Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K081829
    Device Name
    CAPIMAGE
    Date Cleared
    2008-07-11

    (14 days)

    Product Code
    Regulation Number
    892.1100
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CapIMAGE Gamma Camera System is intended to produce planar images depicting the anatomical distributions of single photon emitting radioisotopes in the human body within the energy range of 59kev to 167kev. The system provides software applications for processing, analyzing, and displaying medical images for interpretation by medical personnel.

    The CapIMAGE Gamma Camera System is intended for examination (and diagnosis) of disorders and diseases in the thyroid gland, the heart and other small organs. The device is not interpretative. Data from the device are interpreted by medical personnel who, based on these, make the clinical diagnosis.

    Device Description

    The CapIMAGE is a mobile Small Field of View (SFOV) gamma camera system utilizing a scintillation detector to detect gamma rays. The CapIMAGE is designed to be manually transported and deployed for scanning of patients in laying, sitting or standing positions. The system can operate powered from mains or from the internal battery source for at least 60 min.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the CapIMAGE Gamma Camera System, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the CapIMAGE device are based on compliance with industry standards, specifically NEMA NU 1-2001 for performance specifications. The reported device performance metrics are presented as meeting or being within the specified limits of these standards. The document doesn't provide specific quantitative acceptance thresholds for each performance characteristic but implies compliance.

    Performance CharacteristicAcceptance Criteria (Implied by Standards Compliance)Reported Device Performance
    Energy rangeNEMA NU 1-2001 compliant59kev – 167kev
    Detector UFOVNEMA NU 1-2001 compliantØ 210 mm (8.3")
    Intrinsic Spatial Res.NEMA NU 1-2001 compliant≤ ± 3.7mm @ UFOV / FWHM
    Energy ResolutionNEMA NU 1-2001 compliant≤ 9.4% @ Tc-99m
    Spatial LinearityNEMA NU 1-2001 compliant≤ ± 0.5mm @ UFOV
    Flood Field UniformityNEMA NU 1-2001 compliant≤ ± 2.7% @ UFOV Intrinsic
    Intrinsic Uniformity, UFOV, DifferentialNEMA NU 1-2001 compliant< 1.5%
    Count rate PerformanceNEMA NU 1-2001 compliant>180k cps @ 20% loss
    Spatial Resolution, FWHM, LEGP Tc-99m.NEMA NU 1-2001 compliant< 9.4mm

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states "None." for "Clinical Performance Data." This indicates that no clinical test set with patient data was used to demonstrate the device's performance in a clinical setting. The device's performance was evaluated through non-clinical testing against established physical and safety standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable, as no clinical test set was used.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set was used.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. The document describes a gamma camera system and does not mention any AI components or MRMC studies.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Not applicable, as the device is a gamma camera system, not an algorithm. Its performance is physical and independent of human interpretation for these specifications.

    7. The Type of Ground Truth Used

    For the non-clinical performance data, the ground truth was based on the standards and methodologies prescribed by NEMA NU 1-2001 and other relevant electrical safety standards (UL 60601-1, IEC601-1, IEC601-1-1, IEC601-1-2). These standards define the procedures and expected outcomes for measuring the physical performance characteristics of gamma cameras.

    8. The Sample Size for the Training Set

    Not applicable. This device is a medical imaging hardware system, not a software algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1