Search Results
Found 1 results
510(k) Data Aggregation
(259 days)
CORDIS ENDOVASCULAR TEMPORARY OCCLUSION BALLOON CATHETER
The Temporary Occlusion Balloon Catheter is intended for the temporary occlusion of blood vessels in neuro and peripheral vasculature.
Not Found
The provided 510(k) summary (K984214) describes a Temporary Occlusion Balloon Catheter and outlines performance testing done to establish substantial equivalence to predicate devices.
This is not an AI/ML device. The performance testing described is for a physical medical device. Therefore, many of the requested categories related to AI/ML device studies (e.g., sample sizes for test and training sets, number of experts for ground truth, adjudication methods, MRMC studies) are not applicable.
Here's an analysis based on the information provided, focusing on what is relevant for a non-AI/ML device:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria / Performance Benchmarks (Implied) | Reported Device Performance |
---|---|
In-vitro Testing: | |
Tip linear stiffness similar to predicates | Performs similar to predicate devices |
Catheter kink resistance similar to predicates | Performs similar to predicate devices |
Catheter preparation time similar to predicates | Performs similar to predicate devices |
Balloon inflation/deflation times similar to predicates | Performs similar to predicate devices |
Deflatability similar to predicates | Performs similar to predicate devices |
Animal Study Testing: | |
Device works as intended (temporary occlusion of blood vessels) | Device works as intended |
Biocompatibility: | |
All appropriate biocompatibility tests passed | All appropriate biocompatibility tests were successfully performed on the materials used |
2. Sample size used for the test set and the data provenance
- Sample Size: Not explicitly stated for in-vitro tests (e.g., number of catheters tested for stiffness, kink resistance). For animal studies, the number of animals used is not specified.
- Data Provenance: Not applicable in terms of country of origin or retrospective/prospective data as this pertains to physical device performance testing rather than clinical data collection.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. Ground truth for a physical device's performance is established through direct measurements and observations during in-vitro and animal testing, not typically through expert consensus on medical images or clinical outcomes.
4. Adjudication method for the test set
Not applicable. Adjudication methods are typically used when interpreting ambiguous clinical data or images, which is not the primary focus of this device's performance evaluation.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI/ML device.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI/ML device.
7. The type of ground truth used
- In-vitro testing: Physical measurements and functional performance observations.
- Animal study testing: Direct observation of device function (temporary vessel occlusion) and potential biological response in a living model. Biocompatibility results would be based on laboratory testing against established standards.
8. The sample size for the training set
Not applicable. This is not an AI/ML device.
9. How the ground truth for the training set was established
Not applicable. This is not an AI/ML device.
Ask a specific question about this device
Page 1 of 1