Search Results
Found 2 results
510(k) Data Aggregation
(15 days)
ARSTASIS ACCESS SYSTEM
The Arstasis®16 Access System is intended to provide access for the percutaneous introduction of devices into the peripheral vasculature and to promote hemostasis at the arteriotomy site as an adjunct to manual compression. The Arstasis 000 Access System is indicated for use in patients undergoing diagnostic femoral artery catheterization procedures using 5F or 6F introducer sheaths.
Arstasisone is a device that is comprised of a sheath, anchor mechanism, shaft and handle with control features.
The provided text describes a medical device, the "Arstasisone Access System," and its submission for 510(k) clearance. However, the document primarily focuses on regulatory approval and substantial equivalence to a predicate device, rather than a clinical study designed to establish acceptance criteria for device performance in terms of metrics like sensitivity, specificity, or accuracy.
The information given relates to the safety and performance of the device, not an AI or algorithm's performance. Therefore, many of the requested points regarding AI device criteria, training sets, ground truth establishment, and expert involvement are not applicable to this document.
Here's an analysis based on the information provided:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state quantitative acceptance criteria in terms of performance metrics (e.g., success rate for hemostasis, access rate) with specific thresholds as would be found in a clinical trial report with defined endpoints. Instead, the "acceptance criteria" appear to be implicit in demonstrating substantial equivalence to an existing predicate device for safety and effectiveness.
Acceptance Criteria (Inferred from regulatory context) | Reported Device Performance (Summary) |
---|---|
Premarket Acceptance: Substantial Equivalence to Predicate Device (Arstasisone Access System K100615) regarding: | The cumulative data (including bench testing, animal studies, cadaver assessments, and clinical evaluations) demonstrates that the Arstasis 006 Access System is substantially equivalent to its predicate in providing access to the arterial lumen and facilitating introduction/placement of devices and achievement of hemostasis. |
- Functionality | Bench testing performed (functionality, torque loading, flexibility, tensile strength of core wire assembly and sheath). Prior bench testing on predicate included deployment forces, flexibility, tensile, compression, and torque loading. |
- Torque Loading | Covered by bench testing. |
- Flexibility | Covered by bench testing. |
- Tensile Strength | Covered by bench testing. |
- Biocompatibility | Prior testing on predicate included biocompatibility testing pursuant to ISO-10993-1. |
- Safety (short-term & long-term) | "The short term safety and clinical performance of the device were established." "The long term safety, as well as the ability to access and re-access, was retrospectively studied in a smaller cohort of patients." |
- Clinical Performance (hemostasis, access) | "The cumulative data provided herein demonstrates that the Arstasis 006 Access System is substantially equivalent to its predicate in providing access to the arterial lumen and facilitating the introduction and placement of devices into the peripheral vasculature and achievement of hemostasis." |
2. Sample Size Used for the Test Set and Data Provenance
The document refers to "Multiple clinical evaluations" and a "smaller cohort of patients" for retrospective long-term safety studies. Specific numerical sample sizes for these clinical evaluations are not provided.
- Data Provenance: The document does not specify the country of origin.
- Retrospective/Prospective: "Multiple clinical evaluations were conducted" (implies prospective or mixed). "The long term safety, as well as the ability to access and re-access, was retrospectively studied in a smaller cohort of patients."
Given that this is a medical device (catheter introducer) and not an AI algorithm, the concept of a "test set" in the context of machine learning performance metrics is not directly applicable. The "test set" here refers to the patients involved in the clinical evaluations.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts
This information is not provided in the document. For a medical device like this, "ground truth" would typically refer to clinical outcomes (e.g., successful hemostasis, complication rates), which are observed directly or documented by clinicians, rather than established by expert consensus on interpretations of data.
4. Adjudication Method for the Test Set
This information is not provided. Clinical outcomes are generally observed and recorded, rather than adjudicated in the same way imaging interpretations might be.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What was the effect size of how much human readers improve with AI vs without AI assistance
Not Applicable. This document describes a physical medical device (catheter introducer), not an AI algorithm for interpretation or assistance. Therefore, an MRMC study related to AI performance is not relevant.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not Applicable. This is a physical medical device, not a software algorithm.
7. The Type of Ground Truth Used
Based on the nature of the device and the statements:
- Clinical Outcomes: The "ground truth" for the clinical evaluations would be actual patient outcomes, such as successful hemostasis, ability to access/re-access the vasculature, and safety events (complications). The short-term and long-term safety and clinical performance were established through these evaluations.
8. The Sample Size for the Training Set
Not Applicable. This document describes a physical medical device, not an AI algorithm that requires a training set. The "training" for such a device would involve bench testing, animal studies, and cadaver assessments to refine the device design.
9. How the Ground Truth for the Training Set Was Established
Not Applicable. As per point 8, there's no "training set" in the AI sense. However, the "ground truth" for the development of the device was established through:
- Bench Testing: Evaluation against engineering specifications and performance standards (e.g., torque, tensile strength).
- Animal Studies (non-GLP): Observation of device function and biological response in animal models.
- Cadaver Assessments: Evaluation of device performance in human cadavers.
These methods served to inform and refine the device design prior to human trials.
Ask a specific question about this device
(26 days)
ARSTASIS ACCESS SYSTEM
The Arstasis 000 Access System is intended to provide access for the percutaneous introduction of devices into the peripheral vasculature and to promote hemostasis at the arteriotomy site as an adjunct to manual compression. The Arstasis006 Access System is indicated for use in patients undergoing diagnostic femoral artery catheterization procedures using 5F or 6F introducer sheaths.
Arstasisone is a device that is comprised of a sheath, anchor mechanism, shaft and handle with control features.
The provided text describes the Arstasisone Access System, a catheter introducer. The information focuses on its regulatory submission for substantial equivalence to a predicate device. There is no information in the provided text about acceptance criteria or a study proving the device meets specific performance criteria in the format requested.
Here's a breakdown of what can be extracted based on your request, and what cannot:
Missing Information:
The provided text does not contain any information regarding:
- A table of acceptance criteria and reported device performance.
- Sample sizes used for a test set, data provenance, ground truth experts, or adjudication methods for any study.
- Multi-reader multi-case (MRMC) comparative effectiveness studies or effect sizes for human readers with/without AI assistance.
- Standalone algorithm performance.
- Types of ground truth used (expert consensus, pathology, outcomes data).
- Sample size for training sets.
- How ground truth for training sets was established.
Information that can be inferred/extracted about the study (though not presented as a direct acceptance criteria study):
The submission mentions various types of testing and studies conducted to establish substantial equivalence, rather than to specifically meet predefined acceptance criteria in the typical sense of a performance study for a diagnostic AI device.
1. Table of acceptance criteria and the reported device performance:
- Not provided. There is no explicit table of acceptance criteria or quantitative performance metrics within the document.
2. Sample sized used for the test set and the data provenance:
- Test Set Sample Size: "Multiple clinical evaluations were conducted" and "retrospectively studied in a smaller cohort of patients." No specific numbers are given.
- Data Provenance: Not specified (e.g., country of origin). "Clinical investigations" and "animal studies (non GLP) and cadaver assessments" are mentioned.
- Retrospective/Prospective: Clinical investigations are generally prospective, but the long-term safety study was "retrospectively studied."
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not provided. The text does not detail how ground truth was established for the clinical evaluations.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not provided.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This device is a physical medical device (catheter introducer), not an AI/diagnostic software. Therefore, an MRMC study or AI assistance is not applicable.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No. This device is a physical medical device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not explicitly stated for clinical evaluations. The studies aimed to establish "short term safety and clinical performance" and "long term safety, as well as the ability to access and reaccess." This implies clinical outcomes and observations rather than a singular "ground truth" for a diagnostic task.
8. The sample size for the training set:
- Not applicable/Not provided. This is a physical device, not an AI algorithm requiring a training set.
9. How the ground truth for the training set was established:
- Not applicable/Not provided. This is a physical device, not an AI algorithm.
Summary of available information related to performance/testing:
The Arstasisone Access System's substantial equivalence was established through a combination of:
- Bench testing: Functionality, deployment and release forces, flexibility, torque loading, tensile, compressive, and torque strengths. (No specific acceptance criteria or results are detailed).
- Biocompatibility testing: Conducted pursuant to ISO-10993-1. (Results reported as "successfully conducted").
- Preliminary animal studies (non GLP) and cadaver assessments: Conducted using prototypes of a similar design.
- Multiple clinical evaluations: Conducted on an earlier device version of similar design. These established short-term safety and performance.
- Retrospective study: On a smaller cohort for long-term safety and reaccess ability.
The overall conclusion is that the "cumulative data provided herein demonstrate that the Arstasis Access System is substantially equivalent to its predicate in providing access to the arterial lumen and facilitating the introduction and placement of devices into the peripheral vasculature and achievement of hemostasis."
Ask a specific question about this device
Page 1 of 1