993, SEE = 0.456
Interfering substancesNo interference at or above specific concentration limits for bilirubin, hemoglobin, lipemia, biotin, and rheumatoid factor.No interference at:
Bilirubin: 25.0 mg/dL
Hemoglobin: 1 g/dL
Lipemia: 1500 mg/dL
Biotin: 30 ng/mL
Rheumatoid Factor: 1500 U/mL
Specificity (Cross-reactivity)Low or no cross-reactivity with FSH, HCG, TSH, HGH, HPL (e.g.,

§ 862.1485 Luteinizing hormone test system.

(a)
Identification. A luteinizing hormone test system is a device intended to measure luteinizing hormone in serum and urine. Luteinizing hormone measurements are used in the diagnosis and treatment of gonadal dysfunction.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.