Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K221271
    Date Cleared
    2022-07-27

    (86 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Disposable Nitrile Powder-Free Examination Gloves (Tested for Use with Chemotherapy Drugs)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.

    Device Description

    The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in four sizes: extra-small, small, medium, large, and extra-large.

    AI/ML Overview

    The provided document describes the FDA 510(k) premarket notification for "Disposable Nitrile Powder-Free Examination Gloves (Tested for Use with Chemotherapy Drugs)". This document is for a medical device (gloves), not an AI/Software as a Medical Device (SaMD). Therefore, the concepts of acceptance criteria for AI performance, training/test sets, ground truth establishment by experts, MRMC studies, standalone algorithm performance, or a sample size for training data are not applicable to this submission.

    The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the physical and chemical performance standards for the gloves, particularly their resistance to chemotherapy drugs and biocompatibility.

    Here's the information extracted from the provided text, tailored to the context of a physical medical device:


    Acceptance Criteria and Device Performance for Disposable Nitrile Powder-Free Examination Gloves

    The acceptance criteria for the Disposable Nitrile Powder-Free Examination Gloves are based on compliance with established international and national standards for medical gloves, particularly regarding their physical properties, freedom from holes, residual powder content, and resistance to permeation by chemotherapy drugs.

    1. Table of Acceptance Criteria and Reported Device Performance

    No.Test Methodology / StandardPurposeAcceptance CriteriaReported Device Performance
    1ISO 10993-10:2010 (Irritation & Skin Sensitization)Assess possible contact hazards from chemicals released from medical devices (skin/mucosal irritation, eye irritation, skin sensitization).Skin Sensitization Test: Grades less than 1, otherwise sensitization.
    Skin Irritation Test: Primary irritation index 0-0.4 (Negligible).Skin Sensitization Test: All grades are 0. All animals survived with no abnormal signs.
    Skin Irritation Test: Primary irritation index is 0. Response categorized as negligible.
    2ISO 10993-5:2009 (In Vitro Cytotoxicity)Assess in vitro cytotoxicity of medical devices.Viability reduced to 240 Minutes
    • Cyclophosphamide (Cytoxan) 20.0 mg/ml: > 240 Minutes
    • Dacarbazine 10 mg/ml: > 240 Minutes
    • Doxorubicin 2.0 mg/ml: > 240 Minutes
    • Etoposide 20.0 mg/ml: > 240 Minutes
    • Fluorouracil 50.0 mg/ml: >240 Minutes
    • Paclitaxel 6.0 mg/ml: >240 Minutes
    • Thio Tepa 10.0 mg/ml: 46.8Minutes (min.) | Reported Breakthrough Detection Times: (Matches acceptance criteria values exactly, indicating compliance).
    • Carmustine (BCNU) 3.3 mg/ml: 22.8(23.7, 22.8, 23.1) Minutes
    • Cisplatin 1.0 mg/ml: > 240 Minutes
    • Cyclophosphamide (Cytoxan) 20.0 mg/ml: > 240 Minutes
    • Dacarbazine 10 mg/ml: > 240 Minutes
    • Doxorubicin 2.0 mg/ml: > 240 Minutes
    • Etoposide 20.0 mg/ml: > 240 Minutes
    • Fluorouracil 50.0 mg/ml: >240 Minutes
    • Paclitaxel 6.0 mg/ml: >240 Minutes
    • Thio Tepa 10.0 mg/ml: 46.8(48.2, 48.6, 46.8) Minutes.
      Warning: Low permeation times for Carmustine (BCNU) and Thio Tepa are noted, with a warning not to use with these specific drugs given their breakthrough times. |

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes material and performance testing, not a "test set" in the sense of an AI model's dataset. The "samples" refer to physical gloves selected for testing.

    The provenance of the data is from laboratory testing conducted on the manufactured gloves. The specific country of origin for the testing itself is not explicitly stated, but the manufacturer is Shandong Maida Medical Technology Co.,Ltd. in China. The testing is prospective in the sense that newly manufactured gloves are tested to ensure they meet the standards.

    • ASTM D5151 (Detection of Holes): Sample size for water leakage test: 125 gloves.
    • ASTM D6319 (Dimensions & Physical Properties): Results are shown for multiple lots (e.g., Lot no.: 210515, 210518, 210520), indicating multiple samples were tested across different manufacturing batches. The exact number of gloves per lot for each physical test (e.g., tensile strength, elongation) is not specified but would be defined by the ASTM standard.
    • ASTM D6978 (Chemotherapy Permeation): The results provide triplet values for Carmustine and Thio Tepa (e.g., 22.8(23.7, 22.8, 23.1)), suggesting three independent measurements were taken for these drugs. The standard itself specifies the number of replicates.
    • Biocompatibility Testing: The number of animal subjects used for the animal irritation, sensitization, and systemic toxicity tests is not specified in the summary but would conform to the respective ISO standards.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This section is not applicable as the evaluation involves standardized laboratory testing against defined physical and chemical properties, not human expert interpretation of subjective data or images. The "ground truth" is defined by the objective measurement results according to the specified ASTM and ISO standards.

    4. Adjudication Method for the Test Set

    This is not applicable. Adjudication methods (like 2+1, 3+1) are for resolving discrepancies in human expert assessments (e.g., in medical image interpretation). Here, numerical test results from standardized methods are directly compared against established quantitative or qualitative acceptance criteria.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    This is not applicable. MRMC studies are used to evaluate diagnostic imaging systems or AI tools where human readers are involved. This submission is for a physical medical glove.

    6. Standalone (Algorithm Only) Performance

    This is not applicable. This concept applies to AI/software performance. The device is a physical glove, not an algorithm.

    7. Type of Ground Truth Used

    The ground truth used is based on objective, quantitative measurements derived from standardized laboratory tests (bench testing) and biological responses observed in validated biocompatibility assays according to international standards (ASTM and ISO). This includes:

    • Quantitative Thresholds: e.g., minimum tensile strength, maximum residual powder, minimum breakthrough time for chemotherapy drugs.
    • Qualitative Observations (Biological): e.g., absence of irritation or sensitization, absence of systemic toxicity.
    • Conformity to Standards: The device's performance is compared directly to the specified requirements within each standard (e.g., ASTM D6319-19 dimensions, physical properties).

    One point of note is the cytotoxicity result (17.1% viability) which is below the 70% threshold, indicating cytotoxicity. The document addresses this by stating: "Analysis3: The proposed device is potentially cytotoxic, but all proposed devices are conducted the systemic toxicity test." And the systemic toxicity test showed "no evidence of systemic toxicity." This implies that despite the in vitro cytotoxicity, the manufacturer argues that in vivo systemic toxicity (which is generally considered a more critical endpoint for patient safety) was not observed, suggesting the findings are acceptable for the device's intended use according to the biocompatibility evaluation framework.

    8. Sample Size for the Training Set

    This is not applicable. There is no "training set" in the context of a physical device. The manufacturing process is controlled to produce gloves consistently, and batch testing ensures quality.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable. As there is no training set, the establishment of ground truth for it is irrelevant. The manufacturing process is designed and verified to produce products meeting the specified design and performance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221192
    Date Cleared
    2022-06-22

    (58 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Disposable Nitrile Powder-Free Examination Gloves

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Nitrile Powder-Free Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The proposed device is Powder Free Disposable Nitrile Powder-Free Examination Gloves. The proposed device is blue. The design of proposed device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D6319. The proposed device is non-sterile.

    AI/ML Overview

    The document provided is a 510(k) summary for Disposable Nitrile Powder-Free Examination Gloves. It details the safety and effectiveness information, comparing the proposed device to a legally marketed predicate device. The information provided heavily focuses on non-clinical performance testing rather than studies involving human readers or AI.

    Based on the provided text, here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance:

    No.Name of the Test Methodology / StandardAcceptance CriteriaResults
    1ISO 10993-10:2010 (Skin Sensitization)Skin Sensitization Test: provided grades less than 1, otherwise sensitization.All grades are 0. All animals were survived and no abnormal signs were observed during the study.
    2ISO 10993-10:2010 (Skin Irritation)Skin Irritation Test: If the primary irritation index is 0-0.4, the response category is Negligible. 0.5-1.9 means slight, 2-4.9 means moderate, 5-8 means severe.The primary irritation index is 0. The response of the proposed device was categorized as negligible under the test condition.
    3ISO 10993-5:2009 (In Vitro Cytotoxicity)The viab.% of the 100% extract of the test article is the final result, and if viability is reduced to
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    Disposable Nitrile Powder-Free Examination Gloves (Tested for Use with Chemotherapy Drugs)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.

    Device Description

    The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in four sizes: small, medium, large, and extra-large.

    AI/ML Overview

    This document describes the acceptance criteria and the study proving that the "Disposable Nitrile Powder-Free Examination Gloves (Tested for Use with Chemotherapy Drugs)" meet these criteria.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The device was tested against several standards, as detailed in the "Summary of Non-Clinical Testing" section (pages 8-13). Here's a summary:

    No.Test/StandardAcceptance CriteriaReported Device Performance
    1ISO 10993-10:2010 (Skin Sensitization)Grades less than 1.All grades were 0. All animals survived with no abnormal signs. The device is not a sensitizer.
    2ISO 10993-10:2010 (Skin Irritation)Primary irritation index of 0-0.4 for "Negligible" response.The primary irritation index was 0. The response was categorized as negligible. The device is not an irritant.
    3ISO 10993-5:2009 (In Vitro Cytotoxicity)Viability ≥ 70% of the blank (if viability is reduced to 240 Minutes
    Cyclophosphamide (Cytoxan) 20.0 mg/ml: > 240 Minutes
    Dacarbazine 10 mg/ml: > 240 Minutes
    Doxorubicin HCl 2.0 mg/ml: > 240 Minutes
    Etoposide 20.0 mg/ml: > 240 Minutes
    Fluorouracil 50.0 mg/ml: >240 Minutes
    Methotrexate, 25 mg/ml: >240 Minutes
    Mitomycin C, 0.5 mg/ml: >240 Minutes
    Paclitaxel 6.0 mg/ml: >240 Minutes
    Thio Tepa 10.0 mg/ml: 37.4 Minutes (min.)
    Vincristine Sulfate, 1mg/ml: >240 Minutes

    2. Sample Size Used for the Test Set and Data Provenance:

    The document provides the following details:

    • ASTM D5151 (Freedom from Holes): Sample size of 125 gloves. The provenance of this data (e.g., country of origin, retrospective/prospective) is not explicitly stated but is implied to be from laboratory testing conducted as part of the regulatory submission process.
    • For other tests (Biocompatibility, Physical Properties, Chemotherapy Drug Permeation), specific sample sizes are not detailed in the provided text. However, testing was stated to be "in accordance with the following standards" (ISO and ASTM), which would implicitly define minimum sample sizes for each test method. Data provenance is implied to be from laboratory testing.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    This information is not provided in the document. The tests conducted are laboratory-based and follow established international and national standards (ISO, ASTM). The "ground truth" for these tests is defined by the methodologies and performance requirements within these standards, not by expert consensus in a clinical sense.

    4. Adjudication Method for the Test Set:

    This information is not applicable as the tests are laboratory performance tests following well-defined standards, not clinical studies requiring independent adjudication of outcomes.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    This information is not applicable as the device is a medical glove, and the studies are pre-market non-clinical performance and biocompatibility evaluations. It is not an AI-assisted diagnostic device or a device where human reader performance would be a relevant metric.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    This information is not applicable as the device is a medical glove, not an algorithm or AI system. The performance evaluated is the physical and chemical resistance of the glove itself.

    7. Type of Ground Truth Used:

    The ground truth for the evaluations is based on:

    • Standardized Laboratory Test Results: Defined by the methodologies and acceptance criteria specified in the referenced ASTM and ISO international standards (e.g., specific tensile strength values, breakthrough times, chemical concentrations).
    • Biocompatibility Endpoints: As defined by ISO 10993 standards (e.g., absence of irritation, sensitization, systemic toxicity).

    8. Sample Size for the Training Set:

    This information is not applicable. This is a physical medical device, not a machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211360
    Date Cleared
    2022-01-15

    (257 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    Disposable Nitrile Powder Free Examination Gloves (Non-sterile)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder-free nitrile examination gloves is a disposable device intended for medical purposes that is worn on the examiner 's hand or finger to prevent cross-contamination between patient and examiner.

    Device Description

    Disposable Nitrile Powder Free Examination Gloves (Non-sterile)

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for "Disposable Nitrile Powder Free Examination Gloves (Non-sterile)." It is not a study describing acceptance criteria or performance data for an AI/ML medical device.

    Therefore, I cannot provide the requested information about acceptance criteria and study details as it pertains to AI/ML devices. The document concerns traditional medical devices (examination gloves) and focuses on regulatory clearance based on substantial equivalence to predicate devices, rather than performance data against specific acceptance criteria for an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213176
    Date Cleared
    2021-12-13

    (76 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Disposable Nitrile Powder-Free Examination Gloves

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Nitrile Powder-Free Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The proposed device is Powder Free Disposable Nitrile Powder-Free Examination Gloves. The proposed device is blue. The design of proposed device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D6319. The proposed device is non-sterile.

    AI/ML Overview

    Based on the provided text, the device in question is "Disposable Nitrile Powder-Free Examination Gloves," which are Class I medical devices used to prevent contamination between patients and examiners. The document describes non-clinical performance testing rather than a study involving human subjects or AI, as would be typical for software or imaging devices.

    Therefore, many of the requested categories for AI-based device studies (e.g., sample size for test set, number of experts, adjudication method, MRMC studies, standalone performance, training set details) are not applicable to this premarket notification for examination gloves.

    Here's the relevant information based on the provided document:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily based on established ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization) standards for examination gloves.

    Table of Acceptance Criteria and Reported Device Performance for Disposable Nitrile Powder-Free Examination Gloves:

    TestMethodology / StandardPurposeAcceptance CriteriaReported Performance/Results
    Biocompatibility
    Irritation & Skin SensitizationISO 10993-10:2010Assess possible contact hazards from chemicals, which may produce skin and mucosal irritation, eye irritation or skin sensitization.Skin Sensitization Test: Provided grades less than 1, otherwise sensitization.
    Skin Irritation Test: Primary irritation index: 0-0.4 (Negligible), 0.5-1.9 (Slight), 2-4.9 (Moderate), 5-8 (Severe).Skin Sensitization Test: All grades are 0. All animals survived with no abnormal signs observed.
    Skin Irritation Test: Primary irritation index is 0. Response categorized as negligible.
    In Vitro CytotoxicityISO 10993-5:2009Assess the in vitro cytotoxicity of medical devices.Viability reduced to
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1