Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K133325
    Date Cleared
    2014-08-08

    (283 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands of fingers to prevent contamination between patient and examiner.

    Device Description

    Classified by FDA's General and Plastic Surgery Device panel as Class 1, 21 CFR 880.6250, Powder-Free Nitrile Patient Examination Glove. 80LZA, and meets all requirement of ASTM Standard D6319-10.

    AI/ML Overview

    This document describes the premarket notification for the "Blue Nitrile Powder-Free Exam Glove" by Shandong Xinshi Medical Products Co., Ltd. The submission aims to demonstrate substantial equivalence to a predicate device.

    Here's the breakdown of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriterionPredicate Device RequirementDevice Performance (Blue Nitrile Powder-Free Exam Glove)
    Mechanical Properties
    Before Aging: Tensile Strength (Mpa)Substantially equivalentAverage 28.77 Mpa
    Before Aging: Ultimate Elongations (%)Substantially equivalentAverage 759.23%
    After Aging: Tensile Strength (Mpa)Substantially equivalentAverage 24.00 Mpa
    After Aging: Ultimate Elongations (%)Substantially equivalentAverage 712.31%
    Physical Dimensions
    Overall Length on Medium SizeSubstantially equivalentAverage 231 mm
    Width of Palm on Medium SizeSubstantially equivalentAverage 95.15 mm
    Palm ThicknessSubstantially equivalentAverage 0.08 mm
    Finger ThicknessSubstantially equivalentAverage 0.09 mm
    Barrier Integrity
    Free of PinholeMeets ASTM D5151-06 (AQL 2.5)Testing result indicates pinhole were found less than two pieces gloves out of 125 pieces gloves. AQL 2.5 is met.
    Material/Processing Residuals
    Residue PowderMeets ASTM D6124-06 (
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1