Search Results
Found 2 results
510(k) Data Aggregation
(30 days)
Advantage Ultra System, Advantage Fit Ultra System, Lynx Ultra System
Advantage Ultra System and Advantage Fit Ultra System
The mesh implant is intended for use as a suburethral sling for the treatment of female stress urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency.
The Advantage Ultra System and the Advantage Fit Ultra System delivery devices are intended for use as an aid in insertion, placement, fixation, and anchoring of the Advantage Ultra surgical mesh during urogynecological procedures.
Lynx Ultra System
The mesh implant is intended for use as a suburethral sling for the treatment of female stress urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency.
The Lynx Ultra System delivery device is intended for use as an aid in insertion, placement, fixation, and anchoring of the Lynx Ultra surgical mesh during urogynecological procedures.
Advantage Ultra System and Advantage Fit Ultra System
The Advantage Ultra System and the Advantage Fit Ultra System are sterile, single-use systems each consisting of one delivery device and one blue mesh assembly. The mesh assembly is comprised of a polypropylene knitted mesh protected by two disposable polymer sleeves, two dilators, and one center tab. The sleeves and dilators are secured to the mesh by two polypropylene leader loops, and the center tab is secured to the mesh by a polyester lead. The two dilators at the distal ends of the mesh assembly are designed to be placed over the needle end of the delivery device.
The Advantage Ultra System and the Advantage Fit Ultra Systems are each packaged with their respective delivery device (Advantage Ultra or Advantage Fit Ultra delivery device). This is a single-use disposable delivery device that consists of a handle with a curved needle and a pusher component. The delivery device is designed to facilitate the passage of the mesh assembly through bodily tissues for retropubic placement.
The Advantage Fit Ultra System includes the same mid-urethral mesh implant as the Advantage Ultra System but differs in the design of the dilators and delivery devices. The needle portion of the Advantage Fit Ultra delivery device has a smaller outer diameter and a smaller bend radius and is offered to provide physicians with surgical options. The Advantage Fit Ultra dilators also have a smaller diameter to accommodate for the different needle size.
Lynx Ultra System
The Lynx Ultra System is a sterile, single-use system consisting of two delivery devices and one blue mesh assembly. The mesh assembly is comprised of a polypropylene knitted mesh protected by two disposable polymer sleeves, two dilators, and one center tab. The sleeves and dilators are secured to the mesh by two polypropylene leader loops, and the center tab is secured to the mesh by a polyester lead. At the distal ends of the mesh assembly there are association loops designed to be placed in the needle slot of the distal end of a delivery device.
This document is a 510(k) summary for the Advantage™ Ultra System, Advantage Fit™ Ultra System, and Lynx™ Ultra System, surgical mesh devices for treating female stress urinary incontinence. The core of this submission is to demonstrate that these devices are substantially equivalent to previously cleared predicate devices (K211223). The primary change in these new submissions is related to procedural steps in the electronic Instructions for Use (eIFU), specifically regarding the release and removal of disposable components of the mesh assembly.
Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context are focused on ensuring that the changes (eIFU modifications) do not negatively impact safety or effectiveness, and that the device remains substantially equivalent to its predicate. There isn't a traditional performance study proving clinical efficacy against a disease endpoint, as this is a 510(k) submission for substantial equivalence.
Here's an interpretation of your request based on the provided document:
Acceptance Criteria and Device Performance (in the context of a 510(k) for eIFU changes)
Since the document explicitly states "there are no changes to the intended use or indications for use, design, materials, sterilization, technological and performance characteristics, or principle of operation," the acceptance criteria are implicitly tied to demonstrating that the eIFU modifications do not introduce new safety or effectiveness concerns and allow the device to function as intended.
Table of Acceptance Criteria and Reported Device Performance
Acceptance Criterion (Implicit) | Reported Device Performance |
---|---|
1. User Needs Met: All user needs impacted by eIFU modifications are adequately addressed. | Design Validation: "objective evidence that all user needs impacted by the eIFU modifications were met and overall performance for the intended clinical use was acceptable." |
2. Functionality Maintained: Ultra Products continue to function as intended with the new eIFU. | Summative Usability Evaluation: Demonstrated that "the Ultra Products continue to function as intended." |
3. No New Safety/Effectiveness Issues: eIFU changes do not introduce new issues of safety or effectiveness compared to predicate devices. | Summative Usability Evaluation: Demonstrated that the Ultra Products "do not raise new issues of safety or effectiveness compared to the predicate devices." |
Overall Conclusion: "concluded that the proposed Ultra Products are substantially equivalent to the established predicate devices." |
Study Details
The studies performed are "Performance bench testing" in the form of "Design Validation" and a "Summative Usability Evaluation."
-
Sample size used for the test set and the data provenance:
- Test Set Sample Size: The document does not specify the exact number of participants (users) or instances for the "Design Validation" and "Summative Usability Evaluation." It refers generally to "performance bench testing."
- Data Provenance: Not specified, but "bench testing" typically implies laboratory or simulated environments, not patient data.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable in the sense of clinical "ground truth" (e.g., pathology, outcomes). The "ground truth" here is the successful and safe use of the device according to the eIFU. The design validation and usability studies themselves aim to verify this by testing user interaction and performance. The document does not specify the number or qualifications of evaluators/experts involved in conducting or assessing these bench tests.
-
Adjudication method for the test set:
- Not explicitly stated. For bench testing and usability studies, adjudication might involve expert review of test results, error analysis, and user feedback, but no specific method like "2+1" or "3+1" is mentioned.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is not an AI/imaging device. The study type is bench testing and usability evaluation for a surgical mesh delivery system with modified instructions for use.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable. This is a medical device (surgical mesh and delivery system), not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- The "ground truth" for this submission's studies is the successful and safe operation of the device according to the updated eIFUs. This is assessed via controlled testing environments (bench testing) and usability studies, rather than clinical endpoints, pathology, or outcomes data, as this 510(k) specifically addresses changes to the instructions for use, not the fundamental device design or clinical performance.
-
The sample size for the training set:
- Not applicable. This is not a machine learning/AI device, so there is no "training set."
-
How the ground truth for the training set was established:
- Not applicable (as above).
Ask a specific question about this device
(89 days)
Advantage Ultra System, Advantage Fit Ultra System, Lynx Ultra System
The mesh implant is intended for use as a suburethral sling for the treatment of female stress urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency.
The Advantage Ultra System is a sterile, single-use system consisting of one delivery device and one mesh assembly. The mesh assembly is comprised of a polypropylene knitted mesh protected by two disposable polymer sleeves, two dilators, and one center tab. The sleeves and dilators are secured to the mesh by two polypropylene leader loops, and the center tab is secured to the mesh by a polyester lead. The two dilators at the distal ends of the mesh assembly are designed to be placed over the needle end of the delivery device. The Advantage Ultra System is packaged with the Advantage™ System Delivery Device previously cleared via K172565. This is a single-use disposable delivery device that consists of a handle with a curved needle and a pusher component. The delivery device is designed to facilitate the passage of the mesh assembly through bodily tissues for retropubic placement.
The Advantage Fit Ultra System is a sterile, single-use system consisting of one delivery device and one mesh assembly. The mesh assembly is comprised of a polypropylene knitted mesh protected by two disposable polymer sleeves, two dilators, and one center tab. The sleeves and dilators are secured to the mesh by two polypropylene leader loops, and the center tab is secured to the mesh by a polyester lead. The two dilators at the distal ends of the mesh assembly are designed to be placed over the needle end of the delivery device. The Advantage Fit Ultra System is packaged with the Advantage Fit™ System Delivery Device previously cleared via K172565. This is a single-use disposable delivery device that consists of a handle with a curved needle and a pusher component. The delivery device is designed to facilitate the passage of the mesh assembly through bodily tissues for retropubic placement.
The Lynx Ultra System is a sterile, single-use system consisting of two delivery devices and one mesh assembly. The mesh assembly is comprised of a polypropylene knitted mesh protected by two disposable polymer sleeves, two dilators, and one center tab. The sleeves and dilators are secured to the mesh by two polypropylene leader loops, and the center tab is secured to the mesh by a polyester lead. At the distal ends of the mesh assembly there are association loops designed to be placed in the needle slot of the distal end of a delivery device. The Lynx Ultra System is packaged with two Lynx System Delivery Devices previously cleared via K172565. Each is a single-use disposable delivery device that consists of a handle with a curved needle with a needle slot for connection to the mesh assembly. The needle is designed to facilitate the passage of the mesh assembly through bodily tissues for suprapubic placement.
The provided text describes the 510(k) summary for the Advantage™ Ultra System, Advantage Fit™ Ultra System, and Lynx™ Ultra System, which are surgical meshes for treating female stress urinary incontinence. This document focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a clinical study or a standalone algorithm performance study.
Therefore, the information required for "a table of acceptance criteria and the reported device performance" and the details of a study proving those criteria are not explicitly present in the provided text in the way one might expect for a typical performance validation study with specific metrics like sensitivity, specificity, or accuracy for an AI/CADe device.
However, based on the provided text, the closest interpretations for acceptance criteria and the "study" demonstrating their fulfillment are related to performance testing to ensure the device functions as intended and meets specifications compared to its predicates.
Here's an analysis based on the available information:
1. A table of acceptance criteria and the reported device performance
The document does not provide a table with specific quantitative acceptance criteria (e.g., minimum tensile strength value) explicitly alongside reported performance results. Instead, it broadly states that performance testing was conducted, and the results "demonstrate that the proposed products function as intended and meet specification requirements."
Acceptance Criteria Category | Reported Device Performance (Summary) |
---|---|
Functional & Mechanical | |
Dimensional Tests | Products function as intended and meet specification requirements. |
Tensile Strength Tests | Products function as intended and meet specification requirements. |
Full Functional Tests | Products function as intended and meet specification requirements. |
Usability | |
User Evaluations | Products function as intended and meet specification requirements. |
Packaging & Sterility | |
Packaging Tests | Products function as intended and meet specification requirements. |
Sterility | Products function as intended and meet specification requirements. |
Biocompatibility | |
Biocompatibility Testing | All patient contacting materials meet applicable biocompatibility standards per ISO 10993-1:2018 and FDA Guidance. |
Shelf Life | |
Accelerated Aging (13 months) | Performance testing on non-aged and accelerated aged samples supports a 1-year shelf life. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document does not specify the sample sizes for the performance tests (dimensional, tensile, functional, user evaluations, packaging, sterility, biocompatibility). It also does not mention the data provenance in terms of country of origin or whether a retrospective or prospective study was conducted for establishing clinical performance (as this submission is focused on substantial equivalence based on materials and design, not clinical outcomes from a trial). The "samples" referred to are likely physical devices or components of the device itself (non-aged and accelerated aged for shelf life testing).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This information is not applicable and not provided in the document. The performance tests described (dimensional, tensile, functional, biocompatibility, sterility) are technical evaluations of the device's physical properties and safety, not assessments requiring expert "ground truth" in a clinical diagnostic sense. "User evaluations" are mentioned, which would involve experts (e.g., surgeons or medical professionals) interacting with the device, but the number and qualifications are not detailed.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
This information is not applicable and not provided. Adjudication methods are typically used in clinical studies or studies involving human interpretation (e.g., radiology reads) to establish ground truth or handle discrepancies, neither of which is described as part of the performance testing for this surgical mesh.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable as the device is a surgical mesh, not an AI/CADe system for diagnostic imaging interpretation. Therefore, no MRMC study or AI-related effectiveness details are provided.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable as the device is a surgical mesh, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the performance testing, the "ground truth" would be the established engineering specifications, material standards (e.g., ISO 10993-1:2018 for biocompatibility), and regulatory requirements for sterility and packaging integrity. These are objective measures and standards rather than clinical "ground truth" derived from patient data or expert consensus on a diagnosis.
8. The sample size for the training set
This is not applicable as the device is a physical surgical mesh, not an AI model requiring a training set.
9. How the ground truth for the training set was established
This is not applicable as the device is a physical surgical mesh, not an AI model.
Ask a specific question about this device
Page 1 of 1