(90 days)
The TOEGRIP® is intended for toes for general use in skeletally mature individuals undergoing surgery limited to interdigital fusion.
The TOEGRIP® device is intended to be permanently implanted without any other additional device as an intramedullary bone fastener device for toes or fingers.
The TOEGRIP® device is indicated for small bone reconstruction limited to interdigital fusion of toes in the following cases, listed in random order:
- . Rigid PIP joints deformities
- . Rigid hammertoes deformities
- . Claw toes deformities (PIP and DIP joints)
- Revision hammertoes surgeries .
- . Shortening osteotomies of the proximal phalanx
The TOEGRIP® device consists of a monobloc implant with three flexible prongs inserted intramedullary into each part of phalanx The prongs are linked by a T-shaped beam. The fixation concept is based on a press-fit contact due to its tapered shape and the sharp macrostructures once impacted.
The TOEGRIP® device is intended for single use only, and is available in a range of 5 sizes with 3 possible degrees: 0°, 10° or 20°.
The TOEGRIP® device is manufactured with material according to the ISO-10993, ZENIVA® PEEK.
The feature design of the TOEGRIP® is substantially equivalent to the design features of other devices previously cleared for market.
The Synchro-Medical Toe Grip Device underwent nonclinical testing to demonstrate its substantial equivalence to predicate devices.
1. Acceptance Criteria and Device Performance:
The document states that the Synchro-Medical Toe Grip Device met the acceptance criteria for the tests performed. However, the specific quantitative acceptance criteria or the reported device performance metrics are not detailed in the provided text.
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Not explicitly stated (e.g., minimum bending strength, fatigue life cycles) | "The results of these studies showed that the Synchro-Medical Toe Grip Device met the acceptance criteria." (Specific numerical results are not provided). |
2. Sample Size and Data Provenance for Test Set:
- Sample Size: The document does not explicitly state the sample size used for the nonclinical tests (Static Four-point Bending Test and Dynamic Four-point Bending Test).
- Data Provenance: The tests were nonclinical (bench testing) and therefore do not have data provenance related to country of origin or retrospective/prospective human data.
3. Number of Experts and Qualifications for Ground Truth:
Not applicable. The study involved nonclinical mechanical testing, not human expert assessment for ground truth.
4. Adjudication Method for Test Set:
Not applicable. The study involved nonclinical mechanical testing, not human expert assessment requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No. The document explicitly states, "No clinical tests were performed." Therefore, an MRMC comparative effectiveness study was not conducted.
6. Standalone Performance Study:
Yes, a standalone study (nonclinical mechanical tests) was performed for the algorithm/device. The device's performance was evaluated through:
- Static Four-point Bending Test
- Dynamic Four-point Bending Test
7. Type of Ground Truth Used:
The "ground truth" for these nonclinical tests would be the established engineering standards and specifications for bone fixation devices, against which the device's mechanical properties (e.g., bending strength, fatigue resistance) were measured.
8. Sample Size for Training Set:
Not applicable. This device is a physical implant, not a software algorithm that requires a training set in the context of machine learning.
9. How Ground Truth for Training Set was Established:
Not applicable. This device is a physical implant, not a software algorithm that requires ground truth for a training set.
{0}------------------------------------------------
K133477
Traditional 510(k) Summary
FEB 1 0 2014
Synchro-Medical Toe Grip Device
| Submitter: | Synchro Medical21 rue des MerisiersFR-68920 Wettolsheim les Erlen |
|---|---|
| Contact Person | Danilo CampaniPresidentPhone: +33(0)3 89 23 75 64Fax:+33(0)3 89 27 14 52Email: dcampani@synchro-medical.com |
| Date Prepared | 2/10/14 |
| Device Class | Class II |
| Trade Name | TOEGRIP®. |
|---|---|
| Common Name | Synchro-Medical Toe Grip Device |
| Classification Nameand Number | Smooth or threaded metallic bone fixation fastener21 CFR 888.3040 |
| Classification Panel: | Orthopedic |
| Product Code | HTY |
| Predicate Devices | Inion OTPS Biodegradable Pin (K031712), Memometal Inc™Smart Toe & X-Fuse (K070598) & New Deal K-Wire(K022599) |
| Previous Submissions | There are no previous submissions |
{1}------------------------------------------------
.
| Device Description | The TOEGRIP® device consists of a monobloc implant withthree flexible prongs inserted intramedullary into each part ofphalanx The prongs are linked by a T-shaped beam. Thefixation concept is based on a press-fit contact due to its taperedshape and the sharp macrostructures once impacted.The TOEGRIP® device is intended for single use only, and isavailable in a range of 5 sizes with 3 possible degrees: 0°, 10°or 20°.The TOEGRIP® device is manufactured with materialaccording to the ISO-10993, ZENIVA® PEEK.The feature design of the TOEGRIP® is substantiallyequivalent to the design features of other devices previouslycleared for market. |
|---|---|
| -------------------- | ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
| Intended Use | The TOEGRIP® is intended for toes for general use inskeletally mature individuals undergoing surgery limited tointerdigital fusion.The TOEGRIP® device is intended to be permanentlyimplanted without any other additional device as anintramedullary bone fastener device for toes or fingers.The TOEGRIP® device is indicated for small bonereconstruction limited to interdigital fusion of toes in thefollowing cases, listed in random order:• Rigid PIP joints deformities• Rigid hammertoes deformities• Claw toes deformities (PIP and DIP joints)• Revision hammertoes surgeries• Shortening osteotomies of the proximal phalanx |
|---|---|
| Materials: | The implant is manufactured from ASTM2026 implant gradePolyetheretherketone (PEEK) Solvay Zeniva ZA-500. |
| August March Auction Address Andrews Andand the commend of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comment | |
|---|---|
{2}------------------------------------------------
| Statement of | The purpose of this submission is to obtain market clearance for the |
|---|---|
| Technological | proposed the Synchro-Medical Toe Grip Device. Synchro-Medical |
| Comparison | Toe Grip Device and its predicate devices have the similar |
| indications for use, have a similar functionality. Both devices are | |
| manufactured using materials with a long history of use in | |
| orthopaedic implants. |
.
| Nonclinical TestSummary | The following tests were performed to demonstrate that theSynchro-Medical Toe Grip Device is substantially equivalent toother predicate devices.Static Four-point Bending Test Dynamic Four-point Bending Test The results of these studies showed that the Synchro-MedicalToe Grip Device met the acceptance criteria. |
|---|---|
| Clinical Test Summary | No clinical tests were performed. |
| Conclusion | |
|---|---|
| The Synchro-Medical Toe Grip Device is substantiallyequivalent to its predicate devices. This conclusion is basedupon the fact the Synchro-Medical Toe Grip Device and itspredicate devices have the same indications for use and have asimilar design. |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 10, 2014
Synchro Medical % Mark F. Schenk Consulting Mr. Mark F. Schenk 505 Berks Place West Lawn, Pennsylvania 19609
Re: K133477
Trade/Device Name: Synchro-Medical Toe Grip Device Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HTY Dated: November 12, 2013 Received: November 12, 2013
Dear Mr. Schenk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Mark F. Schenk
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Vincent J原因evlin -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number K133477 Device Name: Synchro-Medical Toe Grip Device Indications for Use:
The TOEGRIP® is intended for toes for general use in skeletally mature individuals undergoing surgery limited to interdigital fusion.
The TOEGRIP® device is intended to be permanently implanted without any other additional device as an intramedullary bone fastener device for toes or fingers.
The TOEGRIP® device is indicated for small bone reconstruction limited to interdigital fusion of toes in the following cases, listed in random order:
- . Rigid PIP joints deformities
- . Rigid hammertoes deformities
- . Claw toes deformities (PIP and DIP joints)
- Revision hammertoes surgeries .
- . Shortening osteotomies of the proximal phalanx
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth 詞版ank -S
Division of Orthopedic Devices
§ 888.3040 Smooth or threaded metallic bone fixation fastener.
(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.