Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K190953
    Manufacturer
    Date Cleared
    2019-10-16

    (188 days)

    Product Code
    Regulation Number
    888.3030
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Arthrex Minimally Invasive Ankle Fusion Plate

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Arthrex Minimally Invasive Ankle Fusion Plate is intended to be used for fixation of bone fractures and fusions in the foot and ankle. The Arthrex Minimally Invasive Ankle Fusion Plate is to be used with the Arthrex Low Profile Screws.

    Device Description

    The Arthrex Minimally Invasive Ankle Fusion Plate is a Titanium bone fixation device intended to be permanently implanted. The implant is contoured with a tapered proximal end and contoured bend on the distal end. The implant has a 3-hole design and is 1.74 inches in length. The implant is single-use and may be available in sterile and non-sterile versions.

    AI/ML Overview

    The Arthrex Minimally Invasive Ankle Fusion Plate is a bone fixation device intended for permanent implantation to fix bone fractures and fusions in the foot and ankle, to be used with Arthrex Low Profile Screws. The device is made of Titanium, has a 3-hole design, and is 1.74 inches in length.

    Here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Study Type)Standard/MethodDevice PerformanceConclusion
    Static Bending StrengthASTM F382-17Statistically equivalent to predicate device (K141735)Met
    Fatigue Bending StrengthASTM F382-17Statistically equivalent to predicate device (K141735)Met
    MRI Safety (Force, Torque, Image Artifact)FDA Guidance: Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) EnvironmentTesting conducted for force, torque, and image artifact. Specific values not reported in this summary, but the implication is that it meets safety guidelines.Met
    Bacterial Endotoxins Test (BET)ANSI/AAMI ST72:2011/(R)2016, USP , USP , EP 2.6.14Meets pyrogen limit specificationsMet

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated in the provided document for each specific test. However, the document refers to "testing conducted" generally.
    • Data Provenance: Not explicitly stated. The studies are laboratory-based performance tests evaluating the physical and biological properties of the device, rather than clinical data from human subjects.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable. The studies conducted are mechanical and material property tests against established engineering and biological standards (ASTM F382-17, FDA Guidance for MRI Safety, ANSI/AAMI ST72, etc.), not studies relying on expert interpretation of medical data. Therefore, there is no "ground truth" in the clinical sense established by medical experts for these tests. The "ground truth" is defined by the objective performance criteria outlined in the standards.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies, particularly for diagnostic devices where multiple readers might interpret data to establish a consensus ground truth. The studies performed here are objective laboratory tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs. without AI assistance

    No. An MRMC comparative effectiveness study is not discussed. This device is a bone fixation plate, not an AI-powered diagnostic or assistive technology.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No. This information is not applicable as the device is a physical implant, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance tests conducted (static bending, fatigue bending, MRI safety, bacterial endotoxins) is based on established engineering and biological standards and specifications. For instance, the mechanical tests are compared against the performance of a predicate device cleared under K141735, implying that the predicate's performance against ASTM F382-17 served as the benchmark ("ground truth") for equivalence. The Bacterial Endotoxins Test has specific pyrogen limit specifications as its "ground truth."

    8. The Sample Size for the Training Set

    This information is not applicable. The device is a physical implant and does not involve AI or machine learning models that require training sets.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1