Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K190855
    Date Cleared
    2019-11-01

    (213 days)

    Product Code
    Regulation Number
    880.5200
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    BD Acute Central Line

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Acute central venous catheters are indicated to provide short-term access (

    Device Description

    A family of power injectable central venous catheters constructed of medical grade polyurethane and is designed for insertion into the central venous system. BD power injectable acute central lines are radiopaque, and have a soft tip that is more pliable than the catheter body. Each catheter is provided in a sterile package with applicable insertion kit accessories.

    AI/ML Overview

    The provided text describes the BD Acute Central Line, a medical device, and its substantial equivalence to a predicate device, the Arrow Central Venous Catheter (K071538). The document outlines various performance tests conducted to establish this equivalence, along with the criteria used.

    Here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly provide a table listing "acceptance criteria" alongside "reported device performance" with specific numerical results for each test. Instead, it states that "All testing passed the predetermined acceptance criteria." The acceptance criteria are implicitly defined by the standards and the purpose of each test.

    Here’s a summary of the tests performed and their implied acceptance criteria:

    Test NameImplied Acceptance CriteriaReported Device Performance
    Biocompatibility Testing (ISO 10993-1:2009)Catheter free from biological hazard; materials acceptable for intended purpose (based on Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Pyrogenicity, Subchronic Systemic Toxicity, Genotoxicity, Hemocompatibility, Implantation)Passed
    Clamp EngagementCatheter assembly will not leak when clamp is engaged.Passed
    Leak TestCatheter assembly will not leak when the distal end of the catheter is occluded.Passed
    Dimensional TestOD and ID/lumen area comply with dimensional specifications.Passed
    Implantable LengthUseful length complies with dimensional specifications.Passed
    Extension Leg LengthExtension leg length complies with dimensional specifications.Passed
    Burst TestCatheter burst pressure exceeds peak pressure in catheter at maximum flow conditions when distal end is occluded.Passed
    Hydraulic Catheter Burst TestCatheter burst pressure exceeds peak pressure in catheter at maximum flow conditions when distal end is occluded.Passed
    Power Injection ConditioningCatheter does not leak or burst as a result of power injections at maximum indicated flow rate.Passed
    Gravity FlowMeets specified gravity flow performance for a full-length catheter.Passed
    Luer to Extension Leg Tensile TestPeak tensile force exceeds minimum peak tensile force.Passed
    Extension Leg to Trifurcation Tensile TestPeak tensile force exceeds minimum peak tensile force.Passed
    Trifurcation to Shaft Tensile TestPeak tensile force exceeds minimum peak tensile force.Passed
    Shaft Tensile TestMaximum catheter strain and modulus at break meet specified criteria.Passed
    Radiopacity (ASTM F640-12)Catheter is radio-detectable.Passed
    Tip Tensile (ISO 10555-3:2013)Peak tensile force exceeds minimum peak tensile force.Passed
    Catheter Collapse TestDemonstrates aspiration flow rate and that catheter will not collapse under vacuum.Passed
    Suture Wing Integrity TestMaximum force catheter junction suture wing can withstand prior to break meets specified criteria.Passed
    Priming VolumeVolume required to prime full-length catheter meets specified criteria.Passed
    OD SwellCatheter does not swell beyond twice the size of the labeled OD during power injection.Passed
    Tip Stability TestCatheter tip remains in same orientation during power injection (tip pointing in direction of venous flow) at maximum indicated flow rate.Passed
    Guidewire Drag TestGuidewire can be removed without difficulty.Passed
    Luer Testing (ISO 594: Part 1 & 2)Luer connectors meet requirements for Leak, Leak Decay, Stress Cracking, Resistance to Separation from Axial Load, Resistance to Separation from Unscrewing, Resistance to Overriding, Gauging.Passed
    Particulate Testing (USP )Particulate matter on catheter post-manufacture does not exceed prescribed particle sizes.Passed
    MR SafetyDevice is safe for use in an MR environment.Passed

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the specific sample sizes used for each of the performance tests. It mentions that testing was conducted "by or for BAS per guidance documents and standards in conjunction with in-house protocols." The data provenance (country of origin, retrospective/prospective) is also not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This section is not applicable. The device is a physical medical device (central venous catheter), not an AI or diagnostic imaging device that typically requires expert-established ground truth for performance evaluation in the context of diagnostic accuracy. The "ground truth" for this device's performance is established through physical and material property testing against engineering specifications and industry standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable for a physical medical device. Performance is determined by objective measurements against established standards, not by agreement among adjudicators.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The BD Acute Central Line is a physical medical device and not an AI or diagnostic imaging device for which MRMC comparative effectiveness studies with human readers would be performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is a physical medical device and does not involve algorithms or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the BD Acute Central Line's performance is based on established industry consensus standards and in-house protocols for performance testing of medical devices, specifically intravascular catheters. This includes physical, chemical, and biological testing as outlined in standards like ISO 10993-1, ISO 10555, ASTM F640-12, USP , and FDA guidance documents. The acceptance criteria for these tests serve as the "ground truth" against which the device's performance is measured.

    8. The sample size for the training set

    This is not applicable. The device is a physical medical device, not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    This is not applicable as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1