Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K183399
    Date Cleared
    2019-03-20

    (103 days)

    Product Code
    Regulation Number
    880.5200
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BD Nexiva closed IV catheter systems are intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when access ports not suitable for use with power injectors are removed.

    Device Description

    BD Nexiva closed IV catheter systems are over-the-needle, intravascular catheters. These devices have a radiopaque BD Vialon catheter, needle, needle shield, septum, stabilization platform, integrated extension tubing, clamp, Luer adapter (single or dual port), vent plug, and pre-attached needleless connector (BD Q-Syte or MaxZero) (dual port configurations only). The needle and catheter are protected by a needle cover. A BD Q-Syte, MaxZero, or end cap with protective cover is provided in the unit package (not available with all configurations).

    The closed system is designed to keep blood contained within the device throughout the insertion process. The septum is designed to wipe visible blood from the needle surface as the needle is withdrawn from the catheter, further reducing the risk of blood exposure. The needle tip is passively protected when the needle is removed, reducing the risk of accidental needlestick injury.

    These devices have BD Instaflash needle technology, allowing for immediate visualization of blood along the catheter. Continuous blood return is seen in the extension tubing. The vent plug prevents blood leakage from the extension tubing during insertion. Both the stabilization platform and Luer connector are color coded to indicate catheter gauge size (24 GA (0.7 mm) = Yellow, 22 GA (0.9 mm) = Blue, 20 GA (1.1 mm) = Pink, 18 GA (1.3 mm) = Green).

    AI/ML Overview

    This document is a 510(k) summary for the BD Nexiva™ Closed IV Catheter System. It describes the device, its intended use, a comparison to predicate devices, and a summary of performance testing to demonstrate substantial equivalence.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that "Per the design control requirements specified in 21 CFR §820.30, the subject device met all predetermined acceptance criteria for the above-listed performance tests, demonstrating substantial equivalence to the predicate devices." However, it does not explicitly list the quantitative acceptance criteria for each test. It only lists the tests performed and states that the device met the criteria.

    Test PerformedReported Device Performance
    MaxZero Retention Rate (BD internal specification)Met predetermined acceptance criteria
    MaxZero Removal Torque (BD internal specification)Met predetermined acceptance criteria
    Flow Control Plug Retention Rate (BD internal specification)Met predetermined acceptance criteria
    Packaging Integrity (ASTM F2096)Met predetermined acceptance criteria
    Damage to Device (ISO 11607-1 §6.3.5)Met predetermined acceptance criteria (testing conducted to ensure packaging system provides adequate protection to the product through the hazards of handling, distribution and storage)
    Biocompatibility: CytotoxicityMet predetermined acceptance criteria (leveraged from predicate devices)
    Biocompatibility: SensitizationMet predetermined acceptance criteria (leveraged from predicate devices)
    Biocompatibility: Intracutaneous ReactivityMet predetermined acceptance criteria (leveraged from predicate devices)
    Biocompatibility: Systemic Toxicity (Acute)Met predetermined acceptance criteria (leveraged from predicate devices)
    Biocompatibility: Pyrogenicity (Material-Mediated Rabbit Pyrogen)Met predetermined acceptance criteria (leveraged from predicate devices)
    Biocompatibility: Pyrogenicity (LAL)Met predetermined acceptance criteria (leveraged from predicate devices)
    Biocompatibility: Subchronic Toxicity (subacute toxicity)Met predetermined acceptance criteria (leveraged from predicate devices)
    Biocompatibility: GenotoxicityMet predetermined acceptance criteria (leveraged from predicate devices)
    Biocompatibility: ImplantationMet predetermined acceptance criteria (leveraged from predicate devices)
    Biocompatibility: HaemocompatibilityMet predetermined acceptance criteria (leveraged from predicate devices)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for the individual performance tests listed.
    The provenance of the data is not explicitly stated as a country of origin or retrospective/prospective. The tests were likely conducted in-house by Becton Dickinson or by a contracted laboratory.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This study does not involve clinical data or "ground truth" established by experts in a diagnostic sense. The tests performed are engineering and biocompatibility tests for a medical device. Therefore, this question is not applicable.

    4. Adjudication Method for the Test Set

    Not applicable, as this is not a study requiring adjudication of expert opinions or diagnostic findings.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a 510(k) submission for a physical medical device (IV catheter system), not an AI/Software as a Medical Device (SaMD).

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a 510(k) submission for a physical medical device.

    7. The Type of Ground Truth Used

    Not applicable as this is a physical medical device and the performance tests are based on engineering and biocompatibility standards, not diagnostic "ground truth."

    8. The Sample Size for the Training Set

    Not applicable. This is a physical medical device, not an AI/machine learning model that requires a training set. The performance tests are for device validation.

    9. How the Ground Truth for the Training Set was Established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170336
    Date Cleared
    2017-03-10

    (36 days)

    Product Code
    Regulation Number
    880.5200
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BD Nexiva Closed IV Catheter System - Single Port with MaxZero Needleless Connector devices are intended to be inserted into a patient's vascular system for short term use to sample blood pressure, or administer fluids. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when access ports not suitable for use with power injectors are removed.

    Device Description

    BD Nexiva Closed IV Catheter System - Single Port with MaxZero Needleless Connector devices are over-the-needle, intravascular catheters. These devices have a radiopaque BD Vialon catheter, needle, needle shield, septum, stabilization platform, integrated extension tubing, clamp, Luer adapter, and vent plug. The needle and catheter are protected by a needle cover. A MaxZero device with protective cover is provided in the unit package.

    The closed system is designed to keep blood contained within the device throughout the insertion process. The septum is designed to wipe visible blood from the needle surface as the needle is withdrawn from the catheter, further reducing the risk of blood exposure. The needle tip is passively protected when the needle is removed, reducing the risk of accidental needlestick injury.

    These devices have BD Instaflash needle technology, allowing for immediate visualization of blood return along the catheter. Continuous blood return is seen in the extension tubing. The vent plug prevents blood leakage from the extension tubing during insertion. Both the stabilization platform and Luer connector are color coded to indicate catheter gauge size (24 GA (0.7 mm)=Yellow, 22 GA (0.9 mm)=Blue, 20 GA (1.1 mm)=Pink, 18 GA (1.3 mm)=Green).

    AI/ML Overview

    The provided text describes specific performance tests for the BD Nexiva Closed IV Catheter System - Single Port with MaxZero Needleless Connector. However, it does not include the acceptance criteria or reported device performance for these tests in a detailed format that would allow for a direct creation of the requested table. It only states that the device "met all predetermined acceptance criteria."

    Therefore, I cannot provide a table of acceptance criteria and reported device performance.

    Furthermore, the document does not detail the following information regarding a hypothetical study proving the device meets acceptance criteria:

    • Sample size used for the test set or the data provenance.
    • Number of experts used to establish the ground truth or their qualifications.
    • Adjudication method for the test set.
    • Whether a multi-reader multi-case (MRMC) comparative effectiveness study was done, or the effect size of human readers improving with AI vs without AI assistance.
    • Whether a standalone (algorithm only without human-in-the-loop performance) was done.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.).
    • The sample size for the training set.
    • How the ground truth for the training set was established.

    This document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed study results with specific performance metrics and acceptance criteria for this particular device. It describes the types of tests performed (e.g., MaxZero Insertion Force, Cytotoxicity, Sensitization) and states that "the subject device met all predetermined acceptance criteria," but it doesn't quantify those criteria or the device's exact performance against them.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102520
    Date Cleared
    2011-01-06

    (126 days)

    Product Code
    Regulation Number
    880.5200
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    As indicated in 21 CFR Part 880.5200, The Nexiva™ intravascular catheter is inserted into a patient's vascular system for a short-term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravascularly. The needle-shielding feature and luer access port, aid in the prevention of needle-stick injuries. Blood is contained within the device during the catheter insertion process aiding in the prevention of blood exposure. This catheter may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure.

    The 18-22 gauge Nexiva™ catheters are suitable for use with power injectors rated for a maximum of 300 psi when the luer access port(s) is removed and a direct connection is made.

    Device Description

    The BD Nexiva™ Closed IV Catheter System consists of an over-the needle, peripheral intravascular catheter made from Vialon™ polyurethane, integrated extension tubing with a Y adapter and clamp, BD Q-Syte™ luer access port, and a passive needleshielding mechanism.

    The design of the Nexiva™ IV catheter can be described as a closed system since it protects clinicians from blood exposure during the catheter insertion procedure. Since the needle is withdrawn through a septum that seals after the needle has been removed and both ports of the Y adapter are closed, blood is contained within the NexivaTM device during catheter insertion. The pressure exerted on the needle as it passes through the septum wipes blood from the needle, further reducing potential blood exposure. The slide clamp on the integrated extension tubing is provided to eliminate blood exposure when the vent plug is replaced with an infusion set connection of a BD Q-Syte™ luer access port.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the BD Nexiva™ Closed IV Catheter System, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The nonclinical tests compare the modified BD Nexiva™ Closed IV Catheter System to its predicate device (K032843). The reported device performance for all listed characteristics is "Pass," indicating that the device met the specified acceptance criteria.

    Associated ModificationProduct Performance Characteristics/Verification TestingAcceptance Criteria for Product VerificationReported Device Performance
    1Premature Decouple ForceDPPM < 280Pass
    2Catheter Pull ForceDPPM < 60Pass
    2Catheter Burst StrengthDPPM < 60Pass
    2Catheter Penetration ForceDPPM <1350Pass
    2Catheter DragDPPM <1350Pass
    2Tip AdhesionDPPM <1350Pass
    1System Drag -Cannula DragDPPM < 500Pass
    1System Drag -Ferrule Retraction ForceDPPM < 500Pass
    1Adapter Release ForceDPPM < 500Pass
    3Needle Cover RemovalDPPM USL < 18000Pass
    3(Needle Cover Removal)DPPM LSL < 200000Pass
    3(Needle Cover Removal)-20%/+25%, DPPM < 5000Pass
    4Flow Rate Label Claim-10%/+15%, DPPM < 5000Pass
    5Pinch Clamp Fluid SealDPPM < 10000Pass
    5Pinch Clamp Engagement Disengagement ForceDPPM < 2000Pass
    8Wing deflection forceDPPM USL < 2000Pass
    8(Wing deflection force)DPPM LSL < 500Pass
    6Bond Strength -Tubing Bond to AdapterDPPM < 0.1Pass
    4Extension Tube Burst PressureDPPM < 60Pass
    10Unit Package Seal Peel Force1.33 CpkPass
    10Seal Width1.33 CpkPass
    10Unit Package Material Thickness - Post Formed (Bottom Wed)1.33 CpkPass
    10Package Integrity (Water Leak)0 out of 2301Pass
    10Ship Testing (Drop/Vibration)0 out of 2301Pass
    7Y-Adapter Wishbone LoadingDPPM < 0.1Pass
    7ISO Liquid LeakageDPPM < 500Pass
    7ISO Air LeakageDPPM < 500Pass
    7ISO Unscrewing TorqueDPPM < 500Pass
    7ISO Resistance to OverridingDPPM < 500Pass
    7ISO Thread SeparationDPPM < 500Pass
    7Vent Plug Removal Force/TorqueDPPM USL < 6500Pass
    7(Vent Plug Removal Force/Torque)DPPM LSL <10000Pass
    7Q-Syte Removal TorqueDPPM USL < 6500Pass
    7(Q-Syte Removal Torque)DPPM LSL <10000Pass
    11ISO Thread Seperation -Both Y female luerDPPM < 500Pass
    11ISO Thread Seperation -Straight female luerDPPM < 500Pass
    11ISO Unscrewing Torque -Both Y female luerDPPM < 500Pass
    11ISO Unscrewing Torque -Straight female luerDPPM < 500Pass
    11ISO Resistance to Override -Both Y female luerDPPM < 500Pass
    11ISO Resistance to Override -Straight female luerDPPM < 500Pass
    11ISO Liquid LeakageDPPM < 500Pass
    11ISO Air LeakageDPPM < 500Pass
    12Vent Plug Torque RemovalDPPM USL < 10000, DPPM LSL< 20000Pass
    11Bond Strength - Tubing Bond to AdapterDPPM < 60Pass
    11Flow Rate Label Claim-10%/+15%, DPPM < 5000Pass
    14, 15Proximal Re-exposureDPPM < 100Pass
    17System Drag -Maximum Adhesion ForceDPPM < 1350Pass
    17System Drag - Average Drag ForceDPPM < 10,000Pass
    17System Drag - Offset Peak ForceDPPM < 2000Pass
    17Flow Rate Label Claim-20%/+25%, DPPM < 5000, -10%/+15%, DPPM < 5000Pass
    17Minimum FlowDPPM < 5000Pass
    17FlashbackDPPM < 1000Pass
    17Cannula PenetrationDPPM < 1350Pass
    17Catheter PenetrationDPPM < 1350Pass
    17, 19Premature DecoupleDPPM < 280Pass
    20Needle Hub PullDPPM < 500Pass

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document specifies a sample size of 2301 units for "Package Integrity (Water Leak)" and "Ship Testing (Drop/Vibration)" tests. For other tests, specific sample sizes are not explicitly stated, but the acceptance criteria are generally expressed in "DPPM" (Defective Parts Per Million) or "Cpk" (Process Capability Index), suggesting statistical process control and quality assurance testing.
    • Data Provenance: The tests are described as "Nonclinical Tests" and "design verification activities," conducted by Becton Dickinson Infusion Therapy Systems, Inc. The data is retrospective in the sense that it was generated during device development and verification to support the 510(k) submission. There is no information on the country of origin of the data, but it can be assumed to be related to the manufacturer's testing facilities, potentially in the US (headquarters in Sandy, UT).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This document primarily describes physical and mechanical performance testing of a medical device, not diagnostic or clinical performance involving expert interpretation of data (like radiology images). Therefore, the concept of "experts" establishing ground truth in this context doesn't directly apply as it would for a software device. The ground truth for these tests is based on established engineering standards (e.g., ANSI/AAMI/ISO 10993-1 2003 (E) for biocompatibility, and presumably other relevant engineering and ISO standards for mechanical performance), internal product specifications, and quality control methodologies.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, these are physical and mechanical tests with quantitative acceptance criteria (e.g., DPPM, Cpk, force measurements). There is no "adjudication" in the sense of reconciling differing expert opinions or classifications. The results are objective measurements compared against predefined thresholds.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. This document does not describe an MRMC comparative effectiveness study. The submission relates to a physical medical device (intravascular catheter) and its modifications, not an AI or software device that would typically involve human readers interpreting cases.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No. This concept is not applicable as the submission is for a physical medical device, not an algorithm or AI. The tests are focused on the device's physical and mechanical properties.

    7. The Type of Ground Truth Used

    The ground truth used for these tests is based on:

    • Engineering Standards: Adherence to established industry and international standards (e.g., ANSI/AAMI/ISO 10993-1 2003 (E), ISO Liquid Leakage, etc.).
    • Product Specifications: Internal design requirements and performance targets defined by the manufacturer.
    • Predicate Device Performance: The modified device is benchmarked against its predicate device (K032843) to demonstrate "substantial equivalence."

    8. The Sample Size for the Training Set

    Not applicable. This is not a machine learning or AI device that requires a training set. The "training" for such physical devices involves design, prototyping, and iterative testing processes, not data-driven algorithmic training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1