Search Results
Found 2 results
510(k) Data Aggregation
(189 days)
Carescape B850, E-musb
The CARESCAPE B850 is a multi-parameter patient monitor intended for use in multiple areas within a professional healthcare facility.
The CARESCAPE B850 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE B850 is indicated for monitoring of:
- · hemodynamic (including ECG, ST segment, arthythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, regional oxygen saturation, total hemoglobin concentration, cardiac output (thermodilution and pulse contour), temperature, mixed venous oxygen saturation, and central venous oxygen saturation),
- · respiratory (impedance respiration, airway gases (CO2, O2, N2O, and anesthetic agents), spirometry, gas exchange), and
- · neurophysiological status (including electroencephalography, Entropy, Bispectral Index (BIS), and neuromuscular transmission).
The CARESCAPE B850 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network.
The CARESCAPE B850 is able to detect and generate alarms for ECG arrhythmias: atrial fibrillation, accelerated ventricular rhythm, asystole, bigeminy, bradycardia, ventricular couplet, missing beat, multifocal premature ventricular contractions (PVCs), pause, R on T, supra ventricular tachycardia, trigeminy, ventricular bradycardia, ventricular fibrillation/ventricular tachycardia, ventricular tachycardia, and VT>2. The CARESCAPE B850 also shows alarms from other ECG sources.
The CARESCAPE B850 also provides other alarms, trends, snapshots and calculations, and can be connected to displays, printers and recording devices.
The CARESCAPE B850 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
Contraindications for using the monitor
The CARESCAPE B850 is not intended for use in a controlled MR environment.
CARESCAPE B850 is a new version of a modular multi- parameter patient monitoring system. The monitor includes a separate 19-inch touchscreen display, the central processing unit (also called CPU), the CARESCAPE Software, and a module frame F5 or F7. CARESCAPE B850 is equipped with the ePort interface that supports use of PDM or CARESCAPE ONE patient data acquisition modules for patient monitoring. In addition, the F5 module frame has five module slots, and the F7 module frame has seven module slots where patient data acquisition modules (E-Module type), can be connected to perform patient monitoring. The CARESCAPE B850 includes features and subsystems that are optional or configurable.
This looks like a 510(k) summary for the GE Healthcare CARESCAPE B850 patient monitor. I will extract information related to the acceptance criteria and study that proves the device meets them.
Based on the provided text, the CARESCAPE B850 is a multi-parameter patient monitor. The 510(k) submission is for a new version with updated software and minor hardware modifications. The submission refers to a primary predicate device, also named CARESCAPE B850 (K191323), and additional predicate/reference devices for specific parameters (INVOS PM7100 and MASIMO RADICAL Y PULSE CO-OXIMETER).
The key takeaway is that the device did not require clinical studies to support substantial equivalence because it is a modified version of an already cleared device and incorporates previously cleared parameters. Therefore, the "study that proves the device meets the acceptance criteria" primarily refers to non-clinical bench testing.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of acceptance criteria with corresponding performance metrics for the new CARESCAPE B850 compared to a specific threshold. Instead, it relies on the concept of substantial equivalence to a predicate device.
The "acceptance criteria" are implied to be that the performance of the new device is "as safe, as effective, and the performance to be substantially equivalent to the predicate device." The reported "device performance" is primarily that it passed various non-clinical tests.
Implied Acceptance Criteria (based on substantial equivalence concept):
Performance Aspect | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Software | Meets specifications and validated as per design requirements. | Bench testing related to software design, development, verification, validation and traceability was conducted. |
Hardware | Meets specifications, including safety and compatibility. | Bench testing related to electromagnetic compatibility, electrical safety, environmental, and usability was conducted. |
Overall Safety & Effectiveness | As safe and effective as the predicate device (K191323). | The device is considered as safe, as effective, and its performance is substantially equivalent to the predicate device. |
Note: The document states that the fundamental function and operation of the proposed CARESCAPE B850 monitor are unchanged compared to its predicate (K191323), except for the addition of an E-musb Interface module and the capability to display previously cleared hemodynamic parameters from OEM devices (regional oxygen saturation and total hemoglobin concentration).
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not explicitly stated. The document refers to "bench testing related to software, hardware and performance." This typically involves testing prototypes or production units, but a "sample size" in the context of patient data is not applicable here as no clinical studies were performed for this submission.
- Data Provenance: Not applicable, as no external data (e.g., patient data from a specific country, retrospective or prospective) was used for this 510(k) submission to demonstrate substantial equivalence. The testing was internal bench testing.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Number of Experts: Not applicable. For bench testing of hardware and software, "ground truth" is typically established by engineering specifications, validated test protocols, and adherence to consensus standards, rather than expert clinical consensus on patient data.
- Qualifications of Experts: Not applicable.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable. This concept applies to clinical studies where discrepancies in observations or diagnoses need to be resolved. For bench testing, test results are typically compared against predefined specifications.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
- MRMC Study: No. The device is a multi-parameter patient monitor, not an AI-assisted diagnostic tool that would typically involve human readers. The new version mostly focuses on software updates, minor hardware changes, and display of previously cleared parameters from other OEM devices.
6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Standalone Performance: The core functionality of the device (e.g., ECG, arrhythmia detection, various physiological measurements) operates in a "standalone" fashion in that the algorithms process patient data collected by the sensors. The document doesn't detail specific "algorithm-only" performance metrics as would be seen for a novel AI algorithm. Instead, it relies on the previous clearance of the predicate device and the fact that the algorithms (like EK-Pro arrhythmia detection algorithm V14) are identical. The newly added parameters (regional oxygen saturation and total hemoglobin concentration) are sourced from OEM devices that would have their own standalone performance data from their original clearances.
7. The Type of Ground Truth Used
- Type of Ground Truth: For the non-clinical bench testing, the ground truth would be the engineering specifications of the device and adherence to relevant consensus standards (e.g., for electromagnetic compatibility, electrical safety, environmental performance). For the physiological parameters, the "ground truth" for the algorithms (e.g., arrhythmia detection) was established during the development and clearance of the predicate device (K191323) and the OEM devices for rSO2 and SpHb.
8. The Sample Size for the Training Set
- Sample Size for Training Set: Not applicable. As this is not an AI/ML device that requires a distinct "training set" for model development for this 510(k) submission, this information is not relevant here. The update involves existing algorithms and integration of existing cleared parameters.
9. How the Ground Truth for the Training Set Was Established
- How Ground Truth for Training Set Was Established: Not applicable, for the same reason as point 8.
Ask a specific question about this device
(259 days)
Carescape B850
The CARESCAPE B850 is a multi-parameter patient montor intended for use in multiple areas within a professional healthcare facility.
The CARESCAPE B850 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE B850 is indicated for monitoring of:
· hemodynamic (including ECC, ST segment, arthythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, cardiac output (thermodilution and pulse contour), temperature, mixed venous oxygen saturation, and central venous oxygen saturation),
· respiratory (impedance respiration, airway gases (CO2, O2, N2O, and anesthetic agents), spirometry, gas exchange), and
· neurophysiological status (including electroencephalography, Entropy, Bispectral Index (BIS), and neuromuscular transmission).
The CARESCAPE B850 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network.
The CARESCAPE B850 is able to detect and generate alarms for ECG arrhythmias: atrial fibrillation, accelerated ventricular rhythm, asystole, bigeminy, bradycardia, ventricular couplet, missing beat, multifocal premature ventricular contractions (PVCs), pause, R on T, supra ventricular tachycardia, trigeminy, ventricular bradycardia, ventricular fibrillation/ventricular tachycardia, ventricular tachycardia, and VT>2. The CARESCAPE B850 also shows alarms from other ECG sources.
The CARESCAPE B850 also provides other alarms, trends, snapshots and calculations, and can be connected to displays, printers and recording devices.
The CARESCAPE B850 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
Contraindications for using the monitor
The CARESCAPE B850 is not intended for use in a controlled MR environment.
The CARESCAPE B850 is a new version of a modular multi-parameter patient monitoring system. The monitor includes a separate 19-inch touchscreen display, the central processing unit (also called CPU) and the module frame F5 or F7. CARESCAPE B850 V3 is equipped with an "ePort" interface that supports use of PDM or CARESCAPE ONE patient data acquisition modules for patient monitoring. In addition, the F5 module frame has five module slots, and the F7 module frame has seven module slots where patient data acquisition modules (E-Module type), can be connected to perform patient monitoring. The CARESCAPE B850 V3 includes features and subsystems that are optional or configurable.
The provided document, [K191323](https://510k.innolitics.com/search/K191323)
, is a 510(k) premarket notification for the GE Healthcare CARESCAPE B850 patient monitor. It describes the device and compares it to a predicate device ([K131414](https://510k.innolitics.com/search/K131414) CARESCAPE Monitor B850
). This document outlines non-clinical testing and explicitly states that clinical studies were not required.
Therefore, I cannot fully answer your request for acceptance criteria and a study proving the device meets those criteria from this specific document, as it indicates clinical studies were not performed. The document focuses on demonstrating substantial equivalence through non-clinical bench testing, including software, hardware, and performance tests, and usability studies.
However, I can extract the information related to the device's performance and the types of tests conducted, even if they aren't presented as a structured "acceptance criteria" table with "reported device performance" against them in a clinical study context.
Here's a breakdown of what can be extracted and what cannot:
What Can Be Inferred/Extracted from the Document:
-
Device Performance (as demonstrated by non-clinical testing):
- The CARESCAPE B850's hardware, software, and overall performance met specifications through bench testing.
- The device demonstrated electromagnetic compatibility, electrical safety, environmental resilience, and usability.
- Software testing covered design, development, verification, validation, and traceability.
- The device addressed patient safety, security, and privacy risks, including cybersecurity measures.
- It can withstand network storm conditions without rebooting.
-
Test Set Description (for non-clinical testing):
- Data Provenance: Not explicitly stated for specific datasets, but testing was conducted on the aformentioned device, manufactured by GE Healthcare Finland Oy. This implies internal testing. The nature of "bench testing" suggests controlled environments rather than real-world patient data for performance validation against specific clinical thresholds.
- Retrospective/Prospective: Not specified. Bench testing is typically a controlled laboratory process.
-
Ground Truth (for non-clinical testing):
- The "ground truth" for the technical specifications and performance of the device (e.g., accuracy of parameter measurements, robust alarms) would have been established by engineering and quality standards, internal specifications, and applicable consensus standards (e.g., IEC 60601-1-8 for alarms).
What Cannot Be Answered From the Document (specifically regarding clinical studies/acceptance criteria as implied by clinical outcomes or human reader performance):
-
A table of acceptance criteria and the reported device performance (in a clinical context): The document explicitly states "CARESCAPE B850 did not require clinical studies to support substantial equivalence." Therefore, no clinical performance acceptance criteria or reported clinical performance is provided. The acceptance criteria for the non-clinical tests are implicit in the statement "demonstrating the design meets the specifications" for hardware, software, and performance.
-
Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective) for a clinical study: Not applicable, as no clinical study was performed.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience): Not applicable, as no clinical study was performed. Ground truth for technical performance would be against engineering benchmarks, not expert clinical assessment in a study.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable, as no clinical study was performed.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a patient monitor, not an AI-assisted diagnostic tool for image interpretation or similar. The "EK-Pro arrhythmia detection algorithm" is an internal algorithm, not a separate AI for human assistance.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The EK-Pro V14 arrhythmia analysis algorithm operates within the device as a standalone component for its intended function. Its performance would have been validated against a set of ECG waveforms with established ground truth for arrhythmias, but details of such testing (like sample size or specific metrics) are not in this 510(k) summary. The document states "EK-Pro V14 arrhythmia analysis algorithm compared to the EK-Pro V13 used in the predicate monitors," implying internal validation of the updated algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the arrhythmia detection algorithm, the ground truth would typically be established by expert cardiologists reviewing ECG recordings. However, the details of how EK-Pro V14's ground truth was established are not provided in this 510(k). For general device performance, the ground truth is adherence to technical specifications and relevant standards.
-
The sample size for the training set: Not applicable, as this is a 510(k) submission for a patient monitor, not an AI/ML device requiring a training set in the typical sense. While the EK-Pro algorithm may have been "trained" or developed using data, the document does not mention details about this.
-
How the ground truth for the training set was established: Not applicable for the same reason as above.
Summary Table of Available Information (based on non-clinical context):
Acceptance Criterion (Type of Test) | Reported Device Performance / Outcome (Non-Clinical) |
---|---|
Hardware Performance | Meets specifications. Includes electromagnetic compatibility, electrical safety, environmental, and usability testing. |
Software Performance | Meets specifications. Includes software design, development, verification, validation, and traceability. The new EK-Pro V14 arrhythmia analysis algorithm is considered equivalent to the predicate's V13. |
Usability | Extensive usability work completed, including critical task identification through use-based hazard analysis, multiple rounds of formative usability testing, and summative testing. The overall user interface layout, structure, operations, and workflow remain the same as the predicate, with only a new color scheme and slight visual adjustments aligned with a new design guideline. |
Cybersecurity / Risks | Patient safety, security, and privacy risks addressed through design and development, including a Security Risk Assessment and Threat model (system integrity controls, access controls, audit controls, network controls, remote service controls) mapping to FDA Guidance for Cybersecurity in Medical Devices (2014). The device can withstand network storm conditions without rebooting when connected by wire to specific networks. |
Alarm Functionality | Alarm management core functionalities (classification, notification, adjustment of settings, On/Off, audio silencing) are implemented and equivalent to the predicate. Added flexibility for users to adjust alarm priorities, criteria for additional parameter alarms, alarm delay options, critical alarm options, alarm acknowledgement, and latched alarm indicators. Supports tailored/specific alarm management schemes to reduce alarm fatigue while maintaining safety. Default settings are according to IEC 60601-1-8. The CARESCAPE RAD (new accessory for remote alarms) replaces the prior Remote Alarm Box and communicates one-way, not affecting the B850's alarm functionality. |
Networking Capability | Supports CARESCAPE Network LAN/VLAN. New single-wire network configuration supported, simplifying installation and maintenance without impact on clinical monitoring. |
Processor | Updated from Intel Celeron-M 423 ULV to Freescale ARM Cortex-A9, considered equivalent. |
Modules Supported | Equivalent. Supports newer acquisition modules, covering the same parameters as the predicate, with some legacy modules removed and new ones (e.g., CARESCAPE ONE, E-EEGX, E-sCAiOVX, E-sCOVX) added. |
General Equivalence | "The fundamental function and operation of the proposed CARESCAPE B850 V3 monitor are unchanged compared to CARESCAPE Monitor B850 with ESP V2 software (K131414). There are no new types of monitored parameters introduced compared to the predicate B850 monitor." The device is found to be substantially equivalent to the predicate device(s) for the intended users, uses and use environments based on non-clinical testing. |
Conclusion from Document:
The acceptance criteria for the CARESCAPE B850, as presented in this 510(k) submission, were met through extensive non-clinical bench testing, hardware and software verification/validation, and usability studies. The submission asserts the device's substantial equivalence to its predicate based on these non-clinical activities, stating explicitly that clinical studies were not required or performed.
Ask a specific question about this device
Page 1 of 1