Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    Why did this record match?
    Device Name :

    ACE ALBUMIN REAGENT, ACE TOTAL PROTEIN REAGENT, ACE CALCIUM-ARSENAZO REAGENT AND ACE PHOSPHORUS REAGENT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ACE Albumin Reagent is intended for the quantitative determination of albumin concentration in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    ACE Total Protein Reagent is intended for the quantitative determination of total protein concentration in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Total protein measurements are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone marrow as well as other metabolic or nutritional disorders. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    ACE Calcium-Arsenazo Reagent is intended for the quantitative determination of calcium concentration in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    ACE Inorganic Phosphorus U.V. Reagent is intended for the quantitative determination of inorganic phosphorus concentration in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Measurements of inorganic phosphorus are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases and vitamin D imbalance. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    Device Description

    In the ACE Albumin Reagent assay, Bromcresol green binds specifically to albumin to form a green colored complex, which is measured bichromatically at 629 nm/692 nm. The intensity of color produced is directly proportional to the albumin concentration in the sample.

    In the ACE Total Protein Reagent assay, cupric ions react with the peptide bonds of proteins under alkaline conditions to form a violet colored complex, which is measured bichromatically at 544 nm/692 nm. The intensity of color produced is directly proportional to the total protein concentration in the sample.

    In the ACE Calcium-Arsenazo Reagent assay, calcium reacts with Arsenazo III in an acidic solution to form a blue-purple colored complex, which is measured bichromatically at 647 nm/692 nm. The intensity of color produced is directly proportional to the calcium concentration in the sample.

    In the ACE Inorganic Phosphorus U.V. Reagent assay, under acidic conditions, inorganic phosphorus in serum reacts with ammonium molybdate to form an unreduced phosphomolybdate complex, which absorbs strongly at 340 nm. The increase in absorbance, measured bichromatically at 340 nm/378 nm, is directly proportional to the amount of phosphorus in the sample.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the ACE Albumin Reagent, ACE Total Protein Reagent, ACE Calcium-Arsenazo Reagent, and ACE Inorganic Phosphorus U.V. Reagent, based on the provided text.


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided documentation does not explicitly state formal "acceptance criteria" with specific thresholds for each performance metric. However, it presents detailed performance data, particularly precision (within-run and total %CV) and method comparison (regression analysis, correlation coefficient), comparing the new reagents on various ACE clinical chemistry systems (ACE, ACE Alera, ACE Axcel) against existing predicate devices and among themselves. The implied acceptance is that the new reagents perform comparably to, or as effectively as, the predicate devices and demonstrate acceptable precision and linearity for clinical use.

    Below is a summary of the reported device performance based on the "In-House Precision" and "In-House Matrix Comparison" tables. Since explicit acceptance criteria are not given, the performance data itself is presented as the evidence of meeting implied clinical utility and equivalence to predicate devices.

    ACE Albumin Reagent

    MetricAcceptance Criteria (Implied)Reported Performance (Range across ACE, Alera, Axcel systems)
    Precision (%CV)Clinically acceptableSerum: Within-Run: 0.5-1.6%, Total: 0.6-1.8%
    Plasma: Within-Run: 0.8-1.7%, Total: 1.1-1.7%
    Matrix Comparison (Serum vs. Plasma)Slope close to 1, Intercept close to 0, High CorrelationSlope: 0.956 - 1.002
    Intercept: -0.01 - 0.20
    Correlation: 0.9850 - 0.9905
    LinearityBroad clinical range, r^2 close to 1Linear to 7.6 g/dL
    y = 0.980x + 0.01, r^2 = 0.9982
    Detection Limits (ACE Alera)Low enough for clinical utilityLoB: 0.08 g/dL, LoD: 0.09 g/dL, LoQ: 0.09 g/dL
    Interferences (ACE Alera)No significant interference at clinically relevant levelsIcterus: 60 mg/dL, Hemolysis: 250 mg/dL, Lipemia: 1000 mg/dL, Ascorbic Acid: 6 mg/dL

    ACE Total Protein Reagent

    MetricAcceptance Criteria (Implied)Reported Performance (Range across ACE, Alera, Axcel systems)
    Precision (%CV)Clinically acceptableSerum: Within-Run: 0.7-1.3%, Total: 0.8-1.4%
    Plasma: Within-Run: 0.5-1.3%, Total: 0.7-1.4%
    Matrix Comparison (Serum vs. Plasma)Slope close to 1, Intercept close to 0, High CorrelationSlope: 0.994 - 1.001
    Intercept: 0.12 - 0.34
    Correlation: 0.9798 - 0.9885
    LinearityBroad clinical range, r^2 close to 1Linear to 15.1 g/dL
    y=0.991x + 0.04, r^2 = 0.9979
    Detection Limits (ACE Alera)Low enough for clinical utilityLoB: 0.08 g/dL, LoD: 0.13 g/dL, LoQ: 0.20 g/dL
    Interferences (ACE Alera)No significant interference at clinically relevant levelsIcterus: 56.8 mg/dL, Hemolysis: 250 mg/dL, Lipemia: 929 mg/dL, Ascorbic Acid: 6 mg/dL

    ACE Calcium-Arsenazo Reagent

    MetricAcceptance Criteria (Implied)Reported Performance (Range across ACE, Alera, Axcel systems)
    Precision (%CV)Clinically acceptableSerum: Within-Run: 0.7-1.6%, Total: 0.9-2.7%
    Plasma: Within-Run: 0.5-1.9%, Total: 1.1-2.0%
    Matrix Comparison (Serum vs. Plasma)Slope close to 1, Intercept close to 0, High CorrelationSlope: 0.978 - 1.008
    Intercept: -0.06 - 0.33
    Correlation: 0.9793 - 0.9911
    LinearityBroad clinical range, r^2 close to 1Linear to 16.5 mg/dL
    y=0.992x +0.27, r^2 = 0.9990
    Detection Limits (ACE Alera)Low enough for clinical utilityLoB: 0.09 mg/dL, LoD: 0.11 mg/dL, LoQ: 0.23 mg/dL
    Interferences (ACE Alera)No significant interference at clinically relevant levelsIcterus: 58.8 mg/dL, Hemolysis: 1000 mg/dL, Lipemia: 1000 mg/dL, Ascorbic Acid: 6 mg/dL

    ACE Inorganic Phosphorus U.V. Reagent

    MetricAcceptance Criteria (Implied)Reported Performance (Range across ACE, Alera, Axcel systems)
    Precision (%CV)Clinically acceptableSerum: Within-Run: 0.3-4.4%, Total: 0.5-5.0%
    Plasma: Within-Run: 0.9-5.1%, Total: 0.9-6.1%
    Matrix Comparison (Serum vs. Plasma)Slope close to 1, Intercept close to 0, High CorrelationSlope: 0.999 - 1.049
    Intercept: -0.28 - 0.04
    Correlation: 0.9927 - 0.9950
    LinearityBroad clinical range, r^2 close to 1Linear to 21 mg/dL
    y=1.001x +0.03, r^2 = 0.9995
    Detection Limits (ACE Alera)Low enough for clinical utilityLoB: 0.25 mg/dL, LoD: 0.35 mg/dL, LoQ: 0.35 mg/dL
    Interferences (ACE Alera)No significant interference at clinically relevant levelsIcterus: 11.5 mg/dL, Hemolysis: 250 mg/dL, Lipemia: 306 mg/dL, Ascorbic Acid: 6 mg/dL

    2. Sample Sizes Used for the Test Set and Data Provenance

    The studies mentioned are "In-House Precision," "In-House Matrix Comparison: Serum vs. Plasma," "POL - Precision," and "POL – Method Comparison."

    • In-House Precision (Serum vs. Plasma):
      • Sample Size: Not explicitly stated for each "low, mid, high" concentration level, but implies multiple replicates for each level tested across the three systems (ACE, Alera, Axcel). For example, the ACE Alera precision table (pg. 16) shows 3 levels (low, mid, high) for serum, with reported mean, within-run SD, and total SD. Typically, precision studies involve running samples multiple times a day over several days.
      • Data Provenance: "In-House" suggests it was conducted by Alfa Wassermann Diagnostic Technologies, LLC, likely at their own facilities. It is a prospective study as they are performing experiments to generate data.
    • In-House Matrix Comparison: Serum vs. Plasma:
      • Sample Size:
        • Albumin: ACE: 55 pairs, ACE Alera: 56 pairs, ACE Axcel: 56 pairs
        • Total Protein: ACE: 56 pairs, ACE Alera: 56 pairs, ACE Axcel: 81 pairs
        • Calcium-Arsenazo: ACE: 56 pairs, ACE Alera: 56 pairs, ACE Axcel: 81 pairs
        • Inorganic Phosphorus: ACE: 100 pairs, ACE Alera: 102 pairs, ACE Axcel: 56 pairs
      • Data Provenance: "In-House" suggests it was conducted by Alfa Wassermann Diagnostic Technologies, LLC, likely at their own facilities. The comparison between serum and plasma samples implies these were collected from human subjects. This is a prospective study.
    • POL (Physician Office Laboratory) - Precision:
      • Sample Size: For each reagent and each system (ACE and ACE Alera), there are 3 "samples" (representing different concentration levels) tested at 3 different POL sites. Each sample/site combination has "Within-Run" and "Total" precision reported, implying multiple replicates for each measurement.
      • Data Provenance: Conducted at "POL 1," "POL 2," and "POL 3" sites, indicating external collection and testing beyond the manufacturer's immediate facilities. This is a prospective study.
    • POL (Physician Office Laboratory) - Method Comparison:
      • Sample Size:
        • Albumin: 50 samples for each POL site (x3 POLs)
        • Total Protein: 51 samples for each POL site (x3 POLs)
        • Calcium-Arsenazo: 50 samples for each POL site (x3 POLs)
        • Inorganic Phosphorus: 50 samples for POL 1 & 3, 48 samples for POL 2
      • Data Provenance: Comparisons between "ACE In-House (x)" and "ACE POL (y)" or "ACE In-House (x)" and "ACE Alera POL (y)". This indicates the data for these studies was collected at both in-house facilities and external Physician Office Laboratories. This is a prospective study design, comparing results from different testing environments.
    • Detection Limits & Linearity (ACE Alera):
      • Sample Size: Not specified for these specific studies, but typically involves a series of diluted and concentrated samples to define the measuring range.
      • Data Provenance: In-House, prospective.
    • Interference (ACE Alera):
      • Sample Size: Not specified, but involves spiking samples with various interferents at different concentrations.
      • Data Provenance: In-House, prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    For these types of in vitro diagnostic (IVD) assays, the "ground truth" is typically established by reference methods or validated comparative methods, often using certified calibrators and controls. The documentation does not mention the use of human experts to establish ground truth for the test set in the traditional sense of medical image interpretation (e.g., radiologists interpreting images). Instead, the studies rely on quantitative measurements and statistical comparisons with established methods (the predicate devices or in-house reference measurements) to demonstrate performance. Therefore, no information is provided on the number or qualifications of experts for ground truth establishment.


    4. Adjudication Method for the Test Set

    Not applicable. As described in point 3, the "ground truth" for these quantitative chemical assays is not established through expert consensus or adjudication in the way it would be for qualitative or interpretive diagnostic devices like medical imaging. Performance is evaluated by statistical comparison of numerical results.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This device consists of chemical reagents for laboratory measurement, not an AI-assisted diagnostic tool interpreted by human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI is not relevant to this submission.


    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

    The performance presented for these reagents is inherently "standalone" in the sense that it reflects the direct analytical performance of the assays on the specified automated clinical chemistry systems. The results are quantitative measurements produced by the device without human interpretation of raw data beyond reading the numerical output. The "without human-in-the-loop" aspect applies here as the device itself performs the measurement and outputs a numerical value of concentration. The method comparison studies demonstrate the standalone performance of the candidate devices compared to predicate devices.


    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for these assays is established through reference methods and comparison to legally marketed predicate devices.

    • For precision, the "ground truth" for each replicate is assumed to be the true concentration within the sample, and the study assesses the reproducibility of the device in measuring that concentration.
    • For method comparison studies (e.g., In-House vs. POL, or ACE vs. ACE Alera), one method's results (often the predicate or an established in-house method) serve as the comparative 'truth' to evaluate the new method's agreement. The reference method would itself be calibrated against known standards.
    • For linearity, samples of known, graded concentrations are used.
    • For detection limits, the ground truth involves samples with very low, known concentrations.

    These are established analytical chemistry principles rather than "expert consensus" or "pathology" in the diagnostic interpretation sense.


    8. The Sample Size for the Training Set

    The concept of a "training set" is primarily relevant for machine learning or AI algorithms which are iteratively developed and optimized using data. These reagents are chemical assays with a defined photometric measurement principle. While there is a development phase that involves optimizing reagent formulations and instrument parameters, there isn't a "training set" in the computational sense. The data presented here are from formal "verification and validation studies" to demonstrate performance characteristics (precision, linearity, accuracy/comparison, interference, detection limits).


    9. How the Ground Truth for the Training Set Was Established

    As noted in point 8, the concept of a "training set" is not directly applicable to these chemical reagents. The "ground truth" for establishing and validating the performance of such assays is based on:

    • Reference materials/calibrators: Solutions with precisely known concentrations of the analyte (albumin, total protein, calcium, phosphorus) traceable to international standards.
    • Validated comparison methods: Measurements made by existing, legally marketed predicate devices or other well-established and accurate laboratory methods.
    • Controlled spiking experiments: Adding known amounts of substance to samples to assess recovery, linearity, and interference.

    These methods establish the quantitative "truth" against which the performance of the new reagents is measured.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    ACE ALBUMIN REAGENT ACE TOTAL PROTEIN REAGENT ACE CALCIUM-ARSENAZO REAGENT ACE PHOSPHORUS U.V.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE Albumin Reagent is intended for the quantitative determination of albumin concentration in serum using the ACE Axcel Clinical Chemistry System. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Total Protein Reagent is intended for the quantitative determination of total protein concentration in serum using the ACE Axcel Clinical Chemistry System. Total protein measurements are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone marrow as well as other metabolic or nutritional disorders. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Calcium-Arsenazo Reagent is intended for the quantitative determination of calcium concentration in serum using the ACE Axcel Clinical Chemistry System. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Inorganic Phosphorus U.V. Reagent is intended for the quantitative determination of inorganic phosphorus concentration in serum using the ACE Axcel Clinical Chemistry System. Measurements of inorganic phosphorus are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases and vitamin D imbalance. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    Device Description

    In the ACE Albumin Reagent assay, Bromcresol green binds specifically to albumin to form a green colored complex, which is measured bichromatically at 629 nm/692 nm. The intensity of color produced is directly proportional to the albumin concentration in the sample.

    In the ACE Total Protein Reagent assay, cupric ions react with the peptide bonds of proteins under alkaline conditions to form a violet colored complex which is measured bichromatically at 544 nm/692 nm. The intensity of color produced is directly proportional to the total protein concentration in the sample.

    In the ACE Calcium-Arsenazo Reagent assay, calcium reacts with Arsenazo III in an acidic solution to form a blue-purple colored complex, which is measured bichromatically at 647 nm/692 nm. The intensity of color produced is directly proportional to the calcium concentration in the sample.

    In the ACE Inorganic Phosphorus U.V. Reagent assay, under acidic conditions, inorganic phosphorus in serum reacts with ammonium molybdate to form an unreduced phosphomolybdate complex, which absorbs strongly at 340 nm. The increase in absorbance, measured bichromatically at 340 nm/378 nm, is directly proportional to the amount of phosphorus in the sample.

    The ACE Albumin Reagent consists of a single reagent bottle. The reagent contains Bromcresol green and acetate buffer.

    The ACE Total Protein Reagent consists of a single reagent bottle. The reagent contains copper sulfate, sodium potassium tartrate, potassium iodide and sodium hydroxide.

    The ACE Calcium-Arsenazo Reagent consists of a single reagent bottle. The Reagent contains Arsenazo III.

    The ACE Inorganic Phosphorus U.V. Reagent consists of a single reagent bottle. The reagent contains ammonium molybdate and sulfuric acid.

    AI/ML Overview

    Acceptance Criteria and Device Performance Study for ACE Reagents

    The provided 510(k) summary (K113374) describes the performance of four reagents: ACE Albumin Reagent, ACE Total Protein Reagent, ACE Calcium-Arsenazo Reagent, and ACE Inorganic Phosphorus U.V. Reagent, when used with the Alfa Wassermann ACE Axcel Clinical Chemistry System. The study establishes the substantial equivalence of these devices to their predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated in numerical terms (e.g., "CV must be

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    ACE AXCEL CLINICAL CHEMISTRY SYSTEM,ACE ALBUMIN REAGENT,ACE TOTAL PROTEIN REAGENT,ACE BUN/UREANITROGEN

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE Axcel Clinical Chemistry System is an automated, discrete, bench-top, random access analyzer that is intended for in vitro diagnostic use in the quantitative measurement of general chemistry assays for clinical use in physician office laboratories or clinical laboratories.

    The ACE Axcel Clinical System includes an Ion Selective Electrode (ISE) module for the measurement of sodium, potassium and chloride in serum. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    • Sodium measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance
    • Potassium measurements are used to monitor electrolyte balance and in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
    • Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

    The ACE Glucose Reagent is intended for the quantitative determination of glucose concentration in serum using the ACE Axcel Clinical Chemistry System. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    Device Description

    The ACE Axcel Clinical Chemistry System consists of two major components, the chemistry instrument and an integrated Panel PC. The instrument accepts the physical patient samples, performs the appropriate optical or potentiometric measurements on those samples and communicates that data to an integral Panel PC. The Panel PC uses keyboard or touch screen input to manually enter a variety of data, control and accept data from the instrument, manage and maintain system information and generate reports relative to patient status and instrument performance. The Panel PC also allows remote download of patient requisitions and upload of patient results via a standard interface.

    In the ACE Glucose Reagent assay, glucose in serum reacts with adenosine triphosphate in the presence of hexokinase and magnesium with the formation of glucose-6-phosphate and adenosine diphosphate. Glucose-6-phosphate dehydrogenase catalyzes the oxidation of glucose-6-phosphate with NAD+ to form 6-phosphogluconate and NADH. NADH absorbs strongly at 340 nm, whereas NAD+ does not. The total amount of NADH formed is proportional to the concentration of glucose in the sample. The increase in absorbance is measured bichromatically at 340 nm/378 nm.

    The ACE Ion Selective Electrode (ISE) Module is used with ACE CAL A and CAL B Calibration Solutions in the performance of a two-point calibration in order to measure concentrations of sodium, potassium and chloride in undiluted serum. The ISE module uses a potentiometric method to simultaneously measure sodium, potassium and chloride in undiluted serum. Each electrode uses an ion-specific membrane to measure the difference in ionic concentration between an inner electrolyte solution and the sample. This difference causes an electro-chemical potential to form on the membrane of the active electrode. The connection of the amplifier and ground (reference electrode) to the ion selective electrode forms the measuring system. The two-point calibration with CAL A and CAL B with precisely known ion concentrations (two-point calibration) and the measured voltage difference of the sample and CAL A are used to determine the ion concentration in the sample.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the Alfa Wassermann ACE Axcel Clinical Chemistry System, ACE Ion Selective Electrode (ISE) Module, and ACE Glucose Reagent, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the "Accuracy" data, where the device (y) is compared to a predicate device (x). The close correlation, with slopes near 1 and intercepts near 0, along with high correlation coefficients, indicates acceptable accuracy. Precision is evaluated by the Coefficient of Variation (CV%).

    Measurement (Reagent/Module)Performance MetricAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance (ACE Axcel Clinical Chemistry System)
    ACE Glucose ReagentPrecision (Within-run CV)(Not explicitly stated, but low CV is desirable)1.0 to 1.4% (lab setting, 4 levels, 22 days); 0.3 to 2.2% (POL sites, 3 locations, 5 days)
    Precision (Total CV)(Not explicitly stated, but low CV is desirable)1.0 to 1.9% (lab setting); 0.5 to 2.2% (POL sites)
    Accuracy (Correlation Coefficient)Close to 1.00.9998 (lab setting, 122 samples); 0.9992 to 0.9998 (POL sites, 3 locations)
    Accuracy (Slope)Close to 1.01.001 to 1.009 (lab setting); 0.972 to 1.021 (POL sites)
    Accuracy (Intercept)Close to 0-1.5 to 0.1 (lab setting); -3.0 to 4.1 (POL sites)
    Detection Limit(Not explicitly stated, but low is desirable)3.1 mg/dL
    ACE Axcel Sodium ISEPrecision (Within-run CV)(Not explicitly stated, but low CV is desirable)0.4 to 1.0% (lab setting, 4 levels, 21 days); 0.6 to 1.0% (POL sites, 3 locations, 5 days)
    Precision (Total CV)(Not explicitly stated, but low CV is desirable)0.8 to 1.4% (lab setting); 0.8 to 1.4% (POL sites)
    Accuracy (Correlation Coefficient)Close to 1.00.9963 (lab setting, 113 samples); 0.9917 to 0.9995 (POL sites, 3 locations)
    Accuracy (Slope)Close to 1.00.992 to 1.024 (lab setting); 0.989 to 1.067 (POL sites)
    Accuracy (Intercept)Close to 0-3.60 to 0.92 (lab setting); -8.85 to 2.30 (POL sites)
    ACE Axcel Potassium ISEPrecision (Within-run CV)(Not explicitly stated, but low CV is desirable)0.6 to 3.5% (lab setting, 4 levels, 21 days); 1.0 to 1.6% (POL sites, 3 locations, 5 days)
    Precision (Total CV)(Not explicitly stated, but low CV is desirable)1.3 to 3.5% (lab setting); 1.1 to 1.6% (POL sites)
    Accuracy (Correlation Coefficient)Close to 1.00.9974 (lab setting, 115 samples); 0.9973 to 0.9996 (POL sites, 3 locations)
    Accuracy (Slope)Close to 1.00.989 to 1.015 (lab setting); 0.960 to 1.035 (POL sites)
    Accuracy (Intercept)Close to 0-0.050 to 0.095 (lab setting); -0.194 to 0.216 (POL sites)
    ACE Axcel Chloride ISEPrecision (Within-run CV)(Not explicitly stated, but low CV is desirable)0.5 to 1.0% (lab setting, 4 levels, 21 days); 0.9 to 1.5% (POL sites, 3 locations, 5 days)
    Precision (Total CV)(Not explicitly stated, but low CV is desirable)1.1 to 1.5% (lab setting); 1.1 to 2.6% (POL sites)
    Accuracy (Correlation Coefficient)Close to 1.00.9855 (lab setting, 111 samples); 0.9885 to 0.9996 (POL sites, 3 locations)
    Accuracy (Slope)Close to 1.00.939 to 1.002 (lab setting); 0.976 to 1.088 (POL sites)
    Accuracy (Intercept)Close to 0-1.07 to 5.63 (lab setting); -8.16 to 2.22 (POL sites)

    2. Sample Size Used for the Test Set and Data Provenance

    • ACE Glucose Reagent:

      • Accuracy (Correlation Study): 122 samples (ranging from 6 to 729 mg/dL).
      • Accuracy (Patient Correlation Studies): Data from three separate Physician Office Laboratory (POL) sites. Specific sample numbers per POL site are not provided, but the combined sites yielded multiple correlation coefficients, standard error estimates, and confidence intervals for slope and intercept.
      • Provenance: Not explicitly stated, but likely from a laboratory setting and Physician Office Laboratories, presumably within the US given the submission to the FDA. The nature of the samples (e.g., patient samples, control materials) is not specified as prospective or retrospective.
    • ACE Axcel Sodium ISE:

      • Accuracy (Correlation Study): 113 samples (ranging from 45.1 to 194.0 mmol/L).
      • Accuracy (Patient Correlation Studies): Data from three separate Physician Office Laboratory (POL) sites.
      • Provenance: Same as Glucose – likely US lab/POL, nature of samples not specified.
    • ACE Axcel Potassium ISE:

      • Accuracy (Correlation Study): 115 samples (ranging from 1.57 to 14.20 mmol/L).
      • Accuracy (Patient Correlation Studies): Data from three separate Physician Office Laboratory (POL) sites.
      • Provenance: Same as Glucose – likely US lab/POL, nature of samples not specified.
    • ACE Axcel Chloride ISE:

      • Accuracy (Correlation Study): 111 samples (ranging from 63.4 to 176.0 mmol/L).
      • Accuracy (Patient Correlation Studies): Data from three separate Physician Office Laboratory (POL) sites.
      • Provenance: Same as Glucose – likely US lab/POL, nature of samples not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. For this type of in vitro diagnostic device, "ground truth" is established by comparing the performance of the new device to a legally marketed predicate device (Alfa Wassermann ACE Clinical Chemistry System and ACE Reagents) using quantitative measurements, not by expert interpretation. The predicate device itself acts as the reference method in these correlation studies.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, this is a quantitative comparison against a predicate device, not an interpretation-based ground truth requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a clinical chemistry analyzer, an automated system for measuring analytes in samples. It does not involve human readers interpreting images or data where AI assistance would be relevant in the context of MRMC studies.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the performance studies described are inherently "standalone" in the sense that they evaluate the performance of the ACE Axcel Clinical Chemistry System (including its reagents and ISE module) in producing quantitative results independently. The accuracy studies compare its output directly against a predicate device's output. There isn't a "human-in-the-loop" aspect to the core measurement performance itself.

    7. The Type of Ground Truth Used

    The "ground truth" for the accuracy studies is the quantitative result obtained from the predicate device (Alfa Wassermann ACE Clinical Chemistry System and ACE Reagents). The studies are correlation studies comparing the new device's measurements (y) to the predicate device's measurements (x).

    8. The Sample Size for the Training Set

    Not applicable. This document describes a traditional medical device (clinical chemistry analyzer), not a machine learning or AI-based device that typically has a "training set." The device is intended to perform measurements based on established chemical and electrochemical principles.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of this device's development as described in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101089
    Manufacturer
    Date Cleared
    2011-07-13

    (450 days)

    Product Code
    Regulation Number
    866.5040
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    EASYRA MICRO-ALBUMIN REAGENT AND CALIBRATOR

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EasyRA µALB reagent is intended for the quantitative determination of micro-albumin in human urine, using the MEDICA "EasyRA Chemistry Analyzer" in clinical laboratories. Micro-albumin measurements using immunological tests aid in the diagnosis of kidney diseases.

    The EasyRA micro-albumin (µALB) calibrator facilitates measurements of µALB on the EasyRA clinical chemistry analyzer when used in conjunction with Medica's µALB reagent. The µALB calibrator is used to establish points of reference that are used in the determination of values in the measurement of µALB in human urine.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called "EasyRA micro-albumin (uALB) Reagent and EasyRA microalbumin (uALB) Calibrator." It describes the intended use of the device and confirms its substantial equivalence to predicate devices. However, this document does not contain any information regarding acceptance criteria, study details, sample sizes, ground truth establishment, or expert involvement as requested in the prompt.

    Therefore, I cannot fulfill the request to describe the acceptance criteria and the study that proves the device meets them based on the provided text. The document is primarily a regulatory clearance letter.

    Ask a Question

    Ask a specific question about this device

    K Number
    K030006
    Manufacturer
    Date Cleared
    2003-03-07

    (64 days)

    Product Code
    Regulation Number
    862.1035
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    ATAC PAK ALBUMIN REAGENT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ATAC PAK Albumin Reagent Kit is intended for use with the ATAC Calibrator and the ATAC 8000 Random Access Chemistry System as a system for the quantitative determination of albumin in serum and plasma. Albumin results are used for the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.

    Device Description

    The ATAC PAK Albumin Reagent Kit is intended for the quantitative determination of albumin in serum and plasma. Albumin results are used for the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys. The ATAC PAK Albumin Reagent determines albumin by the selective binding of bromcresol green. The resulting increase in absorbance at 630 nm is proportional to the albumin concentration in the sample.

    AI/ML Overview

    The ATAC PAK Albumin Reagent Kit is intended for the quantitative determination of albumin in serum and plasma. The study presented demonstrates its substantial equivalence to the Roche Albumin Reagent Kit (product no. 1970569).

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (Implied by equivalence to predicate)Reported Device PerformanceStudy Type
    Linearity RangeEquivalent to predicate (e.g., span useable range)0.1 to 8 g/dLLinearity/Recovery
    Recovery RegressionHigh correlation (e.g., r > 0.99) and low Sy.xATAC = 0.04 + 0.902x (Ref), r = 0.999, Sy.x = 0.11 g/dLLinearity/Recovery
    Within-Run Precision%CV typically
    Ask a Question

    Ask a specific question about this device

    K Number
    K021636
    Device Name
    ALBUMIN REAGENT
    Date Cleared
    2002-07-17

    (61 days)

    Product Code
    Regulation Number
    862.1035
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    ALBUMIN REAGENT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    An albumin test system is an in vitro device intended to measure albumin concentration quantitatively in serum or plasma. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text does not contain information about acceptance criteria, device performance, study details, or ground truth establishment for a medical device. The document is a 510(k) clearance letter from the FDA for an "Albumin Reagent" device, indicating that it has been found substantially equivalent to a legally marketed predicate device. It defines the intended use of the device but does not include any performance data or a description of a study.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    SYNCHRON LX 20 CLINICAL CHEMISTRY SYSTEM, SYNCHRON LX ALBUMIN REAGENT, SYNCHRON LX CREATININE PICRIC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SYNCHRON LX20 Clinical Chemistry System is a fully automated, computer controlled, clinical chemistry analyzer intended for the in vitro quantitative measurement of a variety of analytes of clinical interest in biological fluids, such as, serum, plasma, urine, and cerebral spinal fluid (sample type is chemistry dependent).

    Device Description

    The LX20 Clinical Chemistry System is a fully automated, computer controlled, clinical chemistry analyzer intended for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries of clinical interest in biological fluids such as serum, plasma, urine, and cerebral spinal fluid (sample type is chemistry dependent). The analyzer operates in conjunction with reagents, calibrators, and controls designed for use with the system. The instrument features bar code identification of samples and reagents. It automatically dilutes samples and delivers them to the reaction cuvette along with readents and reaction constituents. The system analyzes up to 100 samples per run with up to 41 analytes per sample. Major hardware components include a reagent compartment, sample and reagent cranes, cartinge chemistry section, modular chemistry section, sample carousel and crane, hydropneumatics, electronics, and power supplies.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the SYNCHRON LX™20 Clinical Chemistry System, based on the provided text:

    Acceptance Criteria and Device Performance

    The acceptance criteria for the SYNCHRON LX™20 Clinical Chemistry System appear to be defined by its substantial equivalence to existing predicate devices. This is demonstrated through method comparison and imprecision experiments, where the device's results are compared to those obtained from selected predicate methods. The performance is considered acceptable if it shows good correlation (high 'r' value, close to 1) and a slope close to 1 and intercept close to 0 in the method comparison studies, along with acceptable levels of imprecision (low %C.V.).

    Table of Acceptance Criteria and Reported Device Performance:

    Since specific acceptance thresholds for slope, intercept, r, SD, and %C.V. are not explicitly stated as numerical criteria in the document, I will infer them based on what would typically be considered good performance for substantial equivalence in clinical chemistry, and present the reported performance against general expectations.

    AnalyteAcceptance Criteria (General Expectation)Reported Device Performance (Method Comparison: Slope, Intercept, r)Reported Device Performance (Within-run Imprecision: Mean, SD, %C.V., N)
    Method ComparisonSlope close to 1.0 (e.g., 0.95-1.05)
    Intercept close to 0.0 (e.g., within ± a small absolute value)
    Correlation coefficient (r) > 0.97 (ideally > 0.99)--
    Albumin-0.970, 0.070, 0.997Level 1: 2.21 g/dL, 0.04, 1.9%, 80
    Level 2: 4.81 g/dL, 0.03, 0.6%, 80
    Creatinine-1.010, 0.04, 0.999Level 1: 0.59 mg/dL, 0.07, 12.2%, 80
    Level 2: 8.24 mg/dL, 0.18, 2.1%, 80
    Glucose-0.991, -0.250, 0.996Level 1: 43.7 mg/dL, 1.3, 2.9%, 80
    Level 2: 397.1 mg/dL, 1.7, 0.4%, 80
    Phosphorus-1.023, 0.27, 0.996Level 1: 1.80 mg/dL, 0.04, 2.5%, 80
    Level 2: 7.04 mg/dL, 0.05, 0.7%, 80
    Total Protein-0.962, -0.01, 0.993Level 1: 3.54 g/dL, 0.06, 1.8%, 80
    Level 2: 7.52 g/dL, 0.05, 0.7%, 80
    Urea Nitrogen-0.949, 1.13, 0.999Level 1: 7.7 mg/dL, 0.4, 5.0%, 80
    Level 2: 58.7 mg/dL, 0.5, 0.9%, 80
    Carbon Dioxide-1.015, -0.70, 0.992Level 1: 10.58 mmol/L, 0.26, 2.5%, 80
    Level 2: 30.16 mmol/L, 0.29, 1.0%, 80
    Calcium-0.990, -0.92, 0.998Level 1: 7.5 mg/dL, 0.08, 1.1%, 80
    Level 2: 13.5 mg/dL, 0.11, 0.8%, 80
    Chloride-0.974, 1.04, 0.997Level 1: 81.11 mmol/L, 0.47, 0.6%, 80
    Level 2: 120.53 mmol/L, 0.84, 0.7%, 80
    Potassium-1.008, -0.01, 0.999Level 1: 2.4 mmol/L, 0.02, 0.9%, 80
    Level 2: 7.29 mmol/L, 0.04, 0.6%, 80
    Sodium-1.028, -4.22, 0.993Level 1: 111.1 mmol/L, 0.52, 0.5%, 80
    Level 2: 170.2 mmol/L, 1.01, 0.6%, 80
    BenzodiazepineConcordance = 100% agreementConcordance = 100% agreementCutoff Calibrator: 293 mA/min, 2.4, 0.8%, 20
    High Control: 362 mA/min, 3.0, 0.8%, 20
    Iron-0.976, 9.03, 0.997Level 1: 51.2 ug/dL, 1.2, 2.4%, 80
    Level 2: 262.6, 2.2, 0.9%, 80
    Magnesium-0.971, 0.09, 0.997Level 1: 1.12 mg/dL, 0.03, 3.0%, 80
    Level 2: 3.63 mg/dL, 0.06, 1.5%, 80
    Phenobarbital-0.976, 0.050, 0.992Level 1: 9.32 ug/mL, 0.22, 2.3%, 80
    Level 2: 65.86 ug/mL, 1.42, 2.2%, 80
    Uric Acid-0.977, -0.02, 0.999Level 1: 2.42 mg/dL, 0.03, 1.1%, 80
    Level 2: 10.48 mg/dL, 0.05, 0.5%, 80
    Lactate Dehydrogenase-1.018, 0.20, 0.999Level 1: 49.9 IU/L, 1.6, 3.2%, 80
    Level 2: 363.5 IU/L, 3.4, 1.0%, 80
    Imprecision%C.V. generally below 5%, with some exceptions for very low concentrations or specific analytes--

    Study Details

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: The sample sizes are explicitly stated for the "Estimated Serum/Plasma Within-run Imprecision" for each analyte and level. For most analytes, the imprecision study used N=80 samples (likely runs or replicates) for each of two levels (Level 1 and Level 2). For Benzodiazepines, it was N=20 for each of the Cutoff Calibrator and High Control.
    • Data Provenance: The document does not specify the country of origin of the data. The studies described are performance validation studies for a new medical device, implying them to be prospective in nature, conducted to evaluate the device against established predicate methods. The samples were "Serum/Plasma" samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • The concept of "ground truth" established by experts (like radiologists) is typically relevant for interpretative diagnostic tests, not for quantitative clinical chemistry analyzers. For this type of device, the "ground truth" or reference standard for comparison is the results obtained from the predicate device itself. Therefore, specific human experts were not explicitly used to establish a subjective "ground truth" for the test set in the way they would be for image interpretation. The predicate device's results are taken as the validated reference.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This concept of "adjudication" is also generally applied to subjective diagnostic evaluations where multiple human readers interpret data, and discrepancies need to be resolved. For a quantitative clinical chemistry analyzer, adjudication between human readers is not applicable. The comparison is objective, between the results of the new device and the predicate device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This type of study assesses improvements in human performance with AI assistance and is irrelevant for a highly automated quantitative clinical chemistry analyzer. The device performs the measurements automatically without direct human interpretation of raw data in a way that would be "assisted" by AI.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Yes, this entire study represents standalone performance. The SYNCHRON LX™20 Clinical Chemistry System is an automated analyzer, and the reported performance data (method comparison, imprecision) reflects the algorithm and hardware's ability to measure analytes independently. Human involvement is in operating the instrument and interpreting the final quantitative results, but the measurement itself is automated.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the method comparison studies, the "ground truth" as a reference for comparison was the results obtained from the specified predicate devices. These predicate devices are already cleared for commercial distribution and their performance is established. For imprecision studies, the ground truth is statistical reliability of the device's own measurements.

    8. The sample size for the training set

    • Not Applicable / Not Provided. Clinical chemistry analyzers like the SYNCHRON LX™20 typically do not rely on "training sets" in the way machine learning algorithms do. Their performance is based on established chemical reactions, photometric principles, and instrument design. Calibration is performed using specific calibrators (e.g., SYNCHRON LX™ AQUA CAL 1, 2, 3) which are part of routine operation, but this is distinct from an AI "training set." The document does not mention any machine learning or AI components that would require a dedicated training set.

    9. How the ground truth for the training set was established

    • Not Applicable. As explained above, for this type of device, there is no "training set" in the context of AI/machine learning, and therefore no ground truth established for such a set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1