Search Results
Found 2 results
510(k) Data Aggregation
(83 days)
CLIQ DV-300 ASPIRATOR
The device is intended to be used for aspiration during flexible endoscopy.
The Cliq Aspirator model DV-300 is an AC-powered high-vacuum / low-flow suction unit used for endoscopic aspiration. The DV-300 creates a negative pressure (vacuum) that draws fluids through disposable tubing, and into a collection container. The fluids are trapped within the collection container for proper disposal. The device is comprised of a maintenance-free pump unit; power cord, on/off switch, pressure relief valve, pressure adjustment knob, pressure gauge, microbial filter, and intermediate tubing.
The Cliq Aspirator: model DV-300 requires 110-120 V and 60 Hz. It has Class II with Type BF applied part protection against electric shock. The DV-300 suction pump operates via a piston pump type, and has a maximum vacuum pressure of 620 mm Hg. The pump produces a flow rate of up to 18 Liters per minute.
The Cliq Aspirator Model DV-300 underwent performance testing to establish substantial equivalence to the predicate device, Olympus Suction Pump Model KV-5. The evaluation focused on electrical safety, electromagnetic compatibility, and overall performance, comparing key specifications between the two devices.
1. Acceptance Criteria and Reported Device Performance
Criterion | Acceptance Criteria (Predicate Device: Olympus Suction Pump Model KV-5) | Reported Device Performance (Cliq Aspirator Model DV-300) | Meets Acceptance Criteria |
---|---|---|---|
Electrical Requirements | 100-120 V, 50/60 Hz | 110-120 V, 60 Hz | Yes |
Protection against electric shock | Class I with Type BF applied part | Class II with Type BF Applied part | No (Different Class) |
Maximum Vacuum Pressure | 638 mm Hg | 620 mm Hg | Yes (Comparable) |
Vacuum Pump Type | Piston | Piston | Yes |
Flow | Up to 20 L/min | Up to 18 L/min | Yes (Comparable) |
Sound Level | Not known (for predicate) | 53 dBA | NA (No direct comparison) |
Weight | 12.7 kg | 3.5 kg | No (Significant Difference) |
Dimensions | 31 x 25.5 x 22 cm | 30 x 16.5 x 19 cm | Yes (Comparable) |
Operating Environment (Temperature) | 10 - 40 °C | 0 - 40 °C | Yes (Wider Range) |
Operating Environment (Humidity) | 0 - 95 % | 0 - 90 % | Yes (Comparable) |
Operating Environment (Atm. Pressure) | 70 - 106 kPa | 70 - 106 kPa | Yes |
Storage Environment (Temperature) | -40 - 70 °C | -20 - 50 °C | No (Narrows Range) |
Storage Environment (Humidity) | 0 - 95 % | 0 - 95 % | Yes |
Storage Environment (Atm. Pressure) | 23.5 - 106 kPa | 50 - 106 kPa | Yes (Comparable) |
Filter | Microbial, hydrophobic | Microbial, hydrophobic | Yes |
Suction Tube length | 2 m | 1.8 m | Yes (Comparable) |
Filter Tube length | 900 mm | 250 mm | No (Significant Difference) |
Power Cable | Yes | Yes | Yes |
Suction Jar Volume | 2 liter (one liter optional) | 1.5 L or 2.4 L | Yes (Comparable) |
Performance Standards adherence | EN IEC 60601-1, EN IEC 60601-1-2, EN ISO10079-1, UL 2601-1, CAN/CSA Std. No. C22.2 No.601.1-M90 | EN IEC 60601-1, EN IEC 60601-1-2, EN ISO10079-1 | Yes (subset met) |
Note on "Meets Acceptance Criteria": The statement "Yes (Comparable)" or "Yes (Wider Range)" indicates that the Cliq Aspirator's performance is either directly equivalent or within an acceptable range compared to the predicate device, supporting substantial equivalence. Differences in Class protection, weight, and filter tube length while noted, were evidently not deemed critical enough by the FDA to prevent substantial equivalence, given the overall conclusion. The "Not known" for predicate sound level means no direct comparison could be made for that specific metric.
2. Sample Size Used for the Test Set and Data Provenance:
The document does not explicitly state a specific "test set" sample size for a particular number of devices or clinical cases. The testing reported is related to general device performance, electrical safety, and electromagnetic compatibility standards. These types of tests typically involve a sample of manufactured devices (e.g., a few units from a production run) rather than patient data.
- Data Provenance: The testing was conducted in the context of regulatory submission for a device manufactured by Ding Hwa Co., Ltd in Taiwan. The tests are likely laboratory-based and performed by the manufacturer or a contracted testing facility to meet international standards. No information is provided regarding country of origin of data in terms of patient population or clinical studies, as this is not a clinical performance study.
- Retrospective or Prospective: Not applicable, as this is not a clinical study involving retrospective or prospective data collection from patients.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
Not applicable. The "ground truth" for this type of device (an aspirator) is its adherence to predefined engineering specifications and safety standards. This is objectively measured and verified according to scientific and engineering principles, not through expert consensus on clinical findings. Therefore, no experts in a clinical sense were used to establish ground truth for a test set.
4. Adjudication Method for the Test Set:
Not applicable. As noted above, the "tests" refer to measurements against engineering and safety standards, not clinical assessments requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:
No. This document describes the regulatory submission for a medical device (an aspirator pump), not an AI/CAD system. Therefore, an MRMC comparative effectiveness study, which is relevant for evaluating human reader performance with and without AI assistance in interpretation tasks, was not performed.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:
No. This is a physical, electromechanical device, not an algorithm or AI. Standalone performance as typically understood for AI was not applicable or performed.
7. The Type of Ground Truth Used:
The ground truth used was adherence to established international and national engineering standards for medical devices, specifically:
- EN ISO 10079-1:2009: Electrically powered suction equipment safety requirements.
- IEC/EN 60601-1-2:2007 (3rd Ed.): Electromagnetic Compatibility.
- IEC 60601-1: General basic safety and essential performance of medical electrical equipment.
- Performance Verification Testing: Testing against predefined functional specifications (e.g., maximum vacuum pressure, flow rate).
The comparison against the predicate device (Olympus Suction Pump Model KV-5) also served as a basis for evaluating substantial equivalence in performance characteristics.
8. The Sample Size for the Training Set:
Not applicable. This is not an AI/machine learning device that utilizes a training set.
9. How the Ground Truth for the Training Set Was Established:
Not applicable. As this device does not use a training set, the establishment of ground truth for such a set is irrelevant.
Ask a specific question about this device
(158 days)
CLIQ DV-300 ASPIRATOR
The device is to be used to remove fluids from the treated tracheotomy patient airway.
The predicate devices and the subject device have the same technological characteristics. Those devices are intended to be used to remove fluids from the treated tracheotomy patient airway. It creates a negative pressure (vacuum) that draws the fluids through disposable tubing that is connected to a collection jar. The fluids are trapped in the collection jar for the proper disposal. The devices are for use on the order of a physician only. If practiced out of the hospital, the care giver has to be trained and recorded.
The predicate devices and the subject device are the portable AC powered suction pumps. Each one consists of an on/off switch, a pump unit, a non-detachable flexible power cord, collection jar, relief valve, pressure gauge, pressure adjustment knob, bacteria filter, suction tubing. The major differences between the predicate devices and the subject device are the capacities and dimensions.
Based on the provided text, here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance:
The document describes the device, the Cliq Aspirator, DV-300, and compares its performance characteristics to two predicate devices (SPARMAX Aspirator TC-2000V and VC-701), both manufactured by Ding Hwa Co., Ltd. The acceptance criteria are implicitly defined by these performance characteristics, and the "reported device performance" refers to the values listed for the subject device.
Performance Characteristic | Predicate Device 1 (TC-2000V) | Predicate Device 2 (VC-701) | Acceptance Criteria (Implicit) | Reported Device Performance (Cliq Aspirator DV-300) |
---|---|---|---|---|
Sound level | 32 LPM | > 16 LPM | Air Flow Rate > 16 LPM (as per predicate 2) | > 18 LPM |
Electrical requirements | 110-120 VAC, 60Hz, 1.2A(max) | 110-120 VAC, 60Hz, 0.9A(max) | N/A (Comparison only) | 110-120 VAC, 60Hz, 0.5A(max) |
Size (LWH) (mm) | 380165240 | 340155210 | N/A (Comparison only) | 300165190 |
Weight | 5.2kg | 4.25kg | N/A (Comparison only) | 3.5 kg |
Collection jar | 1800c.c. | 800c.c. | N/A (Comparison only) | 800c.c. |
Other shared characteristics (Implicit acceptance criteria: "Same" as predicate, meaning meets predicate's criteria) | ||||
Intended Use | Same | Same | To remove fluids from treated tracheotomy patient airway. | Same |
Collection tubing Length / material | 180cm / PVC | 180cm/ PVC | "Same" (180cm / PVC) | Same |
Connection tubing length / material | 400mm / Silicon | 400mm / Silicon | "Same" (400mm / Silicon) | Same |
Internal battery | None | None | "Same" (None) | Same |
Adaptor | None | None | "Same" (None) | Same |
Bacteria filter | Appointed | Appointed | "Same" (Appointed) | Same |
Thermal switch | Appointed | Appointed | "Same" (Appointed) | Same |
Operating temp. | 0 ~ 40°C | 0 ~ 40°C | "Same" (0 ~ 40°C) | Same |
Operating relative humidity | 0 ~ 95% | 0 ~ 95% | "Same" (0 ~ 95%) | Same |
Storage & transport temp | -40°C ~ 70°C | -40°C ~ 70°C | "Same" (-40°C ~ 70°C) | Same |
Storage & transport humidity | 0 ~ 95% | 0 ~ 95% | "Same" (0 ~ 95%) | Same |
Where used | Hospital | Hospital | "Same" (Hospital) | Same |
Standards complied | IEC 60601-1-2, UL 60601, CAN/CSA-C22.2 No.68~92, UL 1450 | IEC 60601-1-2, UL 60601, CAN/CSA-C22.2 No.68~92, UL 1450 | Compliance with relevant medical electrical equipment and suction standards. | IEC 60601-1-2:2007, IEC 60601-1:2005 +C1:2006+C2:2007, ISO 10079-1:2009, IEC 60601-1-11:2010 |
The study proving the device meets these criteria is a substantial equivalence comparison to legally marketed predicate devices. The document explicitly states: "The major difference between the subject device and the predicate devices is the electrical power. The subject device has a less electrical power than the predicate devices. This leads to the differences of the sound levels, maximum vacuum levels, air flow rates, sizes, weights, and the collection jars. The differences are not related to the safety and effectiveness."
This statement, combined with the detailed comparison table, implies that the manufacturer performed tests on the Cliq Aspirator DV-300 to determine its specifications (sound level, vacuum, air flow rate, etc.) and compared these specifications to those of the predicate devices. The conclusion of the 510(k) summary is that the new device "does not raise any safety and effectiveness compared to the predicate devices, thus the subject device is substantially equivalent to the predicate devices."
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify a sample size for a "test set" in the context of clinical or performance data from human/patient cases. This 510(k) summary describes a comparison of device specifications and compliance with recognized standards, not a clinical trial with a patient test set. The data provenance is based on engineering specifications and compliance testing, likely conducted by the manufacturer (Ding Hwa Co., Ltd. in Taiwan). It's a technical comparison, not a clinical data comparison.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable and not provided in the document. The "ground truth" here is the engineering specifications and safety/performance standards. These are established through recognized industry standards (e.g., IEC, ISO, UL, CSA) and the technical expertise of the device manufacturer's engineers, rather than through expert clinical review of a test set.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable as there is no "test set" in the sense of patient data requiring adjudication in this 510(k) summary. The comparison is based on technical specifications and standard compliance, which are typically verified through engineering tests and documentation.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This document describes a medical device (an aspirator), not an AI-powered diagnostic tool or image analysis system. Therefore, an MRMC comparative effectiveness study involving human readers or AI assistance is not applicable and was not done.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. The device is a physical, powered medical suction pump, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" in this context is based on engineering specifications and compliance with recognized national and international safety and performance standards. The manufacturer compared the subject device's technical characteristics (sound level, vacuum, airflow, etc.) and its adherence to standards like IEC 60601-1-2, IEC 60601-1, ISO 10079-1, and IEC 60601-1-11, to established values and predicate devices.
8. The sample size for the training set
This is not applicable. This document describes a physical medical device submission, not a machine learning or AI model that requires a training set.
9. How the ground truth for the training set was established
This is not applicable as there is no training set for this device.
Ask a specific question about this device
Page 1 of 1