Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K202167
    Manufacturer
    Date Cleared
    2021-02-26

    (207 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BRITE TIP RADIANZ™ Guiding Sheath is intravascular introduction of interventional and/or diagnostic devices into the peripheral vasculature through the radial artery.

    Device Description

    Each BRITE TIP RADIANZ™ device consists of a 6F compatible guiding sheath, a vessel dilator (0.035" guidewire compatible) and a removable hemostasis valve. The guiding sheath has a lubricious hydrophilic coating to enhance entry and withdrawal during vessel access, and a low-profile outer diameter with a smooth transition to the vessel dilator to minimize or prevent radial vasculature trauma and spasm of the radial artery. The sheath is available in two lengths, 110 cm and 135 cm. The sheath hub is an over-molded hub with a luer and is compatible with the removable hemostasis valve provided (described below) and with other hemostasis valves or common vascular accessories of the user's choice. The vessel dilator is compatible with 0.035" guidewires (not included) and is tapered such that it creates a smooth transition between the guiding sheath tip and the guidewire. The vessel dilator hub is over molded and includes a locking feature located in the distal end of the hub to facilitate locking the vessel dilator to the sheath cannula proximal hemostasis valve. The system contains a removable hemostasis valve to facilitate entry and withdrawal of intravascular devices through the guiding sheath and minimize the backflow of blood. It also has a side-port arm with attached 3-way stopcock for flushing or aspiration. The preassembled hub connector is removable, such that it can be exchanged for a hemostasis valve or other common vascular accessory of the user's preference.

    AI/ML Overview

    The provided FDA 510(k) summary for the BRITE TIP RADIANZ™ Guiding Sheath does not contain information about acceptance criteria or a study proving that the device meets those criteria in the context of an AI/ML device. This document describes a traditional medical device (a catheter introducer) and its substantial equivalence to a predicate device based on bench testing and biocompatibility, not AI/ML performance.

    Therefore, many of the requested categories related to AI/ML device evaluations (such as sample sizes for test/training sets, expert qualifications, ground truth establishment, MRMC studies, or standalone algorithm performance) are not applicable to this document.

    However, I can extract information related to the device's performance testing and the general conclusion of substantial equivalence.

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present "acceptance criteria" in a quantitative format for specific performance metrics in the way an AI/ML study would. Instead, it states that various tests were "successfully performed" or "passed" according to relevant ISO standards and internal test methods, indicating that the device met the requirements of those standards.

    Acceptance Criteria & Reported Device Performance (as implied by successful testing against standards):

    Test CategoryAcceptance Criteria (Implied)Reported Device Performance
    BiocompatibilityMeets requirements of ISO 10993-1 and specific tests.Biocompatible for intended use based on successful testing.
    SterilizationAchieves a sterility assurance level (SAL) of 10^-6.Validated per ISO 11135:2014+A1:2018 (SAL of 10^-6 achieved).
    Residuals (EO, ECH)Meets requirements for limited exposure devices (< 24 hrs) in acuerdo con ISO 10993-7:2008.Residue levels will not exceed 4 mg EO /device or 9 mg ECH/device.
    Sheath PerformanceMeets applicable sections of ISO 11070:2014, 10555-1:2013+A1:2017, ISO 594-1:1986, ISO 594-2:1998, USP 788 and internal test methods.Successfully performed; provides reasonable assurance device meets intended use.
    Dilator PerformanceMeets applicable sections of ISO 11070:2014, 10555-1:2013+A1:2017, USP 788 and internal test methods.Successfully performed; provides reasonable assurance device meets intended use.
    Hemostasis Valve PerformanceMeets applicable sections of ISO 80369-7:2016, USP 788 and internal test methods.Successfully performed; provides reasonable assurance device meets intended use.
    Packaging IntegrityMeets requirements of ISO 11607-1:2019 and ISO 11607-2:2019.Successfully performed; provides reasonable assurance device meets intended use.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided as the device was evaluated through non-clinical (bench) testing and biocompatibility assessments, not through a clinical study with a "test set" of patient data. The provenance of the bench test data would be the Cordis Corporation's testing facilities.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. Ground truth as typically defined for AI/ML models (e.g., expert consensus on images) was not established for this device's evaluation. Its performance was assessed against engineering standards and specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No expert adjudication for a "test set" was performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's evaluation was defined by established international and internal engineering standards, specifications, and validated test methods (e.g., ISO 10993, ISO 11135, ISO 11070, USP 788). The successful completion of these tests served as the evidence of meeting performance requirements.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1