Search Results
Found 2 results
510(k) Data Aggregation
(70 days)
Disposable Nitrile Examination Gloves (Tested for Use with Chemotherapy Drugs) - Blue/Black/Purple
A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.
The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue/black/purple colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in six sizes: S, M, L, XL. The subject device is non-sterile.
Here's a breakdown of the acceptance criteria and study information based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance:
Test Method | Purpose | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
ASTM D6319 | Physical Dimensions Test | Length(mm): S: ≥220; M/L/XL: ≥230. | |
Width(mm): S: 80±10; M: 95±10; L: 110±10; XL: 120±10. | |||
Thickness: Finger: ≥0.05; Palm: ≥0.05 | Length(mm): S: ≥220; M/L/XL: ≥230 (Pass) | ||
Width(mm): S: 84-87/Pass; M: 95-98/ Pass; L: 105-109/ Pass; XL:113-117/ Pass | |||
Thickness (mm): Finger: 0.10-0.12/Pass; Palm: 0.07-0.09/Pass | |||
ASTM D5151 | Watertightness Test for Detection of Holes | Meet the requirements of ASTM D5151 AQL 2.5 | 0/125/Pass |
ASTM D6124 | Powder Content | Meet the requirements of ASTM D6124 240 Min | |
Cyclophosphamide: >240 Min | |||
Dacarbazine: >240 Min | |||
Doxorubicin HCl: >240 Min | |||
Etoposide: >240 Min | |||
Fluorouracil: >240 Min | |||
Methotrexate: >240 Min | |||
Mitomycin C: >240 Min | |||
Paclitaxel: >240 Min | |||
ThioTepa: 58.8 Min | |||
Vincristine Sulfate: >240 Min | |||
Black Gloves: | |||
Carmustine: 27.3 Min | |||
Cisplatin: >240 Min | |||
Cyclophosphamide: >240 Min | |||
Dacarbazine: >240 Min | |||
Doxorubicin HCl: >240 Min | |||
Etoposide: >240 Min | |||
Fluorouracil: >240 Min | |||
Methotrexate: >240 Min | |||
Mitomycin C: >240 Min | |||
Paclitaxel: >240 Min | |||
ThioTepa: 77.8 Min | |||
Vincristine Sulfate: >240 Min | |||
Purple Gloves: | |||
Carmustine: 23.5 Min | |||
Cisplatin: >240 Min | |||
Cyclophosphamide: >240 Min | |||
Dacarbazine: >240 Min | |||
Doxorubicin HCl: >240 Min | |||
Etoposide: >240 Min | |||
Fluorouracil: >240 Min | |||
Methotrexate: >240 Min | |||
Mitomycin C: >240 Min | |||
Paclitaxel: >240 Min | |||
ThioTepa: 56.1 Min | |||
Vincristine Sulfate: >240 Min |
Note on Cytotoxicity: The table on page 11 explicitly states "Under conditions of the study, device extract is cytotoxic." for ISO 10993-5. This appears to contradict the general conclusion that the device is "as safe, as effective." However, the predicate comparison table on page 8 indicates that the predicate device also stated "ISO 10993-5 Under conditions of the study, device extract is cytotoxic," and the subject device has a "/" for this criterion without a definitive pass/fail statement in that specific comparison table. In the non-clinical testing summary, it lists ISO 10993-5 as one of the standards conducted. If the device genuinely exhibited cytotoxicity, it would typically be a concern. Given the overall clearance, there might be further explanation or context not fully detailed in this summary (e.g., the level of cytotoxicity was deemed acceptable, or the "is cytotoxic" statement might be an error in the summary table on page 11, or the nature of the cytotoxicity and its relevance to direct patient contact for a glove was evaluated and deemed non-critical).
2. Sample size used for the test set and the data provenance:
- Physical Dimensions (ASTM D6319): Not explicitly stated, but the results show ranges across sizes (S, M, L, XL).
- Watertightness Test (ASTM D5151): The results specify "0/125/Pass," indicating a sample size of 125 items (gloves) were tested, all passing (0 failures).
- Powder Content (ASTM D6124): Not explicitly stated.
- Physical Properties (ASTM D412): Not explicitly stated.
- Biocompatibility (ISO 10993-5, -10, -11): Not explicitly stated.
- Chemotherapy Drug Permeation (ASTM D6978-05): Not explicitly stated, however, the standard ASTM D6978 mandates three specimens be tested for each drug exposure for each material tested (so, 3 specimens per drug per color of glove). There are 12 drugs tested for 3 colors, so at least 3 x 12 x 3 = 108 specimens would have been tested as a minimum requirement for this standard.
The data provenance is from non-clinical bench testing conducted to relevant ASTM and ISO standards, indicating laboratory testing. The country of origin of the test data is not explicitly stated, but given the manufacturer is Jiangsu Bytech Medical Supplies Co.,Ltd. in China, it is likely the testing was performed either in China or by a certified lab contracted by the manufacturer. The data is reported retrospectively as part of the 510(k) submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This device is a physical product (gloves) and the tests conducted are bench-top performance tests and biocompatibility tests. Therefore, human experts are typically not used to establish a "ground truth" in the way they would for an AI diagnostic device (e.g., radiologists reviewing images). The ground truth is established by the methods and specifications defined in the referenced industry standards (ASTM, ISO). Laboratory technicians and chemists, qualified to perform these specific tests, would have conducted the studies.
4. Adjudication method for the test set:
Not applicable, as this involves objective laboratory measurements against established standards, not human interpretation requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This device is a medical glove, not an AI diagnostic or assistance tool. The evaluation focuses on its physical and chemical properties.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This is not an AI algorithm.
7. The type of ground truth used:
The ground truth is based on:
- Standardized Test Methods: Specifications and methodologies outlined in ASTM (e.g., D6319, D5151, D6124, D412, D6978) and ISO standards (e.g., 10993-5, 10993-10, 10993-11).
- Objective Measurements: These standards define how properties like dimensions, watertightness, powder content, tensile strength, elongation, and drug permeation breakthrough times are measured objectively.
- Pre-defined Acceptance Criteria: The pass/fail criteria for these measurements are often specified within the standards or derived from regulatory guidance for the device type.
8. The sample size for the training set:
Not applicable. This is a physical device, not a machine learning model, so there is no training set in this context.
9. How the ground truth for the training set was established:
Not applicable. There is no training set.
Ask a specific question about this device
(92 days)
Disposable Nitrile Examination Gloves (Tested for Use with Chemotherapy Drugs)
A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.
The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in four sizes: small, medium, large, and extra-large. The subject device is non-sterile.
Acceptance Criteria and Device Performance for Disposable Nitrile Examination Gloves
This document outlines the acceptance criteria and details the studies conducted to demonstrate that the "Disposable Nitrile Examination Gloves (Tested for Use with Chemotherapy Drugs)" meet these criteria.
1. Table of Acceptance Criteria and Reported Device Performance
Test Method | Purpose | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
ASTM D6319 | Physical Dimensions Test | Length (mm): S: ≥220; M/L/XL: ≥230 | |
Width (mm): S: 80±10; M: 95±10; L: 110±10; XL: 120±10 | |||
Thickness (mm): Finger: ≥0.05; Palm: ≥0.05 | Length (mm): > 240/Pass (for all sizes) | ||
Width (mm): S: 88-89 /Pass; M: 94-97/ Pass; L: 103-107/Pass; XL: 114-116/Pass | |||
Thickness (mm): Finger: 0.10-0.11/Pass; Palm: 0.07-0.08/Pass | |||
ASTM D5151 | Watertightness Test for Detection of Holes | Meet the requirements of ASTM D5151 AQL 2.5 | 0/125/Pass |
ASTM D6124 | Powder Content | Meet the requirements of ASTM D6124 20 Minutes (Predicate's acceptance criteria was >20 minutes, though the predicate showed 11.8 min and 31.6 min depending on color - subject device had 1.9 min, leading to a warning for this drug). | |
Cisplatin 1.0 mg/ml: > 20 Minutes | |||
Cyclophosphamide (Cytoxan) 20.0 mg/ml: > 240 Minutes | |||
Dacarbazine (DTIC) 10.0 mg/ml: > 240 Minutes | |||
Doxorubicin HCl 2.0 mg/ml: > 240 Minutes | |||
Etoposide 20.0 mg/ml: > 240 Minutes | |||
Fluorouracil 50.0 mg/ml: > 240 Minutes | |||
Paclitaxel 6.0 mg/ml: > 20 Minutes (Predicate's acceptance criteria was >20 minutes, though the predicate showed >240 min). | |||
ThioTepa 10.0 mg/ml: > 20 Minutes (Predicate's acceptance criteria was >20 minutes though the subject device had 11.3 min, leading to a warning for this drug). | Carmustine (BCNU) 3.3 mg/ml: 1.9 Minutes (Warning: Do not use) | ||
Cisplatin 1.0 mg/ml: > 240 Minutes | |||
Cyclophosphamide (Cytoxan) 20.0 mg/ml: > 240 Minutes | |||
Dacarbazine 10.0 mg/ml: > 240 Minutes | |||
Doxorubicin HCL 2.0 mg/ml: > 240 Minutes | |||
Etoposide 20.0 mg/ml: > 240 Minutes | |||
Fluorouracil 50.0 mg/ml: > 240 Minutes | |||
Paclitaxel 6.0 mg/ml: > 240 Minutes | |||
ThioTepa 10.0 mg/ml: 11.3 Minutes (Warning: Do not use) |
Note: For Carmustine (BCNU) and ThioTepa, the device performance did not meet the implicitly accepted longer breakthrough times of the predicate device for safe use, and as such, a warning was issued for non-use with these specific drugs. For other drugs, the device demonstrated comparable or superior performance to the predicate.
2. Sample Size and Data Provenance
The document does not explicitly state the specific sample sizes used for each test. However, for the Watertightness Test (ASTM D5151), the result "0/125/Pass" indicates a sample size of 125 gloves were tested, and none failed.
The data provenance is implied to be prospective bench testing conducted specifically for the submission of this device, presumably within China, given the manufacturer's and correspondent's addresses. There is no information provided regarding country of origin of data beyond the manufacturer's location.
3. Number of Experts and Qualifications for Ground Truth
This section is not applicable as the studies are primarily bench testing of physical and chemical properties of a medical device (gloves). The "ground truth" for these tests is established by published ASTM and ISO standards and their specified methodologies, not by expert consensus on diagnostic images or clinical outcomes.
4. Adjudication Method for the Test Set
This is not applicable as the studies involve objective physical and chemical testing against established standards, not interpretation of data requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
This is not applicable. The device is a medical glove, and the evaluation does not involve diagnostic imaging or human-in-the-loop performance with AI.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
This is not applicable. The device is a medical glove, not an algorithm, and its performance is assessed directly through physical and chemical testing.
7. Type of Ground Truth Used
The ground truth used for these studies is based on:
- Established ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization) standards: These standards define the test methodologies, performance criteria (e.g., minimum tensile strength, maximum powder content, minimum breakthrough times for chemotherapy drugs, irritation/sensitization levels), and acceptable limits.
- Objective laboratory measurements: Physical dimensions, tensile strength, elongation, watertightness, powder content, and chemical permeation breakthrough times were measured directly according to the specified standard test methods.
8. Sample Size for the Training Set
This is not applicable. The device is a physical product (medical gloves), not an AI algorithm that requires a training set. The performance is based on direct testing of the product itself.
9. How the Ground Truth for the Training Set Was Established
This is not applicable as there is no training set for this type of device. The "ground truth" for the performance evaluation is established by the relevant ASTM and ISO standards as mentioned in point 7.
Ask a specific question about this device
Page 1 of 1