Search Results
Found 1 results
510(k) Data Aggregation
(316 days)
DUAFIT INTERGPHALANGEAL IMPLANT, PEEK-FUSE INTERPHALANGEAL IMPLANT
The DUAFIT® Interphalangeal implant is intended for fixation of proximal interphalangeal joint arthrodesis of the lesser toes. Examples include: Rigid or semi-rigid deformity of the PIP joint Revision of failed arthroplasty or arthrodesis 2nd toe shortening.
The DUAFIT® interphalangeal implant is an intramedullary implant, designed to act as a bone fastener for proximal interphalangeal arthrodesis of the lesser rays. Design is a combination of a proximal taper with barbs and a distal blade for enhanced stabilization. It is made available in multiple lengths and diameters, and in 3 different angles. The implant is manufactured from PEEK-OPTIMA®, polymer from Invibio®, and is designed for single use only.
The provided text describes a medical device (DUAFIT® interphalangeal implant) and its regulatory submission. It focuses on demonstrating substantial equivalence to predicate devices rather than directly presenting acceptance criteria for a new AI/ML device or a comparative effectiveness study with human readers.
Therefore, the requested information elements related to AI/ML device performance, ground truth establishment, expert adjudication, and comparative effectiveness studies are not applicable or cannot be extracted from the given document.
The document primarily covers the engineering and regulatory aspects of a physical medical implant.
However, I can extract information related to the device's performance demonstration.
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Device is substantially equivalent to predicate devices for its intended use and design. | The technological characteristics of the DUAFIT® interphalangeal implant are the same as the characteristics of predicate devices in terms of intended use and design, including features like insertion into bone, tapered barded design, blade design, cannulated design, straight/angled design, PEEK material, and equivalent size range. |
Mechanical properties are at least equivalent to the predicate device (Merete MetaToe). | Results from testing confirmed that the DUAFIT® interphalangeal implant is at least equivalent to the Merete MetaToe EndoSorb (K100414) in terms of mechanical properties. |
Performance in specific mechanical tests (static and cyclic four-point bending) meets criteria. | The DUAFIT® interphalangeal implant met the acceptance criteria for the static four-point bending test and the cyclic four-point bending test. The implants performed as expected for each test. |
Biocompatibility (implied from material and predicate equivalence) | The DUAFIT® interphalangeal implant is manufactured from PEEK-OPTIMA®, polymer from Invibio®, as per ASTM 2026, a material used in predicate devices (e.g., Smith and Nephew Bioraptor 2.3 PK suture anchor (K071586), Parcus PEEK CF push-in suture anchor (K102326), and Synchro Toegrip (K133477)), suggesting established biocompatibility. |
Sterilization effective (implied) | The DUAFIT® interphalangeal implant is supplied sterile, using gamma irradiation. |
2. Sample size used for the test set and the data provenance
The document does not specify the sample size for the mechanical tests (static and cyclic four-point bending). Data provenance is not explicitly stated beyond "testing performed by the test lab," but it is implicitly from product-specific laboratory testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This device is a physical implant, not an AI/ML diagnostic device requiring expert interpretation for ground truth. Performance was demonstrated through engineering bench testing.
4. Adjudication method for the test set
Not applicable. Performance was determined through objective mechanical testing, not by human adjudication of interpreted data.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI-assisted device.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical medical implant, not an algorithm.
7. The type of ground truth used
The ground truth for the device's performance was established through engineering bench testing (static four-point bending test and cyclic four-point bending test) against established mechanical performance criteria, and comparison to predicate devices' mechanical properties.
8. The sample size for the training set
Not applicable. This is a physical medical implant, not a machine learning model that requires a training set.
9. How the ground truth for the training set was established
Not applicable. This is a physical medical implant, not a machine learning model.
Ask a specific question about this device
Page 1 of 1