Search Results
Found 3 results
510(k) Data Aggregation
(89 days)
The Disposable Medical Nitrile gloves intended to be worn on the hands of examiners to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.
The proposed devices are powder-free nitrile examination gloves, provided as non-sterile and disposable devices. The proposed devices are mainly made from nitrile and there are four sizes, including small (S), medium (M), large (L), X-large (XL) for optional. The gloves are composed of Butadiene-Acrylonitrile Copolymers (CAS No.9003-18-3) and blue colorant Pigment Blue (CAS No.147-14-8). The examination glove is a smooth surface and a rolled rim at the cuff edge.
The Medical nitrile examination gloves meet the specifications in ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application.
The document provided is a 510(k) summary for Disposable Medical Nitrile gloves (K213851) and thus describes non-clinical performance testing rather than a study for an AI-powered device. Therefore, many of the requested categories are not applicable.
Here's the information that can be extracted from the provided text, focused on the acceptance criteria and the tests performed for the Disposable Medical Nitrile gloves:
1. Table of Acceptance Criteria and Reported Device Performance
| Test Method (Reference Standard) | Acceptance Criteria | Reported Device Performance (All three lots: 20210825B, 20210827A, 20210830A) | Conclusion |
|---|---|---|---|
| ASMT D6319-19 - Physical Dimensions (Width) | S: 80±10 mm; M: 95±10 mm; L: 110±10 mm; XL: 120±10 mm | S: 83-86mm (Lot B), 83-85mm (Lot A), 83-85mm (Lot A); M: 91-92mm (Lot B), 91-93mm (Lot A), 90-92mm (Lot A); L: 104-106mm (Lot B), 104-106mm (Lot A), 104-106mm (Lot A); XL: 112-114mm (Lot B), 111-113mm (Lot A), 112-115mm (Lot A) | Passed |
| ASMT D6319-19 - Physical Dimensions (Length) | S: ≥220 mm; M/L/XL: ≥230 mm | S: 234-236mm; M: 235-237mm; L: 237-239mm (Lot B,A), 237-240mm (Lot A); XL: 237-239mm | Passed |
| ASMT D6319-19 - Physical Dimensions (Thickness) | Finger: ≥0.05mm; Palm: ≥0.05mm | Finger: S/M: Pass at 0.11mm (Lot B, A), S/M/L: Pass at 0.11mm (Lot A), L/XL: Pass at 0.12mm (Lot B), M/L/XL: Pass at 0.12mm (Lot A), XL: Pass at 0.12mm (Lot A); Palm: S/M/L/XL: Pass at 0.08mm | Passed |
| ASMT D6319-19 - Physical Properties (Before Aging) | Tensile Strength: ≥14Mpa; Ultimate Elongation: ≥500% | Tensile Strength: ≥14Mpa; Ultimate Elongation: ≥500% | Passed |
| ASMT D6319-19 - Physical Properties (After Aging) | Tensile Strength: ≥14Mpa; Ultimate Elongation: ≥400% | Tensile Strength: ≥14Mpa; Ultimate Elongation: ≥400% | Passed |
| ASTM D6319-19 (ASTM D5151-11) - Freedom from Holes | AQL 2.5 | Pass at AQL 2.5 | Passed |
| ASMT D6319-19 (ASTM D6124-11) - Residual Powder | < 2.0 mg/glove | Lot 20210825B: Pass at 0.22 mg/glove; Lot 20210827A: Pass at 0.19 mg/glove; Lot 20210830A: Pass at 0.20 mg/glove | Passed |
| ISO 10993-10 - Skin Irritation | Under the conditions of study not an irritation | Under the conditions of study not an irritation | Passed |
| ISO 10993-10 - Skin Sensitization | Under the conditions of the study not a sensitization | Under the conditions of the study not a sensitization | Passed |
| ISO 10993-11:2017 - Acute Systemic Toxicity | Under the conditions of the study no systemic toxicity | Under the condition of acute systemic toxicity test, the test article did not show acute systemic toxicity in vivo. | Passed |
2. Sample Size Used for the Test Set and Data Provenance
The document refers to "three lots" (Lot 20210825B, Lot 20210827A, Lot 20210830A) for the performance testing. However, the specific sample sizes (number of gloves) tested within each lot for each test are not explicitly stated in the provided text. The data provenance (e.g., country of origin, retrospective or prospective) is also not specified for these tests. This is a non-clinical device (gloves), so data provenance details like for AI/software would not typically apply.
3. Number of Experts Used to Establish Ground Truth and Qualifications
This information is not applicable as the device is Disposable Medical Nitrile gloves, not an AI or diagnostic imaging device that requires expert ground truth establishment. The performance is based on physical and biological property testing against recognized standards.
4. Adjudication Method
This information is not applicable for the type of product and testing described.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, Effect Size of Human Reader Improvement
This information is not applicable as the device is Disposable Medical Nitrile gloves, not an AI or diagnostic imaging device. No MRMC study was mentioned.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done
This information is not applicable as the device is Disposable Medical Nitrile gloves, not an AI algorithm.
7. The Type of Ground Truth Used
The "ground truth" for this device's performance is established by conformance to recognized international standards such as ASMT D6319-19, ASTM D5151-11, ASTM D6124-11, ISO 10993-10, and ISO 10993-11:2017. These standards define the acceptable physical dimensions, tensile strength, elongation, freedom from holes, residual powder, and biocompatibility (skin irritation, sensitization, acute systemic toxicity).
8. The Sample Size for the Training Set
This information is not applicable as the device is Disposable Medical Nitrile gloves and does not involve AI or machine learning models that require training sets.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable for the same reason as above (no training set for this type of device).
Ask a specific question about this device
(92 days)
Disposable Medical Nitrile Examination Gloves are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
Disposable Medical Nitrile Examination Gloves
This document pertains to the 510(k) premarket notification for Disposable Medical Nitrile Examination Gloves. The information provided focuses on the regulatory clearance process for this medical device.
Crucially, this document does not describe an AI/ML-based medical device. Therefore, it does not contain the information needed to answer your specific questions regarding acceptance criteria, study methodologies (like MRMC, standalone performance), ground truth establishment, or sample sizes related to AI model evaluation.
The provided text from the FDA 510(k) clearance letter confirms that the device (disposable medical gloves) is substantially equivalent to legally marketed predicate devices. The "Indications for Use" section simply describes the intended use of the gloves (to prevent contamination).
To answer your questions, the input document would need to be focused on the clinical validation or performance study of an AI/ML medical device.
Therefore, I cannot extract the requested information from the provided text because it describes a Class I medical device (gloves) and not an AI/ML medical device.
Ask a specific question about this device
(75 days)
Disposable Medical Nitrile Gloves are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
The subject device in this 510(k) Notification is Disposable Medical Nitrile Gloves. The subject device is a patient examination glove made from nitrile compound, blue color, powder free and nonsterile (Per 21 CFR 880.6250, Class I). The device meets the specifications in ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application.
The provided document describes the acceptance criteria and study results for "Disposable Medical Nitrile Gloves." Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
| Test Method | Purpose | Acceptance Criteria | Reported Device Performance |
|---|---|---|---|
| ASTM D6319-19 - Physical Dimensions Test | To determine the width, length, and thickness of the gloves | Width: 80 ± 10 mm (small)95 ± 10 mm (medium)110 ± 10 mm (large)Length: (Minimum)220 mm (small)230 mm (med, large)Thickness: (Minimum)Finger - 0.05 mm (all sizes)Palm - 0.05 mm (all sizes) | Passed |
| ASTM D6319-19 - Physical Requirements Test | To determine the tensile strength and ultimate elongation before and after acceleration aging | Before Aging:Tensile Strength: 14 MPa (Min)Ultimate Elongation: 500% (Min)After Aging:Tensile Strength: 14 MPa (Min)Ultimate Elongation: 400% (Min) | Passed |
| ASTM D6319-19 (ASTM D5151-11) - Detection of Holes in Medical Gloves | To determine the holes in the gloves | (Mean) AQL 2.5 | Passed |
| ASTM D6319-19 (ASTM D6124-11) - Residual Powder on Medical Gloves | To determine the residual powder in the gloves | ≤ 2.0 mg/glove | 0.1 mg/glove, Passed |
| ISO 10993-10 - Skin Irritation | To determine the potential of the material to produce skin irritation in rabbits | Under the condition of study not an irritant | Under the condition of study not an irritant |
| ISO 10993-10 - Skin Sensitization | To determine the skin sensitization potential of the material in guinea pigs | Under the conditions of the study not a sensitizer. | Under the conditions of the study not a sensitizer. |
| ISO 10993-5 - In Vitro Cytotoxicity | To evaluate the in vitro cytotoxic potential of the test item extracts in L-929 mouse fibroblasts | Under the conditions of study, no more than grade 2 cytotoxic reaction | Mild (Grade 2) cytotoxicity reaction observed. |
| ISO 10993-11:2017 - Acute Systemic Toxicity (in ICR Mice) | The test item was evaluated for acute systemic toxicity in ICR Mice | Under the conditions of the study no systemic toxicity | Under the conditions of the study no systemic toxicity |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes for each specific test (e.g., number of gloves tested for physical dimensions, number of animals for biocompatibility). It generally refers to the standards (e.g., ASTM D6319-19, ISO 10993 series) which inherently define sample size requirements for those tests. The data provenance is from tests conducted on the subject device, "Disposable Medical Nitrile Gloves," by Jiangsu Health-Peace Medical Technology Co. Ltd. It is retrospective in the sense that the tests were completed and reported to demonstrate compliance for premarket notification. The country of origin for the manufacturing and testing is not explicitly stated as separate from the company's address in China, but the submitting company is based in China.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. This device is a medical glove, and the "ground truth" for its performance is established through standardized laboratory testing (physical, chemical, and biological tests) rather than expert interpretation of medical images or clinical outcomes. The performance is objectively measured against specified criteria in established standards like ASTM and ISO.
4. Adjudication Method for the Test Set
Not applicable. As described above, the testing involves objective measurements against predefined criteria, not subjective interpretation requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. This type of study is relevant for diagnostic or interpretive devices where human readers (e.g., radiologists) interact with the device's output. Medical gloves are not diagnostic devices, and there is no "human reader" component in their performance evaluation in this context.
6. Standalone Performance Study
Yes, standalone performance studies were done. The "PERFORMANCE DATA" section and "BIO-COMPATIBILITY DATA" section detail the results of various tests conducted on the "Disposable Medical Nitrile Gloves" alone. These tests directly measure the device's characteristics against the specified acceptance criteria of recognized standards without human intervention in the loop for assessment.
7. Type of Ground Truth Used
The ground truth used for these tests is defined by established International Standards and Specifications. Specifically:
- ASTM D6319-19: Standard Specification for Nitrile Examination Gloves for Medical Application.
- ASTM D5151-11: Standard Test Method for Detection of Holes in Medical Gloves.
- ASTM D6124-11: Standard Test Method for Residual Powder on Medical Gloves.
- ISO 10993-10: Biological evaluation of medical devices — Part 10: Tests for skin irritation and skin sensitization.
- ISO 10993-5: Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity.
- ISO 10993-11:2017: Biological evaluation of medical devices — Part 11: Tests for acute systemic toxicity.
These standards provide objective, measurable criteria that serve as the "ground truth" for the device's performance and safety.
8. Sample Size for the Training Set
Not applicable. This is a conventional medical device (physical product) and not an AI/machine learning algorithm. Therefore, there is no "training set" in the context of AI model development.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this device.
Ask a specific question about this device
Page 1 of 1