Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K010534
    Date Cleared
    2001-08-06

    (164 days)

    Product Code
    Regulation Number
    880.2800
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K003533
    Date Cleared
    2001-02-07

    (83 days)

    Product Code
    Regulation Number
    880.2800
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K972300
    Device Name
    CROSS-CHECKS P
    Date Cleared
    1998-03-25

    (279 days)

    Product Code
    Regulation Number
    880.2800
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cross-Checks P chemical process indicator strips are designed to be used in the Sterrad sterilizer. Cross-Checks P Process Indicator Strips will exhibit a blue to red color change when exposed to hydrogen peroxide during the dillusion phase of the Sterrad cycle. When used as directed, the Cross Checks P strips give visible confirmation that the item has been exposed to hydrogen peroxide during the Sterrad Cycle.

    Device Description

    SteriTec Cross-Checks P strips are 4" x 5/8" process indicators with indicator ink printed onto Tyvek plastic.

    AI/ML Overview

    The SteriTec Cross-Checks P process indicator strips are designed to change color from blue to red when exposed to hydrogen peroxide during the diffusion phase of the Sterrad sterilizer cycle.

    Acceptance Criteria and Device Performance:

    Exposure Time to Hydrogen PeroxideAcceptance Criteria (Expected Outcome)Reported Device Performance (Number of strips meeting criteria)
    Less than 25 minutesIncomplete color change200/200 incomplete color change (at 10 and 15 minutes)
    25 minutesPossible pass (indicating some color change, but not necessarily complete)200/200 possible pass
    30 minutesComplete color change200/200 complete color change

    Study Details:

    1. Sample Size and Data Provenance:

      • Test Set Sample Size: 200 strips were used for each of the four different exposure timings (10 minutes, 15 minutes, 25 minutes, and 30 minutes), totaling 800 strips.
      • Data Provenance: The study was conducted using a Sterrad sterilizer, implying a controlled laboratory environment rather than a retrospective analysis of field data. The country of origin is not explicitly stated, but the submission is to the U.S. FDA by a U.S. company.
    2. Number of Experts and Qualifications: Not applicable. The assessment of color change for process indicators is typically objective and determined visually, often against a reference color chart, rather than through expert interpretation.

    3. Adjudication Method: Not applicable. The color change is a direct physical observation.

    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Not performed. This type of study is relevant for diagnostic devices requiring human interpretation.

    5. Standalone Performance: Yes, a standalone performance study was conducted. The device's color change response was evaluated directly under controlled exposure times without human intervention influencing the outcome.

    6. Type of Ground Truth: The ground truth was established by defining the expected color change (or lack thereof) at specific exposure durations to hydrogen peroxide within the Sterrad cycle. This is a direct measure of the device's functional response to the sterilizing agent.

    7. Training Set Sample Size: The document does not specify a separate training set. The study appears to be a validation study of the final product with the reported 800 strips.

    8. Ground Truth for Training Set: Not applicable, as a separate training set is not described. The performance criteria against which the device was evaluated are inherent to the intended function of a chemical process indicator.

    Ask a Question

    Ask a specific question about this device

    K Number
    K951113
    Device Name
    CROSS-CHECKS
    Date Cleared
    1996-08-19

    (528 days)

    Product Code
    Regulation Number
    880.2800
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SteriTer CROSS-CHECKS sterilization chemical monitoring strips are designed to be utilized in steam sterilizers operating at 132" (270°F to 276°F to 276°F). When used as directed, the SteriToc Cross.Checks indicator strips give visible indication that sterilizing conditions were met. During steritization, the check mark at the end of the strip changes from white to black, becoming as dark or darker in intensity than the black reference arrow printed on the strip.

    Device Description

    The basic formula and concept of black to white chemical steam sterilization indicators has not changed dramatically since they were invented in 1932.

    AI/ML Overview

    Below is the information regarding the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the "CROSS-CHECKS Steam Sterilization Indicator Strip" are primarily based on the visual color change of the strip from white to black, becoming as dark or darker than a reference arrow, indicating proper sterilization, and an incomplete color change for improper sterilization. This performance is validated against biological spore strips.

    Acceptance CriteriaReported Device Performance
    Test 1: Completed Sterilization Cycle (3 Minutes @ 132°C (270°F)) Indicator strips change color to black. Biological spore strips show complete kill.Results: All indicator strips changed color to black. All test spore strips showed a complete kill. All indicator strips passed this test.
    Test 2: Incomplete Sterilization Cycle (20 Seconds @ 132°C (270°F)) Indicator strips show incomplete color change (none turn black). Biological spore strips show all survivors.Results: All indicator strips showed incomplete color change (none turned to black). All test spore strips and the control strip showed all survivors. All indicator strips passed this test.
    Test 3: Performance Data for Storage of Cross-Check StripsA. Normal Storage: Strips appeared unchanged (white). B. Normal Storage then Sterilized: All indicator strips changed color to black, indicating proper functioning and no change in performance due to normal storage. C. Extreme Temperature Storage (57°C (134°F) for 8 days): Color of the strip had NOT changed from its original normal white color. D. Extreme Temperature Storage then Sterilized: All indicator strips changed color to black, indicating proper functioning and no change due to extreme heat storage. E. Extreme Humidity Storage (75% humidity for 7 days): Color of the strip had NOT changed from its original normal white color. F. Extreme Humidity Storage then Sterilized: All indicator strips changed color to black, indicating proper functioning and no change due to extreme humidity storage. G. Extreme Temperature and Humidity Storage (57°C (134°F) and 75% humidity for 8 days): Color of the strip had changed from its original white color to a light tan. H. Extreme Temperature and Humidity Storage then Sterilized: All indicator strips changed color to black, indicating proper functioning and no change due to extreme heat and humidity conditions.
    Test 4: Performance Data for Cross-Checks Strips at Various Temperatures and Times Indicators should correlate with biological kill/survivor rates. "Passed as specified, and in all cases completed a color change only after the biological spore strips tested at the same time under the same conditions, had expired."Results: Detailed in Tables 1, 2, 3, and 4 (presented across pages 6-9 in the document, but referred to as Tables 1-4 for clarity). The tables show the number of positive (Pos) and negative (Neg) reads for the indicator strips, along with the biological spore strip results (Bio +/-). The overall conclusion was that the product performed as specified.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test 1 & 2 (Completed/Incomplete Sterilization Cycle): The text mentions "All indicator strips" and one "spore strip" per run. The exact number of SteriTec and ATI strips per run is not explicitly stated, but it implies a small, controlled sample for these specific tests.
      • Provenance: Not explicitly stated, but implies laboratory testing performed by an "independent testing laboratory."
      • Retrospective/Prospective: Prospective (laboratory testing).
    • Test 3 (Storage Tests): "SteriTec CROSS-CHECKS sterilization indicator strips" were used. The exact number is not stated, but the results are qualitative observations ("All strips appeared unchanged," "All indicator strips changed color").
      • Provenance: Not explicitly stated, but implies laboratory testing performed by an "independent testing laboratory."
      • Retrospective/Prospective: Prospective (laboratory testing).
    • Test 4 (Various Temperatures and Times):
      • Sample Size:
        • For each temperature range (124°C, 128°C, 130°C, 132°C, 134°C): Approximately 300 strips (SteriTec Cross-Checks CI 102 and CI 111 combined) along with 12 biological spore strips.
        • For each time duration (1 min, 2 min, 2.5 min, 3 min, 3.5 min at 132°C): Approximately 300 strips (SteriTec Cross-Checks CI 102 and CI 111 combined) along with 12 biological spore strips.
      • Provenance: Not explicitly stated, but implies laboratory testing performed by an "independent testing laboratory."
      • Retrospective/Prospective: Prospective (laboratory testing).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Test 1, 2, 3 (Initial Interpretation): Implied that the "independent testing laboratory" personnel made the initial observations and recorded the results. Specific number and qualifications are not mentioned.
    • Test 4 (Performance Data):
      • First Interpretation: A panel of three people selected by the independent laboratory were used to make the PASS/FAIL determinations. Their specific qualifications are not mentioned.
      • Second Interpretation: Another set of three interpreters were hired by SteriTec. Their qualifications are explicitly stated as nurses, who "generally have more experience in chemical indicator interpretations."

    4. Adjudication Method for the Test Set

    • Test 4 (Performance Data): The text indicates decisions were made by a panel of three for both rounds of interpretation, but it does not specify an adjudication method like 2+1 or 3+1. It simply states they were "used to make the PASS/FAIL determinations." The fact that a second panel was brought in due to a few results from the first set being "considered outside the product specifications" suggests a lack of consensus or issues with the first panel's interpretation, leading to a re-evaluation by the second panel.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not explicitly done in the way typically understood for comparing AI to human-assisted AI.
    • The study involved panels of human readers interpreting the chemical indicators. There was a comparison between two sets of human interpreters (first panel vs. nurses) for Test 4, noting that the "interpretations of the second set of interpreters (Nurses) were more uniform." This focuses on inter-rater variability and expertise among human readers, not on AI assistance.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was done

    • No. The device is a physical chemical indicator strip that requires visual interpretation by a human. There is no algorithm or AI involved for standalone performance. The "performance" of the device itself (color change) is assessed by human observation.

    7. The Type of Ground Truth Used

    • For the indicator strips' performance: The primary ground truth for the effectiveness of the sterilization process itself (which the indicator is supposed to reflect) was established using biological spore strips. The indicator's color change was correlated directly with the kill/survival status of these biological indicators.
    • For the interpretation of the color change: Expert consensus (from the panels of interpreters) was used to determine if the color change of the chemical indicator strip qualified as a "pass" or "fail."

    8. The Sample Size for the Training Set

    • Not applicable. This document describes the validation of a physical chemical indicator, not an AI or algorithm. Therefore, there is no "training set" in the context of machine learning. The "learning" for the product's design would have happened during its development, but this document focuses on its performance validation against established standards.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. As there is no training set mentioned for an AI/algorithm, there is no ground truth establishment for a training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K952244
    Date Cleared
    1996-08-05

    (451 days)

    Product Code
    Regulation Number
    880.2800
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sterilier CROSS-CHECKS EO gas process indicator monitoring strips are a yellow to brown chemical indicator strip that changes color during Ethylenc Oxide (EO) processing. They indicate that the item has been processed in Ethylene Oxide.

    Device Description

    SteriTec CROSS-CHECKS EO process indicators are a pale yellow to brown chemical indicator strip that changes color during ethylene oxide processing.

    AI/ML Overview

    The provided text describes a 510(k) application for a chemical process indicator and includes limited information about performance criteria and study details. Based on the available text, here's an attempt to answer the questions, highlighting what is present and what is missing.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implied)Reported Device Performance
    Changes color during ethylene oxide processing."a pale yellow to brown chemical indicator strip that changes color during ethylene oxide processing."
    Indicates that the item has been processed in Ethylene Oxide."They indicate that the item has been processed in Ethylene Oxide."
    Performs safely and effectively."performed in a safe and effective manner"
    Equivalent to the predicate device (Surgicot 20 EO Gas Indicator)."performed...equivalently to the Surgicot 20 predicate device."

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not specified. The document states "Laboratory tests were performed" but does not give a number of indicators tested.
    • Data Provenance: Not explicitly stated, but the tests were performed by "an independent testing laboratory." The location of this laboratory and whether the data is retrospective or prospective is not mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided. For a chemical indicator, ground truth would likely be established by exposure to EO gas under controlled conditions, not by expert interpretation.

    4. Adjudication method for the test set

    Not applicable/Not specified. Given the nature of a chemical indicator, "adjudication" in the sense of human interpretation of complex outputs is not relevant. The "ground truth" for whether the indicator changed color would be directly observable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a chemical process indicator, not an AI-assisted diagnostic tool for human readers. Therefore, an MRMC study and AI assistance are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept is not directly applicable. The device itself is a "standalone" chemical indicator that changes color. There is no algorithm or human-in-the-loop interaction described beyond the observation of the color change. The "laboratory tests" inherently represent the standalone performance of the indicator.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for this type of device would be whether the indicator was actually exposed to ethylene oxide gas under conditions that should trigger a color change, and conversely, whether it remained unchanged when not exposed. This would be established by controlled laboratory exposure to EO (or lack thereof) rather than expert consensus or pathology.

    8. The sample size for the training set

    Not applicable. Chemical process indicators typically do not involve "training sets" in the machine learning sense. Their performance is based on chemical reactions.

    9. How the ground truth for the training set was established

    Not applicable (see point 8).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1