Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K981909
    Manufacturer
    Date Cleared
    1998-11-24

    (176 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    BAXTER HEMOSTASIS VALVE INTRODUCERS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Single Lumen Introducers: The Baxter Healthcare Single Lumen Hemostasis Valve Introducers are indicated for use in patients requiring access of the venous system and to facilitate catheter insertion (e.g. pulmonary artery or infusion catheter). Multiple Lumen Access Products: The Baxter Healthcare Multiple Lumen Access Products are indicated for use in patients requiring access of the venous system, to facilitate catheter insertion (e.g., pulmonary artery or infusion catheter) and central venous pressure monitoring.

    Device Description

    The Baxter Hemostasis Valve Introducers are used to access the venous system and to facilitate catheter insertion. The introducers are composed of a housing body to which a sheath is attached distally and a side port/extension tube is connected proximally. A valve is located in the housing body to provide a seal around a catheter when inserted through the Introducer and to prevent backflow when no catheter is present. A dilator is provided with the Introducer to ease insertion of the device into the vessel. The Baxter Hemostasis Valve Introducers are provided with and without AMC Thromboshield™ which is used on Baxter's catheters and introducer products. The device will be packaged in a tray sealed with a tyvek lid and sterilized using 100% ethylene oxide.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary regarding the Baxter Hemostasis Valve Introducers:

    Missing Information:

    It's crucial to note that the provided 510(k) summary is for a medical device (introducer sheath), not for an AI/algorithm-based diagnostic device. Therefore, many of the requested categories (e.g., sample sizes for test/training sets, number of experts, adjudication methods, MRMC studies, standalone performance, ground truth types) are not applicable to this type of device submission and are not present in the document.

    Acceptance Criteria and Device Performance for Baxter Hemostasis Valve Introducers

    Given the nature of this medical device, the "acceptance criteria" are based on comparative technological equivalence to predicate devices and successful results in a series of in-vitro and biocompatibility tests. There are no performance metrics like sensitivity, specificity, or AUC that would be associated with an AI/diagnostic algorithm.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryDescriptionReported Device Performance
    Technological EquivalenceDevice construction and physical specifications are comparable to legally marketed predicate devices."The Baxter Hemostasis Valve Introducers are technologically comparable to the predicate devices in construction and physical specifications."
    Functional Integrity/Performance (In-vitro)Device demonstrates integrity and performance for its intended use through functional testing."Functional testing was performed on the Baxter Hemostasis Valve Introducers to evaluate the integrity and performance of the device. Based upon the results of this testing, the Baxter CardioVascular Group, Edwards Critical-Care has determined that the Baxter Hemostasis Valve Introducers are safe and effective and are acceptable in design and construction for its intended use."
    BiocompatibilityDevice meets established biocompatibility standards and is non-toxic."Biocompatability testing was performed on the Baxter Hemostasis Valve Introducers in accordance with the requirements specified in International Standards Organization (ISO) 10993-1-1994 Biological Evaluation of Medical Devices - Part 1: Guidance on Selection of Tests and the FDA General Program Memorandum No. G95-1. The Baxter Hemostasis Valve Introducers were found to be biocompatable and nontoxic and acceptable for its intended use."
    Clinical Safety/EfficacyClinical performance demonstrates safety and efficacy for the stated indications for use."Clinical testing was not performed on the subject catheter because the intended use and indications are the same as the predicate devices. Furthermore, the Baxter Hemostasis Valve Introducers exhibited comparable design characteristics to the predicate devices in the in vitro testing, thus clinical testing was not performed." This demonstrates equivalence was based on non-clinical data.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not applicable in the context of an AI/diagnostic algorithm. For an in-vitro device, "test set" refers to the units of the device subjected to functional and biocompatibility tests. The document does not specify a numerical sample size for these tests, which is common for this type of submission. It implies that a sufficient number of devices were tested to demonstrate performance and establish comparability.
    • Data Provenance: The in-vitro testing and biocompatibility testing were conducted directly on the manufactured Baxter Hemostasis Valve Introducers. No external patient data (country of origin, retrospective/prospective) was used, as clinical testing was explicitly waived due to equivalence with predicate devices.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not Applicable. This document pertains to a physical medical device, not an AI/diagnostic algorithm where expert-established ground truth on a test set would be relevant. The assessment was based on engineering and laboratory testing protocols.

    4. Adjudication Method for the Test Set

    • Not Applicable. There was no "test set" in the sense of patient cases requiring adjudication for ground truth. Device performance was evaluated against technical specifications and established standards through specific tests.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

    • No. An MRMC study is not relevant for this type of purely mechanical medical device.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable. This is a physical introducer sheath, not an algorithm.

    7. The Type of Ground Truth Used

    • Not Applicable in the traditional sense of AI/diagnostic algorithms. For this device, the "ground truth" for the in-vitro and biocompatibility testing involved objective measurements and adherence to pre-defined engineering specifications and international standards (e.g., ISO 10993-1-1994 guidance). For "equivalence," the predicate devices served as the benchmark.

    8. The Sample Size for the Training Set

    • Not Applicable. There is no "training set" for a physical medical device. Manufacturing processes and design iterations, while involving testing, are not analogous to training an AI model.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As there is no training set for an AI algorithm, no ground truth was established in this context.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1