Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K990814
    Manufacturer
    Date Cleared
    1999-06-07

    (88 days)

    Product Code
    Regulation Number
    864.7290
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cryov Check™ Factor 2 Deficient Plasma is recommended for use as a substrate in clot-based factor II assays using the one stage prothrombin time (PT) assay.
    Cryov Check™ Factor 2 Deficient Plasma is human plasma deficient in the Factor II coagulation protein while having all other coagulation factors within normal limits. It is recommended for use as a substrate in clot-based Factor II assays using the one stage prothrombin time (PT).

    Device Description

    CryovCheck™ Factor 2 Deficient Plasma is frozen human plasma deficient in the Factor II coagulation protein. It is prepared from citrated pooled normal human plasma which has been depleted of Factor II by immunoadsorption. Activity levels of Factor II are assayed at less than 1% normal levels while all other coagulation factors are within normal levels.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Factor II activity levelLess than 1% normal levels
    All other coagulation factorsWithin normal limits
    24-hour open vial stability (PT and APTT assays)No anomalous results detected
    24-hour open vial stability (Factor II recovery, Mean (% recovery), 2 SD, CV%)Mean Factor II recovery: 83.07% (Reference Value = 85%). Max: 87%, Min: 80%. Acceptable values are: Mean (+/-) 5% of reference value (80.75% - 89.25%); and %C.V.