(517 days)
This is a single use sterile insulin pen-injector needle to be used by diabetic patients for the purpose of injecting insulin.
CPL insulin pen needle consist of a sterile cap, needle, hub and blister paper. The sterile cap and blister paper function to sustain sterilization of the product. The hub can be connected with pen type insulin syringe. The needle cap protects the needle. This device is single use.
The document describes a 510(k) premarket notification for the "CPL Insulin Pen Needle" and its substantial equivalence to a predicate device. It details performance data to demonstrate this equivalence.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of "acceptance criteria" with numerical or categorical targets followed by "reported device performance" in the format typically used for AI/ML device evaluations. Instead, it lists various tests performed and states that the device "demonstrated equivalent performance to the predicate devices" or "is substantially equivalent" to the predicate.
However, we can infer the acceptance criteria and performance from the tests conducted against international standards (ISO).
| Acceptance Criteria (Inferred Standard Compliance) | Reported Device Performance |
|---|---|
| Biocompatibility (ISO 10993-1 series) | |
| Cytotoxicity | Performed and suitable |
| Sensitization | Performed and suitable |
| Intracutaneous reactivity (Acute) | Performed and suitable |
| Systemic toxicity (Acute) | Performed and suitable |
| Pyrogenicity | Performed and suitable |
| Hemolysis | Performed and suitable |
| Physical/Mechanical Performance (ISO 11608-2 & ISO 9626) | |
| Inner/outside and structure | Performed and demonstrated equivalent performance |
| Dimension | Performed and demonstrated equivalent performance |
| Draw test | Performed and demonstrated equivalent performance |
| Elasticity test | Performed and demonstrated equivalent performance |
| Flexural rigidity | Performed and demonstrated equivalent performance |
| Inner and outside of needle | Performed and demonstrated equivalent performance |
| Size designation | Performed and demonstrated equivalent performance |
| Elasticity of the needle tube | Performed and demonstrated equivalent performance |
| Pull | Performed and demonstrated equivalent performance |
| Lubricant | Performed and demonstrated equivalent performance |
| Limits for acidity and alkalinity | Performed and demonstrated equivalent performance |
| Stiffness | Performed and demonstrated equivalent performance |
| Resistance to breakage (ISO 7864) | Performed and demonstrated equivalent performance |
| Resistance to corrosion | Performed and demonstrated equivalent performance |
| Determination of flow rate through the needle (ISO 9626) | Performed and demonstrated equivalent performance |
| Bond between hub and needle tube | Performed and demonstrated equivalent performance |
| Freedom from defects | Performed and demonstrated equivalent performance |
| Penetration resistance | Performed and demonstrated equivalent performance |
| Cap-hub fitting strength | Performed and demonstrated equivalent performance |
| Dislocation of measuring point at patient end | Performed and demonstrated equivalent performance |
| Needle dose accuracy test | Performed and demonstrated equivalent performance |
| Needle hub torque removal | Performed and demonstrated equivalent performance |
| Ease of assembly and disassembly test | Performed and demonstrated equivalent performance |
| Compatibility with needle-based injection systems | Performed and demonstrated equivalent performance and a list of compatible pens is provided |
| Sterilization effectiveness | EO Gas sterilization, stated as "Same" as predicate |
| Packaging integrity | Sterile cap(PP)+sterile paper, stated as "Similar" |
The key acceptance criterion is that the "CPL Insulin Pen Needle" demonstrates substantially equivalent performance to the predicate device (K080904 Feel Fine Insulin Pen Needle) based on compliance with relevant ISO standards and a series of bench tests.
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample size used for the bench tests. It also does not specify the data provenance (e.g., country of origin or whether the data was retrospective or prospective) for these performance tests. The tests are described as "bench tests," implying laboratory testing of the physical devices.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This section is not applicable in the context of this 510(k) submission. The CPL Insulin Pen Needle is a physical medical device, not an AI/ML diagnostic or predictive algorithm where expert ground truth for a test set would be established. The "ground truth" here is defined by the technical specifications and performance requirements outlined in the ISO standards and the comparative performance against the predicate device.
4. Adjudication Method for the Test Set
This section is not applicable for the same reasons as point 3. Adjudication methods like "2+1" or "3+1" are relevant for expert consensus in cases of subjective interpretation (e.g., medical image reading), not for objective physical device performance testing against defined standards.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The CPL Insulin Pen Needle is a physical medical device (an insulin pen needle), not an AI-assisted diagnostic or therapeutic system. Therefore, an MRMC study related to human readers improving with AI assistance would not be performed for this type of device.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. As explained for point 5, this is a physical medical device, not an algorithm.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" in this context is established by international standards (e.g., ISO 10993-1, ISO 11608-2, ISO 7864, ISO 9626) and the demonstrated performance of the legally marketed predicate device (K080904 Feel Fine Insulin Pen Needle). The product's performance is compared against these objective benchmarks and the predicate device's established performance to determine substantial equivalence.
8. The Sample Size for the Training Set
This is not applicable. As this is a physical medical device, there is no "training set" in the context of machine learning. The device's design, manufacturing, and quality control processes are informed by engineering principles and compliance with standards, not by AI model training.
9. How the Ground Truth for the Training Set was Established
This is not applicable for the same reasons as point 8.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the perimeter. Inside the circle is an image of three human profiles facing to the right. The profiles are stacked on top of each other, with the first profile being the most prominent.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 21, 2016
CPL Co., Ltd. c/o Mr. Peter Chung President Plus Global 300 Atwood Street Pittsburgh, Pennsylvania 15213
Re: K151090
Trade/Device Name: CPL Insulin Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: August 16, 2016 Received: August 23, 2016
Dear Mr. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang
-S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K151090
Device Name
CPL Insulin Pen Needle
Indications for Use (Describe)
This is a single use sterile insulin pen-injector need by diabetic patients for the purpose of injecting insulin.
Type of Use (Select one or both, as applicable)
الص Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
{3}------------------------------------------------
510(k) Summary
[as required by 807.92(c)] K151090
1. Applicant
-
- Company : CPL Co.,Ltd.
-
- Address : 36, Yongteurim-gil, Danwon-gu, Ansan-si, Gyeonggi-do, Korea
-
- Tel : 82-31-483-7301
-
- Fax : 82-31-483-7351
-
- Contact person : Peter Chung, 412-687-3976
-
- Contact person address : 300, Atwood Street, Pittsburgh, PA, 15213, USA
-
- Date prepared : September 16, 2016
2. Device Information
-
- Trade Name : CPL Insulin Pen Needle
-
- Common Name : Hypodermic single lumen needle
-
- Classification Name : Needle, Hypodermic, Single Lumen
-
- Product Code : FMI
-
- Regulation Number : 880.5570
-
- Class of device : Class II
-
- Panel : General Hospital
3. The legally marketed device to which we are claiming equivalence
K080904 Feel Fine Insulin Pen Needle (Company : Feel Tech)
4. Device description
CPL insulin pen needle consist of a sterile cap, needle, hub and blister paper. The sterile cap and blister paper function to sustain sterilization of the product. The hub can be connected with pen type insulin syringe. The needle cap protects the needle. This device is single use.
| Model | Needle length (mm) : | ||
|---|---|---|---|
| Patient-side tip | CPLPN-29 | 8mm, 10mm, 12mm | |
| CPLPN-30 | 5mm, 8mm, 10mm, 12 mm | ||
| CPLPN-31 | 4mm, 5mm, 6mm, 8mm | ||
| CPLPN-32 | 4mm, 5mm, 6mm, 8mm | ||
| Cartridge-side tip | CPLPN-29 | ||
| CPLPN-30 | 5mm, | ||
| CPLPN-31 | |||
| CPLPN-32 | |||
| Needle taper | 3.5mm | ||
| Wall type(Thin-walled) | Needle gauge | Range of out diameters | Inside diameter of tubing |
| CPLPN-29 | 29G(0.324mm~0.351mm) | 29G (0.190) | |
| CPLPN-30 | 30G (0.298mm~0.320mm) | 30G (0.165) | |
| CPLPN-31 | 31G (0.254mm~0.267mm) | 31G (0.125) | |
| CPLPN-32 | 32G (0.229mm~0.241mm) | 32G (0.105) |
{4}------------------------------------------------
5. Intended Use :
This is a single use sterile insulin pen-injector needle to be used by diabetic patients for the purpose of injecting insulin.
6. Performance data:
In accordance with ISO 10993-1, CPL Insulin Pen Needle is classified as : Externally Communicating Device, Blood Path Indirect, Shor Term (<24 hours) Use, as the cannula is immediately withdrawn after injection into the body. This classification was chosen as "worst case scenario."
PER ISO 10993-1, the following tests were performed for this classification:
-
- Cytotoxicity
-
- Sensitization
-
- Intracutaneous reactivity (Acute)
-
- Systemic toxicity (Acute)
-
- Pyrogenicity
-
- Hemolysis
Bench tests relating to the performance of the "CPL Insulin Pen Needle" were conducted. The principal device demonstrated equivalent performance to the predicate devices during bench testing. Bench testing consisted of:
| Inner/outside and structure | |
|---|---|
| Dimension | |
| Draw test | |
| Elasticity test | |
| Flexual rigidity | |
| Inner and outside of needle | |
| Size designation | |
| Elasticity of the needle tube | |
| Pull | Test standard |
| Lubricant | |
| Limits for acidity and alkalinity | ISO 11608-2 |
| Stiffness | |
| Resistance to breakage | ISO 7864 |
| Resistance to corrosion | |
| Determination of flow rate through the needle | ISO 9626 |
| Bond between hub and needle tube | |
| Freedom from defects | |
| Penetration resistance | |
| Cap-hub fitting strength | |
| Dislocation of measuring point at patient end | |
| Needle dose accuracy test | |
| Needle hub torque removal | |
| Ease of assembly and disassembly test | |
| Compatibility with needle-based injection systems |
The performance tests demonstrated that CPL Insulin Pen Needle is performes in a substantially equivalent manner to the predicate device.
{5}------------------------------------------------
Compatible pens
| Victoza Liraglutide injection | HumaLog KwikPen |
|---|---|
| Apidra SoloStar | HumaLog Mix 75/25 KwikPen |
| Lantus SoloStar | HumaLog Mix 50/50 KwikPen |
| OptiClik for Lantus and Apidra | Humalog Pen |
| SymlinPen 120 (pramlintide acetate) pen injector | Humalog Mix 75/25 Pen |
| SymlinPen 60 (pramlintide acetate) pen injector | Humalog Mix 50/50 Pen |
| Byetta exenatide injection 10mog | Humulin N Pen |
| Byetta exenatide injection 5mog | Humulin 70/30 Pen |
| Levemir FlexPen | HumaPen LUXURA HD |
| Novolog FlexPen | HumaPen MEMOIR |
| Novolog Mix 70/30 FlexPen | AutoPen |
| NovoPen Junior | Forteo teriparatide (rDNA origin) injection |
| NovoPen 3 |
7. Predicate device comparison table
| Manufacturer | CPL Co.,Ltd. | Feel Tech | Results | |||
|---|---|---|---|---|---|---|
| 510(k) No. | K151090 | K080904 | N/A | |||
| Indication for use | This is a single use sterileinsulin pen-injector needle to beused by diabetic patients for thepurpose of injecting insulin. | These disposable sterile insulin penneedles are intended forsubcutaneous injection of insulin inthe treatment of diabetes. | Same intended use | |||
| Product name | Hypodermic single lumen needle | Hypodermic single lumen needle | Identical | |||
| Trade name | CPL Insulin Pen Needle | Feel Fine Insulin Pen Needle | N/A | |||
| Model/type | CPLPN-29, CPLPN-30, CPLPN-31,CPLPN-32 | Feel Fine Insulin Pen Needle-29Feel Fine Insulin Pen Needle-30Feel Fine Insulin Pen Needle-31 | N/A | |||
| Appearance | Image: CPL Insulin Pen Needle | Image: Feel Fine Insulin Pen Needle | Similar design | |||
| Productconfiguration | HubNeedle capSterile capNeedleSterile paper | HubNeedle capSterile capNeedleSterile paper | Similar devicecomponenets | |||
| Material | Sterile cap : PolypropyleneNeedle cap : PolyethyleneNeedle : Stainless steel 304Hub : Polypropylene | Sterile cap : PolypropyleneNeedle cap : PolyethyleneNeedle : Stainless steel 304Hub : Polypropylene | Similar material | |||
| Length of parts (mm) | ||||||
| Total length ofassembly | 29.8 | 28.8 | ||||
| Side length ofassembly(max) | 15.2 | 15.5 | Similar dimensions | |||
| Inner diameter ofsterile cap | 11.2 | 12.95 | ||||
| Gauge | 29G | 30G | 31G | 32G | 29, 30, 31G | Similar |
| Length of needle | 8 mm10 mm12 mm | 5 mm8 mm10 mm12 mm | 4 mm5 mm6 mm8 mm | 4 mm5 mm6 mm8 mm | 5 mm8 mm12.7 mm | Similar needle length |
| Sterilization | EO Gas sterilization | EO Gas sterilization | Same | |||
| Packagine | Sterile cap(PP)+sterile paper | Sterile cap(PP)+sterile paper | Similar | |||
| Performance test | Accordance with ISO 11608-2 | N/A | Suitable | |||
| Biocompatibility | Accordance with ISO 10993 series | N/A | Suitable |
{6}------------------------------------------------
The device is investigated for function and effectiveness to compare the operation of function between CPL Co.,Ltd. and Feeltech (K080904). Comparison results demonstrate that the specifications and performance of the device are same as functional and effective as the legally marketed predicate device. Therefore, it is concluded that is Insulin pen needle of CPL Co.,Ltd. substantially equivalent to the legally marketed predicate device.
Intended use statement is similar with that of the predicate device. They are both single use only, used for insulin injection and for diabetes treatment. The difference in wording of these intended use does not change the intended use or raise questions of safety and effectiveness per 807.92(a)(5).
The bench tests of Needles with 32G demonstrated conformances to ISO 11608-2 and ISO 7864. Therefore the differences do not raise new concerns to establish substantial equivalence to the predicate.
9. Conclusion:
The subject device has been tested according to ISO 11608-2:2012 (Needle-based injection systems for medical use-Requirements and test methods-Part 2:Needles) as well as ISO 9626 (stainless steel tubing for manufacture of medical devices) and test results demonstrated it is substantial equivalent to as the cited predicate device.
§ 880.5570 Hypodermic single lumen needle.
(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).