510(k) Data Aggregation

    K Number
    K162538
    Date Cleared
    2016-11-17

    (66 days)

    Product Code
    Regulation Number
    862.1150
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ADVIA Centaur Insulin (IRI) Master Curve Material (MCM), ADVIA Centaur (IRI) Calibrator

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ADVIA Centaur® Insulin (IRI) Master Curve Material (MCM) is for in vitro diagnostic use in the verification of calibration and reportable range of the ADVIA Centaur® Insuin assay.

    The ADVIA Centaur® Insulin (IRI) Calibrator is for in vitro diagnostic use in calibrating the ADVIA Centaur® systems Insulin assays.

    Device Description

    ADVIA Centaur® Insulin (IRI) Master Curve Material is a set of ten vials of material containing various levels of insulin in buffered saline with casein, potassium thiocyanate (3.89%), sodium azide (

    AI/ML Overview

    The provided text describes modifications to the "ADVIA Centaur® Insulin (IRI) Master Curve Material (MCM)" and "ADVIA Centaur® Insulin (IRI) Calibrator" and outlines the studies conducted to demonstrate their substantial equivalence to predicate devices.

    Here's an analysis of the acceptance criteria and the study information, extracting what is available:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state quantitative acceptance criteria in a table format with corresponding performance metrics for the modified devices. Instead, it states that "Verification testing was conducted and results show that the modified devices meet pre-determined acceptance criteria." and that a "method comparison study demonstrates that human clinical sample results are equivalent with both calibrators (modified and unmodified)."

    However, based on the context, the implicit acceptance criteria and reported performance would be:

    Acceptance Criteria CategoryImplicit Acceptance Criteria (based on "same" characteristics and equivalence claims)Reported Device Performance (based on "same" characteristics and equivalence claims)
    ADVIA Centaur® Insulin (IRI) Master Curve Material (MCM)
    Intended UseSame as predicateSame as predicate
    InstrumentSame as predicate (ADVIA Centaur Systems)Same as predicate (ADVIA Centaur Systems)
    FormSame as predicate (Liquid)Same as predicate (Liquid)
    MatrixSame as predicate (Buffered saline with casein, potassium thiocyanate, sodium azide and preservatives)Same as predicate (Buffered saline with casein, potassium thiocyanate, sodium azide and preservatives)
    Number of LevelsSame as predicate (10)Same as predicate (10)
    Target ConcentrationsSame as predicate (0.0, 2.5, 4.5, 10.0, 20.0, 39.0, 79.0, 158, 225, 330 mU/L)Same as predicate (0.0, 2.5, 4.5, 10.0, 20.0, 39.0, 79.0, 158, 225, 330 mU/L)
    Fill VolumeSame as predicate (1.0 mL)Same as predicate (1.0 mL)
    Shelf Life/StabilitySame as predicate (12 months at 2-8 degrees C)Same as predicate (12 months at 2-8 degrees C)
    On board StabilitySame as predicate (8 hours on board)Same as predicate (8 hours on board)
    ADVIA Centaur® Insulin (IRI) Calibrator
    Intended UseSame as predicateSame as predicate
    InstrumentSame as predicate (ADVIA Centaur Systems)Same as predicate (ADVIA Centaur Systems)
    FormSame as predicate (Liquid)Same as predicate (Liquid)
    MatrixSame as predicate (Buffered saline with casein, potassium thiocyanate, sodium azide and preservatives)Same as predicate (Buffered saline with casein, potassium thiocyanate, sodium azide and preservatives)
    Number of LevelsSame as predicate (2)Same as predicate (2)
    Target ConcentrationsSame as predicate (Low: