Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    Why did this record match?
    Device Name :

    ACUSON S1000, S2000, S3000 Diagnostic Ultrasound Systems, ACUSON P200 Diagnostic Ultrasound System, ACUSON
    P500 Diagnostic Ultrasound System, ACUSON NX3, NX3 Elite Diagnostic Ultrasound System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACUSON Sequoia ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Fetal, Abdominal, Pediatric, Neonatal Cephalic, Small Parts, OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures), Cardiac, Pelvic, Vascular, Musculoskeletal and Peripheral Vascular applications.

    The system also provides the ability to measure anatomical structures for fetal, abdominal, pediatric, small organ, cardiac, transrectal, transvaginal, peripheral vessel, musculoskeletal and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

    The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transesophageal, Adult Cephalic, Peripheral Vessel, Abdominal, Intraoperative Abdominal, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

    The typical examinations performed using the SC2000 Ultrasound System are: Cardiac Imaging Applications and Analysis, Vascular Imaging Applications and Analysis, Superficial Imaging Applications, Intraoperative Imaging Applications, Transcranial Imaging Applications.

    The ACUSON Freestyle Ultrasound System is intended for diagnostic imaging or fluid flow analysis of the human body performed by an appropriately wained healthcare professional in a healthcare setting for the following conditions: Abdominal, Pediatric, Small Organ, Peripheral Vessel, Musculoskeletal (Conventional), Musculoskeletal (Superficial).

    The ultrasound imaging systems are intended for the following applications: Fetal. Abdominal. Intraoperative, Pediatric. Small Parts, Transcranial, OB/GYN (including monitoring of the ovarian follicle development), Cardiac. Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.

    The system also provides the ability to measure anatomical structures {fetal, abdominal, intraoperative, small organ, neonatal cephalic, adult cephalic, trans-esophageal, transvaginal, peripheral vessel, musculoskeletal (conventional), musculo-skeletal (superficial) and neonatal cardiac} and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

    The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement: Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging".

    The ACUSON AcuNav Ultrasound Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients.

    The ACUSON P200 ultrasound imaging system is intended for the following applications: Fetal, Abdominal, Pediatric, Small Parts, OB/GYN (useful for visualization of the ovaries, follicles, uterus, and other pelvic structures), Adult, Pediatric and Neonatal Cardiac, Pelvic, Neonatal Cephalic, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.

    The system also provides the ability to measure anatomical structures; fetal, abdominal, pediatric, small organ, neonatal cephalic, cardiac (adult, pediatric and neonatal), trans-esophageal, transrectal, transvaginal, peripheral vessel, musculoskeletal (conventional), musculosketal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

    The ACUSON P500 ultrasound imaging system is intended for the following applications: Fetal, Abdominal (including live), Pediatric, Small Parts, Transcranial, OB/GYN(useful for visualization of the ovaries, follicles, uterus and other pelvic structures), Pelvic. Neonatal, Cardiac, Vascular(including Peripheral Vessel), Musculoskeletal, Superficial Musculoskeletal, and Urology applications.

    The system also provides the ability to measure anatomical structures fetal, abdominal, small organ, transrectal, transvaginal, cardiovascular, peripheral vessel, musculoskeletal (conventional), and musculoskeletal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

    The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system.

    This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Society of Echocardiography Carotid Intima-Media Thickness Task Force. Endorsed by the Society for Vascular Imaging."

    The ACUSON Acunav Ultrasound Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other of adult and pediatric patients.

    For ACUSON NX3: The ACUSON NX3 ultrasound imaging system is intended for the following applications: Fetal Abdominal (including liver), Pediatric, Small Parts (Small Organ), Neonatal Cephalic, Transcranial, OB/GYN, Pelvic, Neonatal, Cardiac, Vascular (including Peripheral Vessel), Musculoskeletal and Urology applications. The systems also provide for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

    For ACUSON NX3 Elite: The ACUSON NX3 Elite ultrasound imaging system is intended for the following applications: Fetal, Abdominal (including liver, intra-operative), Pediatric, Small Parts (Small Organ including intra-operative), Neonatal Cephalic, Adult Cephalic, Transcranial, OB/GYN, Pelvic, Neonatal, Cardiac(including Transesophageal), Vascular (including Peripheral Vessel, intra-operative), Musculoskeletal, Superficial Musculoskeletal and Urology applications. The systems also provide for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

    The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system.

    Note: This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Society of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Medicine."

    Device Description

    The ACUSON Sequoia Diagnostic Ultrasound System is a multi-purpose mobile, software controlled, diagnostic ultrasound system with an on-screen display of thermal and mechanical indices related to potential bio-effect mechanisms. Its function is to transmit and receive ultrasound echo data and display it in B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Color M Mode, Doppler Tissue Mode, Amplitude Doppler Mode, a combination of modes, Panoramic Imaging, Contrast agent Imaging, Virtual Touch Strain Imaging, Virtual Touch - pSWE Imaging, Virtual Touch - SWE Imaging, syngo Velocity Vector Imaging, Custom Tissue Imaging, 3D/4D Volume Imaging and Harmonic Imaging on a Display.

    All of the transducers and the catheter-based transducers will follow Track 3 acoustic labeling (AIUM 1004, IEC 2007, AIUM/NEMA 2004a) and remain unchanged from the currently cleared ACUSON systems.

    Note: Some ACUSON Diagnostic Ultrasound configurations of components, accessories and/or software may vary from device brands and within device families. There are no modifications, additions or labeling changes to components, accessories and/or software for each device indicated.

    The ACUSON Diagnostic Ultrasound systems are intended to be used by trained medical professionals at various locations where patient care, evaluation, monitoring or research is performed.

    AI/ML Overview

    The provided document is a 510(k) Pre-market Notification for several Siemens ACUSON Diagnostic Ultrasound Systems, seeking clearance based on substantial equivalence to previously marketed devices. The document explicitly states that no new features, modifications, or changes were made to the hardware or software of the devices themselves. The entire submission focuses on an "expanded labeling" for lung ultrasound imaging, supported by existing clinical literature.

    Therefore, the acceptance criteria and the study proving the device meets these criteria are not related to a new device or a new AI/algorithm feature being validated. Instead, it is about demonstrating that the existing, cleared devices can be marketed with expanded indications for use, supported by references to previously conducted tests and existing clinical literature.

    Given this, I cannot provide information on:

    • Reported device performance (as no new performance was claimed or tested)
    • Sample size used for the test set or data provenance
    • Number of experts used to establish ground truth or their qualifications
    • Adjudication method
    • MRMC comparative effectiveness study or effect size
    • Standalone algorithm performance
    • Sample size for training set or how ground truth for training set was established

    The document explicitly states: "clinical studies were not required to support substantial equivalence. The expanded Labeling for Lung Ultrasound imaging (LUS) is supported by clinical literature." and "There are no software or hardware modifications to any device under this review and the expanded labeling has no impact on the individual ACUSON ultrasound device, therefore, testing results are not included in this Special 510(k) submission."

    Below is a table summarizing the acceptance criteria and "device performance" in the context of this 510(k) submission, emphasizing that the "performance" here refers to the device meeting the previously established safety and effectiveness for its predicate indications, which is being extended to a new recognized clinical use case via labeling.


    Description of Acceptance Criteria and Evidence for ACUSON Diagnostic Ultrasound Systems (K202683)

    This 510(k) submission (K202683) by Siemens Medical Solutions, USA, Inc., is a Special 510(k). This type of submission is used when changes to a cleared device do not alter its fundamental scientific technology or intended use, but may expand its labeling or functionality within its established safety and effectiveness profile. In this specific case, the submission focuses on expanded labeling for Lung Ultrasound Imaging (LUS), utilizing the existing capabilities of the cleared ACUSON Diagnostic Ultrasound Systems and their associated transducers.

    Crucially, the document states that no modifications were made to the software or hardware of the devices. Therefore, the "acceptance criteria" and "proof" are based on demonstrating substantial equivalence to previously cleared predicate devices and leveraging existing validation data for those predicates, rather than presenting new performance metrics for a novel algorithm or device.

    1. Acceptance Criteria and Reported Device "Performance" (Meeting Substantial Equivalence)

    Acceptance Criteria (from 510(k) perspective)Reported Device "Performance" / Justification for Meeting Criteria
    Identical Intended Use/Indications for Use: The devices, with expanded labeling, align with the fundamental intended use of previously cleared predicate devices.The Indications for Use statements for all listed ACUSON systems (Sequoia, SC2000, Freestyle, S1000/S2000/S3000, P200, P500, NX3/NX3 Elite) are presented as identical to their respective primary predicate devices, with the exception of the added use case for lung imaging explicitly mentioned in the "Indications for Use" section descriptions (e.g., "Abdominal (including liver, intra-operative)" for NX3 Elite). The core functionality and diagnostic purpose remain the same. The submission argues that lung ultrasound imaging falls within the scope of general diagnostic ultrasound.
    Identical Technological Characteristics: The device hardware, software, and operating principles are unchanged from their predicate versions, ensuring safety and effectiveness.The document explicitly states: "the subject devices are substantially equivalent to the cleared primary predicate devices and employs the same fundamental scientific technology and intended use/indications for use."

    Detailed comparison tables (Table 2 in the original document, not fully reproduced here due to length) consistently list "Identical" for all compared technological characteristics: Reusability, Duration of Use, Scientific Technology, Operating Principles, Type of Previously Cleared Transducers, Acoustic Outputs, Previously Cleared Imaging Modes, and Biocompatibility (ISO 10993-1). |
    | Safety and Effectiveness: There are no new safety or effectiveness concerns introduced by the expanded labeling. | - Software Verification and Validation: The CRS (Customer Requirement Specifications) for each device were verified/validated during original development. No software/hardware modifications for this submission.

    • Standards Compliance: Conforms to IEC 62304, IEC 62366-1, NEMA PS 3.1 - 3.20 (DICOM).
    • Risk Management: Risk analysis (ISO 14971:2007) was performed on predicate devices, and mitigation controls were implemented. The expanded labeling does not impact these.
    • Cybersecurity: Cybersecurity measures are addressed and conform to requirements to prevent unauthorized access, modification, misuse, or denial of use.
    • Clinical Literature Support: The expanded labeling for Lung Ultrasound Imaging (LUS) is supported by clinical literature, indicating that the use of ultrasound for lung imaging is an established and safe practice. Clinical studies were not required for this 510(k) as the technology and operating principles are unchanged. |

    2. Sample Size and Data Provenance for Test Set

    • Test Set Description: Not applicable. For this Special 510(k), no new test set data for device performance was generated or presented. The submission relies on existing validation data for the predicate devices.
    • Data Provenance: The underlying data for previously cleared predicate devices would have been generated through rigorous internal testing and validation processes during their initial development and clearance. The document implies that this data exists and supports the current device. The document explicitly states: "Non-clinical Testing has been conducted during product development for each of the predicate ACUSON Diagnostic Ultrasound device described in this submission."

    3. Number, Qualifications, and Adjudication Method for Experts

    • No new expert assessment: Not applicable for this Special 510(k). Since no new AI/algorithm features or changes to device performance were introduced, there was no independent expert review of a new test set for the purpose of this submission. The "ground truth" for the expanded labeling (Lung Ultrasound) is based on established clinical recognition and existing peer-reviewed literature.
    • Adjudication method: Not applicable.

    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No new MRMC study: Not applicable. This submission does not involve a new AI or algorithm that would require an MRMC study to demonstrate human improvement with AI assistance.

    5. Standalone (Algorithm Only) Performance Study

    • No standalone algorithm: Not applicable. The submission is for diagnostic ultrasound systems, not a standalone algorithm. The core functionality is the ultrasound imaging, and the "expansion" is in the labeling for an existing modality.

    6. Type of Ground Truth Used for Device Functions

    • Existing Validated Performance: For the predicate devices, the ground truth and validation would have been established through a combination of:
      • Direct measurement and engineering specifications: For image quality, acoustic output, and system performance.
      • Clinical validation: Through various means including phantom studies, retrospective/prospective clinical data, and expert consensus for the initial indications.
    • Clinical Literature/Consensus for Expanded Indication: For the new labeling related to Lung Ultrasound Imaging (LUS), the "ground truth" or justification is based on:
      • Established clinical practice: Recognition within the medical community for the utility of ultrasound in lung assessment.
      • Peer-reviewed studies and professional guidelines: As explicitly stated, the LUS labeling is "supported by clinical literature." This effectively implies that the medical community's consensus on the utility of LUS (as documented in such literature) serves as the "ground truth" for its clinical relevance.

    7. Sample Size for Training Set

    • No new training set: Not applicable. As no new AI/algorithm components were introduced or modified, there was no "training set" for the purpose of this specific 510(k) submission.

    8. How Ground Truth for Training Set was Established

    • No new training set: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K150050
    Date Cleared
    2015-02-09

    (28 days)

    Product Code
    Regulation Number
    892.1550
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ACUSON P500 Diagnostic Ultrasound System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACUSON P500 ultrasound system is intended for diagnostic ultrasound imaging and fluid flow analysis for the following applications: Fetal, Abdominal, Small Parts, OB/GYN, Pelvic, Cardiovascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.

    The system also provides the ability to measure anatomical structures fetal, abdominal, small organ, transrectal, transvaginal, cardiovascular, peripheral vessel, musculoskeletal (conventional), and musculoskeletal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

    Device Description

    The ACUSON P500 is a multi-purpose mobile, software controlled, diagnostic ultrasound system with an on-screen display for thermal and mechanical indices related to potential bioeffect mechanisms. Its function is to acquire harmonic ultrasound echo data and display it in B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Color M mode, Tissue Doppler Image, Power (Amplitude) Doppler Mode and a combination of modes on a Flat Panel Display.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from Siemens Medical Solutions USA, Inc. for their ACUSON P500 Ultrasound System, submitted to the FDA on January 12, 2015. The purpose of the submission is to demonstrate substantial equivalence to legally marketed predicate devices, the ACUSON X300 ultrasound system (K121699) and the ACUSON S2000 ultrasound system (K140959).

    Based on the provided document, here's an analysis of the acceptance criteria and study information:

    1. A table of acceptance criteria and the reported device performance:

    The document does not explicitly present a table of "acceptance criteria" with quantitative performance metrics for the ACUSON P500 ultrasound system itself. Instead, it demonstrates substantial equivalence to predicate devices based on:

    • Identical Indications for Use: The ACUSON P500 shares the same intended uses as the predicate devices across various applications (Fetal, Abdominal, Small Organ, Cardiac, Transrectal, Transvaginal, Peripheral vessel, Musculo-skeletal (conventional), Musculo-skeletal (superficial), Emergency Medicine).
    • Similar Technological Characteristics: The document provides a detailed comparison table (Page 5) showing that the ACUSON P500 largely shares:
      • Supported Center Frequencies (2.0 MHz to 11.0 MHz across the board for all three devices).
      • Imaging Modes (B, M, PWD, SCW, Color Doppler, Power Doppler, Combined modes, THI, M-THI are all present in the P500 and the predicates).
      • Many Features (e.g., Dual-Beam Processing, Quad-Beam Processing, Clip Capture, DIMAQ, Multiple Frequency Imaging, Digital Architecture, DICOM, Dynamic TCE, Wireless functionality).
    • Compliance with Safety Standards: The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety, conforming to applicable medical device safety standards like UL 60601-1, IEC 60601-2-37, CSA C22.2 No. 601-1, AIUM/NEMA UD-3, AIUM/NEMA UD-2, 93/42/EEC Medical Devices Directive, EN/IEC 60601 series, EN/IEC 62304, EN/IEC 62366, and ISO 10993-1.

    Reported Device Performance (by inference of substantial equivalence):

    Based on the assertion of substantial equivalence, the implied "performance" of the ACUSON P500 is considered to be equivalent to that of the predicate devices. The document does not provide specific numerical performance metrics (e.g., sensitivity, specificity, resolution) for the ACUSON P500 or its predicates. It focuses on functional equivalence.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    The document explicitly states under "F. A summary discussion of the clinical tests submitted, referenced, or relied on for a determination of substantial equivalence": "Since the P500 uses the same technology and principles as existing devices, clinical data is not required."

    Therefore, no clinical test set was required or used for this 510(k) submission. The entire evaluation for substantial equivalence was based on non-clinical tests (engineering, safety, and performance characteristic comparisons to predicates).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    Not applicable. As no clinical data/test set was required, no experts were needed to establish ground truth for this submission.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable. No clinical test set was used.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. The document is for an ultrasound system, not an AI-assisted diagnostic device. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. The device is a diagnostic ultrasound system, not a standalone AI algorithm. Its performance is inherent to its hardware and software for image acquisition and display, which are deemed substantially equivalent to existing devices.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    Not applicable. As no clinical data/test set was required, no ground truth was established for this submission. The "ground truth" for the FDA's substantial equivalence determination here relies on the established safety and effectiveness of the predicate devices and the P500's compliance with recognized safety standards through non-clinical testing.

    8. The sample size for the training set:

    Not applicable. The document describes a medical device seeking 510(k) clearance based on substantial equivalence to existing predicate devices, not an AI/machine learning device that would typically involve a training set.

    9. How the ground truth for the training set was established:

    Not applicable. No training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1