Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    Why did this record match?
    Device Name :

    Aristotle 14 Guidewire; Aristotle 18 Guidewire; Aristotle 24 Guidewire; Zoom Wire 14 Guidewire; Aristotle
    Colossus Guidewire

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aristotle 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    The Aristotle 18 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    The Aristotle 24 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    The Zoom Wire 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    The Aristotle Colossus Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    Device Description

    The Scientia Vascular's Aristotle 14, Aristotle 18, Aristotle 24, Zoom Wire 14, and Aristotle Colossus Guidewires ("the guidewires") are steerable guidewires with a shapeable tip to aid in accessing the neuro and peripheral vasculatures. The guidewires are supplied sterile, for single use only.

    The distal portion of each guidewire's tip includes a radiopaque platinum wire marker coil to facilitate fluoroscopic visualization. All guidewires have a hydrophilic polymer coating on the distal portion and a polytetrafluoroethylene (PTFE) coating on the proximal portion to reduce friction during manipulation in vessels.

    The guidewires are provided with an accessory kit consisting of an introducer (to aid with the insertion of the guidewire into a catheter hub and/or a hemostasis valve), a torque device (to attach to the proximal portion of the guidewire to facilitate gripping and manipulation of the guidewire during use), and a shaping mandrel (to aid in shaping the flexible tip of the guidewire). These accessory devices are included to facilitate use of the guidewires and are not intended to contact the patient's body.

    AI/ML Overview

    This document describes a 510(k) clearance for several guidewire devices, not an AI/ML medical device. Therefore, the information requested about acceptance criteria and studies related to AI/ML device performance (such as sample sizes for test/training sets, expert ground truth establishment, MRMC studies, etc.) is not applicable and cannot be extracted from this document.

    The document primarily focuses on establishing substantial equivalence for physical medical devices—guidewires—by comparing their technological characteristics and performing non-clinical performance tests (biocompatibility, sterilization, bench performance).

    Here's a breakdown of why this document does not contain the requested information about an AI/ML device:

    • Device Type: The clearance is for "Aristotle 14 Guidewire; Aristotle 18 Guidewire; Aristotle 24 Guidewire; Zoom Wire 14 Guidewire; Aristotle Colossus Guidewire" which are physical catheter guide wires. These are traditional medical devices, not software or AI/ML devices.
    • Study Types: The studies described are "Biocompatibility," "Sterilization," and "Bench Performance Testing" (Visual Inspection, Coating Integrity, Corrosion Resistance, Agent Compatibility, Simulated Use, Particulate). These are mechanical and biological tests relevant to physical devices, not algorithmic performance.
    • Missing Information: All points you requested regarding AI/ML performance are absent because they are not relevant to this type of device clearance:
      • No acceptance criteria for AI performance metrics (e.g., sensitivity, specificity, AUC).
      • No test or training set sample sizes for AI data.
      • No information about experts establishing ground truth for images or data used by an AI.
      • No adjudication methods.
      • No MRMC comparative effectiveness studies.
      • No standalone algorithm performance.
      • No ground truth type related to AI (e.g., imaging labels, pathology).

    The document explicitly states:

    • "No animal testing was deemed necessary to support the substantial equivalence of the subject devices."
    • "No clinical testing was deemed necessary to support the substantial equivalence of the subject devices."

    This indicates that the clearance relies on non-clinical data and comparison to predicate devices, which is common for guidewires, especially when the changes are minor (like a different PTFE coating).

    Ask a Question

    Ask a specific question about this device

    K Number
    K222437
    Date Cleared
    2023-01-26

    (167 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Aristotle Colossus Guidewire

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aristotle Colossus Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    Device Description

    Scientia Vascular's Aristotle Colossus Guidewire is a modification of Scientia Vascular's Aristotle 24 Guidewire. It is a 0.035" diameter, steerable guidewire with a shapeable tip to aid in accessing vasculature. The guidewire is supplied sterile and is for single use only. It is provided in a single stiffness profile (standard) and is available in a range of lengths from 150 cm to 300 cm. The distal portion of the guidewire tip includes a radiopaque platinum wire marker coil to facilitate fluoroscopic visualization. The guidewire has a hydrophilic polymer coating on the distal portion and a polytetrafluoroethylene (PTFE) coating on the proximal portion to reduce friction during manipulation in tortuous vessels.

    The guidewire is provided with a shaping mandrel, an introducer (to aid with the insertion of the guidewire into a catheter hub and/or a rotating hemostasis valve (RHV)), and a torque device (to attach to the proximal portion of the guidewire to facilitate gripping and manipulation of the guidewire during use). The shaping mandrel, introducer, and torque device accessories are included to facilitate use of the guidewire and are not intended to contact the patient's body.

    AI/ML Overview

    The provided text describes a medical device, the Aristotle Colossus Guidewire, and its clearance process with the FDA. It details a comparison to a predicate device and reports on non-clinical performance tests. However, the document does NOT contain information about any study involving human readers, AI assistance, or the establishment of ground truth by experts for such a study. It specifically focuses on the physical and functional characteristics of a guidewire.

    Therefore, I cannot provide the detailed information requested regarding acceptance criteria and studies that prove the device meets those criteria in the context of an AI/human-in-the-loop performance evaluation. The document describes a traditional medical device clearance process based on material properties, functional performance, and biocompatibility.

    Here's a breakdown of why I cannot fulfill your request based on the provided text, and what is present:

    Information NOT present in the document which your request asks for:

    • A table of acceptance criteria and reported device performance related to AI/human reader performance.
    • Sample size used for a test set in the context of AI/human reader performance.
    • Data provenance (country of origin, retrospective/prospective) for AI/human reader performance data.
    • Number of experts used to establish ground truth for an AI/human reader test set or their qualifications.
    • Adjudication method for an AI/human reader test set.
    • Results of a multi-reader multi-case (MRMC) comparative effectiveness study (effect size of human readers with/without AI).
    • Results of a standalone (algorithm only) performance study.
    • Type of ground truth used (expert consensus, pathology, outcomes data) in the context of AI interpretation.
    • Sample size for a training set (for an AI model).
    • How ground truth for a training set was established (for an AI model).

    Information that IS present in the document (but is for a physical medical device, not an AI/Software as a Medical Device):

    The document describes the acceptance criteria and performance study for the Aristotle Colossus Guidewire, which is a physical medical device (a guidewire for vascular procedures). The studies described are non-clinical (laboratory and simulated use) to demonstrate the device's physical and biological safety and effectiveness.

    1. Acceptance criteria and reported device performance (for the physical guidewire):

    The document states: "Results of tests performed on the Aristotle Colossus Guidewire demonstrate that it met the test acceptance criteria and meets the requirements of relevant standards and FDA guidance documents."

    The "Table 3: Summary of Subject Device Functional Testing" lists the tests performed and consistently reports: "The Aristotle Colossus Guidewires met test acceptance criteria."

    Here's a summary of the stated "acceptance criteria" through the lens of successful test outcomes for the physical guidewire:

    Acceptance Criteria (Implied by Test Name)Reported Device Performance (Always "Met Acceptance Criteria")
    Biocompatibility:
    Non-cytotoxicNon-cytotoxic
    Non-hemolyticNon-hemolytic
    No effect on Partial Thromboplastin TimeNo effect on the PTT. The two samples are considered similar.
    Similar or lower Complement ActivationSimilar or lower potential to activate the complement system when compared to the predicate.
    Similar Thrombogenic risk potentialThrombogenic risk potential similar to the predicate.
    Functional Testing:
    Dimensional conformanceThe Aristotle Colossus Guidewires met test acceptance criteria.
    Sufficient Tensile StrengthThe Aristotle Colossus Guidewires met test acceptance criteria.
    Resistance to Fracture and FlexingThe Aristotle Colossus Guidewires met test acceptance criteria.
    Acceptable Torqueability and Torque StrengthThe Aristotle Colossus Guidewires met test acceptance criteria.
    Appropriate Tip FlexibilityThe Aristotle Colossus Guidewires met test acceptance criteria.
    Tip Shape RetentionThe Aristotle Colossus Guidewires met test acceptance criteria.
    Acceptable Coating Lubricity and DurabilityThe Aristotle Colossus Guidewires met test acceptance criteria.
    Coating IntegrityThe Aristotle Colossus Guidewires met test acceptance criteria.
    Low Particulate GenerationThe Aristotle Colossus Guidewires met test acceptance criteria.
    Performance in Simulated Use ModelThe Aristotle Colossus Guidewires met test acceptance criteria.
    Favorable Usability Evaluation by PhysiciansThe Aristotle Colossus Guidewires met test acceptance criteria.
    Sterilization & Shelf Life:
    Sterilization assurance levels (SAL) of at least 10^-6Not explicitly stated 'met', but implied by final conclusion for clearance.
    Acceptable EO and ECH residualsNot explicitly stated 'met', but implied by final conclusion for clearance.
    Acceptable bacterial endotoxin levelsNot explicitly stated 'met', but implied by final conclusion for clearance.
    Packaging integrity maintained through shelf-lifeNot explicitly stated 'met', but implied by final conclusion for clearance and new 1-year shelf life.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified quantitatively for each test, but it refers to the physical guidewires tested. Standard engineering and biocompatibility testing typically involves a specific number of units per lot or batch, but these numbers are not detailed in this summary.
    • Data Provenance: The studies are non-clinical performance tests conducted in a laboratory setting, likely by or for the manufacturer, Scientia Vascular, Inc. (West Valley City, Utah). They are simulations or bench tests rather than data derived from patients. The studies are prospective in the sense that they were designed and executed to evaluate this specific device for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not applicable in the context of this device's testing. The "Usability Evaluation" involved "Physicians [who] evaluated subject and predicate guidewires for various performance characteristics in a human cadaver." The number and specific qualifications of these physicians are not provided. This is a usability assessment, not an expert panel for establishing ground truth for diagnostic accuracy.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. This is for a physical device, and the "usability evaluation" doesn't describe a ground truth adjudication process in the manner seen for diagnostic AI.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a guidewire, not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a guidewire, not an AI algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

    The "ground truth" for this device's performance is established by engineering specifications, material science standards, and in vitro/ex vivo (cadaver) performance. For biocompatibility, it's established via standardized biological assays. There is no "ground truth" in the clinical diagnostic sense with expert consensus or pathology involved.

    8. The sample size for the training set:

    Not applicable. This device does not use a "training set" in the context of machine learning or AI.

    9. How the ground truth for the training set was established:

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1