Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K122998
    Date Cleared
    2012-11-06

    (40 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Nitrile Powdered patient examination glove

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for Nitrile Powdered Examination Gloves. It details the device's characteristics, intended use, and a comparison to a predicate device, focusing on nonclinical data.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    CharacteristicAcceptance Criteria (Standard)Reported Device Performance
    DimensionASTM Standard D6319-10Meets
    Physical PropertiesASTM Standard D6319-10Meets (Elongation 500%/500%, Tensile Strength 14MPa/14MPa for before/after aging)
    Freedom from Pinholes21 CFR 800.20; ASTM D5151-06Meets
    Powder ResidualASTM Standard D6124-06Meets (Results generated values below 10 mg/dm² of powder)
    Biocompatibility:
    Primary Skin IrritationISO 10993-10:2010 (in rabbits)Gloves are non-irritating
    Dermal SensitizationISO 10993-10:2010 (in guinea pig)Gloves do not display any potential for sensitization

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each test set that led to the "Meets" determination. It references ASTM and ISO standards for testing, which would outline the appropriate sample sizes and methodologies for each specific test (e.g., how many gloves were tested for pinholes, how many animals for biocompatibility).

    The data provenance is from nonclinical testing conducted by the manufacturer (Northstar Healthcare Holdings) or a designated testing facility on their behalf, designed to demonstrate compliance with established standards. No specific country of origin or retrospective/prospective nature of the data is mentioned beyond the standards themselves.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as the device is a patient examination glove. The "ground truth" for this type of device is established by compliance with standardized physical performance and biocompatibility tests, not by expert interpretation of images or clinical data. The standards (ASTM, ISO) themselves define the criteria for "ground truth" or acceptable performance.

    4. Adjudication method for the test set

    This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies where multiple human readers interpret data, often with conflicting opinions, requiring a consensus mechanism. For the nonclinical testing of gloves, the results are quantitative measurements against predefined thresholds specified in the ASTM and ISO standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This device is a physical medical device (examination glove), not an AI diagnostic or assistance tool. Therefore, MRMC studies and AI assistance are irrelevant to its evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is not an algorithm or an AI system. Its performance is evaluated through physical and biocompatibility testing.

    7. The type of ground truth used

    The ground truth for this device is based on established industry standards (ASTM and ISO), which define objective, measurable criteria for physical properties (dimensions, tensile strength, elongation, freedom from pinholes, powder residual) and biological safety (biocompatibility). The tests measure these characteristics directly against the specified thresholds.

    8. The sample size for the training set

    This information is not applicable. This device is not an AI/ML device that requires a training set. Its performance is assessed through direct testing against predetermined criteria.

    9. How the ground truth for the training set was established

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1