(156 days)
The Zenith Flex System, including the Zetheter, Aspiration Tubing Set, and VC-701 Cliq Aspirator Pump, is indicated in the revascularization of patients with acute secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment.
The InNeuroCo 074 Zenith Flex Catheter is a variable stiffness catheter that has a catheter shaft reinforced with Stainless-Steel and Nitinol to provide support. The Stainless-Steel is wound as a double coil in the proximal section and the Nitinol is a single coil in the distal section. It has a radiopaque Platinum/lridium marker band on the distal 25 cm of the 074 Zenith Flex Catheter has a hydrophilic coating. The 074 Zenith Flex Catheter is available with an internal diameter of 0.074 inches. The outer diameter is of 0.089 inches along its proximal shaft, and 0.083 inches along its distal shaft. The 074 Zenith Flex Catheter is available in three workinq lengths: 115 cm, 125 cm, and 132 cm. The 074 Zenith Flex Catheter has a PTFE lined lumen. Accessories included with the device are a Tuohy-Borst Hemostasis Valve, two peel-away Introducers, and a Scout introducer. The Scout may be used to introduce the 074 Zenith Flex Catheter into distal vasculature, thereby helping the device reach the target anatomy.
The provided text describes the Zenith Flex System (074 Zenith Flex Catheter) and its non-clinical data for demonstrating substantial equivalence. It does not describe a study involving an AI/Machine Learning device or algorithm. Therefore, I cannot provide details regarding acceptance criteria, sample size, ground truth, or adjudication methods in the context of an AI study.
However, I can extract the acceptance criteria and observed performance described for the non-clinical (bench and animal) testing of the Zenith Flex Catheter.
1. Table of Acceptance Criteria and Reported Device Performance
| Test | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Zenith Flex Catheter & its accessories Biocompatibility | ||
| Material Mediated Pyrogen | The test article extracts must not cause a febrile reaction greater than 0.5°C in any individual subject. | Test articles met the acceptance criteria (leveraged from K172167, materials and manufacturing processes equivalent). |
| Cytotoxicity MEM Elution | The cultures treated with the test article must not have a reactivity grade greater than 2. | Test articles met the acceptance criteria (leveraged from K172167, materials and manufacturing processes equivalent). |
| Hemolysis (ASTM Method, extract human blood) | The hemolytic index above the negative control article must be less than 5%. | Test articles met the acceptance criteria (leveraged from K172167, materials and manufacturing processes equivalent). |
| Unactivated Partial Thromboplastin Time (UPTT) | There must be no statistical decrease between the UPTT of plasma exposed to the test article and to the negative or untreated control. | Test articles met the acceptance criteria (leveraged from K172167, materials and manufacturing processes equivalent). |
| Complement Activation | There must be no statistical increase between either the C3a or SC5b-9 concentrations in plasma exposed to the test article as compared to the negative and untreated controls. | Test articles met the acceptance criteria (leveraged from K172167, materials and manufacturing processes equivalent). |
| Dog Thromboresistance | The test articles must receive a thrombus formation score less than or equal to that of the control. | Test articles met the acceptance criteria (leveraged from K172167, materials and manufacturing processes equivalent). |
| Maximization Sensitization | The test article must elicit a positive response in less than 10% of the test animals. | Test articles met the acceptance criteria (leveraged from K172167, materials and manufacturing processes equivalent). |
| Intracutaneous Toxicity/Reactivity | The test article extracts must not induce a significantly greater biological reaction than the control. | Test articles met the acceptance criteria (leveraged from K172167, materials and manufacturing processes equivalent). |
| Acute Systemic Toxicity Test | The test article extracts must not induce a significantly greater biological reaction than the control. | Test articles met the acceptance criteria (leveraged from K172167, materials and manufacturing processes equivalent). |
| Chemical Compatibility | Product shall withstand exposure to chemicals without degradation. | Test articles met the acceptance criteria (leveraged from K172167, materials and manufacturing processes equivalent). |
| Zenith Flex Catheter Performance Testing | ||
| Working Length | Test samples should be within existing working length specification. | Zenith Flex Catheter test samples met the acceptance criteria. |
| Hub compatibility | Hub shall meet existing Luer specifications. | Test articles met the acceptance criteria (leveraged from K171672, materials and manufacturing processes equivalent). |
| Outside diameter | Test samples should be within existing outside diameter specification. | Zenith Flex Catheter test samples met the acceptance criteria. |
| Leak - Air | Test samples should be within existing Air - Leak specifications. | Zenith Flex Catheter test samples met the acceptance criteria. |
| Leak - Liquid | Test samples should be within existing Leak - Liquid specifications. | Zenith Flex Catheter test samples met the acceptance criteria. |
| Particulates | Test samples should be within existing Particulate specifications. | Test articles met the acceptance criteria (leveraged from K172167, materials and manufacturing processes equivalent). |
| Simulated Use - Bench | Test samples must meet predetermined user needs. | Zenith Flex Catheter test samples met the acceptance criteria. |
| Catheter Burst | Test sample burst pressures must meet or exceed existing minimum burst pressure specification. | Zenith Flex Catheter test samples met the acceptance criteria. |
| Tensile | Test sample ultimate tensile strength must meet or exceed existing tensile strength specifications. | Zenith Flex Catheter test samples met the acceptance criteria. |
| Corrosion | Test samples shall exhibit no evidence of corrosion. | Test articles met the acceptance criteria (leveraged from K152202, materials and manufacturing processes equivalent). |
| PTFE Liner inspection | Test sample liner adhesion must meet or exceed existing PTFE Liner inspection specifications. | Zenith Flex Catheter test samples met the acceptance criteria. |
| Hydrophilic Coating Integrity | Test sample results must meet or exceed existing Hydrophilic Coating Integrity specifications. | Test articles met the acceptance criteria (leveraged from K172167, materials and manufacturing processes equivalent). |
| Simulated Use Testing (Aspiration) | Successfully removed the thrombi. | Zenith Flex System test samples met the acceptance criteria. |
| Labeling Legibility | Test samples shall demonstrate text legibility. | Test articles met the acceptance criteria (leveraged from K172167, materials and manufacturing processes equivalent). |
| Barcode | Test samples shall demonstrate readily readable barcodes. | Test articles met the acceptance criteria (leveraged from K172167, materials and manufacturing processes equivalent). |
| Sterilization (Catheter & Aspiration Tubing Set) | Sterilization load shall pose an equal or lesser challenge to sterilize than existing sterile product loads. | Sterilization loads met the acceptance criteria. |
| Shelf Life (Catheter & Aspiration Tubing Set) | Aged test samples must meet or exceed existing specifications. | Zenith Flex Catheter & Aspiration Tubing Set test samples met the acceptance criteria. |
| Torque Strength | Test sample results must meet or exceed existing torque specifications. | Zenith Flex Catheter test samples met the acceptance criteria. |
| Kink Resistance | Test sample results must meet or exceed existing kink resistance specifications. | Zenith Flex Catheter test samples met the acceptance criteria. |
| Lumen Patency | Test sample results must meet or exceed existing lumen patency specifications. | Zenith Flex Catheter test samples met the acceptance criteria. |
| Tip Flexibility | Zenith Flex tip flexibility results were compared to predicate device results. | Zenith Flex Catheter test samples performed comparatively against the predicate device. |
| Friction Force | Zenith Flex friction force results were compared to predicate device results. | Zenith Flex Catheter test samples performed comparatively against the predicate device. |
| Simulated Use Testing - Usability | Test samples must meet predetermined user needs. | Zenith Flex System test samples met the acceptance criteria. |
| Aspiration Tubing Set Performance Testing | ||
| Visual Inspection | Test samples should meet visual inspection specifications. | Test articles met the acceptance criteria (leveraged from K172167, aspiration tubing is the same). |
| Tensile | Test sample ultimate tensile strength must meet or exceed existing tensile strength specifications. | Test articles met the acceptance criteria (leveraged from K172167, aspiration tubing is the same). |
| Leak - Liquid | Test samples should be within existing Leak - Liquid specifications. | Test articles met the acceptance criteria (leveraged from K172167, aspiration tubing is the same). |
| Leak - Air, Tubing and Control Switch | Test samples should be within existing Air - Leak specifications. | Test articles met the acceptance criteria (leveraged from K172167, aspiration tubing is the same). |
| Luer Compatibility | Hub shall meet existing Luer specifications. | Test articles met the acceptance criteria (leveraged from K172167, aspiration tubing is the same). |
| Suction Connector Separation Force | Test sample results must meet or exceed existing force specifications. | Test articles met the acceptance criteria (leveraged from K172167, aspiration tubing is the same). |
| Vacuum Drop / Suction Connector Secure Attachment | Test sample results must meet or exceed existing pressure specifications. | Test articles met the acceptance criteria (leveraged from K172167, aspiration tubing is the same). |
| Lumen Patency | Test sample results must meet or exceed existing lumen patency specifications. | Test articles met the acceptance criteria (leveraged from K172167, aspiration tubing is the same). |
| Dimensions | Test samples should be within existing dimensional specifications. | Test articles met the acceptance criteria (leveraged from K172167, aspiration tubing is the same). |
| Packaging (Catheter and Aspiration Tubing Set) Testing | ||
| Packaging - Dye Leak | Test sample shall not exhibit any visual leaks or channels. | Test articles met the acceptance criteria (leveraged from K172167, packaging is the same). |
| Packaging - Peel | Test sample tensile strength must meet or exceed existing tensile strength specifications. | Test articles met the acceptance criteria (leveraged from K172167, packaging is the same). |
| Aspiration Tubing Set Biocompatibility | ||
| Cytotoxicity MEM Elution | The cultures treated with the test article must not have a reactivity grade greater than 2. | Test articles met the acceptance criteria (leveraged from K172167, aspiration tubing is the same). |
| Maximization Sensitization | The test article must elicit a positive response in less than 10% of the test animals. | Test articles met the acceptance criteria (leveraged from K172167, aspiration tubing is the same). |
| Intracutaneous Toxicity/Reactivity | The test article extracts must not induce a significantly greater biological reaction than the control. | Test articles met the acceptance criteria (leveraged from K172167, aspiration tubing is the same). |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify exact sample sizes for each benchtop or animal test. It refers to "test samples" for bench testing and does not provide numerical counts.
For animal testing: "Clot was used to create occlusions within several arteries of swine. One side of each swine was treated with the subject Zenith Flex system and the contralateral side was treated with the reference device." The exact number of swine used is not provided.
Data provenance: The non-clinical data is generated from laboratory (bench) and animal (swine) studies. The country of origin for the studies is not specified.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable in this context. The non-clinical studies rely on objective measurements and observations in laboratory and animal settings, not on expert ground truth for interpretation of images or clinical outcomes. For 'Simulated Use - Bench' and 'Simulated Use Testing - Usability', a physician's assessment was used, but details on the number or qualifications of these physicians are not provided.
4. Adjudication Method for the Test Set
Not applicable, as the tests involve objective measurements and direct observation rather than subjective interpretations requiring adjudication.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a medical device (catheter), not an AI/ML device for diagnostic interpretation. There was no MRMC study involving human readers and AI assistance.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done
Not applicable. This is a medical device (catheter), not an AI/ML device.
7. The Type of Ground Truth Used
For bench testing, the ground truth is based on physical and chemical properties measured against established specifications (e.g., ISO, ASTM standards).
For animal testing, the ground truth for effectiveness was "angiographic assessment of revascularization," and for safety, "angiographic assessment and histopathological assessment."
8. The Sample Size for the Training Set
Not applicable. This is a medical device (catheter); there is no "training set" in the context of an AI/ML algorithm.
9. How the Ground Truth for the Training Set was Established
Not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 25, 2018
InNeuroCo. Inc. Marianne Grunwaldt Director, Quality Assurance & Regulatory Affairs 4635 NW 103rd Avenue Sunrise, Florida 33351
Re: K181354
Trade/Device Name: Zenith Flex System (074 Zenith Flex Catheter) Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY. DTL. DYB Dated: September 21, 2018 Received: September 25, 2018
Dear Marianne Grunwaldt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Xiaolin Zheng -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
K181354
Device Name
Zenith Flex System (074 Zenith Flex Catheter)
Indications for Use (Describe)
The Zenith Flex System, including the Zetheter, Aspiration Tubing Set, and VC-701 Cliq Aspirator Pump, is indicated in the revascularization of patients with acute secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) SUMMARY
K181354
Submitter's Name and Address
InNeuroCo, Inc. 4635 NW 103rd Avenue Sunrise, Florida 33351 Telephone: 1-954-742-5988 Facsimile: 1-954-742-5989
Contact Information
Marianne Grunwaldt Director, Quality Assurance & Requlatory Affairs InNeuroCo, Inc 4635 NW 103rd Avenue Sunrise, Florida 33351 Telephone: 1-305-495-3883 Facsimile: 1-954-742-5989 E-Mail: Marianne@InNeuroCo.com
Secondary Contact
Ming Cheng Chew Regulatory Consultant, Libra Medical, Inc. 8401 73rd Ave N, Suite 63 Brooklyn Park, MN 55428 Telephone: 763-232-3701 E-Mail: mcchew@libramed.com
Date Prepared
October 22, 2018
Device Trade or Proprietary Name
Zenith Flex System (074 Zenith Flex Catheter)
{4}------------------------------------------------
Device Common or Classification Name:
Percutaneous Catheter, 21 CFR 870.1250, Class II Hemostasis Valve, 21 CFR 870.4290, Class II Catheter Introducer, 21 CFR 870.1340, Class II
Product Code:
NRY (Catheter, Thrombus Retriever) DTL (Hemostasis Valve) DYB (Catheter Introducer)
| Name of Predicate Device | Name of Manufacturer | 510(k) Number |
|---|---|---|
| Zenith Flex Catheter | InNeuroCo, Inc | K172167 |
| Name of Reference Device | Name of Manufacturer | 510(k) Number |
ldentification of the Legally Marketed Devices to which Equivalence is Being Claimed
Device Description
The InNeuroCo 074 Zenith Flex Catheter is a variable stiffness catheter that has a catheter shaft reinforced with Stainless-Steel and Nitinol to provide support. The Stainless-Steel is wound as a double coil in the proximal section and the Nitinol is a single coil in the distal section. It has a radiopaque Platinum/lridium marker band on the distal 25 cm of the 074 Zenith Flex Catheter has a hydrophilic coating. The 074 Zenith Flex Catheter is available with an internal diameter of 0.074 inches. The outer diameter is of 0.089 inches along its proximal shaft, and 0.083 inches along its distal shaft. The 074 Zenith Flex Catheter is available in three workinq lengths: 115 cm, 125 cm, and 132 cm. The 074 Zenith Flex Catheter has a PTFE lined lumen. Accessories included with the device are a Tuohy-Borst Hemostasis Valve, two peel-away Introducers, and a Scout introducer. The Scout may be used to introduce the 074 Zenith Flex Catheter into distal vasculature, thereby helping the device reach the target anatomy.
{5}------------------------------------------------
Indications for Use
The Zenith Flex System, including the Zenith Flex Catheter, Aspiration Tubing Set, and VC-701 Cliq Aspirator Pump, is indicated in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment.
| Predicate DeviceInNeuroCo, Inc.071 Zenith Flex | Reference DevicePenumbra Ace 68 | Subject DeviceInNeuroCo, Inc.074 Zenith Flex | |
|---|---|---|---|
| 510(k) Number | K172167 | K161064 | K181354 |
| Classification | Class II | Class II | Same |
| Product Code | NRY | NRY | Same |
| Review Panel | Neurology | Neurology | Same |
| Indications ForUse | The Zenith Flex System,including the Zenith FlexCatheter, AspirationTubing Set, and VC-701Cliq Aspirator Pump, isindicated in therevascularization ofpatients with acuteischemic strokesecondary to intracraniallarge vessel occlusivedisease (within theinternal carotid, middlecerebral - M1 and M2segments, basilar, andvertebral arteries) within8 hours of symptomonset. Patients who areineligible for intravenoustissue plasminogenactivator (IV t-PA) or whofailed IV t-PA therapy arecandidates for treatment. | The PenumbraSystem is intendedfor use in therevascularization ofpatients with acuteischemic strokesecondary tointracranial largevessel occlusivedisease (within theinternal carotid,middle cerebral- M1 and M2segments, basilar,and vertebralarteries) within 8hours of symptomonset. | Same as 071Zenith Flex |
| Predicate DeviceInNeuroCo, Inc.071 Zenith Flex | Reference DevicePenumbra Ace 68 | Subject DeviceInNeuroCo, Inc.074 Zenith Flex | |
| ComponentsSupplied | Zenith Flex Catheter,Peel Away Introducer,Hemostasis Valve, ScoutIntroducer | PenumbraReperfusionCatheter, AspirationPump,Pump/Canister | Same as 071Zenith Flex |
| Catheter ShaftMaterial | Polyether Block Amide(PEBAX),Polycarbonate/Urethane,Nylon | Pebax, Nylon,Urethane | Same as 071Zenith Flex |
| Inner Liner | PTFE | PTFE | Same |
| Hub Material | Polycarbonate | Grilamid | Same as 071Zenith Flex |
| Strain Relief | Polyolefin | Polyolefin, PETYellow, Grilamid | Same as 071Zenith Flex |
| Catheter ShaftReinforcement | Stainless Steel/Nitinol | Stainless Steel/Nitinol | Same |
| Reinforcementpattern | Coil | Coil | Same |
| LubriciousCoating | Hydrophilic Coating | Hydrophilic Coating | Same |
| RadiopaqueMarker Band | Platinum/ Iridium | Platinum/ Iridium | Same |
| CatheterPackaging | Tyvek/Nylon Pouch,polyethylene supporttube, packaging card,SBS carton | Tyvek/Nylon Pouch,polyethylene supporttube, packaging card,SBS carton | Same |
| Sterilization | Ethylene Oxide | Ethylene Oxide | Same |
| Pyrogenicity | Nonpyrogenic | Nonpyrogenic | Same |
| Working Lengths | 115, 125, 132 cm | 115, 120, 125, 127,132 cm | Same as 071Zenith Flex |
| Inside Diameter(ID) | 0.071 in | 0.068 in min. | 0.074 in |
| Predicate DeviceInNeuroCo, Inc.071 Zenith Flex | Reference DevicePenumbra Ace 68 | Subject DeviceInNeuroCo, Inc.074 Zenith Flex | |
| Proximal OuterDiameter | 0.085 in max | 0.084 in max | 0.089 in max |
| Distal OuterDiameter | 0.082 in nominal | 0.084 in max | 0.083 in nominal |
| Shelf Life | 3 years | 3 years | Same |
| Introducer | Peel away to aid incatheter tip introductioninto hemostasis valveScout to help with thenavigation | Included as part ofthe Separator with atorque device | Same as 071Zenith Flex |
| Hemostasis Valve | Polycarbonate, SiliconeO-Ring, Side Port | Polycarbonate,Silicone O-Ring, SidePort | Same |
| Luer Tapered Hub | Yes | Yes | Same |
| CompatibleGuidewire | 0.038 inches | 0.038 inches | Same |
| Scout Introducer | Yes | No | Same as 071Zenith Flex |
| Aspiration Method | Pump | Pump | Same |
| AspirationPressure | 22-28 in Hg | 20-29 in Hg | Same as 071Zenith Flex |
Comparison to Predicate Device
{6}------------------------------------------------
{7}------------------------------------------------
Summary of Non-Clinical Data
The subject device has the same materials, packaging, manufacturing process, and sterilization process as the predicate device. Therefore, testing was only conducted for specifications that were affected by the catheter size change.
Animal Testing
Clot was used to create occlusions within several arteries of swine. One side of each swine was treated with the subject Zenith Flex system and the contralateral side was treated with the reference device. The study included two follow up evaluations: 3 days and 30 days.
{8}------------------------------------------------
There were 3 endpoints of the study: angiographic assessment of revascularization to establish effectiveness, angiographic assessment and histopathological assessment to demonstrate safety. As testing included the reference device, results were compared to demonstrate substantial equivalence.
Zenith Flex Performance Testing
To demonstrate substantial equivalence between the subject 074 Zenith Flex Catheter and the predicate 071 Zenith Flex Catheter and the reference Penumbra ACE 68 Catheter, performance testing was conducted. The testing covered verification and validation testing, including biocompatibility, bench testing, sterilization and shelf life. Additionally, some tests were leveraged from previous testing conducted by InNeuroCo. The tests were performed using standard test methods and pre-determined acceptance criteria and all samples passed. Therefore, this test data supports the argument that the 074 Zenith Flex Catheter has similar performance characteristics as the predicate device and the reference device. All of the testing conducted to demonstrate substantial equivalence are presented in the following table.
| Test | Test MethodSummary | Acceptance Criteria | Conclusions |
|---|---|---|---|
| Zenith Flex Catheter& its accessoriesBiocompatibility-Material MediatedPyrogen | Testing completed perISO 10993-11 | The test article extracts mustnot cause a febrile reactiongreater than 0.5°C in anyindividual subject. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subject ZenithFlex Catheter as the materialsand manufacturing processesare equivalent. Test articles metthe acceptance criteria. |
| Zenith Flex Catheter& its accessoriesBiocompatibility-Cytotoxicity MEMElution | Testing completed perISO 10993-5 | The cultures treated with thetest article must not have areactivity grade greater than2. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subject ZenithFlex Catheter as the materialsand manufacturing processesare equivalent. Test articles metthe acceptance criteria. |
| Zenith Flex Catheter & itsaccessoriesBiocompatibility- HemolysisASTM Method, extracthuman blood | Testing completed perISO 10993-4 | The hemolytic indexabove the negative controlarticle must be less than5%. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as thematerials and manufacturingprocesses are equivalent. Testarticles met the acceptance |
| Test | Test MethodSummary | Acceptance Criteria | Conclusions |
| Zenith Flex Catheter &its accessoriesBiocompatibility-Hemolysis, ASTM | Testing completed perISO 10993-4 | The hemolytic indexabove the negative controlarticle must be less than5%. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as thematerials and manufacturingprocesses are equivalent. Testarticles met the acceptancecriteria. |
| Zenith Flex Catheter &its accessoriesBiocompatibility-Unactivated PartialThromboplastin Time | Testing completed perISO 10993-4 | There must be nostatistical decreasebetween the UPTT ofplasma exposed to thetest article and to thenegative or untreatedcontrol. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as thematerials and manufacturingprocesses are equivalent. Testarticles met the acceptancecriteria. |
| Zenith Flex Catheter &its accessoriesBiocompatibility-Complement Activation | Testing completed perISO 10993-4 | There must be nostatistical increasebetween either the C3a orSC5b-9 concentrations inplasma exposed to thetest article as compared tothe negative anduntreated controls | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as thematerials and manufacturingprocesses are equivalent. Testarticles met the acceptancecriteria. |
| Zenith Flex Catheter &its accessoriesBiocompatibility-DogThromboresistance | Testing completed perISO 10993-4 | The test articles mustreceive a thrombusformation score less thanor equal to that of thecontrol. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as thematerials and manufacturingprocesses are equivalent. Testarticles met the acceptancecriteria. |
| Zenith Flex Catheter &its accessoriesBiocompatibility-MaximizationSensitization | Testing completed perISO 10993-10 | The test article must elicita positive response in lessthan 10% of the testanimals. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as thematerials and manufacturingprocesses are equivalent. Testarticles met the acceptancecriteria. |
| Test | Test MethodSummary | Acceptance Criteria | Conclusions |
| Zenith Flex Catheter &its accessoriesBiocompatibility-IntracutaneousToxicity/Reactivity | Testing completed perISO 10993-10 | The test article extractsmust not induce asignificantly greaterbiological reaction thanthe control. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as thematerials and manufacturingprocesses are equivalent. Testarticles met the acceptancecriteria. |
| Zenith Flex Catheter &its accessoriesBiocompatibility-AcuteSystemic Toxicity Test | Testing completed perISO 10993-11 | The test article extractsmust not induce asignificantly greaterbiological reaction thanthe control. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as thematerials and manufacturingprocesses are equivalent. Testarticles met the acceptancecriteria. |
| Chemical Compatibility(Zenith Flex Catheter) | Catheter exposed tochemicals readilyavailable in a clinicalsetting. | Product shall withstandexposure to chemicalswithout degradation. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as thematerials and manufacturingprocesses are equivalent. Testarticles met the acceptancecriteria. |
| Working Length(Zenith Flex Catheter) | Testing completed perISO 10555-1 | Test samples should bewithin existing workinglength specification. | Zenith Flex Catheter testsamples met the acceptancecriteria for Working Length todemonstrate that the ZenithFlex Catheter is substantiallyequivalent to the predicatedevice. |
| Hub compatibility(Zenith Flex Catheter) | Testing completed perISO 594-1 and ISO594-2 | Hub shall meet existingLuer specifications. | Test results for the ZenithCatheter (K171672) wereleveraged for the subjectZenith Flex Catheter as thematerials and manufacturingprocesses are equivalent. Testarticles met the acceptancecriteria. |
| Test | Test MethodSummary | Acceptance Criteria | Conclusions |
| Outside diameter(Zenith FlexCatheter) | Testing completed perISO 10555-1 | Test samples should bewithin existing outsidediameter specification. | Zenith Flex Catheter testsamples met the acceptancecriteria for Outside diameter todemonstrate that the ZenithFlex Catheter is substantiallyequivalent to the predicatedevice. |
| Leak - Air(Zenith Flex Catheter) | Testing completed perISO 10555-1 | Test samples should bewithin existing Air - Leakspecifications. | Zenith Flex Catheter testsamples met the acceptancecriteria for Leak - Air todemonstrate that the ZenithFlex Catheteris substantiallyequivalent to the predicatedevice. |
| Leak - Liquid(Zenith Flex Catheter) | Testing completed perISO 10555-1 | Test samples should bewithin existing Leak -Liquid specifications. | Zenith Flex Catheter testsamples met the acceptancecriteria for Leak – Liquid todemonstrate that the ZenithFlex Catheter is substantiallyequivalent to the predicatedevice. |
| Particulates(Zenith Flex Catheter) | Samples conditioned inbenchtop anatomicalmodel.Testing completed perUSP 788 | Test samples should bewithin existing Particulatespecifications. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as thematerials and manufacturingprocesses are equivalent. Testarticles met the acceptance |
| Simulated Use - Bench(Zenith Flex Catheter) | Zenith Flex underwentsimulated use testing bya physician in abenchtop model | Test samples must meetpredetermined user needs | Zenith Flex Catheter testsamples met the acceptancecriteria for Simulated Use- Bench to demonstrate thatthe Zenith Flex Catheter issubstantially equivalent to thepredicate device. |
| Catheter Burst(Zenith Flex Catheter) | Testing completed perISO 10555-1 | Test sample burstpressures must meet orexceed existing minimumburst pressurespecification. | Zenith Flex Catheter testsamples met the acceptancecriteria for Catheter Burst todemonstrate that the ZenithFlex Catheter is substantiallyequivalent to the predicatedevice. |
| Test | Test MethodSummary | Acceptance Criteria | Conclusions |
| Tensile(Zenith Flex Catheter) | Testing completed perISO 10555-1 | Test sample ultimatetensile strength must meetor exceed existing tensilestrength specifications. | Zenith Flex Catheter testsamples met the acceptancecriteria for Tensile todemonstrate that the ZenithFlex Catheter is substantiallyequivalent to the predicatedevice. |
| Corrosion(Zenith Flex Catheter) | Testing completed perISO 10555-1 | Test samples shall exhibitno evidence of corrosion. | Test results for theIntermediate Catheter(K152202) were leveraged forthe subject Zenith FlexCatheter as the materials andmanufacturing processes areequivalent. Test articles metthe acceptance criteria. |
| PTFE Liner inspection(Zenith Flex Catheter) | Zenith Flex waschallenged todemonstrate lineradherence. | Test sample lineradhesion must meet orexceed existing PTFELiner inspectionspecifications. | Zenith Flex Catheter testsamples met the acceptancecriteria for PTFE Linerinspection to demonstrate thatthe Zenith Flex Catheter issubstantially equivalent to thepredicate device. |
| Hydrophilic CoatingIntegrity(Zenith Flex Catheter) | Conditioned sampleswere repeatedly exposedto friction and introducedinto an anatomical modelto demonstrate that thehydrophilic coating is notaffected. | Test sample results mustmeet or exceed existingHydrophilic CoatingIntegrity specifications. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as thematerials and manufacturingprocesses are equivalent.Test articles met theacceptance criteria. |
| Simulated Use Testing(Zenith Flex System) | Using a simulated model,the samples retrievedthrombi with anaspiration pump | Successfully removed thethrombi | Zenith Flex System testsamples met the acceptancecriteria for Aspiration todemonstrate that the ZenithFlex System is substantiallyequivalent to the predicatedevice |
| Test | Test MethodSummary | Acceptance Criteria | Conclusions |
| Labeling Legibility | Label is legible afterprinting. | Test samples shalldemonstrate text legibility. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as thematerials and manufacturingprocesses are equivalent.Test articles met theacceptance criteria. |
| Barcode | Barcode is readable witha standard barcodereader. | Test samples shalldemonstrate readilyreadable barcodes | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex as the materialsand manufacturing processesare equivalent. Test articlesmet the acceptance criteria. |
| Sterilization(Zenith Flex Catheter &Aspiration Tubing Set) | ISO 11135 andAAMI TIR 28 | Sterilization load shallpose an equal or lesserchallenge to sterilize thanexisting sterile productloads. | Sterilization loads met theacceptance criteria forsterilization to demonstratethat the Zenith Flex and theAspiration Tubing Set issubstantially equivalent to thepredicate device. |
| Shelf Life(Zenith Flex Catheter &Aspiration Tubing Set) | ASTM F1980 | Aged test samples mustmeet or exceed existingspecifications | Zenith Flex Catheter &Aspiration Tubing Set testsamples met the acceptancecriteria for shelf life todemonstrate that the ZenithFlex Catheter & AspirationTubing Set are substantiallyequivalent to the predicatedevice. |
| Toque Strength(Zenith Flex Catheter) | Samples were placed inanatomical model andtorqued until failure | Test sample results mustmeet or exceed existingtorque specifications. | Zenith Flex Catheter testsamples met the acceptancecriteria for torque todemonstrate that the subjectZenith Flex Catheter issubstantially equivalent to thepredicate device |
| Test | Test MethodSummary | Acceptance Criteria | Conclusions |
| Kink Resistance(Zenith Flex Catheter) | Reduction in flow wasevaluated while sampleswere exposed toappropriate tortuosity | Test sample results mustmeet or exceed existingkink resistancespecifications. | Zenith Flex Catheter testsamples met the acceptancecriteria for kink resistance todemonstrate that the subjectZenith Flex Catheter issubstantially equivalent to thepredicate device |
| Lumen Patency(Zenith Flex Catheter) | Samples were placed inbenchtop anatomicalmodel and evaluated forlumen collapse duringaspiration | Test sample results mustmeet or exceed existinglumen patencyspecifications. | Zenith Flex Catheter testsamples met the acceptancecriteria for lumen patency todemonstrate that the subjectZenith Flex Catheter issubstantially equivalent to thepredicate device |
| Tip Flexibility(Zenith Flex Catheter) | Samples were deflected,and associated forceswere measured | Zenith Flex tip flexibilityresults were compared topredicate device results. | Zenith Flex Catheter testsamples performedcomparatively against thepredicate devicedemonstrating that the subjectZenith Flex Catheter issubstantially equivalent to thepredicate device |
| Friction Force(Zenith Flex Catheter) | Samples were trackedthrough benchtopanatomical model andfrictional forces weremeasured | Zenith Flex friction forceresults were compared topredicate device results. | Zenith Flex Catheter testsamples performedcomparatively against thepredicate devicedemonstrating that the subjectZenith Flex Catheter issubstantially equivalent to thepredicate device |
| Simulated Use Testing- Usability(Zenith Flex System) | Zenith Flex underwentsimulated use testing by aphysician in a benchtopmodel | Test samples must meetpredetermined user needs | Zenith Flex System testsamples met the acceptancecriteria for Simulated Use- Usability to demonstrate thatthe Zenith Flex System issubstantially equivalent to thepredicate device. |
| Test | Test MethodSummary | Acceptance Criteria | Conclusions |
| Visual Inspection (AspirationTubing Set) | Finished Devices wereinspected for damagevisually | Test samples should meetvisual inspectionspecifications. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as theaspiration tubing is the same.Test articles met theacceptance criteria. |
| Tensile(Aspiration Tubing Set) | Finished devices wereelongated until failure | Test sample ultimatetensile strength must meetor exceed existing tensilestrength specifications. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as theaspiration tubing is the same.Test articles met theacceptance criteria. |
| Leak - Liquid(Aspiration TubingSet) | Tubing pressurized withfluid and inspected for leak | Test samples should bewithin existing Leak -Liquid specifications. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as theaspiration tubing is the same.Test articles met theacceptance criteria. |
| Leak - Air, Tubing andControl Switch(Aspiration Tubing Set) | Tubing samples evaluatedfor air leak duringaspiration | Test samples should bewithin existing Air - Leakspecifications. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as theaspiration tubing is the same.Test articles met theacceptance criteria. |
| Luer Compatibility(Aspiration Tubing Set) | Testing completed per ISO80369-7 | Hub shall meet existingLuer specifications. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as theaspiration tubing is the same.Test articles met theacceptance criteria. |
| Test | Test MethodSummary | Acceptance Criteria | Conclusions |
| Suction ConnectorSeparation Force (AspirationTubing Set) | Force to separate suctionconnector from cannisterwas evaluated using forcegauge | Test sample results mustmeet or exceed existingforce specifications. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as theaspiration tubing is the same.Test articles met theacceptance criteria. |
| Vacuum Drop / SuctionConnector SecureAttachmentAspiration Tubing Set) | Vacuum pressuremeasured at source andtip to evaluate pressuredifference | Test sample results mustmeet or exceed existingpressure specifications. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as theaspiration tubing is the same.Test articles met theacceptance criteria. |
| Lumen Patency(Aspiration Tubing Set) | Samples were evaluatedfor lumen collapse duringaspiration | Test sample results mustmeet or exceed existinglumen patencyspecifications. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as theaspiration tubing is the same.Test articles met theacceptance criteria. |
| Dimensions(Aspiration Tubing Set) | Critical dimensions weremeasured. | Test samples should bewithin existing dimensionalspecifications. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as theaspiration tubing is the same.Test articles met theacceptance criteria. |
| Packaging - Dye Leak(Zenith Flex Catheter andAspiration Tubing Set) | Testing completed perASTM F1929-12 | Test sample shall notexhibit any visual leaks orchannels | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as thepackaging is the same. Testarticles met the acceptancecriteria. |
| Test | Test MethodSummary | Acceptance Criteria | Conclusions |
| Packaging - Peel(Zenith Flex Catheter andAspiration Tubing Set) | Testing completed perASTM F88-09 | Test sample tensilestrength must meet orexceed existing tensilestrength specifications. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as thepackaging is the same. Testarticles met the acceptancecriteria. |
| Aspiration Tubing SetBiocompatibility-Cytotoxicity MEM Elution | Testing completed per ISO10993-5 | The cultures treated withthe test article must nothave a reactivity gradegreater than 2. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as theaspiration tubing is the same.Test articles met theacceptance criteria. |
| Aspiration Tubing SetBiocompatibility-Maximization Sensitization | Testing completed per ISO10993-10 | The test article must elicita positive response in lessthan 10% of the testanimals. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as theaspiration tubing is the same.Test articles met theacceptance criteria. |
| Aspiration Tubing SetBiocompatibility-IntracutaneousToxicity/Reactivity | Testing completed per ISO10993-10 | The test article extractsmust not induce asignificantly greaterbiological reaction thanthe control. | Test results for the Zenith FlexCatheter (K172167) wereleveraged for the subjectZenith Flex Catheter as theaspiration tubing is the same.Test articles met theacceptance criteria. |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
{14}------------------------------------------------
{15}------------------------------------------------
{16}------------------------------------------------
{17}------------------------------------------------
Performance Data - Clinical
No clinical study was conducted as bench and animal testing was determined sufficient for verification and validation purposes. A review was conducted considering published clinical study articles that featured the Reference Device and other devices with similar dimensions used for direct aspiration. The literature review was used to support the proposed indications for use under the NRY product code by leveraging clinical outcomes from devices that are considered technologically similar.
Summary of Substantial Equivalence
The performance characteristics and the test results demonstrate that the 074 Zenith Flex Catheter meets the acceptance criteria to determine that the 074 Zenith Flex Catheter is substantially equivalent to the predicate device. Furthermore, the intended use, the operating principles, the design, materials and the results of biocompatibility testing, shelf-life, packaging and sterilization processes are all equivalent and support the all devices are
{18}------------------------------------------------
technologically similar. Lastly, the review of published clinical data also supports the use of the 074 Zenith Flex Catheter under the NRY product code based on the similar technological characteristics to devices currently available.
§ 870.1250 Percutaneous catheter.
(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).