Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K053440
    Manufacturer
    Date Cleared
    2005-12-30

    (21 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AMPLATZER® Sizing Balloon II is indicated for use in those patients with cardiovascular defects wherein accurate measurement of the defect is important to select the appropriately sized occluder device.

    Device Description

    The AMPLATZER® Sizing Balloon II is a triple lumen balloon catheter with three (3) radiopaque marker bands located inside the balloon to allow for radiographic measurement. The center of the balloon contains a pair of marker bands 0.4mm apart (inside to inside), and one (1) marker band 15 mm proximal of that pair.

    AI/ML Overview

    The provided document is a 510(k) summary for the AMPLATZER® Sizing Balloon II, which is a medical device. This type of submission focuses on demonstrating substantial equivalence to a predicate device rather than proving novel efficacy through detailed acceptance criteria and a comprehensive study with specific performance metrics for a new AI device.

    Therefore, the information required to populate the tables and answer the questions related to acceptance criteria, AI performance, and AI-specific ground truth methodologies is not present in the provided text. The document describes a non-clinical evaluation of a physical medical device, not a software algorithm or AI.

    However, I can extract the relevant information from the document as it pertains to the device's "performance" in the context of a 510(k) submission for a physical device:

    Device Performance and Acceptance Criteria (Non-AI device)

    Acceptance Criteria (Non-Clinical Tests)Reported Device Performance (Summary)
    Inflation timePerformed (Results not detailed, but deemed acceptable)
    Deflation timePerformed (Results not detailed, but deemed acceptable)
    Burst volumePerformed (Results not detailed, but deemed acceptable)
    Bifurcation separationPerformed (Results not detailed, but deemed acceptable)
    Tip separationPerformed (Results not detailed, but deemed acceptable)
    Sidearm stopcock to tube separationPerformed (Results not detailed, but deemed acceptable)
    Sidearm luer to tube separationPerformed (Results not detailed, but deemed acceptable)
    Marker band placementPerformed (Results not detailed, but deemed acceptable)
    Biological testingPerformed (Results not detailed, but deemed acceptable)

    Study Details (Based on the provided K053440 document, for a physical device)

    1. Sample size used for the test set and the data provenance: Not specified in the provided summary. The tests are "non-clinical" and likely involved a batch of manufactured devices, but the exact sample size for each test is not detailed.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This is not an AI device where human experts establish ground truth. The "ground truth" for these tests would be established by engineering specifications and measurements.
    3. Adjudication method for the test set: Not applicable. This is not an AI device.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the non-clinical tests, the ground truth would be based on engineering specifications, physical measurements, and established testing standards for medical devices.
    7. The sample size for the training set: Not applicable. This is not an AI device.
    8. How the ground truth for the training set was established: Not applicable. This is not an AI device.

    Conclusion from The Document:

    The document concludes that "The AMPLATZER® Sizinq Balloon II is substantially equivalent to currently marketed balloon catheters used for temporary vessel occlusion," based on design modifications, a new supplier, and an additional catheter size, without altering the intended use or fundamental scientific technology of the predicate product. The non-clinical tests listed were performed to support this claim of substantial equivalence.

    Ask a Question

    Ask a specific question about this device

    K Number
    K993248
    Manufacturer
    Date Cleared
    2000-07-12

    (288 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AMPLATZER® Sizing Balloon is intended for use in those patients with cardiovascular defects wherein accurate measurement of the defect is important to select the appropriately sized occluder device.

    Device Description

    The AMPLATZER® Sizing Balloon is a 7F double lumen balloon catheter made from radiopaque nylon. The balloon is compliant and made from a very thin stretchable plastic membrane. The balloon-carrying segment is angled to 45° to the shaft providing a more or less right-angled position in relation to the atrial septum.

    AI/ML Overview

    The acceptance criteria and study that proves the device meets the acceptance criteria are described below:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    In Vitro Functional Tests
    Inflation/Deflation Time calculation (cc)Not explicitly stated (implied suitable performance)Performed
    Leakage and Burst TestingNot explicitly stated (implied suitable performance)Performed
    Pull Test Bifurcation and Port TubingNot explicitly stated (implied suitable performance)Performed
    Pull Test Bifurcation and ShaftNot explicitly stated (implied suitable performance)Performed
    Biocompatibility Testing
    HemocompatibilityNot explicitly stated (implied suitable performance)Performed
    CytotoxicityNot explicitly stated (implied suitable performance)Performed
    Dermal SensitizationNot explicitly stated (implied suitable performance)Performed
    Intracutaneous InjectionNot explicitly stated (implied suitable performance)Performed
    Systemic ToxicityNot explicitly stated (implied suitable performance)Performed
    PyrogenicityNot explicitly stated (implied suitable performance)Performed
    Other Tests
    Animal TestingNot explicitly stated (implied suitable performance)Performed

    Note: The document states that these tests were "performed" but does not provide specific quantitative acceptance criteria or the results of these tests, beyond the implication that they were successfully met to establish substantial equivalence.

    2. Sample Size and Data Provenance

    • Test Set Sample Size: Not explicitly stated for any of the performed tests (in vitro functional, biocompatibility, or animal testing).
    • Data Provenance: The tests are described as "in vitro functional tests," "biocompatibility testing," and "animal testing," implying laboratory-based and animal-based studies. The country of origin is not specified but is presumably the location of AGA Medical Corporation (Golden Valley, MN, USA). The studies appear to be prospective for the purpose of demonstrating safety and performance.

    3. Number of Experts and Qualifications for Ground Truth Establishment (Test Set)

    Not applicable. The document describes laboratory, in-vitro, and animal testing. There is no mention of human-expert-established ground truth for a test set in the context of diagnostic performance.

    4. Adjudication Method (Test Set)

    Not applicable. There is no mention of expert adjudication for a test set as the studies involve functional, biocompatibility, and animal testing.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. The submission focuses on the safety and performance of the device itself, not on the improvement of human readers with or without AI assistance. The device is a physical sizing balloon, not an AI diagnostic tool.

    6. Standalone (Algorithm Only) Performance Study

    No, a standalone performance study in the context of an algorithm's performance was not done. The device is a physical medical device (balloon catheter), not a software algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for the tests performed would be based on:

    • In vitro functional tests: Engineering and materials science standards for catheter performance (e.g., measuring inflation/deflation, material strength, leakage limits).
    • Biocompatibility tests: Established ISO standards and guidelines for biological evaluation of medical devices (e.g., absence of cytotoxicity, sensitization, systemic toxicity).
    • Animal testing: Physiological and pathological observations in animal models to assess safety and performance in a living system.

    8. Sample Size for the Training Set

    Not applicable. The document does not describe a training set for an algorithm. The "training set" concept is relevant for machine learning or AI models, which is not the nature of this medical device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an algorithm. The device's design and manufacturing processes are likely informed by established medical device development guidelines and previous knowledge, rather than a data-driven training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1