Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K151090
    Manufacturer
    Date Cleared
    2016-09-21

    (517 days)

    Product Code
    Regulation Number
    880.5570
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This is a single use sterile insulin pen-injector needle to be used by diabetic patients for the purpose of injecting insulin.

    Device Description

    CPL insulin pen needle consist of a sterile cap, needle, hub and blister paper. The sterile cap and blister paper function to sustain sterilization of the product. The hub can be connected with pen type insulin syringe. The needle cap protects the needle. This device is single use.

    AI/ML Overview

    The document describes a 510(k) premarket notification for the "CPL Insulin Pen Needle" and its substantial equivalence to a predicate device. It details performance data to demonstrate this equivalence.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" with numerical or categorical targets followed by "reported device performance" in the format typically used for AI/ML device evaluations. Instead, it lists various tests performed and states that the device "demonstrated equivalent performance to the predicate devices" or "is substantially equivalent" to the predicate.

    However, we can infer the acceptance criteria and performance from the tests conducted against international standards (ISO).

    Acceptance Criteria (Inferred Standard Compliance)Reported Device Performance
    Biocompatibility (ISO 10993-1 series)
    CytotoxicityPerformed and suitable
    SensitizationPerformed and suitable
    Intracutaneous reactivity (Acute)Performed and suitable
    Systemic toxicity (Acute)Performed and suitable
    PyrogenicityPerformed and suitable
    HemolysisPerformed and suitable
    Physical/Mechanical Performance (ISO 11608-2 & ISO 9626)
    Inner/outside and structurePerformed and demonstrated equivalent performance
    DimensionPerformed and demonstrated equivalent performance
    Draw testPerformed and demonstrated equivalent performance
    Elasticity testPerformed and demonstrated equivalent performance
    Flexural rigidityPerformed and demonstrated equivalent performance
    Inner and outside of needlePerformed and demonstrated equivalent performance
    Size designationPerformed and demonstrated equivalent performance
    Elasticity of the needle tubePerformed and demonstrated equivalent performance
    PullPerformed and demonstrated equivalent performance
    LubricantPerformed and demonstrated equivalent performance
    Limits for acidity and alkalinityPerformed and demonstrated equivalent performance
    StiffnessPerformed and demonstrated equivalent performance
    Resistance to breakage (ISO 7864)Performed and demonstrated equivalent performance
    Resistance to corrosionPerformed and demonstrated equivalent performance
    Determination of flow rate through the needle (ISO 9626)Performed and demonstrated equivalent performance
    Bond between hub and needle tubePerformed and demonstrated equivalent performance
    Freedom from defectsPerformed and demonstrated equivalent performance
    Penetration resistancePerformed and demonstrated equivalent performance
    Cap-hub fitting strengthPerformed and demonstrated equivalent performance
    Dislocation of measuring point at patient endPerformed and demonstrated equivalent performance
    Needle dose accuracy testPerformed and demonstrated equivalent performance
    Needle hub torque removalPerformed and demonstrated equivalent performance
    Ease of assembly and disassembly testPerformed and demonstrated equivalent performance
    Compatibility with needle-based injection systemsPerformed and demonstrated equivalent performance and a list of compatible pens is provided
    Sterilization effectivenessEO Gas sterilization, stated as "Same" as predicate
    Packaging integritySterile cap(PP)+sterile paper, stated as "Similar"

    The key acceptance criterion is that the "CPL Insulin Pen Needle" demonstrates substantially equivalent performance to the predicate device (K080904 Feel Fine Insulin Pen Needle) based on compliance with relevant ISO standards and a series of bench tests.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the bench tests. It also does not specify the data provenance (e.g., country of origin or whether the data was retrospective or prospective) for these performance tests. The tests are described as "bench tests," implying laboratory testing of the physical devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This section is not applicable in the context of this 510(k) submission. The CPL Insulin Pen Needle is a physical medical device, not an AI/ML diagnostic or predictive algorithm where expert ground truth for a test set would be established. The "ground truth" here is defined by the technical specifications and performance requirements outlined in the ISO standards and the comparative performance against the predicate device.

    4. Adjudication Method for the Test Set

    This section is not applicable for the same reasons as point 3. Adjudication methods like "2+1" or "3+1" are relevant for expert consensus in cases of subjective interpretation (e.g., medical image reading), not for objective physical device performance testing against defined standards.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The CPL Insulin Pen Needle is a physical medical device (an insulin pen needle), not an AI-assisted diagnostic or therapeutic system. Therefore, an MRMC study related to human readers improving with AI assistance would not be performed for this type of device.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. As explained for point 5, this is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in this context is established by international standards (e.g., ISO 10993-1, ISO 11608-2, ISO 7864, ISO 9626) and the demonstrated performance of the legally marketed predicate device (K080904 Feel Fine Insulin Pen Needle). The product's performance is compared against these objective benchmarks and the predicate device's established performance to determine substantial equivalence.

    8. The Sample Size for the Training Set

    This is not applicable. As this is a physical medical device, there is no "training set" in the context of machine learning. The device's design, manufacturing, and quality control processes are informed by engineering principles and compliance with standards, not by AI model training.

    9. How the Ground Truth for the Training Set was Established

    This is not applicable for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1