Search Results
Found 2 results
510(k) Data Aggregation
(29 days)
AXS Vecta Aspiration System
The AXS Vecta Aspiration Catheter, as part of the AXS Vecta Aspiration System, is indicated in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - MI and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment.
The Stryker AXS Vecta Aspiration System includes an aspiration catheter and its accessories, including the Scout Introducer, the Peel-Away Introducers, the Hemostasis Valve, the AXS Universal Aspiration Tubing, the Medela Dominant Flex Pump, and the AXS Universal Liner Set.
This document is a 510(k) summary for the AXS Vecta Aspiration System, which is a percutaneous catheter used for revascularization in patients with acute ischemic stroke. The submission seeks to demonstrate substantial equivalence to a previously cleared predicate device (K190212).
Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
The document provides a "Performance Testing Summary" table (Table 2) which includes the "Test" conducted and the "Conclusion". While it states "All units met the acceptance criteria and passed," the specific acceptance criteria values for each test are not explicitly listed in this document. The table implicitly states that the device met the unlisted criteria.
Test | Reported Device Performance (Conclusion) | Explicit Acceptance Criteria |
---|---|---|
Visual Inspection (Packaging: Pouch Visual) | All units met the acceptance criteria and passed Packaging Visual Inspection. | Not explicitly stated |
Visual Inspection (Packaging: Undamaged Product) | All units met the acceptance criteria and passed Packaging Visual Inspection. | Not explicitly stated |
Tensile Strength | All units met the acceptance criteria and passed Tensile Strength testing. | Not explicitly stated |
PTFE Delamination | All units met the acceptance criteria and passed PTFE Delamination testing. | Not explicitly stated |
Torque Strength | All samples met acceptance criteria and passed Torque Strength testing. | Not explicitly stated |
Catheter Burst | All samples met acceptance criteria and passed Catheter Burst testing. | Not explicitly stated |
Leak (Liquid) | ||
Leak (Air) | All samples met acceptance criteria and passed the Air and Liquid Leakage testing. | Not explicitly stated |
Dimensional (ID, OD, & Working Length) | All samples met acceptance criteria and passed Dimensional testing. | Not explicitly stated |
Kink Resistance | All samples met acceptance criteria and passed Kink Resistance testing. | Not explicitly stated |
Visual Inspection (Transition & Tip) | All samples met acceptance criteria and passed both the transition and tip visual inspections. | Not explicitly stated |
Lumen Patency | All samples met acceptance criteria and passed Lumen Patency testing. | Not explicitly stated |
Vacuum Drop | All samples met acceptance criteria and passed Vacuum Drop testing. | Not explicitly stated |
Tip Flexibility | All samples met acceptance criteria and passed Tip Flexibility testing. | Not explicitly stated |
Friction Force | All samples met acceptance criteria and passed Friction Force testing. | Not explicitly stated |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document states, "All units met the acceptance criteria" or "All samples met acceptance criteria" for each test, implying that the acceptance criteria were met by all tested units/samples. However, the specific number of samples used for each non-clinical bench test is not explicitly mentioned.
No animal or clinical studies were performed for the current submission. The document states that animal testing previously conducted for the AXS Vecta Aspiration System was leveraged (from K190212, K172167, and K181354). The provenance of this leveraged animal data (country of origin, retrospective/prospective) is not specified in this 510(k) summary.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This 510(k) summary primarily relies on non-clinical bench testing. Therefore, there are no human experts establishing ground truth in the context of diagnostic performance or clinical outcomes. The "ground truth" for bench tests is defined by engineering specifications and test protocols.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Since no clinical or human-reader studies were conducted for this submission, an adjudication method for a test set is not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was done as this is a medical device (catheter) and not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This refers to an algorithm's performance without human interaction. This is not applicable as the device is a physical catheter, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the non-clinical bench tests (e.g., tensile strength, burst pressure, dimensions), the ground truth is based on engineering specifications and established test methods that define what constitutes acceptable performance for a medical device of this type.
8. The sample size for the training set
This is not applicable as the device is a physical medical device (catheter) and not a machine learning or AI model that requires a training set.
9. How the ground truth for the training set was established
This is not applicable for the same reason as above.
Ask a specific question about this device
(79 days)
AXS Vecta Aspiration System
The AXS Vecta Aspiration Catheter, as part of the AXS Vecta Aspiration System, is indicated in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - MI and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment.
The Stryker AXS Vecta Aspiration System includes an aspiration catheter and its accessories, including the Scout Introducer, the Peel-Away Introducers, the Hemostasis Valve, the AXS Universal Aspiration Tubing, the Medela Dominant Flex Pump, and the AXS Universal Liner Set.
The acceptance criteria and study proving the device meets them are described in the provided text.
Here's the breakdown:
1. Table of Acceptance Criteria and Reported Device Performance
Test | Test Method Summary | Acceptance Criteria | Conclusions (Reported Device Performance) |
---|---|---|---|
Simulated Use - Bench (AXS Vecta Aspiration System) | AXS Vecta Aspiration System underwent simulated use testing by a physician in a benchtop model. | Test samples must meet predetermined user needs. | AXS Vecta Aspiration System test samples met the acceptance criteria for Simulated Use - Bench to demonstrate that the AXS Vecta System is substantially equivalent to the predicate device. |
Simulated Use Testing – Usability (AXS Vecta Aspiration System) | AXS Vecta Aspiration System underwent simulated use testing by a physician in a benchtop model. | Test samples must meet predetermined user needs. | AXS Vecta Aspiration System test samples met the acceptance criteria for Simulated Use - Usability to demonstrate that the AXS Vecta Aspiration System is substantially equivalent to the predicate device. |
Vacuum Drop / Suction Connector Secure Attachment (Aspiration Tubing Set) | Vacuum pressure measured at source and tip to evaluate pressure difference. | Test sample results must meet or exceed existing pressure specifications. | AXS Vecta Aspiration System test samples met the acceptance criteria for Vacuum Drop to demonstrate that the AXS Vecta Aspiration System is substantially equivalent to the predicate device. |
Lumen Patency (Aspiration Tubing Set) | Samples were evaluated for lumen collapse during aspiration. | Test sample results must meet or exceed existing lumen patency specifications. | AXS Vecta Aspiration System test samples met the acceptance criteria for Lumen Patency to demonstrate that the AXS Vecta Aspiration System is substantially equivalent to the predicate device. |
Note: For Calculated Force at Tip (shown in a previous table within the document), the reported values for the Subject Device (AXS Vecta Aspiration System) are 0.025 kgf for 0.071 in ID and 0.027 kgf for 0.074 in ID. The original document does not explicitly state the acceptance criteria for force at tip, but compares it to values for the predicate and reference devices, implying that similar characteristics are acceptable. | Note: For Calculated Force at Tip (shown in a previous table within the document), the measurement method is not explicitly detailed but involves measuring Catheter Inner Diameter (ID) to calculate force. | Note: For Calculated Force at Tip, the document doesn't explicitly state numerical acceptance criteria. However, the comparison table implicitly suggests that the calculated forces for the subject device (0.025 kgf for 0.071 in ID and 0.027 kgf for 0.074 in ID) are considered acceptable because they are similar to or slightly higher than the predicate device (0.024 kgf for 0.071 in ID, 0.026 kgf for 0.074 in ID) which was already cleared. The text states: "a calculation of the force at the tip demonstrated that the pressure differences and forces exhibited on the product with the different pumps are minimal." | The calculated forces at the tip for the AXS Vecta Aspiration System (0.025 kgf for 0.071 in ID and 0.027 kgf for 0.074 in ID) are shown to be comparable to the predicate device, supporting substantial equivalence. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document repeatedly refers to "test samples" for each test. However, the exact number of samples used for "Simulated Use - Bench," "Simulated Use – Usability," "Vacuum Drop / Suction Connector Secure Attachment," and "Lumen Patency" tests is not explicitly stated.
- Data Provenance: The tests are described as "benchtop model" tests and "Animal testing previously conducted for the Zenith Flex System was used to support this change." This indicates:
- Bench Tests: Performed in a laboratory setting.
- Animal Testing: The previous animal testing for the predicate device (Zenith Flex System) was retrospectively leveraged. The original context of this animal testing (country of origin, etc.) is not detailed in this document.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- For "Simulated Use - Bench" and "Simulated Use – Usability," the testing was conducted "by a physician." The document specifies "a physician" (singular).
- Qualifications of the physician: Not explicitly stated beyond "physician."
4. Adjudication Method for the Test Set
- The document does not describe a specific adjudication method (e.g., 2+1, 3+1). The "Simulated Use" tests were conducted "by a physician," and the "test samples must meet predetermined user needs." It implies a direct assessment by the physician against predefined criteria rather than an adjudication process involving multiple reviewers of outcomes.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC study was done. The document explicitly states: "No clinical study was conducted as bench and animal testing was determined sufficient for verification and validation purposes."
6. Standalone (Algorithm Only) Performance Study
- Not applicable. This device is a medical aspiration system, not an AI algorithm. Therefore, a standalone (algorithm only) performance study would not be relevant.
7. Type of Ground Truth Used
- Expert Assessment/Predicate Device Performance:
- For the "Simulated Use" tests, the ground truth was effectively whether the "test samples met predetermined user needs" as assessed by a physician.
- For the engineering tests (Vacuum Drop, Lumen Patency, Calculated Force at Tip), the ground truth was based on meeting "existing pressure specifications" or "existing lumen patency specifications," which are predefined technical performance standards, and comparison to the performance of the predicate device.
- For the animal testing (leveraged from the predicate), the ground truth would have been safety and efficacy outcomes observed in the animal model.
8. Sample Size for the Training Set
- Not applicable. This device is a physical medical device, not an AI algorithm. Therefore, there is no "training set."
9. How the Ground Truth for the Training Set Was Established
- Not applicable. As above, there is no training set for this type of device.
Ask a specific question about this device
Page 1 of 1