510(k) Data Aggregation

    K Number
    K211310
    Date Cleared
    2021-06-16

    (47 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    AMADEX - Nitrile Powder Free Examination Gloves Non-Sterile, (Blue, Black, Orange) are Class I patient examination gloves bearing the product code Nitrile - LZA (2 ICFR880.6250). AMADEX - Nitrile Powder Free Examnation Gloves Non-Sterile,(Blue, Black, Orange) meet all the requirements of ASTM standard D 6319-10 and FDA 21 CFR 880.6250. The AMADEX - Nitrile Powder Free Examination Gloves Non-Sterile, (Blue, Black, Orange) is single-use device to prevent contamination between patient and examiner.

    AI/ML Overview

    This document describes the acceptance criteria and study results for the AMADEX Nitrile Powder Free Examination Gloves Non-Sterile (Blue, Black, Orange) to demonstrate substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicsStandardPurposeAcceptance CriteriaReported Device PerformanceRemarks
    DimensionsASTM D 6319-10
    LengthTo determine the length of the glovesMin 230 mm (All Sizes)>230 mm (Passes)Same
    Width (XS)70 ± 10 mm76 ± 3 mmSame
    Width (S)80 ± 10 mm84 ± 3 mmSame
    Width (M)95 ± 10 mm94 ± 3 mm (Also reported as 98 ± 2 mm, which passes 95 ± 10 mm)Same
    Width (L)110 ± 10 mm105 ± 3 mmSame
    Width (XL)120 ± 10 mm113 ± 3 mmSame
    Physical PropertiesASTM D 6319-10To determine the physical properties (Tensile strength and Elongation)
    Tensile Strength (Before Aging)14 Mpa min24.5 Mpa (Passes) (Also reported as 14 Mpa in comparison table)Same
    Ultimate Elongation (Before Aging)500% min500% (Implied, as result is 14 Mpa and 500% in comparison table)Same
    Tensile Strength (After Aging)14 Mpa min14 Mpa (Implied)Same
    Ultimate Elongation (After Aging)400% min400% (Implied)Same
    ThicknessASTM D 6319-10To determine the thickness of the glovesPalm 0.05 mm min, Finger 0.05 mm minPalm: 0.09 mm (Passes), Finger: 0.12 mm (Passes)Same
    Powder FreeASTM D 6319-10, D 6124-06To determine the amount of residual powder and non-powder solids found on gloves