Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    Why did this record match?
    Device Name :

    Sensor Enabled™; Advisor™ HD Grid X Mapping Catheter, Sensor Enabled™; Agilis™ NxT Steerable Introducer; Agilis
    ™ NxT Steerable Introducer Dual-Reach™

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ViewFlex™ Xtra ICE Catheter
    The ViewFlex™ Xtra ICE Catheter is indicated for use in adult and adolescent pediatric patients to visualize cardiac, structures, blood flow and other devices within the heart.

    ViewFlex™ Eco Reprocessed ICE Catheter
    The ViewFlex™ Eco Reprocessed Catheter is indicated for use in adult and adolescent pediatric patients to visualize cardiac structures, blood flow and other devices within the heart.

    Advisor™ HD Grid Mapping Catheter, Sensor Enabled™
    The Advisor™ HD Grid Mapping Catheter, Sensor Enabled™, is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart.

    Advisor™ HD Grid X Mapping Catheter, Sensor Enabled™
    The Advisor™ HD Grid X Mapping Catheter, Sensor Enabled™, is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart.

    Agilis™ NxT Steerable Introducer
    The Agilis™ NxT Steerable Introducer is indicated for the introduction of various cardiovascular catheters into the heart, including the left side of the heart, during the treatment of cardiac arrhythmias.

    Agilis™ NxT Steerable Introducer Dual-Reach™
    The Agilis™ NxT Steerable Introducer Dual-Reach™ is indicated for the introduction of various cardiovascular catheters into the heart, including the left side of the heart, during the treatment of cardiac arrhythmias.

    Device Description

    The Agilis™ NxT Steerable Introducer Dual-Reach™ is a sterile, single-use device that con-sists of a dilator and steerable introducer, which is designed to provide flexible catheter positioning in the cardiac anatomy. The inner diameter of the steerable introducer is 13F. The steerable introducer includes a hemostasis valve to minimize blood loss during catheter intro-duction and/or exchange. It has a sideport with three-way stopcock for air or blood aspiration, fluid infusion, blood sampling, and pressure monitoring. The handle is equipped with a rotating collar to deflect the tip clockwise ≥180° and counterclockwise ≥90°. The steerable introducer features distal vent holes to facilitate aspiration and minimize cavitation and a radiopaque tip marker to improve fluoroscopic visualization.

    AI/ML Overview

    This FDA 510(k) clearance letter (K251211) and its accompanying 510(k) summary pertain to a change in workflow for several existing cardiovascular catheters, specifically allowing for a "Zero/Low Fluoroscopy Workflow."

    The key phrase here is "Special 510(k) – Zero/Low Fluoroscopy Workflow". This type of submission is for modifications to a previously cleared device that do not significantly alter its fundamental technology or intended use, but rather introduce a change in how it's used or processed.

    Crucially, this submission does NOT describe a new AI/software device that requires extensive performance testing against acceptance criteria in the manner you've outlined for AI/ML devices. Instead, it's about demonstrating that existing devices, when used with a new, less-fluoroscopy-dependent workflow, remain as safe and effective as before.

    Therefore, many of the questions you've asked regarding acceptance criteria, study details, ground truth, and expert adjudication are not applicable to the information provided in this 510(k) document. The document explicitly states:

    • "Bench-testing was not necessary to validate the Clinical Workflow modifications."
    • "Substantial Equivalence of the subject devices to the predicate devices using the zero/low fluoroscopy workflow has been supported through a summary of clinical data across multiple studies in which investigators used alternative visualization methods."

    This indicates that the "study" proving the device (or rather, the new workflow) meets acceptance criteria is a summary of existing clinical data where alternative visualization methods were already employed, rather than a prospective, controlled study of a new AI algorithm.

    Based on the provided document, here's what can be answered:

    1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria: The implicit acceptance criterion is that the devices, when used with "zero/low fluoroscopy workflow," maintain substantial equivalence to their predicate devices in terms of safety and effectiveness. This means they must continue to perform as intended for visualizing cardiac structures, blood flow, mapping, or introducing catheters.
    • Reported Device Performance: The document states that "Substantial Equivalence... has been supported through a summary of clinical data across multiple studies in which investigators used alternative visualization methods." This implies that the performance (e.g., adequate visualization, successful mapping, successful catheter introduction) was maintained. Specific quantitative metrics of performance (e.g., accuracy, sensitivity, specificity, or inter-reader agreement for a diagnostic AI) are not provided or applicable here as this is not an AI/ML diagnostic clearance.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified. The document refers to "a summary of clinical data across multiple studies." This suggests an aggregation of results from existing (likely retrospective) patient data where alternative visualization techniques (allowing for "zero/low fluoroscopy") were already utilized clinically. It's not a new, single, prospectively designed test set for an AI algorithm.
    • Data Provenance: Not specified regarding country of origin or specific patient demographics. It is implied to be clinical data collected from studies where these types of procedures were performed using alternative visualization. The data would be retrospective as it's a "summary of clinical data" that already exists.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable in the context of this 510(k). Ground truth in an AI/ML context typically refers to adjudicated labels for images or signals. Here, the "ground truth" is inferred from standard clinical practice and outcomes in the historical data summarized. There's no mention of a specific expert panel for new ground truth establishment for a diagnostic AI.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This is not a study requiring adjudication of diagnostic outputs by multiple readers.

    5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI-assisted diagnostic device. The workflow change is about using alternative non-fluoroscopic imaging modalities (e.g., intracardiac echocardiography, electro-anatomical mapping systems), not about AI improving human reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This is not an AI algorithm. The predicate devices are physical catheters.

    7. The type of ground truth used:

    • The "ground truth" is inferred from clinical outcomes and established clinical practice using the devices with alternative visualization methods in real-world scenarios. It's not a specific, adjudicated dataset for an AI algorithm. The performance of the devices (such as successful navigation, visualization, and mapping) under the "zero/low fluoroscopy" workflow is assumed to be equivalent to their performance under full fluoroscopy, as demonstrated by prior clinical use where such methods were employed.

    8. The sample size for the training set:

    • Not applicable. There is no AI model being trained discussed in this document.

    9. How the ground truth for the training set was established:

    • Not applicable. No training set for an AI model.

    In summary:

    This 510(k) is for a workflow modification for existing medical devices (catheters), not for an AI/ML diagnostic or assistive software. Therefore, the detailed data performance evaluation typically required for AI models against specific acceptance criteria (as requested in your template) is not presented or relevant in this clearance letter. The "proof" relies on the concept of substantial equivalence to previously cleared predicate devices, supported by a summary of existing clinical data that used alternative visualization methods, implying that the devices function safely and effectively even with reduced fluoroscopy.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243493
    Manufacturer
    Date Cleared
    2024-12-13

    (31 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Agilis NxT Steerable Introducer Dual-Reach

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Agilis NxT Steerable Introducer Dual-Reach is indicated for the introduction of various cardiovascular catheters into the heart, including the left side of the heart, during the treatment of cardiac arrhythmias.

    Device Description

    The Agilis™ NxT Steerable Introducer Dual-Reach™™ is a sterile, single-use device that consists of a dilator and steerable introducer, which is designed to provide flexible catheter positioning in the cardiac anatomy. The inner diameter of the steerable introducer is 13F. The steerable introducer includes a hemostasis valve to minimize blood loss during catheter introduction and/or exchange. It has a sideport with a three-way stopcock for air or blood aspiration, fluid infusion, blood sampling, and pressure monitoring. The handle is equipped with a rotating collar to deflect the tip clockwise ≥180° and counterclockwise ≥90°. The steerable introducer features distal vent holes to facilitate aspiration and minimize cavitation and a radiopaque tip marker to improve fluoroscopic visualization.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device and, as such, focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study proving the device meets specific acceptance criteria in the way a clinical trial would for an AI/algorithm-based product.

    The information provided describes design verification activities for a hardware medical device (steerable introducer), not an AI algorithm. Therefore, many of the requested categories (e.g., sample size for test/training set, ground truth experts, MRMC studies, standalone algorithm performance) are not applicable.

    Here's an analysis based on the provided text, addressing the relevant points:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission states: "Design verification activities were performed and met the respective acceptance criteria to ensure that the devices in scope of this submission are safe and effective." However, specific numerical acceptance criteria and their corresponding reported device performance values are not detailed in this summary. The document lists the types of testing performed and the standards met.

    Acceptance Criterion (Type of Test)Reported Device Performance (Implied)
    Performance testing of modified designMet acceptance criteria (no specific numerical values provided)
    Biocompatibility testingMet acceptance criteria (no specific numerical values provided)
    Sterility testingMet acceptance criteria (no specific numerical values provided)
    Shelf-life testingMet acceptance criteria (no specific numerical values provided)
    Adherence to ISO 80369-7 (Small-bore connectors)Compliance confirmed
    Adherence to ANSI AAMI ISO 10993-1 (Biological evaluation)Compliance confirmed
    Adherence to ANSI AAMI ISO 11135 (Sterilization)Compliance confirmed
    Adherence to ANSI AAMI ST72 (Bacterial endotoxins)Compliance confirmed
    Adherence to ISO 11607-1 & -2 (Packaging)Compliance confirmed
    Adherence to ISO 15223-1 & ISO 20417 (Symbols & Information to be supplied)Compliance confirmed

    2. Sample Size Used for the Test Set and Data Provenance

    This is not applicable as the document describes physical device testing, not an AI/algorithm-based test set with data provenance. "Test set" here refers to the physical devices subjected to various engineering and biological tests. The sample sizes for each specific test (e.g., number of devices for durability, sterility, or biocompatibility) are not provided in this summary, but would be part of the full design verification report.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. "Ground truth" in the context of expert consensus is relevant for AI algorithms interpreting medical images or data. For a physical medical device like an introducer, "ground truth" is typically established by engineering specifications, validated test methods, and biological safety standards. Experts involved would be engineers, microbiologists, and other technical specialists relevant to the specific tests performed (e.g., biocompatibility specialist, sterilization expert).

    4. Adjudication Method for the Test Set

    Not applicable. This refers to methods for reconciling differing expert opinions on ground truth for AI algorithms. For device performance testing, adjudication involves comparing test results against predefined acceptance criteria from engineering specifications and regulatory standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. without AI Assistance

    Not applicable. This is a hardware device (catheter introducer), not an AI system that assists human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is a hardware device, not a standalone algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is based on established engineering specifications, international and national standards (e.g., ISO, ANSI AAMI listed), and regulatory requirements for mechanical performance, material compatibility, biocompatibility, and sterility. For example, for biocompatibility, the ground truth is that the materials do not elicit an adverse biological response as defined by ISO 10993-1.

    8. The Sample Size for the Training Set

    Not applicable. This refers to AI/machine learning models.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This refers to AI/machine learning models.

    Summary of Study Type:

    The study referenced is a series of non-clinical design verification activities and laboratory tests. These tests were conducted to demonstrate that the modifications made to the Agilis NxT Steerable Introducer Dual-Reach™ (the subject device) did not adversely affect its safety and effectiveness and that it continues to meet established performance, biological safety, and sterility standards, making it substantially equivalent to its predicate device. This is typical for a 510(k) submission for a modified hardware device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241370
    Manufacturer
    Date Cleared
    2024-06-13

    (30 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Agilis™ NxT Steerable Introducer Dual-Reach™ (I-V2-MED)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Agilis NxT Steerable Introducer Dual-Reach™ is indicated for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.

    Device Description

    The Agilis™ NxT Steerable Introducer Dual-Reach™ set consists of a dilator and steerable introducer, which is designed to provide flexible catheter positioning in the cardiac anatomy. The inner diameter of the steerable introducer is 13 F. The steerable introducer includes a hemostasis valve to minimize blood loss during catheter introduction and/or exchange. It has a sideport with three-way stopcock for air or blood aspiration, fluid infusion, blood sampling, and pressure monitoring. The handle is equipped with a rotating collar to deflect the tip clockwise ≥135° and counterclockwise ≥90°. Actual curve angles will vary. The steerable introducer features distal vent holes to facilitate aspiration and minimize cavitation and a radiopaque tip marker to improve fluoroscopic visualization.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for a medical device, the Agilis™ NxT Steerable Introducer Dual-Reach™. This document focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a study proving a device meets specific acceptance criteria for performance in a clinical or diagnostic context.

    The document states that the device is a catheter introducer and its indications for use are "introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum."

    The "study" referenced in the document is a non-clinical testing summary for design verification to ensure substantial equivalence to a predicate device, not a performance study in the sense of accuracy, sensitivity, or specificity for a diagnostic algorithm.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria for clinical performance (e.g., accuracy, sensitivity, specificity, or specific outcomes in patients) or report such performance metrics. Instead, it indicates that "Design verification activities were performed and met the respective acceptance criteria to ensure that the devices in scope of this submission are substantially equivalent to the predicate device."

    The types of testing performed, which would have their own internal acceptance criteria, are:

    Acceptance Criteria Category (Implied)Reported Device Performance (Summary)
    Mechanical/Functional PerformanceMet respective acceptance criteria for design verification.
    BiocompatibilityTesting performed.
    SterilityTesting performed.
    Packaging IntegrityTesting performed.

    2. Sample size used for the test set and the data provenance

    Not applicable. This is a submission for a physical medical device, not a software algorithm tested on a data set. The "testing" refers to non-clinical laboratory or engineering tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a submission for a physical medical device. Ground truth, in the context of clinical expert review of data/images, is not relevant here.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is a submission for a physical medical device.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document is for a physical medical device and does not involve AI or human readers for diagnostic interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a submission for a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable. For this type of physical medical device, "ground truth" would be established through engineering specifications, material standards, and performance benchmarks for the physical properties and function of the device in a controlled environment.

    8. The sample size for the training set

    Not applicable. This is a submission for a physical medical device, not an AI/ML algorithm.

    9. How the ground truth for the training set was established

    Not applicable. This is a submission for a physical medical device, not an AI/ML algorithm.

    Summary from the document:

    The Agilis™ NxT Steerable Introducer Dual-Reach™ is a physical medical device. The "study" mentioned in the document is a series of non-clinical design verification activities, biocompatibility, sterility, and packaging tests conducted in accordance with industry standards and guidance documents. These tests were performed to demonstrate that the modified device is substantially equivalent to its predicate device (Agilis™ NxT Steerable Introducer) and that all associated risks with the modifications were mitigated. The device is not a diagnostic algorithm or AI software, and therefore, typical performance metrics, sample sizes for data sets, expert involvement for ground truth, or MRMC studies are not applicable or described in this document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1