510(k) Data Aggregation

    K Number
    K180992
    Device Name
    Biotex
    Date Cleared
    2019-07-16

    (456 days)

    Product Code
    Regulation Number
    878.5035
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Biotex

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Biotex Non-Absorbable PTFE Suture is indicated for use in dental soft tissue approximation and/or ligation. The device is not indicated for use in cardiovascular, ophthalmic surgery, microsurgery or peripheral neural tissue. Biotex is provided sterile as a single use device and the duration of use or contact with the body is less than 30 days.

    Device Description

    Biotex Non-Absorbable PTFE Suture is comprised of a single-arm, non-absorbable monofilament suture and the stainless-steel surgical needle connected to the suture. It is composed of 100% high-density PTFE provided uncoated, undyed and sterile for one-time use only.

    AI/ML Overview

    The Biotex Non-Absorbable PTFE Suture was evaluated for substantial equivalence to a predicate device through mechanical, physical, and biocompatibility testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test CategorySpecific TestAcceptance CriteriaReported Device Performance
    Mechanical/PhysicalDiameterConforms to USP monograph for non-absorbable suturesConformed to USP monograph
    Tensile StrengthConforms to USP monograph for non-absorbable suturesConformed to USP monograph
    Needle AdhesionConforms to USP monograph for non-absorbable suturesConformed to USP monograph
    BiocompatibilityCytotoxicity (ISO 10993-5)Met acceptance criteriaAll acceptance criteria were met
    Sensitization (ISO 10993-10)Met acceptance criteriaAll acceptance criteria were met
    Intracutaneous Reactivity (ISO 10993-10)Met acceptance criteriaAll acceptance criteria were met
    Acute Systemic Toxicity (ISO 10993-11)Met acceptance criteriaAll acceptance criteria were met
    Genotoxicity (ISO 10993-3)Met acceptance criteriaAll acceptance criteria were met
    Implantation (ISO 10993-6)Met acceptance criteriaAll acceptance criteria were met
    SterilizationPyrogen TestingMet acceptance criteriaAll acceptance criteria were met
    EO Sterilization ResidualsMet acceptance criteriaAll acceptance criteria were met
    StabilityShelf-LifeMet acceptance criteriaAll acceptance criteria were met

    2. Sample Size and Data Provenance

    The document does not specify the exact sample sizes used for each test. The studies were performed in accordance with FDA's Class II Special Controls Guidance Document for Surgical Sutures and ISO 10993-1. These are likely prospective evaluations conducted in a laboratory setting, with data provenance typically from the device manufacturer's testing facilities.

    3. Number of Experts and Qualifications

    Not applicable. This is a medical device for approximation and/or ligation of dental soft tissue, not an AI/CADe device. The ground truth for this device is based on standardized physical, mechanical, and biological testing, not expert interpretation of medical images.

    4. Adjudication Method

    Not applicable. This is a medical device for approximation and/or ligation of dental soft tissue, not an AI/CADe device. The assessment relies on objective measurements against established standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is a medical device for approximation and/or ligation of dental soft tissue, not an AI/CADe system.

    6. Standalone Performance Study

    Yes, standalone (algorithm only) performance was done. The performance of the Biotex Non-Absorbable PTFE Suture was assessed independently against established standards such as the USP monograph for non-absorbable sutures and ISO 10993-1 for biocompatibility.

    7. Type of Ground Truth Used

    The ground truth used for performance evaluation was based on:

    • Established Standards: USP monograph for non-absorbable surgical sutures (for mechanical and physical properties).
    • International Standards: ISO 10993-1 "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing Within a Risk Management Process" (for biocompatibility).
    • Internal Specifications: Derived from the above standards, for tests like pyrogenicity, EO residuals, and shelf-life.

    8. Sample Size for the Training Set

    Not applicable. This is a medical device, not an AI/CADe system that requires a "training set." Performance is evaluated against pre-defined, objective criteria.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for physical medical devices of this nature. The ground truth for the performance evaluation relies on established scientific and regulatory standards.