Search Results
Found 1 results
510(k) Data Aggregation
(213 days)
BD Acute Central Line
Acute central venous catheters are indicated to provide short-term access (
A family of power injectable central venous catheters constructed of medical grade polyurethane and is designed for insertion into the central venous system. BD power injectable acute central lines are radiopaque, and have a soft tip that is more pliable than the catheter body. Each catheter is provided in a sterile package with applicable insertion kit accessories.
The provided text describes the BD Acute Central Line, a medical device, and its substantial equivalence to a predicate device, the Arrow Central Venous Catheter (K071538). The document outlines various performance tests conducted to establish this equivalence, along with the criteria used.
Here's a breakdown of the requested information:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly provide a table listing "acceptance criteria" alongside "reported device performance" with specific numerical results for each test. Instead, it states that "All testing passed the predetermined acceptance criteria." The acceptance criteria are implicitly defined by the standards and the purpose of each test.
Here’s a summary of the tests performed and their implied acceptance criteria:
Test Name | Implied Acceptance Criteria | Reported Device Performance |
---|---|---|
Biocompatibility Testing (ISO 10993-1:2009) | Catheter free from biological hazard; materials acceptable for intended purpose (based on Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Pyrogenicity, Subchronic Systemic Toxicity, Genotoxicity, Hemocompatibility, Implantation) | Passed |
Clamp Engagement | Catheter assembly will not leak when clamp is engaged. | Passed |
Leak Test | Catheter assembly will not leak when the distal end of the catheter is occluded. | Passed |
Dimensional Test | OD and ID/lumen area comply with dimensional specifications. | Passed |
Implantable Length | Useful length complies with dimensional specifications. | Passed |
Extension Leg Length | Extension leg length complies with dimensional specifications. | Passed |
Burst Test | Catheter burst pressure exceeds peak pressure in catheter at maximum flow conditions when distal end is occluded. | Passed |
Hydraulic Catheter Burst Test | Catheter burst pressure exceeds peak pressure in catheter at maximum flow conditions when distal end is occluded. | Passed |
Power Injection Conditioning | Catheter does not leak or burst as a result of power injections at maximum indicated flow rate. | Passed |
Gravity Flow | Meets specified gravity flow performance for a full-length catheter. | Passed |
Luer to Extension Leg Tensile Test | Peak tensile force exceeds minimum peak tensile force. | Passed |
Extension Leg to Trifurcation Tensile Test | Peak tensile force exceeds minimum peak tensile force. | Passed |
Trifurcation to Shaft Tensile Test | Peak tensile force exceeds minimum peak tensile force. | Passed |
Shaft Tensile Test | Maximum catheter strain and modulus at break meet specified criteria. | Passed |
Radiopacity (ASTM F640-12) | Catheter is radio-detectable. | Passed |
Tip Tensile (ISO 10555-3:2013) | Peak tensile force exceeds minimum peak tensile force. | Passed |
Catheter Collapse Test | Demonstrates aspiration flow rate and that catheter will not collapse under vacuum. | Passed |
Suture Wing Integrity Test | Maximum force catheter junction suture wing can withstand prior to break meets specified criteria. | Passed |
Priming Volume | Volume required to prime full-length catheter meets specified criteria. | Passed |
OD Swell | Catheter does not swell beyond twice the size of the labeled OD during power injection. | Passed |
Tip Stability Test | Catheter tip remains in same orientation during power injection (tip pointing in direction of venous flow) at maximum indicated flow rate. | Passed |
Guidewire Drag Test | Guidewire can be removed without difficulty. | Passed |
Luer Testing (ISO 594: Part 1 & 2) | Luer connectors meet requirements for Leak, Leak Decay, Stress Cracking, Resistance to Separation from Axial Load, Resistance to Separation from Unscrewing, Resistance to Overriding, Gauging. | Passed |
Particulate Testing (USP ) | Particulate matter on catheter post-manufacture does not exceed prescribed particle sizes. | Passed |
MR Safety | Device is safe for use in an MR environment. | Passed |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not explicitly state the specific sample sizes used for each of the performance tests. It mentions that testing was conducted "by or for BAS per guidance documents and standards in conjunction with in-house protocols." The data provenance (country of origin, retrospective/prospective) is also not specified.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This section is not applicable. The device is a physical medical device (central venous catheter), not an AI or diagnostic imaging device that typically requires expert-established ground truth for performance evaluation in the context of diagnostic accuracy. The "ground truth" for this device's performance is established through physical and material property testing against engineering specifications and industry standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable for a physical medical device. Performance is determined by objective measurements against established standards, not by agreement among adjudicators.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The BD Acute Central Line is a physical medical device and not an AI or diagnostic imaging device for which MRMC comparative effectiveness studies with human readers would be performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. The device is a physical medical device and does not involve algorithms or AI.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the BD Acute Central Line's performance is based on established industry consensus standards and in-house protocols for performance testing of medical devices, specifically intravascular catheters. This includes physical, chemical, and biological testing as outlined in standards like ISO 10993-1, ISO 10555, ASTM F640-12, USP , and FDA guidance documents. The acceptance criteria for these tests serve as the "ground truth" against which the device's performance is measured.
8. The sample size for the training set
This is not applicable. The device is a physical medical device, not an AI model that requires a training set.
9. How the ground truth for the training set was established
This is not applicable as there is no training set for a physical medical device.
Ask a specific question about this device
Page 1 of 1