Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K173354
    Date Cleared
    2017-12-15

    (51 days)

    Product Code
    Regulation Number
    880.5200
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    BD Nexiva Diffusics Closed IV Catheter Systems

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BD Nexiva™ Diffusics™ Closed IV Catheter Systems are intended to be inserted into a patient's vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. These devices are suitable for use with power injectors set to a maximum pressure of 325 psi (2240 kPa) when access ports not suitable for use with power injectors are removed.

    Device Description

    BD Nexiva Diffusics closed IV catheter systems are over-the-needle, intravascular catheters. These devices have a radiopaque BD Vialon catheter with side holes located near the tip of the catheter which are designed to optimize power injection procedures. These devices also have a needle, needle shield, septum, stabilization platform, integrated extension tubing, clamp, Luer connector, and vent plug. The Luer connector displays gauge-specific maximum flow rate and the maximum power injector pressure limit setting. The needle and catheter are protected by a needle cover. An end cap with protective cover is provided in the unit package. The closed system is designed to keep blood contained within the device

    AI/ML Overview

    This document is a 510(k) summary for the BD Nexiva Diffusics Closed IV Catheter System (K173354). It describes a modification to an existing device (predicate device K170283), specifically a change in the extension tubing polyurethane material formulation and manufacturer. The submission aims to demonstrate substantial equivalence to the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/AttributeAcceptance Criteria (Implicit from "met all predetermined acceptance criteria")Reported Device Performance
    Performance TestingNo Leak after Repeated Bending CyclesNo leaks after specified bending cycles.Met predetermined acceptance criteria.
    Extension Set Pull ForceWithstand specified pull force without failure.Met predetermined acceptance criteria.
    Extension Tube Yield / Rupture PressureWithstand specified pressure without yield or rupture.Met predetermined acceptance criteria.
    Pinch Clamp Vacuum OcclusionAchieve specified occlusion under vacuum conditions.Met predetermined acceptance criteria.
    Pinch Clamp Engagement ForceEngage with specified force.Met predetermined acceptance criteria.
    Biocompatibility Testing (per ISO 10993-1:2009)CytotoxicityNon-cytotoxic.Met predetermined acceptance criteria.
    SensitizationNon-sensitizing.Met predetermined acceptance criteria.
    Intracutaneous ReactivityNon-reactive intracutaneously.Met predetermined acceptance criteria.
    Systemic Toxicity (Acute)Non-systemically toxic in acute exposure.Met predetermined acceptance criteria.
    Material-Mediated PyrogenicityNon-pyrogenic.Met predetermined acceptance criteria.
    Subchronic Toxicity (subacute toxicity)Non-toxic in subchronic exposure.Met predetermined acceptance criteria.
    HaemocompatibilityCompatible with blood.Met predetermined acceptance criteria.
    Material/Physical PropertiesParticulate Analysis (surface and fluid path)Meet clinically relevant limits per USP .Met predetermined acceptance criteria.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the numerical sample sizes for each test. It broadly mentions "Summary of Performance tests completed on the subject device were limited to those tests Performance required to support a determination of substantial equivalence to the predicate Tests devices." It also states, "When technological characteristics between the subject and predicate devices were found to be identical, results of performance testing conducted on the predicate devices were applied to the subject device."

    • Sample size: Not explicitly stated. The tests are industrial-standard tests for medical device components.
    • Data provenance: Not explicitly stated, though it implicitly refers to "design verification" and "risk analysis" conducted by the manufacturer, Becton Dickinson Infusion Therapy Systems Inc. The testing results are presented as having met internal, pre-determined criteria. Given the nature of a 510(k) submission for a material change, the testing would be prospective for the modified component.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable to this type of device and submission. The tests are physical, chemical, and biological performance tests for a medical device, not clinical studies requiring expert interpretation of outcomes. The "ground truth" is defined by the technical specifications and standards (e.g., ISO 10993-1, USP ) that the device must meet.

    4. Adjudication Method for the Test Set

    This information is not applicable. The nature of the tests (e.g., measuring pull force, pressure, or biological response) does not involve adjudication by human readers or experts in the sense of clinical image interpretation or diagnostics. The results are objective measurements compared against pre-defined acceptance criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. This is a 510(k) submission for a Class II medical device (intravascular catheter) involving a material change. MRMC studies are typically associated with diagnostic imaging devices where human interpretation is a key component of the clinical workflow.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    No, a standalone performance study was not done. This device is a physical medical device, not a software algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance tests is derived from engineering specifications, recognized industry standards (e.g., ISO 10993-1 for biocompatibility, USP for particulate analysis), and internal design requirements developed by BD according to 21 CFR 820.30 Design Controls. These objective criteria define what constitutes acceptable performance for the device and its material components.

    8. The Sample Size for the Training Set

    This information is not applicable. This is a submission for a physical medical device, not an AI/ML algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161777
    Date Cleared
    2016-08-29

    (61 days)

    Product Code
    Regulation Number
    880.5200
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    BD Nexiva and BD Nexiva Diffusics Closed IV Catheter Systems

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BD Nexiva closed IV catheter systems are intended to be inserted into a patient's vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when access ports not suitable for use with power injectors are removed.

    BD Nexiva Diffusics closed IV catheter systems are intended to be inserted into a patient's vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. These devices are suitable for use with power injectors set to a maximum pressure of 325 psi (2240 kPa) when access ports not suitable for use with power injectors are removed.

    Device Description

    BD Nexiva closed IV catheter systems are over-the-needle, intravascular catheters. These devices have a radiopaque BD Vialon™ catheter, needle, needle shield, septum, stabilization platform, integrated extension tubing, clamp, Luer adapter, vent plug and, with dual port configurations, a pre-attached BD Q-Syte Luer Access Split Septum. The needle and catheter are protected by a needle cover. For dual port configurations, a BD Q-Syte device or an end cap with protective cover is provided in the unit package. The closed system is designed to keep blood contained within the device throughout the insertion process. The septum is designed to wipe visible blood from the needle surface as the needle is withdrawn from the catheter, further reducing the risk of blood exposure. The needle tip is passively protected when the needle is removed, reducing the risk of accidental needlestick injury. These devices have BD Instaflash™ Needle Technology, allowing for immediate visualization of blood along the catheter. Continuous blood return is seen in the extension tubing. The vent plug prevents blood leakage from the extension tubing during insertion. Both the stabilization platform and Luer connector are color coded to indicate catheter gauge size (24GA (0.7 mm)=Yellow, 22GA (0.9 mm)=Blue, 20GA (1.1 mm)=Pink, 18GA (1.3 mm)=Green).

    BD Nexiva Diffusics closed IV catheter systems are over-the-needle, intravascular catheters. These devices have a radiopaque BD Vialon™ catheter with side holes located near the tip of the catheter which are designed to optimize power injection procedures. These devices also have a needle, needle shield, septum, stabilization platform, integrated extension tubing, clamp, Luer connector, and vent plug. The Luer connector displays gauge-specific maximum flow rate and the maximum power injector pressure limit setting. The needle and catheter are protected by a needle cover. An end cap with protective cover is provided in the unit package. The closed system is designed to keep blood contained within the device throughout the insertion process. The septum is designed to wipe visible blood from the needle surface as the needle is withdrawn from the catheter, further reducing the risk of blood exposure. The needle tip is passively protected when the needle is removed, reducing the risk of accidental needlestick injury. These devices have BD Instaflash™ Needle Technology, allowing for immediate visualization of blood along the catheter. Continuous blood return is seen in the extension tubing. The vent plug prevents blood leakage from the extension tubing during insertion. The stabilization platform and Luer connector are color coded to indicate catheter gauge size (24GA (0.7 mm)=Yellow, 22GA (0.9 mm)=Blue, 20GA (1.1 mm)=Pink, 18GA (1.3 mm)=Green).

    AI/ML Overview

    The document describes the BD Nexiva and BD Nexiva Diffusics Closed IV Catheter Systems and their substantial equivalence to predicate devices, based on performance testing.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test NameAcceptance CriteriaReported Device Performance
    Air LeakagePre-determined acceptance criteria (specific values not detailed in the provided text)All pre-determined acceptance criteria were met.
    Liquid LeakagePre-determined acceptance criteria (specific values not detailed in the provided text)All pre-determined acceptance criteria were met.
    Fluid Leakage at Venous PressurePre-determined acceptance criteria (specific values not detailed in the provided text)All pre-determined acceptance criteria were met.
    Septum Assembly Failure / Separation at Max PressurePre-determined acceptance criteria (specific values not detailed in the provided text)All pre-determined acceptance criteria were met.
    Cannula / Hub Bond StrengthPre-determined acceptance criteria (specific values not detailed in the provided text)All pre-determined acceptance criteria were met.
    Strength of Union between Needle Hub and Needle TubePre-determined acceptance criteria (specific values not detailed in the provided text)All pre-determined acceptance criteria were met.
    Force to Remove Needle from Catheter System (Average System Drag)Pre-determined acceptance criteria (specific values not detailed in the provided text)All pre-determined acceptance criteria were met.
    System Drag - Peak Offset ForcePre-determined acceptance criteria (specific values not detailed in the provided text)All pre-determined acceptance criteria were met.

    2. Sample size used for the test set and the data provenance:
    The document does not explicitly state the sample sizes used for each performance test. It notes that "Performance testing performed to support the design modifications included" a list of tests. The data provenance is industrial, conducted by the manufacturer (Becton Dickinson Infusion Therapy Systems Inc.) for regulatory submission to the FDA. The tests are prospective in nature, assessing the physical and mechanical characteristics of the devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    This section is not applicable. The device is a physical medical device (intravascular catheter system), and its performance is evaluated based on objective physical and mechanical tests, not through human interpretation or expert ground truth establishment for a test set in the same way an AI algorithm for image analysis would be.

    4. Adjudication method for the test set:
    This section is not applicable for the reasons outlined in point 3. Performance is determined by meeting pre-determined objective criteria in physical tests.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    This section is not applicable. The device is not an AI-assisted diagnostic tool or an algorithm that human readers would use. It is a medical device for vascular access.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    This section is not applicable for the reasons outlined in point 5.

    7. The type of ground truth used:
    The "ground truth" for the performance tests are the objective engineering and regulatory standards and specifications for the physical and mechanical properties of intravascular catheter systems. These are not expert consensus, pathology, or outcomes data.

    8. The sample size for the training set:
    Not applicable. This is not an AI/ML model, so there is no "training set."

    9. How the ground truth for the training set was established:
    Not applicable. This is not an AI/ML model, so there is no "training set" or ground truth establishment for it.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1