510(k) Data Aggregation

    K Number
    K142332
    Device Name
    BT-400
    Manufacturer
    Date Cleared
    2015-03-20

    (212 days)

    Product Code
    Regulation Number
    880.5700
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BT-400 is the Phototherapy unit intended for the treatment of Neonatal Hyperbilirubinemia caused by Indirect Hyperbilirubinemia.

    Device Description

    BT-400 Neonatal phototherapy is designed to emit the blue light wave length range 450-470nm spectrum to treat Neonatal Jaundice to use in Bassinets, Incubators, Open beds or Radiant warmer and consists of Head ( Light Source ) Control box with LCD window Roll Stand ( optional)

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the BT-400 Neonatal Phototherapy device. It compares the device to a predicate device, the BILI-THERAPY Spot type (K103828), to establish substantial equivalence.

    Here's an analysis of the acceptance criteria and study information provided:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in the sense of predefined thresholds for a study outcome. Instead, it compares the proposed device's performance to that of a legally marketed predicate device to demonstrate substantial equivalence. The "acceptance criteria" here are implied to be that the proposed device performs comparably to or better than the predicate device across key technical and functional characteristics.

    CharacteristicAcceptance Criteria (Implied: Comparable to Predicate)Reported Device Performance (BT-400)Predicate Device Performance (BILI-THERAPY Spot type K103828)
    Light SourceBlue LEDsBlue LEDsBlue LEDs
    WavelengthPeak 450-475nmPeak 450-475nmPeak 450-475nm
    Intensity30-40 uW/cm² at 30cm distance30-45 uW/cm² at 30cm distance30-40 uW/cm² at 30cm distance
    Variation in intensity in 6 hrs±10% in Illumination area±10% in Illumination areaIdentical
    Effective Surface30x20cm40x20cm30x20cm
    Head output over 6 hrs
    Ask a Question

    Ask a specific question about this device