Search Filters

Search Results

Found 15 results

510(k) Data Aggregation

    K Number
    K230958
    Date Cleared
    2023-05-30

    (56 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    First Glove Blue Nitrile Examination Gloves Powder Free

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    First Glove Blue Nitrile Examination Gloves Powder Free is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    First Glove Blue Nitrile Examination Gloves Powder Free are equivalent to Class I patient examination gloves bearing the product code LZA (21CFR880.6250). They meet all the current specifications listed under the ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application. They are made from Nitrile (NBR)100%. These gloves are blue in color and are powder free. The product is non-sterile, finger textured, ambidextrous with beaded cuff and single use only.

    AI/ML Overview

    This document describes the acceptance criteria and performance of the "First Glove Blue Nitrile Examination Gloves Powder Free" based on non-clinical testing.

    Here's the information requested:

    1. Table of acceptance criteria and the reported device performance:

    CharacteristicStandardAcceptance CriteriaReported Device Performance (Subject Device K230958)
    Dimensions - LengthASTM D6319-19XS: ≥ 220 mm, S: ≥ 220 mm, M: ≥ 230 mm, L: ≥ 230 mm, XL: ≥ 230 mm, XXL: ≥ 230 mmXS: 247.00 mm, S: 247.23 mm, M: 252.54 mm, L: 250.92 mm, XL: 245.54 mm, XXL: 251.00 mm
    Dimensions - WidthASTM D6319-19XS: 70±10 mm, S: 80±10 mm, M: 95±10 mm, L: 110±10 mm, XL: 120±10 mm, XXL: 130±10 mmXS: 74.31 mm, S: 85.08 mm, M: 94.85 mm, L: 105.38 mm, XL: 110.77 mm, XXL: 121.31 mm
    ThicknessASTM D6319-19Palm: ≥ 0.05 mm, Finger: ≥ 0.05 mm (for all sizes)Palm: XS 0.07, S 0.06, M 0.07, L 0.06, XL 0.07, XXL 0.07 (mm); Finger: XS 0.10, S 0.10, M 0.10, L 0.10, XL 0.11, XXL 0.10 (mm)
    Physical Properties - Tensile Strength (Before Ageing)ASTM D6319-19≥ 14 MPa (for all sizes)XS: 35.25 MPa, S: 29.47 MPa, M: 27.72 MPa, L: 30.12 MPa, XL: 36.26 MPa, XXL: 33.60 MPa
    Physical Properties - Tensile Strength (After Ageing)ASTM D6319-19≥ 14 MPa (for all sizes)XS: 35.75 MPa, S: 30.02 MPa, M: 28.75 MPa, L: 29.55 MPa, XL: 31.20 MPa, XXL: 35.66 MPa
    Physical Properties - Ultimate Elongation (Before Ageing)ASTM D6319-19≥ 500% (for all sizes)XS: 537%, S: 525%, M: 525%, L: 523%, XL: 528%, XXL: 533%
    Physical Properties - Ultimate Elongation (After Ageing)ASTM D6319-19≥ 400% (for all sizes)XS: 493%, S: 478%, M: 516%, L: 495%, XL: 485%, XXL: 502%
    Freedom from HolesASTM D5151-19AQL 2.5Passes AQL 2.5
    Powder Free ResidueASTM D6124-06 (Reapproved 2017)≤ 2 mg/gloveXS: 0.04 mg/glove, S: 0.28 mg/glove, M: 0.04 mg/glove, L: 0.04 mg/glove, XL: 0.26 mg/glove, XXL: 0.48 mg/glove
    Biocompatibility - Primary Skin IrritationANSI AAMI ISO 10993-10:2010/(R)2014Not an irritantConsidered a non-irritant
    Biocompatibility - Dermal SensitizationANSI AAMI ISO 10993-10:2010/(R)2014Not a sensitizerNot considered to be a contact sensitizer
    Biocompatibility - In vitro CytotoxicityANSI AAMI ISO 10993-5:2009/(R)2014Non-cytotoxicCytotoxic at extract concentrations of 100%, 66.7%, 44.4%, 29.6% and 19.8%; non-cytotoxic at 13.2%
    Biocompatibility - Acute Systemic ToxicityANSI AAMI ISO 10993-11:2017Device extracts do not pose a systemic toxicity concernDid not induce any systemic toxicity

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the specific sample sizes used for each individual test (e.g., number of gloves tested for dimensions, tensile strength, etc.). However, it indicates the tests were conducted according to the specified ASTM and ISO standards, which typically include defined sampling plans.

    The data provenance is from non-clinical bench testing. The country of origin of the data is not explicitly stated in the provided text, but the applicant's address is Malaysia (First Glove Sdn. Bhd.).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable as the evaluation is based on non-clinical, objective laboratory tests against established international standards (ASTM and ISO). There were no human expert assessments of images or patient data to establish ground truth in this context.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    None. Adjudication methods are typically relevant for studies involving human interpretation or subjective assessments. This submission relies on objective physical, chemical, and biological tests.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a submission for medical gloves, which are physical devices. MRMC comparative effectiveness studies are used for AI-powered diagnostic or interpretive tools, not for devices like gloves.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This is not an algorithm or AI-powered device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the non-clinical tests is based on the objective and quantifiable criteria defined by the relevant ASTM and ISO international standards. For example, for tensile strength, the ground truth is a specific numerical value (e.g., ≥ 14 MPa). For biocompatibility, it's defined by biological endpoints (e.g., "non-irritant," "non-sensitizer," "non-cytotoxic").

    8. The sample size for the training set:

    Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable. As mentioned above, there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230601
    Date Cleared
    2023-05-28

    (86 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Blue Nitrile Examination Gloves Powder Free

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Blue Nitrile Examination Gloves Powder Free is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Blue Nitrile Examination Gloves Powder Free are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250).The gloves are made from acrylonitrile-butadiene copolymer dispersion. These gloves are blue in color, powder free and non sterile. The device is available in Extra Small, Small, Medium, Large and Extra Large sizes.

    AI/ML Overview

    The provided document details the acceptance criteria for "Blue Nitrile Examination Gloves Powder Free" and presents the results of non-clinical testing to demonstrate the device meets these criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicStandardAcceptance Criteria (Predicate / Subject Device)Reported Device Performance (Subject Device)
    Dimensions
    LengthASTM D6319-2019Min 230 mm for all sizesX-Small: 245 mm, Small: 245 mm, Medium: 246 mm, Large: 248 mm, X-Large: 248 mm
    WidthASTM D6319-2019X-small: 70+/-10 mm, Small: 80+/-10 mm, Medium: 95+/-10 mm, Large: 110+/-10 mm, X-Large: 120+/-10 mmX-small: 76 mm, Small: 84 mm, Medium: 94 mm, Large: 105 mm, X-Large: 115 mm
    Physical Properties
    ThicknessASTM D6319-2019Palm min 0.05 mm, Finger min 0.05 mm for all sizesPalm: 0.09 mm (all sizes), Finger: 0.15 mm (all sizes)
    Tensile Strength (Before Aging)ASTM D6319-2019Min 14 Mpa for all sizesX-Small: 22.76 Mpa, Small: 22.77 Mpa, Medium: 24.46 Mpa, Large: 24.51 Mpa, X-Large: 24.59 Mpa
    Tensile Strength (After Aging)ASTM D6319-2019Min 14 Mpa for all sizesX-Small: 20.50 Mpa, Small: 20.50 Mpa, Medium: 21.81 Mpa, Large: 21.95 Mpa, X-Large: 22.05 Mpa
    Ultimate Elongation (Before Aging)ASTM D6319-2019Min 500% for all sizesX-Small: 884%, Small: 885%, Medium: 888%, Large: 891%, X-Large: 892%
    Ultimate Elongation (After Aging)ASTM D6319-2019Min 400% for all sizesX-Small: 867%, Small: 868%, Medium: 870%, Large: 872%, X-Large: 875%
    Water Tight (Detection of Holes)ASTM D5151-2019Passes AQL-1.5Gloves Passes AQL 1.5
    Powder ResidueASTM D6124-06 (Reapproved 2017)≤2 mg/gloveX-small: 0.16 mg/glove, Small: 0.16 mg/glove, Medium: 0.16 mg/glove, Large: 0.16 mg/glove, X-Large: 0.16 mg/glove
    Biocompatibility
    Primary Skin IrritationISO 10993-10Under the condition of study not an irritantUnder the condition of study not an irritant
    Dermal SensitizationISO 10993-10Under the conditions of the study not a sensitizerUnder the conditions of the study not a sensitizer
    In vitro CytotoxicityISO 10993-5:2009Under the conditions of study non cytotoxicUnder the conditions of the study cytotoxic (Note: This result is listed as "cytotoxic" while the acceptance criteria is "non cytotoxic", which could be a typo or indicates further consideration is needed).
    Acute Systemic ToxicityISO 10993-11:2017Under the conditions of study the device extracts do not pose a systemic toxicity concernUnder the conditions of study the device extracts do not pose a systemic toxicity concern
    Material Mediated PyrogenicityISO 10993-11:2017/USP 41\Under the conditions of the study, the device did not demonstrate a material mediated pyrogenicity responseUnder the conditions of the study, the device did not demonstrate a material mediated pyrogenicity response

    From the table, the device performance meets or exceeds the acceptance criteria for all tested characteristics, with the exception of the "In vitro Cytotoxicity" test where the reported result is "cytotoxic" while the acceptance criteria is "non cytotoxic". This discrepancy needs clarification. However, in the "SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS", the predicate also states "cytotoxic" for in vitro cytotoxicity, implying it might be an acceptable characteristic for this type of device or a different interpretation of cytotoxicity specific to gloves.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the specific sample sizes used for each test. For example, for "Water Tight (1000 ml)" test, it mentions "Passes AQL-1.5", which implies a sampling plan was used according to the ASTM standard, but the specific number of gloves tested is not provided. The data provenance (country of origin, retrospective/prospective) is also not specified, beyond the fact that the tests were conducted according to international standards (ASTM, ISO).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable. The device under review is a medical glove, and the evaluation relies on adherence to established physical, chemical, and biological performance standards, not expert interpretation of medical images or cases. Ground truth for these tests is based on objective measurements and established test methodologies defined in the ASTM and ISO standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving interpretation of data (e.g., medical images) where expert consensus is required. The testing for this device involves objective, quantitative measurements against predefined standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The device is a medical glove. MRMC studies are relevant for AI-powered diagnostic tools where human readers (e.g., radiologists) interpret images with or without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a medical glove, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for this device's performance is based on the objective measurements and predefined thresholds within recognized international standards (ASTM D6319-2019, ASTM D5151-2019, ASTM D6124-06, ISO 10993-10, ISO 10993-5:2009, ISO 10993-11:2017, USP 41). For example:

    • Dimensions: Measured length and width against minimum/range requirements.
    • Physical Properties: Measured tensile strength and elongation against minimum percentage requirements.
    • Water Tightness: Statistical sampling and inspection against an Acceptable Quality Level (AQL).
    • Biocompatibility: Observation of biological responses (e.g., irritation, sensitization) following standardized protocols, compared against criteria for "not an irritant" or "not a sensitizer".

    8. The sample size for the training set

    This information is not applicable. The device is a medical glove, not an AI or machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable, as there is no training set for this type of medical device.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs and Fentanyl

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs and Fentanyl Citrate is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. Additionally, the gloves were tested for use with chemotherapy drugs and Fentanyl Citrate in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.

    Device Description

    The subject device in 510(K) notification is a blue nitrile examination gloves powder free tested for use with Chemotherapy drugs and Fentanyl Citrate. The subject device is a patient examination glove made from acrylonitrile-butadiene copolymer dispersion, blue color, powder free and non sterile (as per 21CFR 880.6250, class I). The device is available in Small, Medium, Large and Extra Large sizes. The subject device meets all the current specifications listed under the ASTM Specification D 6319 -2019, Standard Specification for Nitrile Examination Gloves for Medical Application. This device also complies with requirements for standard practice for assessment of resistance of medical gloves to permeation by chemotherapy drugs as per ASTM D6978- 05(2019)

    AI/ML Overview

    The provided document is a 510(k) premarket notification for "Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs and Fentanyl Citrate." This document outlines the acceptance criteria and the results of a study to demonstrate the device's performance.

    Here's the breakdown of the information requested based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document provides extensive tables for the performance of the gloves. Rather than recreating the entire tables found on pages 10-14, a summary of key acceptance criteria and reported performance is presented below. Please refer to pages 10-14 of the original document for the complete lists of chemotherapy drugs and specific results.

    Test Method (Standard)PurposeAcceptance CriteriaReported Device Performance
    ASTM D6319-2019Length of the glovesMin 220 mm for Size Small & Min 230 mm for all other sizesSmall: 404 mm
    Medium: 405 mm
    Large: 405 mm
    X-Large: 406 mm (All exceed minimums)
    ASTM D6319-2019Width of the glovesSmall: 80+/-10 mm; Medium: 95+/-10 mm; Large: 110+/-10 mm; X-Large: 120+/-10 mmSmall: 84 mm
    Medium: 94 mm
    Large: 105 mm
    X-Large: 115 mm (All within +/- 10mm range)
    ASTM D6319-2019Thickness of the glovesPalm 0.05 mm min; Finger 0.05 mm min for all sizesPalm: 0.19mm (all sizes); Finger: 0.21mm (all sizes) (All exceed minimums)
    ASTM D6319-2019Physical properties - Tensile strength (Before Ageing)$\ge$ 14 Mpa for all sizesSmall: 22.77Mpa; Medium: 24.46Mpa; Large: 24.51Mpa; X-Large: 24.59Mpa (All exceed minimum)
    ASTM D6319-2019Physical properties - Tensile strength (After Ageing)$\ge$ 14 Mpa for all sizesSmall: 20.50Mpa; Medium: 21.81Mpa; Large: 21.95Mpa; X-Large: 22.05Mpa (All exceed minimum)
    ASTM D6319-2019Physical properties - Ultimate Elongation (Before Ageing)$\ge$ 500% for all sizesSmall: 885%; Medium: 888%; Large: 891%; X-Large: 892% (All exceed minimum)
    ASTM D6319-2019Physical properties - Ultimate Elongation (After Ageing)$\ge$ 400% for all sizesSmall: 868%; Medium: 870%; Large: 872%; X-Large: 875% (All exceed minimum)
    ASTM D5151-2019Detection of Holes in Medical GlovesAQL 2.5Passes AQL 1.5 (Better than acceptance criteria, indicating fewer defects)
    ASTM D6124-06Residual Powder on Medical Gloves2 Mg/Glove MaxSmall: 0.16 mg/glove; Medium: 0.16 mg/glove; Large: 0.16 mg/glove; X-Large: 0.16mg/glove (All meet or are below maximum)
    ASTM D6978-05 (2019)Permeation by Chemotherapy Drugs and Fentanyl CitrateSpecified minimum breakthrough detection times for each drug (e.g., >240 Minutes for most)Most chemotherapy drugs and Fentanyl Citrate showed >240 Minutes breakthrough detection time.
    Exceptions: Carmustine (3.3 mg/ml) - 35.0 Minutes; Thiotepa (10 mg/ml) - 64.9 Minutes. Warnings are provided for these two drugs. (See pages 2-3 and 12-13 for full list and values).
    ISO 10993-10Dermal Irritation (Rabbits)Non irritantUnder the condition of study not an irritant (Meets criteria)
    ISO 10993-10Skin Sensitization (Guinea Pig)Non sensitizerUnder the conditions of the study not a sensitizer (Meets criteria)
    ISO 10993-5:2009In vitro CytotoxicityNon cytotoxicUnder the conditions of the study cytotoxic. (Does NOT meet criteria for non-cytotoxic)
    ISO 10993-11:2017Acute Systemic ToxicityNot pose systemic toxicity concernUnder the conditions of study the device extracts do not pose acute systemic toxicity concern (Meets criteria)
    ISO 10993-11:2017 / USP 41Material-Mediated PyrogenicityNot pose material mediated pyrogenicity responseUnder the conditions of the study, the device did not demonstrate a material mediated pyrogenicity response (Meets criteria)

    Note on Cytotoxicity: The device itself is reported as cytotoxic under the tested conditions. This is a critical finding that does not meet a "Non cytotoxic" acceptance criterion. However, the FDA has still cleared the device. This implies that either the level of cytotoxicity observed is deemed acceptable for the intended use and risk-benefit profile of examination gloves, or mitigating factors/information were provided elsewhere (not in this summary). It's important to note this discrepancy.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document lists various ASTM and ISO standards for testing, which define the sample sizes required for each test (e.g., AQL levels for watertightness). However, the specific quantity/sample size used for each individual test conducted by the manufacturer is not explicitly stated in this summary document beyond what's implied by the standard (e.g., AQL 1.5 for watertightness performance is a characteristic, not a sample size).

    • Data Provenance: The manufacturing location of the device owner, JR Engineering & Medical Technologies (M) Sdn. Bhd., is in Malaysia (Lot 8 &10, Jalan Zurah 3 & Lot 1&3, Jalan Zurah 3A/1, Pusat Perindustrian 2, 44200 Rasa, Hulu Selangor, Selangor Darul Ehsan, Malaysia). The testing was conducted in accordance with international standards (ASTM, ISO), implying that the data originates from laboratory studies, likely conducted by third-party labs or the manufacturer's own qualified labs, but the specific country of the testing labs is not mentioned.
    • Retrospective or Prospective: The testing described is prospective laboratory performance testing of the device, rather than a retrospective analysis of existing data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This product is a medical glove, not an AI or imaging device requiring expert consensus for ground truth. The "ground truth" for this device's performance is established through objective, standardized laboratory testing (e.g., ASTM D6319 for physical properties, ASTM D6978 for chemical permeation, ISO 10993 for biocompatibility). These tests rely on chemical and physical measurements, not human expert interpretation of images or clinical outcomes. Therefore, the concept of "number of experts used to establish ground truth" with qualifications like "radiologist with 10 years of experience" is not applicable to this type of device. The "experts" are the technicians and scientists performing the standardized validation tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As this is not a study involving human interpretation of data where adjudication would be necessary (like in imaging diagnosis), there is no adjudication method described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physical examination glove, not an AI or imaging diagnostic tool that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI device. The tests performed are "standalone" in the sense that they assess the intrinsic material properties and barrier performance of the glove itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for this device is based on standardized, objective physical, chemical, and biological measurements obtained through laboratory testing, as defined by relevant ASTM and ISO standards. This includes:

    • Physical measurements (e.g., length, width, thickness, tensile strength, elongation).
    • Chemical analysis (e.g., residual powder, breakthrough detection time for drugs).
    • Biological testing (e.g., irritation, sensitization, cytotoxicity, systemic toxicity, pyrogenicity).

    8. The sample size for the training set

    Not applicable. This device is a manufactured product (glove), not an AI algorithm or model that requires a "training set" of data.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221226
    Date Cleared
    2022-06-16

    (49 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    1st Glove Blue Nitrile Examination Gloves Powder Free

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K212584
    Date Cleared
    2022-05-28

    (285 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Blue Nitrile Examination Gloves Powder Free

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Blue Nitrile Examination Gloves Powder Free is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Blue Nitrile Examination Gloves Powder Free are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250). The gloves are made from acrylonitrilebutadiene copolymer dispersion. These gloves are blue in color and are powder free.

    AI/ML Overview

    The provided text is a 510(k) summary for "Blue Nitrile Examination Gloves Powder Free". It details the device's characteristics, intended use, and a comparison to a predicate device. The document explicitly states that clinical testing is not applicable and that clinical data is not needed for gloves or for most devices cleared by the 510(K) process.

    Therefore, there is no information in the provided text regarding a study that proves the device meets acceptance criteria related to AI or human-in-the-loop performance, as this is a physical medical device (examination gloves), not an AI-powered diagnostic or assistive tool.

    However, I can extract the acceptance criteria and reported performance for the non-clinical tests conducted on the device, which are relevant for demonstrating its substantial equivalence to the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical)

    CharacteristicStandardAcceptance CriteriaReported Device Performance
    DimensionsASTM D6319-2019
    LengthMin 230 mm for all sizesX-Small: 303 mm
    Small: 304 mm
    Medium: 305mm
    Large: 306mm
    X-Large: 307mm
    Width (for specific sizes)X-small: 70+/-10 mm
    Small: 80+/-10 mm
    Medium: 95+/-10mm
    Large: 110+/-10 mm
    X-Large: 120+/-10 mmX-small: 74 mm
    Small: 84 mm
    Medium: 94 mm
    Large: 104 mm
    X-Large: 115 mm
    Physical PropertiesASTM D6319-2019
    Before Aging:
    Tensile Strength14 Mpa Min for all sizesX-Small: 18.44 Mpa
    Small: 18.49 Mpa
    Medium: 18.56 Mpa
    Large: 18.62 Mpa
    X-Large: 18.64 Mpa
    Ultimate Elongation500% Min for all sizesX-Small: 683%
    Small: 687%
    Medium: 692%
    Large: 695%
    X-Large: 698%
    After Aging:
    Tensile Strength14 Mpa Min for all sizesX-Small: 17.49 Mpa
    Small: 17.53 Mpa
    Medium: 17.65 Mpa
    Large: 17.71 Mpa
    X-Large: 17.76 Mpa
    Ultimate Elongation400% Min for all sizesX-Small: 656%
    Small: 660%
    Medium: 663%
    Large: 666%
    X-Large: 669%
    ThicknessASTM D6319-2019Palm 0.05 mm min
    Finger 0.05 mm minPalm: 0.10 mm for all sizes
    Finger: 0.12 mm for all sizes
    Water TightnessASTM D5151-2019AQL-1.5Passes AQL 1.5
    Powder ResidueASTM D6124-06 (Reapproved 2017)2 mg/glove MaxX-Small: 0.21 mg/glove
    Small: 0.21 mg/glove
    Medium: 0.22 mg/glove
    Large: 0.22 mg/glove
    X-Large: 0.22 mg/glove
    Biocompatibility
    Primary Skin IrritationISO 10993-10Under the condition of study not an irritantUnder the condition of study not an irritant
    Dermal SensitizationISO 10993-10Under the conditions of the study not a sensitizerUnder the conditions of the study not a sensitizer
    In vitro CytotoxicityISO 10993-5:2009Under the conditions of study non cytotoxicUnder the conditions of the study cytotoxic.
    Acute Systemic ToxicityISO 10993-11:2017Under the conditions of study the device extracts do not pose a systemic toxicity concernUnder the conditions of study the device extracts do not pose a systemic toxicity concern
    Material Mediated PyrogenicityISO 10993-11:2017 / USP 41Under the conditions of the study, the device did not demonstrate a material mediated pyrogenicity response.Under the conditions of the study, the device did not demonstrate a material mediated pyrogenicity response.

    Regarding the other requested information (2-9), as a physical medical device (gloves) and not an AI/software device, these questions are not applicable to the context of the provided document.

    • 2. Sample size used for the test set and the data provenance: Not applicable. (These are physical tests on glove samples, not a test set for AI performance).
    • 3. Number of experts and qualifications: Not applicable. (These are standardized physical and biological tests).
    • 4. Adjudication method: Not applicable. (These are standardized physical and biological tests).
    • 5. Multi-reader multi-case (MRMC) comparative effectiveness study: Not applicable. This is for AI-assisted image interpretation or similar.
    • 6. Standalone (algorithm only without human-in-the-loop performance): Not applicable. This is for AI algorithms.
    • 7. Type of ground truth used: Not applicable in the AI sense. The "ground truth" for these tests comes from established industry standards (ASTM, ISO, USP) for physical and chemical properties and biological responses.
    • 8. Sample size for the training set: Not applicable. Applies to AI/machine learning.
    • 9. How the ground truth for the training set was established: Not applicable. Applies to AI/machine learning.
    Ask a Question

    Ask a specific question about this device

    K Number
    K212850
    Date Cleared
    2021-12-10

    (94 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ComfortPro Blue Nitrile Examination Gloves Powder Free

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ComfortPro Blue Nitrile Examination Gloves Powder free is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    ComfortPro Blue Nitrile Examination Gloves Powder Free is Class I patient examination gloves bearing the product code LZA (21CFR880.6250). The gloves are made from Nitrile (NBR)100%. These gloves are blue in color, non-sterile, ambidextrous, powder free and single use only. Blue Nitrile Examination Gloves Powder Free are available in sizes X-small, Small, Medium, Large and X Large.

    AI/ML Overview

    The provided document is a 510(k) summary for the ComfortPro Blue Nitrile Examination Gloves Powder Free. It describes the device, its intended use, and a comparison to a predicate device based on non-clinical performance data.

    Here's the breakdown of the acceptance criteria and study information as requested:

    1. A table of acceptance criteria and the reported device performance

    Test MethodPurposeAcceptance Criteria (Predicate / Standard)Reported Device Performance (Subject Device)
    Bench Test Data
    ASTM D6319-19 (Length)To determine the length of the glovesMin 230 mm for all sizesX-Small: 250mm, Small: 241mm, Medium: 245mm, Large: 242mm, X-Large: 242mm
    ASTM D6319-19 (Width)To determine the width of the glovesX-Small: 70+/-10 mm, Small: 80+/-10mm, Medium: 95+/-10 mm, Large: 110+/-10 mm, X Large: 120+/-10 mmX-Small: 80 mm, Small: 82mm, Medium: 93mm, Large: 103mm, X-Large: 110mm
    ASTM D6319-19 (Thickness)To determine the thickness of the glovesPalm: 0.05 mm min, Finger: 0.05 mm min (for all sizes)Palm (mm): X-Small: 0.101, Small: 0.081, Medium: 0.091, Large: 0.092, X-Large: 0.091
    Finger (mm): X-Small: 0.121, Small: 0.097, Medium: 0.108, Large: 0.108, X-Large: 0.111
    ASTM D6319-19 (Tensile Strength)To Determine the physical properties - Tensile strengthBefore Ageing: Tensile Strength 14 Mpa Min for all sizes
    After Ageing: Tensile Strength 14 Mpa Min for all sizesBefore Ageing (Mpa): X-Small: 27.2, Small: 23.2, Medium: 24, Large: 26.2, X-Large: 27.4
    After Ageing (Mpa): X-Small: 24.2, Small: 23.8, Medium: 27.6, Large: 27.6, X-Large: 27.9
    ASTM D6319-19 (Ultimate Elongation)To Determine the physical properties - Ultimate ElongationBefore Ageing: Ultimate Elongation 500% Min for all sizes
    After Ageing: Ultimate Elongation 400% Min for all sizesBefore Ageing (%): X-Small: 620, Small: 590, Medium: 570, Large: 570, X-Large: 550
    After Ageing (%): X-Small: 600, Small: 560, Medium: 540, Large: 540, X-Large: 550
    ASTM D5151-19 (Water Tight)To determine the holes in the glovesAQL 2.5Gloves Passes AQL 2.5
    ASTM D6124-06 (Residual Powder)To determine the residual powder in the gloves≤2 Mg/Glove0.49 mg/glove
    Biocompatibility Data
    ISO 10993-10:2010 (Primary Skin Irritation)To evaluate the test item, for skin irritation test in New Zealand White rabbits.Under the condition of study not an irritantUnder the condition of study, not an irritant
    ISO 10993-10:2010 (Dermal Sensitization)To evaluate the test item, for the Skin Sensitization in Guinea pigs by maximization test.Under the conditions of the study, not a sensitizerUnder the conditions of the study, not a sensitizer
    ISO10993-5:2009 (In vitro cytotoxicity)To evaluate the test item, for its ability to induce cytotoxicity using L-929 mouse fibroblast cells by Elution Method.Under the conditions of the study, non-cytotoxicUnder the conditions of the study cytotoxic for 100% test item extract and non-cytotoxic at 50%, 25%, 12.5% and 6.25% test item extracts. Moreover, under the conditions of the study, non acute systemic toxic.
    ISO 10993-11:2017 (Acute Systemic Toxicity Test)To evaluate the test item, for Acute Systemic Toxicity in Swiss Albino Mice.Under the conditions of study, the device extracts do not pose a systemic toxicity concernUnder the condition of study, the device extracts did not reveal any systemic toxicity.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • The document does not specify the exact sample sizes used for each test (e.g., how many gloves were tested for length, width, tensile strength, etc., or how many animals were used for biocompatibility tests). It only states they meet the relevant ASTM and ISO standards, which likely prescribe minimum sample sizes.
    • Data Provenance: The document does not explicitly state the country of origin of the data or whether it was retrospective or prospective. Given that the manufacturer is "THAI RUBBER INDUSTRY COMPANY LIMITED" in Thailand, it is highly probable the testing was conducted in Thailand or by contractors on behalf of the Thai manufacturer. The nature of these tests (bench testing for physical properties and biocompatibility for materials) suggests they are prospective tests conducted on representative samples of the manufactured device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This question is not applicable to this type of device (nitrile examination gloves) and the studies performed. The "ground truth" for these tests are objective, measurable physical and chemical properties and biological responses, determined by standardized test methods (ASTM, ISO), rather than expert interpretation or consensus.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This question is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies, particularly those involving interpretation of medical images or patient outcomes where subjective expert opinion is involved. The tests for these gloves are objective measurements against predefined standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This question is not applicable. MRMC studies are specific to medical imaging devices, often involving AI, to assess the impact of AI assistance on human reader performance. Examination gloves do not involve "readers" or "AI assistance" in this context.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This question is not applicable. This concept applies to AI-driven algorithms. The ComfortPro Blue Nitrile Examination Gloves are physical medical devices.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" is established by the defined specifications and test methods within the ASTM and ISO standards. For example, the "ground truth" for length is "Min 230 mm," and the test result is a direct measurement. Similarly, for biocompatibility, the "ground truth" is "not an irritant" or "non-cytotoxic," determined by specific biological assays.
    • It's a combination of objective measurements compared against pre-defined normative standards.

    8. The sample size for the training set

    • This question is not applicable. The device is not an AI/ML algorithm that requires a "training set." The performance data is derived from testing physical samples of the manufactured gloves.

    9. How the ground truth for the training set was established

    • This question is not applicable for the same reason as point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K212827
    Date Cleared
    2021-12-06

    (94 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    JR Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs is a disposable device intended for medical purpose that is worn on the examination between patient and examiner.

    Additionally, the gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.

    The tested chemotherapy drugs and their breakthrough detection times are as follows:

    Tested Chemotherapy Drug Name & ConcentrationMinimum Breakthrough Detection Time
    Carmustine (BCNU) (3.3 mg/ml)35.0 Minutes
    Carboplatin (10 mg/ml)>240 Minutes
    Cisplatin (1 mg/ml)>240 Minutes
    Cyclophosphamide (Cytoxan) (20 mg/ml)>240 Minutes
    Dacarbazine (10.0 mg/ml)>240 Minutes
    Doxorubicin HCl (2 mg/ml)>240 Minutes
    Etoposide (20 mg/ml)>240 Minutes
    Fluorouracil (50 mg/ml)>240 Minutes
    Ifosfamide (50 mg/ml)>240 Minutes
    Methotrexate (25 mg/ml)>240 Minutes
    Mitomycin C (0.5 mg/ml)>240 Minutes
    Mitoxantrone (2 mg/ml)>240 Minutes
    Paclitaxel (6 mg/ml)>240 Minutes
    Thiotepa (10 mg/ml)64.9 Minutes
    Vincristine Sulfate (1 mg/ml)>240 Minutes
    Device Description

    The subject device in 510(K) notification is JR Medic blue nitrile examination gloves powder free tested for use with Chemotherapy drugs.

    The subject device is a patient examination glove made from acrylonitrile-butadiene copolymer dispersion, blue color, powder free and non sterile (as per 21CFR 880.6250, class I).

    The subject device meets all the current specifications listed under the ASTM Specification D 6319 -2019, Standard Specification for Nitrile Examination Gloves for Medical Application. This device also complies with requirements for standard practice for assessment of resistance of medical gloves to permeation by chemotherapy drugs as per ASTM D6978- 05(2019)

    AI/ML Overview

    I am sorry, but the provided text does not describe a study involving an AI device or algorithm, and therefore all the requested information about acceptance criteria, study design, sample sizes, ground truth, expert involvement, and MRMC studies cannot be extracted.

    The document is a 510(k) premarket notification for "JR Medic Blue Nitrile Examination Powder Free Gloves tested for use with chemotherapy drugs." It details the non-clinical testing performed on these medical gloves to demonstrate their safety and effectiveness, specifically their resistance to permeation by chemotherapy drugs, in accordance with established ASTM (American Society for Testing and Materials) standards.

    The document focuses on the physical, chemical, and biocompatibility properties of the gloves, not on the performance of a software algorithm or AI.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212308
    Date Cleared
    2021-10-29

    (98 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    SHARK GLOVES Blue Nitrile Examination Gloves Powder Free, INCONGLOVE Blue Nitrile Examination Gloves
    Powder Free

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SHARK GLOVES and INCONGLOVE Blue Nitrile Examination Gloves Powder Free are disposable devices intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    SHARK GLOVES and INCONGLOVE Blue Nitrile Examination Gloves Powder Free are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250). The gloves are made from acrylonitrile-butadiene copolymer dispersion. These gloves are blue in color and are powder free

    AI/ML Overview

    This document describes the acceptance criteria and study results for SHARK GLOVES and INCONGLOVE Blue Nitrile Examination Gloves Powder Free.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical ApplicationTo determine the length of the glovesMin 230 mm for all sizesSmall: Pass, Medium: Pass, Large: Pass, X-Large: Pass
    ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical ApplicationTo determine the width of the glovesSmall: 80+/-10 mm, Medium: 95+/-10mm, Large: 110+/-10 mm, X-Large: 120+/-10 mmSmall: Pass, Medium: Pass, Large: Pass, X-Large: Pass
    ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical ApplicationTo determine the thickness of the glovesPalm 0.05 mm min, Finger 0.05 mm min for all sizesAll sizes: Palm Pass, Finger Pass
    ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application (Physical properties - Tensile Strength)To Determine the physical properties Tensile strengthBefore Ageing: Tensile Strength 14Mpa Min for all sizes. After Ageing: Tensile Strength 14Mpa Min for all sizes.All sizes: Pass (Both before and after ageing)
    ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application (Physical properties - Ultimate Elongation)To Determine the physical properties Ultimate ElongationBefore Ageing: Ultimate Elongation 500% Min for all sizes. After Ageing: Ultimate Elongation 400% Min for all sizes.Small, Medium, Large, X-Large: Pass (Both before and after ageing)
    ASTM D5151-19 Standard Test Method for detection of holes in medical glovesTo determine the holes in the glovesAQL 2.5Pass
    ASTM D6124-06 Standard Test Method for Residual Powder on Medical GlovesTo determine the residual powder in the gloves2 Mg/Glove MaxSmall, Medium, Large, X-Large: Residual Powder Content Pass
    ISO 10993-10:2010 Biological Evaluation of Medical Devices Test for Irritation and Skin Sensitization. Test done for irritation.To determine the potential of the material to produce dermal irritation in RabbitsUnder the condition of study not an irritantUnder the condition of study not an irritant
    ISO10993-10:2010 Biological Evaluation of Medical Devices Test for Irritation and Skin Sensitization. Test done Skin sensitization.To determine the skin sensitization potential of the material in Guinea Pig.Under the conditions of the study not a sensitizerUnder the conditions of the study not a sensitizer
    ISO 10993-5:2009 biological evaluation of medical devices - part 5, tests for in vitro cytotoxicity.To evaluate the in vitro cytotoxic potential of the test item extracts in L-929 mouse fibroblasts cells using elution method.Under the conditions of study non cytotoxicUnder the conditions of the study, cytotoxic. Additional testing was performed to determine if this was a systemic toxicity concern.
    ISO 10993-11:2017 biological evaluation of medical devices - part 11, tests for systemic toxicity.To determine the acute systemic toxicity potential of the test item extracts in Swiss Albino mice.Under the conditions of study the device extracts do not pose a systemic toxicity concernUnder the conditions of study the device extracts do not pose a systemic toxicity concern

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the sample sizes used for each specific test listed (e.g., number of gloves tested for length, width, etc.). However, the tests are conducted in accordance with ASTM and ISO standards, which typically specify sampling plans.

    The data provenance is from non-clinical tests conducted for the SHARK GLOVES and INCONGLOVE Blue Nitrile Examination Gloves Powder Free. The manufacturer and exporter is SRIRUNGRUANG GLOBAL CO., LTD, located in Thailand. The studies are prospective as they were conducted specifically for this 510(k) submission to demonstrate substantial equivalence.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    The document does not mention the use of experts or the establishment of ground truth in the context of clinical or diagnostic studies. The tests described are non-clinical, laboratory-based physical and biocompatibility assessments of the gloves. Therefore, the concept of "ground truth" established by experts, as it would be for diagnostic accuracy studies, is not applicable here. The "ground truth" for these tests is defined by the objective measurement criteria within the specified ASTM and ISO standards.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, these are non-clinical laboratory tests with objective pass/fail criteria based on established standards, not studies requiring expert adjudication of results.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic device, nor is it a study involving human readers or cases. It is a submission for patient examination gloves.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm-based device.

    7. The Type of Ground Truth Used

    For the non-clinical tests, the "ground truth" is based on the specified quantitative and qualitative requirements of the referenced ASTM and ISO standards, such as minimum length, tensile strength, maximum powder residue, and the absence of irritation or sensitization. For the biocompatibility tests that showed cytotoxicity but no systemic toxicity concern, the "ground truth" was established through additional testing to confirm the lack of systemic risk.

    8. The Sample Size for the Training Set

    Not applicable. This is a medical device (examination gloves), not a machine learning or AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202384
    Date Cleared
    2020-12-02

    (104 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Palm Care Blue Nitrile Examination Gloves Powder Free

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Blue Nitrile Examination Gloves Powder Free is disposable devices intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Blue Nitrile Examination Gloves Powder Free are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250). The gloves are made from acrylonitrilebutadiene copolymer dispersion. These gloves are blue in color and are powder free.

    AI/ML Overview

    The provided information is a 510(k) summary for "Blue Nitrile Examination Gloves Powder Free". This document describes the device and its equivalence to a predicate device, based on non-clinical performance testing. It does not describe a study involving an AI/Machine Learning driven medical device, human readers, or the establishment of ground truth by expert consensus for medical image interpretation or similar.

    Therefore, most of the requested information regarding AI/ML device studies (e.g., number of experts, adjudication methods, MRMC studies, standalone algorithm performance, training set details) is not applicable to this document.

    However, I can extract the acceptance criteria and reported device performance from the non-clinical testing summary as requested for a physical device:


    Acceptance Criteria and Device Performance for Blue Nitrile Examination Gloves Powder Free

    This document describes the non-clinical performance of the "Blue Nitrile Examination Gloves Powder Free" and its substantial equivalence to a predicate device (K192333). The study performed is a series of laboratory tests to ensure the physical and chemical properties of the gloves meet established standards.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D6319-10 (Reapproved 2015) Standard Specification for Nitrile Examination Gloves for Medical ApplicationTo determine the length of the glovesMin 230 mm for all sizesX-Small: 305 mm
    Small: 305 mm
    Medium: 305mm
    Large: 306mm
    X-Large: 306mm
    ASTM D6319-10 (Reapproved 2015) Standard Specification for Nitrile Examination Gloves for Medical ApplicationTo determine the width of the glovesX-small: 70+/-10 mm
    Small: 80+/-10 mm
    Medium: 95+/-10mm
    Large: 110+/-10 mm
    X-Large: 120+/-10 mmX-small: 74 mm
    Small: 84 mm
    Medium: 94 mm
    Large: 105 mm
    X-Large: 115 mm
    ASTM D6319-10 (Reapproved 2015) Standard Specification for Nitrile Examination Gloves for Medical ApplicationTo determine the thickness of the glovesPalm 0.05 mm min
    Finger 0.05 mm min for all sizesPalm
    X-Small: 0.10mm
    Small: 0.10mm
    Medium: 0.10mm
    Large: 0.10mm
    X-Large: 0.10mm

    Finger
    X-Small: 0.13mm
    Small: 0.13mm
    Medium: 0.14mm
    Large: 0.13mm
    X-Large: 0.14mm |
    | ASTM D5151-06 (Reapproved 2015) Standard Test Method for Detection of Holes in Medical Gloves | To determine the holes in the gloves | AQL 1.5 | Passes AQL 1.5 |
    | ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves | To determine the residual powder in the gloves | 2 Mg/Glove Max | Residual Powder Content
    X-Small: 0.21mg/glove
    Small: 0.21mg/glove
    Medium: 0.22 mg/glove
    Large: 0.22 mg/glove
    X-Large: 0.22 mg/glove |
    | ASTM D6319-10 (Reapproved 2015) Standard Specification for Nitrile Examination Gloves for Medical Application | To Determine the physical properties - Tensile strength | Before Ageing: Tensile Strength 14Mpa Minimal for all sizes
    After Ageing: Tensile Strength 14Mpa Minimal for all sizes | Tensile Strength (Mpa)
    Before Ageing
    X-Small: 18.43
    Small: 18.49
    Medium: 18.54
    Large: 18.57
    X-Large: 18.64

    After Ageing
    X-Small: 17.49
    Small: 17.51
    Medium: 17.65
    Large: 17.67
    X-Large: 17.71 |
    | ASTM D6319-10 (Reapproved 2015) Standard Specification for Nitrile Examination Gloves for Medical Application | To Determine the physical properties - Ultimate Elongation | Before Ageing: Ultimate Elongation 500% Min for all sizes
    After Ageing: Ultimate Elongation 400% Min for all sizes | Ultimate Elongation (%)
    Before Ageing
    X-Small: 685
    Small: 688
    Medium: 692
    Large: 695
    X-Large: 698

    After Ageing
    X-Small: 654
    Small: 658
    Medium: 661
    Large: 665
    X-Large: 669 |
    | Biocompatibility (Primary Skin Irritation - ISO 10993-10:2010(E)) | Biocompatibility | Under the condition of study not an irritant | Under the condition of study not an irritant |
    | Biocompatibility (Dermal Sensitization - ISO 10993-10:2010(E)) | Biocompatibility | Under the conditions of the study not a sensitizer | Under the conditions of the study not a sensitizer |
    | Biocompatibility (In vitro cytotoxicity - ISO10993-5:2009(E)) | Biocompatibility | (Predicate: Under the conditions of the study, cytotoxic) (Subject: Under the conditions of the study, non-cytotoxic) | Under the conditions of the study, non-cytotoxic |
    | Biocompatibility (Acute Systemic Toxicity Test ISO 10993-11:2017(E)) | Biocompatibility | Under the condition of study not systemic toxic | Under the conditions of study, the device extracts do not pose a systemic toxicity concern |
    | Biocompatibility (Material Mediated Pyrogenicity ISO 10993-11:2017(E) / USP 41) | Biocompatibility | Under the conditions of the study non pyrogenic | Under the conditions of the study non pyrogenic |
    | Biocompatibility (Bacterial Endotoxin test USP 42) | Biocompatibility | (Predicate: No data available) (Subject:

    Ask a Question

    Ask a specific question about this device

    K Number
    K192333
    Date Cleared
    2020-01-24

    (150 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Blue Nitrile Examination Gloves Powder Free

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Blue Nitrile Examination Gloves Powder Free is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    JR MEDIC Blue Nitrile Examination Gloves Powder Free are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250). The gloves are made from acrylonitrile-butadiene copolymer dispersion. These gloves are blue in color and are powder free.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for "Blue Nitrile Examination Gloves Powder Free." It details the device's technical characteristics, non-clinical performance, and comparison to a predicate device to demonstrate substantial equivalence.

    This document describes the testing of a medical device (examination gloves), which
    Does not involve AI, machine learning, or complex algorithmic performance in its functionality or assessment. The tests performed are
    physical, chemical, and biological compatibility tests of the material itself.

    Therefore, the requested information regarding acceptance criteria and study proving AI/ML device performance (such as sample size for test/training sets, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, etc.) is not applicable to this document. The document describes the data and testing for a Class I medical device (gloves) and demonstrates its "substantial equivalence" to an existing predicate device, primarily through non-clinical bench testing and biocompatibility studies.

    However, I can extract the acceptance criteria and reported performance for the physical and biological properties of the gloves, as provided in the document.


    Acceptance Criteria and Reported Device Performance for Blue Nitrile Examination Gloves Powder Free

    Since this is a physical medical device (gloves) and not an AI/ML-driven device, the criteria typically associated with AI/ML performance (e.g., AUC, sensitivity, specificity for a diagnostic algorithm) are not relevant here. Instead, the acceptance criteria relate to the physical properties, chemical composition, and biocompatibility of the gloves, demonstrating they meet established standards for medical examination gloves.

    Here's the table of acceptance criteria and reported device performance based on the non-clinical testing summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicTest MethodPurposeAcceptance CriteriaReported Device Performance
    Physical Properties
    LengthASTM D6319-10 (Reapproved 2015)To determine the length of the glovesMin 230 mm for all sizesX-Small: 404 mm
    Small: 404 mm
    Medium: 405 mm
    Large: 404 mm
    X-Large: 404 mm
    WidthASTM D6319-10 (Reapproved 2015)To determine the width of the glovesX-small: 70+/-10 mm
    Small: 80+/-10 mm
    Medium: 95+/-10 mm
    Large: 110+/-10 mm
    X-Large: 120+/-10 mmX-small: 74 mm
    Small: 84 mm
    Medium: 94 mm
    Large: 105 mm
    X-Large: 115 mm
    ThicknessASTM D6319-10 (Reapproved 2015)To determine the thickness of the glovesPalm 0.05 mm min
    Finger 0.05 mm min for all sizesPalm:
    X-Small: 0.20mm
    Small: 0.19mm
    Medium: 0.20mm
    Large: 0.20mm
    X-Large: 0.20mm
    Finger:
    X-Small: 0.22mm
    Small: 0.21mm
    Medium: 0.22mm
    Large: 0.21mm
    X-Large: 0.22mm
    Tensile StrengthASTM D6319-10 (Reapproved 2015)To determine the physical propertiesBefore Ageing: 14Mpa Min for all sizes
    After Ageing: 14Mpa Min for all sizesBefore Ageing:
    X-Small: 22.0Mpa
    Small: 23.0Mpa
    Medium: 25.6Mpa
    Large: 24.0Mpa
    X-Large: 24.5Mpa
    After Ageing:
    X-Small: 18.5Mpa
    Small: 20.5Mpa
    Medium: 22.0Mpa
    Large: 21.0Mpa
    X-Large: 21.7Mpa
    Ultimate ElongationASTM D6319-10 (Reapproved 2015)To determine the physical propertiesBefore Ageing: 500% Min for all sizes
    After Ageing: 400% Min for all sizesBefore Ageing:
    X-Small: 898%
    Small: 896%
    Medium: 868%
    Large: 899%
    X-Large: 874%
    After Ageing:
    X-Small: 872%
    Small: 861%
    Medium: 828%
    Large: 869%
    X-Large: 853%
    Water Tightness (Holes)ASTM D5151-06 (Reapproved 2015)To determine the holes in the glovesAQL 2.5Gloves pass AQL 1.5 (This meets or exceeds the AQL 2.5 criteria, as a lower AQL indicates fewer allowable defects).
    Residual PowderASTM D6124-06 (Reapproved 2017)To determine the residual powder in the gloves2 mg/glove MaxX-small: 0.16 mg/glove
    Small: 0.20 mg/glove
    Medium: 0.21 mg/glove
    Large: 0.20 mg/glove
    X-Large: 0.21 mg/glove
    Biocompatibility-
    Primary Skin IrritationISO 10993-10:2010(E)To determine the potential to produce dermal irritation in RabbitsUnder the condition of study not an irritantUnder the condition of study not an irritant
    Dermal SensitizationISO 10993-10:2010(E)To determine skin sensitization potential in Guinea PigUnder the conditions of the study not a sensitizerUnder the conditions of the study not a sensitizer
    In vitro CytotoxicityISO 10993-5:2009(E)To evaluate in vitro cytotoxic potential in L-929 mouse fibroblasts cellsUnder the conditions of study non cytotoxicUnder the conditions of the study, cytotoxic. Additional testing was performed to determine if this was a systemic toxicity concern. (This indicates an initial result that was followed up to ensure safety).
    Acute Systemic ToxicityISO 10993-11:2017(E)To determine acute systemic toxicity potential in swiss Albino miceUnder the conditions of study the device extracts do not pose a systemic toxicity concernUnder the conditions of study the device extracts do not pose a systemic toxicity concern. (This addresses the concern from the cytotoxicity test).
    Material Mediated PyrogenicityISO 10993-11:2017(E) / USP 41To determine pyrogenic potential in New Zealand white RabbitsUnder the conditions of the study, the device did not demonstrate a material mediated pyrogenicity responseUnder the conditions of the study, the device did not demonstrate a material mediated pyrogenicity response.

    Since the device is a simple medical examination glove and not an AI/ML system, the following points regarding AI/ML study design are explicitly not applicable to the data provided in this document:

    1. Sample sizes used for the test set and the data provenance: Not applicable. Performance is based on physical and chemical tests on manufactured gloves, not a "test set" of patient data for AI.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for gloves is defined by adherence to physical standards (ASTM) and biological safety (ISO, USP), not expert annotation of medical images/data.
    3. Adjudication method: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: Not applicable. This is for performance assessment of diagnostic/AI tools with human readers.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. There is no algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable in the context of AI/ML. The "ground truth" here is compliance with established international standards for physical properties and biocompatibility of medical gloves.
    7. The sample size for the training set: Not applicable. There is no training set as there is no AI/ML model.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2